GRYVOZE GV-DTFM-IN1DP5 1.5 Inch Digital Turbine Flow Meter User Manual

June 16, 2024
GRYVOZE

GRYVOZE GV-DTFM-IN1DP5 1.5 Inch Digital Turbine Flow Meter

Product Information

  • Specifications
    • Flowmeter Types: 1, 1.5, 2, 2.5, and 3
    • Initial Coefficients:
    • 1 Flowmeter: 1000
    • 1.5 Flowmeter: 1850
    • 2 Flowmeter: 4900
    • 2.5 Flowmeter: 1000
    • 3 Flowmeter: 1000

Product Usage Instructions

  • On-Site Calibration
    • After installing the turbine flowmeter at the designated location, calibrate the flowmeter under actual working conditions to improve its measurement accuracy and precision.
  • To perform on-site calibration:
    • Press the CALIBRATE (C) and DISPLAY (D) keys simultaneously to view the coefficient of the flowmeter.
    • Hold down the CALIBRATE (C) key for 5 seconds to enter the coefficient setting mode. The coefficient will blink.
    • Use the DISPLAY (D) key to adjust the coefficient and the CALIBRATE (C) key to switch coefficients.
    • After adjustment, wait for 10 seconds for the mode to automatically exit.
  • The Calibration Formula for the Flowmeter
    • Calibrated Coefficient = Actual Flow Rate / Displayed Flow Rate Coefficient Before Calibration.
    • For example, if the flowmeter displays 98L when testing 100L of liquid, and the coefficient is displayed as 1000, the coefficient should be adjusted to: 100L / 98L * 1000 = 1020.
  • Other Settings Adjustments
    • Reset Single Time Metering
    • Press the DISPLAY (D) key once.
    • Check Total Accumulated
    • Under the Initial Interface Display, press the CALIBRATE (C) key once to enter the total accumulated display (TOTAL).
    • To return to the Initial Interface Display, press the CALIBRATE (C) key again.
  • Reset Total Accumulated
    • Press and hold the DISPLAY (D) key for 5 seconds to clear the total accumulated value.
  • Metering Unit Setting
    • To set the metering unit:
    • Press and hold the DISPLAY (D) key for 5 seconds to enter the unit setting mode.
    • Use the CALIBRATE (C) key to switch between units.
    • Release the DISPLAY (D) key to save the selected unit.
  • FAQ – Frequently Asked Questions
    • No change in flow rate Possible solutions:
    • Increase the flow of the medium.
    • Remove the turbine component and check for any foreign objects blocking it.
    • If the sensor is damaged, replace the chip.
    • Low flow rate
    • Possible solutions:
    • Increase the flow of the medium.
    • Replace with a flow meter of a smaller range.
    • The screen does not display text
    • Possible solutions:
    • Check if the battery is detached or out of power.
    • If the chip is damaged, replace it.

FUNCTION

TWO FORMS OF USE

GRYVOZE-GV-DTFM-IN1DP5-1-5-Inch-Digital-Turbine-Flow-Meter-FIG-1
\(1\)

ON-SITE CALIBRATION

  • After installing the turbine flowmeter at the designated location, calibrate the flowmeter under actual working conditions to improve its measurement accuracy and precision.
  • On-site calibration requires adjusting the coefficient of the flowmeter.
  • Pressing the C and D keys simultaneously allows you to view the coefficient of the flowmeter.
  • Hold down the C key for 5 seconds to enter the coefficient setting mode (the coefficient will blink).
  • Use the D key to adjust the coefficient and the C key to switch coefficients.
  • After adjustment, wait for 10 seconds for the mode to automatically exit.

THE CALIBRATION FORMULA FOR THE FLOWMETER

  • Calibrated Coefficient = Actual Flow Rate / Displayed Flow Rate * Coefficient Before Calibration.
  • For example, if the flowmeter displays 98L when testing 100L of liquid, and the coefficient is displayed as 1000, the coefficient should be adjusted to: 100L / 98L * 1000 ≈ 1020.

INITIAL COEFFICIENTS

  • The initial coefficients for the 1″ flowmeter, 2.5″ flowmeter, and 3″ flowmeter are all 1000; the initial coefficient for the 1.5″ flowmeter is 1850; the initial coefficient for the 2″ flowmeter is 4900.

OTHER SETTINGS ADJUSTMENTS

Reset Single Time Metering

  • Press the D key once.

Check Total Accumulated

  • Under the “Initial Interface Display,” press the C key once to enter the total accumulated display (TOTAL). To return to the “Initial Interface Display,” press the C key again.

Reset Total Accumulated

  • Press and hold the D key for 5 seconds to clear the total accumulated value.

Metering Unit Setting

  • Simultaneously press and hold the C and D keys for more than 5 seconds to enter the metering unit setting.
  • The default metering unit at the factory is ITR (liters), and you can switch between GAL (gallons), QTS (quarts), PTS (pints), L (liters), and m³.
  • Press the C key for 3 seconds to save and exit the metering unit setting.

ANSWERS TO COMMON QUESTIONS

No change in flow rate Increase the flow of the medium.

Remove the turbine component and check for

any foreign objects blocking it.

If the sensor is damaged, replace the chip.
Low flow rate| Increase the flow of the medium.
Replace with a flow meter of a smaller range.
The screen not display text| Check if the battery is detached or out of power.
If the chip is damaged, replace it.