GRYVOZE GV-DTFM-IN1DP5 1.5 Inch Digital Turbine Flow Meter User Manual
- June 16, 2024
- GRYVOZE
Table of Contents
GRYVOZE GV-DTFM-IN1DP5 1.5 Inch Digital Turbine Flow Meter
Product Information
- Specifications
- Flowmeter Types: 1, 1.5, 2, 2.5, and 3
- Initial Coefficients:
- 1 Flowmeter: 1000
- 1.5 Flowmeter: 1850
- 2 Flowmeter: 4900
- 2.5 Flowmeter: 1000
- 3 Flowmeter: 1000
Product Usage Instructions
- On-Site Calibration
- After installing the turbine flowmeter at the designated location, calibrate the flowmeter under actual working conditions to improve its measurement accuracy and precision.
- To perform on-site calibration:
- Press the CALIBRATE (C) and DISPLAY (D) keys simultaneously to view the coefficient of the flowmeter.
- Hold down the CALIBRATE (C) key for 5 seconds to enter the coefficient setting mode. The coefficient will blink.
- Use the DISPLAY (D) key to adjust the coefficient and the CALIBRATE (C) key to switch coefficients.
- After adjustment, wait for 10 seconds for the mode to automatically exit.
- The Calibration Formula for the Flowmeter
- Calibrated Coefficient = Actual Flow Rate / Displayed Flow Rate Coefficient Before Calibration.
- For example, if the flowmeter displays 98L when testing 100L of liquid, and the coefficient is displayed as 1000, the coefficient should be adjusted to: 100L / 98L * 1000 = 1020.
- Other Settings Adjustments
- Reset Single Time Metering
- Press the DISPLAY (D) key once.
- Check Total Accumulated
- Under the Initial Interface Display, press the CALIBRATE (C) key once to enter the total accumulated display (TOTAL).
- To return to the Initial Interface Display, press the CALIBRATE (C) key again.
- Reset Total Accumulated
- Press and hold the DISPLAY (D) key for 5 seconds to clear the total accumulated value.
- Metering Unit Setting
- To set the metering unit:
- Press and hold the DISPLAY (D) key for 5 seconds to enter the unit setting mode.
- Use the CALIBRATE (C) key to switch between units.
- Release the DISPLAY (D) key to save the selected unit.
- FAQ – Frequently Asked Questions
- No change in flow rate Possible solutions:
- Increase the flow of the medium.
- Remove the turbine component and check for any foreign objects blocking it.
- If the sensor is damaged, replace the chip.
- Low flow rate
- Possible solutions:
- Increase the flow of the medium.
- Replace with a flow meter of a smaller range.
- The screen does not display text
- Possible solutions:
- Check if the battery is detached or out of power.
- If the chip is damaged, replace it.
FUNCTION
TWO FORMS OF USE
ON-SITE CALIBRATION
- After installing the turbine flowmeter at the designated location, calibrate the flowmeter under actual working conditions to improve its measurement accuracy and precision.
- On-site calibration requires adjusting the coefficient of the flowmeter.
- Pressing the C and D keys simultaneously allows you to view the coefficient of the flowmeter.
- Hold down the C key for 5 seconds to enter the coefficient setting mode (the coefficient will blink).
- Use the D key to adjust the coefficient and the C key to switch coefficients.
- After adjustment, wait for 10 seconds for the mode to automatically exit.
THE CALIBRATION FORMULA FOR THE FLOWMETER
- Calibrated Coefficient = Actual Flow Rate / Displayed Flow Rate * Coefficient Before Calibration.
- For example, if the flowmeter displays 98L when testing 100L of liquid, and the coefficient is displayed as 1000, the coefficient should be adjusted to: 100L / 98L * 1000 ≈ 1020.
INITIAL COEFFICIENTS
- The initial coefficients for the 1″ flowmeter, 2.5″ flowmeter, and 3″ flowmeter are all 1000; the initial coefficient for the 1.5″ flowmeter is 1850; the initial coefficient for the 2″ flowmeter is 4900.
OTHER SETTINGS ADJUSTMENTS
Reset Single Time Metering
- Press the D key once.
Check Total Accumulated
- Under the “Initial Interface Display,” press the C key once to enter the total accumulated display (TOTAL). To return to the “Initial Interface Display,” press the C key again.
Reset Total Accumulated
- Press and hold the D key for 5 seconds to clear the total accumulated value.
Metering Unit Setting
- Simultaneously press and hold the C and D keys for more than 5 seconds to enter the metering unit setting.
- The default metering unit at the factory is ITR (liters), and you can switch between GAL (gallons), QTS (quarts), PTS (pints), L (liters), and m³.
- Press the C key for 3 seconds to save and exit the metering unit setting.
ANSWERS TO COMMON QUESTIONS
No change in flow rate | Increase the flow of the medium. |
---|
Remove the turbine component and check for
any foreign objects blocking it.
If the sensor is damaged, replace the chip.
Low flow rate| Increase the flow of the medium.
Replace with a flow meter of a smaller range.
The screen not display text| Check if the battery is detached or out of power.
If the chip is damaged, replace it.