provita IV Stand Mobile Accessories
Product Information
- Specifications
- Product: IV Stand Mobile + Accessories
- Model Number: 42601239302J6_EN
- Date: 13.12.2022
- Manufacturer: Provita Medical GmbH & Co. KG
- Address: Auf der Huhfuhr 8, D-42929 Wermelskirchen
- Phone: +49 2193 5105-0
- Fax: +49 2193 5105-269
- Website: www.provita.de.
- Email: info@provita.de.
- General Information
- Notes and their meaning
- DANGER: This symbol in conjunction with the signal word Danger indicates an imminent danger. Failure to observe the safety instructions will result in death or serious injury.
- WARNING: The symbol in connection with the signal word Warning indicates a possibly dangerous situation. Failure to observe the safety instructions may result in death or serious injury.
- CAUTION: The symbol in connection with the signal word Caution indicates a possibly dangerous situation. Failure to observe the safety instructions may result in minor or slight injuries. It may also be used for property damage warnings.
- NOTICE: The symbol indicates a possibly harmful situation. Failure to observe the safety instructions may result in damage to or destruction of the product and/or other system parts.
- NOTE: Here you can find information or references.
- Information on the instructions for use
- The product is a technical aid that requires proper handling and regular care.
- Improper use can result in risks and accidents.
- These instructions for use are provided to help users familiarize themselves with the product and avoid accidents.
- Intended Use
- The IV Stand Mobile + Accessories are used for hanging infusion and nutrient solutions in bag or bottle form.
- The infusion delivery speed is influenced by instructions from the physician, who adjusts the relative hanging height according to medical considerations.
- Medical devices can also be attached to the stationary part of the IV stand, depending on its design.
- Liability and Warranty
- The statutory warranty applies to this product within the scope of Provita Medical GmbH & Co. KG’s general terms and conditions.
- All data and information provided are based on their previous experience and findings to the best of their knowledge.
- Technical changes may be made in line with further development.
- Identification Plate
- The operating instructions can be accessed in digital form by scanning the QR code on the type plate.
- Blind and visually impaired users can download the operating instructions in digital form from the Provita homepage by scanning the QR code on the type plate and have them read aloud with their software.
Safety Information
- Safety Precautions
- DANGER: The product may only be assembled, handled, and used by trained personnel. Careful observation of the safety information is required. Improper operation of the system may result in hazards to persons and objects.
- Assembly
- The assembly instructions for the IV Stand Mobile + Accessories are as follows:
- Ensure that you have all the necessary components and tools required for assembly.
- Identify the base of the IV stand and attach the wheels securely to the designated positions.
- Take the main stand tube and insert it into the base, ensuring a stable connection.
- Attach any additional accessories, such as hooks or trays, according to the provided instructions.
- Check that all connections are secure and tighten any screws or fasteners if necessary.
- Perform a stability check by gently pushing and pulling the IV stand to ensure it remains steady.
- Operation
- To operate the IV Stand Mobile + Accessories, follow these steps:
- Hanging Infusion or Nutrient Solutions
- Ensure that the IV bag or bottle containing the infusion or nutrient solution is properly sealed and intact.
- Identify the appropriate hook or attachment point on the IV stand where the bag or bottle will be hung.
- Carefully lift the bag or bottle and hang it securely on the hook or attachment point, ensuring it is stable.
- If necessary, adjust the relative hanging height according to the instructions from the physician.
- Attaching Medical Devices
- If your IV stand has a stationary part where medical devices can be attached, follow these steps:
- Identify the designated attachment point on the stationary part of the IV stand.
- Ensure that the medical device is compatible with the attachment point and can be securely fastened.
- Attach the medical device to the designated attachment point, following any provided instructions.
- Check that the attachment is secure and perform any necessary adjustments for proper positioning.
- FAQ
- Q: Can I use the IV Stand Mobile + Accessories without proper training?
- A: No, the product may only be assembled, handled, and used by trained personnel. Careful observation of the safety information is required to prevent hazards.
- Q: How can I access the operating instructions for the IV Stand Mobile + Accessories?
- A: The operating instructions can be accessed in digital form by scanning the QR code on the type plate. Blind and visually impaired users can download the operating instructions from the Provita homepage and have them read aloud with their software.
