Electronic Stethoscope Patch
ES-Patch
User’s Manual
Version:V1-4
ES-Patch Electronic Stethoscope Patch
Electronic Stethoscope Patch
Please read the original instruction manual carefully before use and follow
the instructions
Model:ES-Patch
Product Description
This product’s electronic stethoscope patch is a wireless heart and lung sound electronic stethoscope. It is transmitted through wireless Bluetooth with the Apple iOS application, providing a wireless, comfortable and intelligent measurement experience. And built-in Active Noise Cancellation (ANC) function, reduce the impact of ambient sound, improve sound quality.
Statement
This product is not intended to replace the advice of a professional physician. If you have any doubts about the measurement results, please consult your professional physician.
Product Usage
This product is a general hospital medical device that can instantly display and detect sounds related to the heart, arteries, veins, lungs, and other internal organs, and can be used for long-term monitoring of any patient undergoing physical examination.
3.1 User Background
This product is specially designed for Stethoscope by medical personnel. This
instruction manual provides the necessary information so that the user can
operate the stethoscope without additional training.
3.2 Indications for Use
There are no known contraindications for ES-Patch, although care should be
taken when considering using the device according to the warnings and
precautions below.
ES-Patch is not life-supporting or life sustaining.
Safety Information
Please read, understand and follow all safety information instructions in this manual before using this product. Please keep this operation manual for future reference.
4.1 Symbol Description
Description of safety related labels and signs
| Warning.
| Please refer to the instruction manual.
| Type BF (Body Floating) instrument label: This instrument provides
protection against electric shock and leakage. The applied part is a complete
chest piece with diaphragm.
| This product is subject to ambient temperature limitations.
| This product contains electrical and electronic components and must not be
disposed of using standard refuse collection. Please consult local directives
for disposal of electrical and electronic equipment.
| This product uses Bluetooth wireless transmission.
| Please try to avoid contacting this product with object that contains static
electricity.
| Keep product away from moisture.
| Federal (USA) law restricts this device to sale to or on the order of a clinician.
IP24
| Protection class.
2: Dustproof grade 4: Waterproof grade
| This product is FCC certified for RF radiation.b
Description of word signs
NOTICE| It suggests to pay attention to the situation. If not avoided,
it may cause property loss or damage.
WARNING| It suggests that the situation is dangerous. If not avoided, it
may cause minor injuries and/or property loss or damage
4.2 Precautions
Precautions
◎ Keep this product away from direct sunlight, flammable materials and
environments with high temperature and humidity.
◎ Do not modify the design of this product arbitrarily. Please have the
product repaired by authorized service personnel. Any improper modification or
maintenance shall be borne by the user as a result of the modification.
◎ This product is not safe to use in the MRI environment. Do not use this
product in a MRI environment.
◎ Do not wear the device for bathing or swimming.
◎ Do not immerse the product in liquid, or immerse it in any disinfectant, it
may cause damage to the device.
◎ This product should not be used close to or stacked on other instruments.
◎ In order to avoid affecting the medical electrical equipment, this product
has the function of Bluetooth wireless communication. It is recommended to
keep a safe distance of 30 cm for the electronic medical equipment that should
be avoided.
◎ Pay attention to whether the charging induction pin on the charging base is
short-circuited. Damage to the charging cradle or USB charging device may
result.
◎ This product has been tested and found to comply with FCC RF exposure
guidelines and may not be guaranteed to comply with FCC RF exposure guidelines
when used with accessories that contain metal.
4.3 WARNING
WARNING
◎ When using this product for the first time, please attach it for 20 minutes.
If there is any allergic reaction or any discomfort, please remove it
immediately, please consult your professional physician.
◎ In order to reduce unknown problems, please do not use the electronic patch
stethoscope on the patient for more than 12 hours.
◎ When using the electronic patch stethoscope, please avoid wounds, inflamed
or infected areas. Apply it only to intact and clean skin.
◎ Do not swallow and store this product in a place that is not easily
accessible to children.