- Q: What is the intended use of the IV Stand Mobile + Accessories?
- A: The IV Stand Mobile + Accessories are used for hanging infusion and nutrient solutions in bag or bottle form. The hanging height can be adjusted according to medical considerations.
General information
Notes and their meaning
- DANGER
- This symbol in conjunction with the signal word “Danger” indicates an imminent danger.
- Failure to observe the safety instructions will result in death or serious injury.
- WARNING
- The symbol in connection with the signal word “Warning” indicates a possibly dangerous situation.
- Failure to observe the safety instructions may result in death or serious injury.
- CAUTION
- The symbol in connection with the signal word “Caution” indicates a possibly dangerous situation.
- Failure to observe the safety instructions may result in minor or slight injuries.
- It may also be used for property damage warnings.
- NOTICE
- The symbol indicates a possibly harmful situation.
- Failure to observe the safety instructions may result in damage to or destruction of the product and/or other system parts.
- NOTE
- Here you can find information or references.
Information on the instructions for use
- The product is a technical aid. It needs to be explained, requires regular care, and involve risks in the event of improper use.
- It is therefore necessary to learn proper handling.
- These instructions for use are intended to help you familiarize yourself with the handling of the product as well as avoid accidents.
Intended use
- The products from the IV stand product group are used for hanging infusion and nutrient solutions in bag or bottle form.
- The infusion delivery speed is influenced by instructions from the physician according to medical considerations by adjusting the relative hanging height accordingly.
- Depending on the design, it is possible to attach medical devices to the stationary part of the IV stand (on the outer tube).
Liability and warranty
- The statutory warranty applies for this product within the scope of our general terms and conditions.
- All data and information are supplied based on our previous experience and findings to the best of our knowledge.
- We reserve the right to make technical changes in line with further development.
NOTICE
- We accept no liability for any damage caused by failure to observe this user manual.
Identification plate
NOTE
- The operating instructions can be accessed in digital form by scanning the QR code on the type plate.
- Blind and visually impaired users can download the operating instructions in digital form from the Provita homepage by scanning the QR code on the type plate and have them read aloud with their software.
Safety information
DANGER
- The product may only be assembled, handled and used by trained personnel.
- This requires careful observation of the information for safety. If the system is operated improperly, this may result in hazards to persons and objects!
DANGER
- To avoid hazardous situations, it is necessary to practice handling the IV stand.
- Prior to using the IV stand, check that all attached parts are properly secured and not damaged.
- To move the IV stand, the locking casters must be released and the stand must be guided by hand! Particular caution is needed when moving the stand over obstacles, soft floor coverings, and inclines.
- The extension tube of the stand must be fully retracted.
- The tilt resistance (stability) is guaranteed up to a 5° incline. However, never park the IV stand on sloped surfaces!
- The information on the max. system load of the IV stand, the load per hook, and the max. load of the bottle holder can be found on the identification plate on the IV stand.
- It may be possible to secure additional loads, for example, devices, to the outer tube. The permissible system load can be found on the identification plate. (Please also observe the additional information from the device manufacturer)
- Handle the IV stand carefully when it is fully loaded since it may tip!
- Only use accessories approved by the manufacturer.
NOTICE
- The manufacturer accepts no liability for damage caused by insufficient maintenance or by failure to comply with the following instructions.
- Compliance with the user manual and the maintenance instructions is an essential part of the validity of the warranty.
Obligation to report serious incidents
- All serious incidents that have occurred in connection with the product must be reported to the manufacturer and the relevant competent authority per the reporting deadlines of the MDR.
Assembly
Operation
Single-handed height adjustment
- For adjustment of the inner tube the re-lease ring needs to be lifted with the thumb.
- When removing the thumb from the ring the inner tube is automatically fixed at the re-quested position.
CAUTION
- When the IV-pole is loaded, the inner tube lowers when the actuation ring is triggered.
Screw hight-adjustment
- Please do hold the upper tube 1 , when releasing the adjustment screw 2 for adjusting the tube.
- Position the tube into the required height and do tighten the adjustment screw.
- Do not release the upper tube until the clamping force is sufficient for the requested use.
- Please do not forget to include the additional weight of the infusion bottles.