◎ To reduce the risk of incorrect results, personal injury and equipment
damage, please follow the recommended instructions in this manual for storage
and operation of this product.
◎ To reduce the risk of electric shock, do not use this product on the patient
if the Stethoscope patch is damaged in any way.
◎ To reduce the risk of damage to the Stethoscope head, do not let the
Stethoscope head close to strong sound sources.
◎ To reduce the risk of infection, please follow the cleaning instructions in
the manual to set up and follow a cleaning plan.
◎ In order to reduce the risk of extremely strong magnetic fields, when using
this product, please avoid close to strong RF signals or portable and/or
mobile RF devices. If you hear sudden or unexpected sounds, keep away from any
radio transmitting antennas.
◎ Do not touch the USB charging socket of this product with metal objects or
conductors.
◎ This product is not suitable for residential environments as it may not
provide adequate protection against radio reception in this environment.
Usage Notice
Clean the electronic Stethoscope patch with a 75% medical alcohol wipe before
starting the measurement and after the measurement.
The electronic Stethoscope patch can be fixed to different Stethoscope
positions on the patient through commercially available medical adhesive
materials, and the noise-cancelling microphone in the patch can reduce ambient
noise and sound feedback. Stream audio to Apple iOS apps via wireless
Bluetooth, and store digitized audio as audio files.
Remind patients that if the skin becomes itchy or allergic during wearing,
they should remove the electronic patch stethoscope and consult a physician.
If the patient has no skin or other discomfort, the electronic patch
stethoscope should not be removed from the body to avoid incomplete records.
NOTICE: In order to provide smooth operation of the APP, it is
recommended to download and use the latest iOS application system for mobile
devices.
The ideal remote monitoring range is within 5 meters from the mobile device.
Remote monitoring is affected by factors such as obstacles, the performance of
the mobile device, and the position of the product attached to the human body.
Product components
Host:
ES-Patch Electronic stethoscope patch
NOTICE: It is recommended to clean the device with a 75% alcohol pad before each use.
USB TYPE-C charging cable:
TYPE-C charging cable NOTICE: This USB TYPE-C cable is for charging only.
Instructions
7.1 On/Off mode
Turn on: lightly press the button for 1.5 seconds, the orange light is
always on and power on.
Stand-by: Press the button for 1.5 seconds in the power-on state, the
orange light will be off and the machine will be turned off.
7.2 Light display The power indicator light is solid orange: the power is
sufficient.
The power indicator flashes orange quickly 3 times: the power is insufficient.
NOTICE: After that, it flashes 3 times every 300 seconds.
The power indicator light orange slowly flashes 6 times: low battery, ready to
shut down.
No power indicator light: no power to charge or damage.
7.3 Frequency response mode It is full frequency mode when it is turned
on;
When you tap the power button once, the power indicator light flashes twice
slowly and switches to heart sound mode;
When you tap the power button again, the power indicator flashes slowly three
times, and it switches to lung sound mode;
When you tap the power button again, the power indicator flashes slowly once,
and switches to full frequency mode;
Repeat the above steps to switch the frequency response mode.
The sequence is: full frequency mode→heart sound mode→lung sound mode.
Users can choose from three different frequency response modes.
Full frequency mode: It can amplify the sound whose frequency is within
1200Hz.
Heart sound mode: It can amplify and emphasize the sound in the frequency
range of 20~200Hz.
Lung sound mode: It can amplify and emphasize the sound in the frequency
range of 100~1000Hz.
APP Instructions
8.1 Download and install APP
Please go to the Apple App Store and search for “Smart Stethoscope” to
download the APP.
8.2 Turn on bluetooth and bind with mobile device.
When the product is turned on, when the Bluetooth indicator keeps flashing, it
means that the Bluetooth has been turned on;
Turn on Bluetooth on the mobile device to search for the device name (ES-
Patch) of the product and select it for connection binding.
When the Bluetooth indicator is off, it means that it has been successfully
bound to the product.
NOTICE:It is recommended to fully charge this product before using it.
After the product is turned on, if there is no binding for more than three
minutes, it will automatically shut down.
8.3 Apply the ES-Patch Stethoscope patch
After the button position of this product faces the outside of the human body,
it is attached to an appropriate position on the human body for monitoring and
recording. 8.4 The measurement data
is automatically uploaded to the bound mobile device via Bluetooth
Select the device and click ES-Patch.
8.5 Recording using the auscultation recording function
Confirm that [device is connected] is displayed on the screen, and click
[Auscultation Recording] to enter the recording screen for recording
operation. Note: Enter the
auscultation recording, and the real-time auscultation waveform will be
displayed at the bottom of the screen.
For the body part you want to record, press the record button to start
recording, and press it once to end the recording.
The value of this product
will reach a stable voice data transmission state after being worn for 10
seconds.
The sound waveforms related to the heart, arteries, veins, lungs, and other
internal organs can be displayed in real time through the APP.
WARNING: When the Bluetooth of the mobile device is not paired with this
product ES-Patch, it is impossible to operate the APP to record.
8.6 Use the recording to play the recorded waveform and sound
Click [Auscultation Play] to enter the recording screen, select the recording
audio to be played, and press the play button to play the recording
audio. 8.7 About information
related to this APP
Click [About] to enter the About screen, which contains the software version
and company website. NOTICE: Only
supports playback of sound after recording, real-time monitoring cannot be
performed during recording.
Running or any strenuous exercise is not recommended after wearing.
Charging Instructions
9.1 Charge confirmation
Please take out the included accessory USB TYPE-C charging cable. NOTICE: The
ES-Patch Stethoscope patch must be charged immediately when it is out of
power.
9.2 Insert the USB plug into this product
Insert the USB TYPE-C head into the USB socket below the product. 9.3 Start
Charging
Insert the USB charging cable into the mobile phone charger (not included with
this product) to use.
LED light display
Blinking orange: charging
Orange light off: fully charged or charging stopped
Note: When charging via USB, please use an adaptor that conforms to
IEC62368-1 or IEC60601-1.
Other operating precautions
- The operating ambient temperature shall be within 41° to 104°F (5° to 40°C), the relative humidity shall be within the range of 15 to 95%, and the atmospheric pressure shall be within the range of 1002hPa to 1008hPa.
- Storage and Shipping Conditions The temperature range should be -4° to 140°F (-20° to 60°C) and the relative humidity should be 15 to 95%.
- To prolong the life of the electronic patch stethoscope, avoid extreme cold, heat, exposure to solvents or oils.
- Medical personnel recommend wearing medical gloves when using this product.
- Failure to follow the recommended care and maintenance instructions may result in damage to the internal parts of this product. Internal damage can lead to product failure, ranging from minor Stethoscope to complete damage.
- Do not disassemble or reassemble this product by yourself, it may cause damage to this product.
- If you experience any problems with the electronic Stethoscope patch, do not attempt to repair it yourself. Please notify the original customer service center to contact the follow-up matters.
- If the electronic patch stethoscope will not be used for several months, please charge it first to get enough power before using it.
- When the power is turned off, the power will gradually decrease. Please confirm whether the power is sufficient before use.
Parts List
WARNING: Do not use unauthorized accessories to avoid unexpected hazards.
Cleaning and Maintenance
- Before starting cleaning, please check the structure of the product for damage and make sure the charging cover is properly closed to avoid liquid seepage.
- ES-Patch electronic stethoscope patch can be sprayed on the product with 75% alcohol solution, and then cleaned and wiped with cotton cloth. After drying, put it back in its original box and store it under the conditions of “Storage Environment” in point 9 of this instruction manual.
- Regularly check the electronic stethoscope patch and accessories USB TYPE-C charging cable for damage. If damaged, do not continue to use it and contact the original factory designated customer service department for replacement.