WARNING
- Risk of eye injury. Keep a safe distance from the bottle hooks!
- Risk of crushing fingers or hands by operating the height adjustment!
- Risk of uncontrolled lowering of the load by actuating the height adjustment!
Cleaning and maintenance
ATTENTION
- Liquids containing chloride (e.g., blood, physiological saline solutions) must be removed to avoid rust formation. After removal, clean with water to completely remove residues.
ATTENTION
- Observe the application instructions of the detergent and disinfectant manufacturer, as well as the currently valid hygiene regulations.
ATTENTION
- Non-pH-neutral and chemically active disinfectants can lead to surface and material damage depending on the exposure time and “aggressiveness” (pH value, chemical reactivity, aggressiveness of solvents).
- Therefore, the following products/ingredients should not be used for cleaning and disinfection:
- solvents containing hydrocarbons (petrol, thinner). These can damage coatings and plastics.
- Agents with an alcohol content of more than 10 %. These can damage coatings and plastics.
- Strong acids or alkalis
- Halogenides or halogen-releasing compounds (chlorine, bromine, iodine). These can also cause corrosion on stainless steel.
- Products that scratch the surface (e.g., scouring agents, wire brushes, steel wool), water containing iron particles, cleaning sponges and liquids containing hydrochloric acid.
NOTICE
- Do not use any mechanical cleaning and disinfection procedures (washer system, pressure washer)
Maintenance and care
- The product must be checked annually by qualified staff regarding the correct function of the device:
- Further information regarding Maintenance history can be found on page 19.
- In case that a component of the device is damaged, it must be replaced only with original parts by a trained specialist. Contact the manufacturer or your local distributor for more information.
- Dispose the device according to the local and national regulations.
- Apart from the transport and stock-keeping conditions below, no special measures are required for storing the product.
- Before returning to service after an extended period of non-use, the product should be inspected according to the maintenance plan/checklist.
Environmental conditions
For transport and stock-keeping the following stock-keeping conditions have to beheeded:
- Temperature: – 25 bis 70 °C
- Humidity: 30 – 75%
- Atmospheric pressure: 700 – 1060 hPa
At application you have to assure the ambient temperature from 10 °C to 40 °C.
Maintenance history
The operator is obligated to regularly examine the system. If one of the
points is assessed negatively, the affected component must be immediately
taken out of service.
Please confirm with your signature the inspection performed (with positive
result) of the following points:
Intactness
The IV stand should be checked at least once annually with regard to the
full functioning of all components.| Functio
The function of the bottle hooks. The function of the height adjustment.
Clean the rollers to remove any soiling so that the brakes function and the
electrical conductivity is not reduced.| Completeness
If a defect is identified, only original spare parts may be used and they
may be replaced only by a technical specialist or customer service.
---|---|---
Date| Signature| Date| Signature| Date| Signature
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
DANGER
- If there is a problem regarding one or more points, safe operation is no longer guaranteed and suitable measures must be taken to avoid errors.
- The system may NOT be used if proper function is not ensured.
Service life
- The product is expected to have a service life of 7 years, assuming that it is only used for its intended purpose and is subject to regular and correct maintenance.
- The service life largely depends on the frequency of use, the operational environment, and upkeep.
- The service life of the product can be extended by using original spare parts. For more information, contact the manufacturer or the distributor.
- The service life indicated does not constitute an additional warranty.
Service
- Please contact the manufacturer or the local distributor if you have any questions about the product, if a repair is due, or if you require information regarding service partners.
- Address data is provided at the end of this manual.
Technical data
Type plate | on the product |
---|---|
Service life | 7 years |
CONTACT
- provita medical gmbh & co. kg
- Auf der Huhfuhr 8
- 42929 Wermelskirchen
- Germany
- Phone +49 2193 5105-0
- Fax +49 2193 5105-269
- www.provita.de.
- info@provita.de.
- 42601239302J6_EN
- 13.12.2022
- V002
- provita
- provita medical gmbh & co. kg
- Auf der Huhfuhr 8
- D-42929 Wermelskirchen
- Phone +49 2193 5105-0
- Fax +49 2193 5105-269
- www.provita.de.
- info@provita.de.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>