WARNING: To reduce the risk of infection, the ES-Patch must be cleaned after each user.
Warranty and Service Life
If it is operated normally according to the user manual, the service life is
one year, unless improper operation may shorten the service life.
Original Factory Guarantee During the 1 year warranty period, return the
product to the original factory and it will be repaired or replaced free of
charge.
Troubleshooting
Question | Possible reason | Approach |
---|---|---|
Can not boot | low power | Please connect the USB charging cable to charge. |
Bluetooth can’t connect | The mobile device has not turned on the Bluetooth | |
function | Make sure that the Bluetooth on the mobile device is available. | |
ES-Patch has not been opened | Please press the power button of the ES-Patch |
and confirm that the power indicator is flashing normally.
low power| When the power indicator flashes 3 times quickly, it means that the
power is insufficient, please connect the USB charging cable to charge.
Mobile device not receiving signal| ES-Patch is not yet bound to mobile
devices.| Please download the official iOS system APP and perform the binding
operation.
Connection disconnected while monitoring| Out of valid connection range|
Please place your mobile device as close to the ES-Patch electronic
stethoscope as possible.
low power| Please connect the USB charging cable to charge.
NOTICE: If you have tried all the solutions to the problem and still cannot solve the problem, please call the original local service department for assistance.
Maintenance and Repair
For maintenance or repair service, or any questions or comments, please feel
free to contact the original customer service center. We will do our best to
assist you and solve your problems.
NOTICE: Do not disassemble this product without authorization. Only
authorized service personnel should repair this product; if the equipment is
modified by the user, the user is solely responsible.
Name of producer: Sound Land Corp.
Producer address: No. 32, Keji 1st Rd., Guishan Township, Taoyuan County,
33383 Taiwan R.O.C.
Name of manufacturer: Sound Land Corp.
Manufacturer address: No. 32, Keji 1st Rd., Guishan Township, Taoyuan
County, 33383 Taiwan R.O.C.
Email: service@soundland.com.tw
Tel: +886-3-396-1958
Fax: +886-3-396-0231
Website: https://soundland.com.tw/
Transport, storage and disposal
16.1 Transport and storage
General transportation of the equipment is subject to the conditions described
in the “10. Other operating precautions ” section of this manual.
This product needs to be sent to an authorized service center for inspection
and repair.
Storage environmental conditions must comply with the “10. Other operating
precautions ” section of this manual.
16.2 Disposal
Discarded electronic Stethoscope patches and charging bases must be disposed
of in accordance with local general electronic waste recycling methods.
Product Specifications
Product name: | Electronic Stethoscope Patch |
---|---|
Model number: | ES-Patch |
Compatible mobile devices: | Applicable to mobile devices with operating system |
iOS14.0 (inclusive) or above, with Bluetooth version 5.0 or above
Battery Specifications:| Li-ion battery DC3.7V / 110mah
Use current:| 12mA
Charging voltage and current:| 5V/100mA
Battery life cycle:| 500
Estimated battery power:| about 8 hours of use in a fully charged state
Estimated battery charging time:| about 4 hours
Power display:| Through LED or APP
Audio transmission:| Bluetooth, iOS mobile device
Frequency response mode:| Full Frequency / heart sound / breathing (lung)
sound Full Frequency (20~1200Hz)
Frequency Response Mode Frequency Range:| heart sound (20~200Hz)
lung sound (100~1000Hz)
Appearance dimensions:| length 56 X width 30 X thickness 12.7mm
Body weight:| 20g
Dust and Waterproof:| IP26
Safety specifications:| No fire, No combustion, In line with
telecommunications, EMC related regulations
Compliant with medical device standards:|
EN12470-4,IEC60601-1,IEC60601-1-2,IEC/EN62366-1,IEC62304,IEC60650
Compliant with telecom and EMC regulations:| Bluetooth5.1,FCC Part15,NCC
LP0002,Conform to CE-R&TTE
Operating ambient temperature and humidity:| 41° to 104°F (5° to 40°C), 15 to
95% relative humidity
Storage ambient temperature and humidity:| -4° to 140°F (-20° to 60°C), 15 to
95% relative humidiy Bluetooth (Transmitter):
Bluetooth v5.2 specification compliant. 2.402GHz ~ 2.480GHz unlicensed ISM
band. GFSK, pi/4-DQPSK, or 8-DPSK modulated digital signals in polar format
for 1 Mbps, 2 Mbps, and 3 Mbps transmission data rate, respectively.
Digital wirelesstransmission:| Bluetooth (Receiver):
Bluetooth v5.2 specification compliant.2.402GHz ~ 2.480GHz unlicensed ISM
band. GFSK, pi/4-DQPSK or 8-DPSK coded bit stream and provides RSSI
information.
Appendix: Guiding Principles and Manufacturer’s Declaration
Recommended separation distance between
portable and mobile RF communications equipment and the ES-Patch
The ES-Patch is intended for use in an electromagnetic environment (for
Professional healthcare environments) in which radiated RF disturbances are
controlled. The customer or the user of the ES-Patch can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the ES-
Patch as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output power of transmitter
W
|
Separation distance according to frequency of
transmitter
m
---|---
150 kHz to 80 MHz
d =1,2√P
| 80 MHz to 800 MHz
d =1,2√P|
800 MHz to 2,7 GHz
d =2,3√P
0,01| N/A| 0,12| 0,23
0,1| N/A| 0,38| 0,73
1| N/A| 1,2| 2,3
10| N/A| 3,8| 7,3
100| N/A| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Manufacturer's declaration-electromagnetic emissions
The ES-Patch is intended for use in the electromagnetic environment (for
Professional healthcare environments) specified below.
The customer or the user of the ES-Patch should assure that it is used in
such an environment.
Emission test| Compliance| Electromagnetic environment-
guidance
(for Professional healthcare environments)
RF emissions CISPR 11| Group 1| The ES-Patch uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11| Class A| The ES-Patch is suitable for use in all
establishments other than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2| Not applicable
Voltage fluctuations /flicker emissions
IEC 61000-3-3| Not applicable
Manufacturer's declaration-electromagnetic immunity
The ES-Patch is intended for use in the electromagnetic environment (for
Professional healthcare environments) specified below.
The customer or the user of the ES-Patch should assure that it is used in
such an environment.
Immunity test| IEC 60601
test level| Compliance level| Electromagnetic
environment-guidance
(for Professional healthcare environments )
Electrostatic discharge(ESD)
IEC 61000-4-2| Contact:±8 kV
Air±2 kV,±4 kV,±8 kV,±15 kV| Contact:±8 kV
Air±2 kV,±4 kV,±8 kV,±15 kV| Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the relative humidity should be
at least 30%
Electrical fast transient/burst
IEC 61000-4-4| + 2kV for power supply lines
+ 1kV for input/output lines| Not applicable Not applicable| Mains power
quality should be that of a typical Professional healthcare environments
Surge
IEC 61000-4-5| + 0.5kV, +1kV line(s) to line(s)
+ 0.5kV, +1kV,+ 2kV line(s) to earth| Not applicable Not applicable| Mains
power quality should be that of a typical Professional healthcare environments
Voltage Dips, short interruptions and voltage variations on power supply input
lines
IEC 61000-4-11| Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30 cycles
** Voltage interruptions:
0 % UT; 250/300 cycle| Voltage dips:
Not applicable Not applicable Not applicable
Voltage interruptions: Not applicable| Mains power quality should be that of a
typical Professional healthcare environments. If the user of the ES-Patch
requires continued operation during power mains interruptions, it is
recommended that the ES-Patch be powered from an uninterruptible power
supply or abattery.
Power frequency(50, 60 Hz) magnetic field
IEC 61000-4-8| 30 A/m
50 Hz or 60 Hz| 30 A/m
50 Hz and 60 Hz| The ES-Patch power frequency magnetic fields should be at
levels characteristic of a typical location in a typical Professional
healthcare environments.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Manufacturer's declaration-electromagnetic immunity**
The ES-Patch is intended for use in the electromagnetic environment (for
Professional healthcare environments) specified below.
The customer or the user of the ES-Patch should assure that it is used in such
and environment.
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment-guidance
(for Professional healthcare environments)
Conducted RF
IEC 61000-4-6Radiated RF
IEC 61000-4-3| 3 Vrms: 0,15 MHz – 80 MHz
6 Vrms: in ISM bands between
0,15 MHz and 80 MHz 80 % AM at 1 kHz e)3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz| Not applicable
Not applicable3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz| Portable and mobile RF communications
equipment should be used no closer to any part of the
ES-Patch including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.Recommended separation
distance:
d = 1,2√P
d = 1,2√P 80MHz to 800 MHz
d = 2,3√P 800MHz to 2,7 GHzWhere P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Manufacturer’s declaration-electromagnetic immunity
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment
The ES-Patch is intended for use in the electromagnetic environment (for
Professional healthcare environments) specified below.
The customer or the user of the ES-Patch should assure that it is used in
such an environment.
Test frequency (MHz)| Band a)
(MHz)| Service a)| Modulation b)| Maximum power (W)|
Distance (m)| IMMUNITY TEST LEVEL (V/m)| Compliance LEVEL
(V/m)
(for Professional healthcare environments)
385| 380 –390| TETRA 400| Pulse modulation b) 18 Hz| 1,8| 0,3| 27| 27
450| 430 – 470| GMRS 460, FRS 460| FM c)
±5 kHz deviation 1 kHz sine| 2| 0,3| 28| 28
710| 704 – 787| LTE Band 13, 17| Pulse modulation b) 217 Hz| 0,2| 0,3| 9| 9
745
780
810| 800 – 960| GSM 800/900, TETRA 800,iDEN 8 20, CDMA 850, LTE Band 5| Pulse
modulation b) 18 Hz| 2| 0,3| 28| 28
870
930
1720| ****
1700 – 1990
| GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3, 4, 25; UMTS| Pulse modulation b) 217 Hz| 2| 0,3| 28| 28
1845
1970
2450| 2400 – 2570| Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE
Band 7| Pulse modulation b) 217 Hz| 2| 0,3| 28| 28
5240| 5100 – 5800| WLAN 802.11 a/n| Pulse modulation b) 217 Hz| 0,2| 0,3| 9| 9
5500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance
between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be
reduced to 1 m.
The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be
used because while it does not represent actual modulation, it would be worst
case.
FCC and NCC regulations statement
FCC Part 15.19This device complies with Part 15 of the FCC Rules.
ES-Patch Operation is subject to the following two conditions:
- this device may not cause harmful interference.
- this device must accept any interference received, including interference that may cause undesired operation FCC Part 15.21 information for user You are cautioned that changes or modifications not expressly approved by the party responsible for compliance could void your authority to operate the equipment. FCC Section 15.105 Information to the user.
NOTE:
This equipment has been tested and found to comply with the limits for a class
B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
IMPORTANT NOTE: FCC Radiation Exposure Statement :This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment.
NCC Regulation Statement:
According to NCC LP0002 Technical Specification for Low Power Radio Frequency
Equipment Chapter 3.8.2:
For the low-power radio frequency equipment that has obtained the verification
certificate, without approval, the company, firm or user shall not change the
frequency, increase the power or change the features and functions of the
original design without authorization.
The use of low-power radio frequency equipment shall not affect flight safety
and interfere with legal communications; if any interference is found, it
should be stopped immediately, and it can be continued to be used until there
is no interference.
The aforesaid legal communication refers to the radio communication operated
in accordance with the provisions of the Telecommunications Management Act.
Low power radio frequency equipment must endure interference from legal
communication or radio wave radiating electrical equipment for industrial,
scientific and medical use.
NCC form certification number: