SERVOspin H7 800 PLUS Low Speed Benchtop Centrifuge User Manual
- June 15, 2024
- SERVOspin
Table of Contents
SERVOspin H7 800 PLUS Low Speed Benchtop Centrifuge
FOREWORD
Thank you for purchasing our SERVOspin PLUS centrifuge, This Manual for users contains function and operation of the Instrument. In order to use the instrument properly, please read this manual carefully before using the instrument.
OPENING CHECK
Please check the Instrument and appendix with the packing list when you first open the instrument packing case. If you find there is something wrong with the instrument and the appendix, do contact the vendor or the producer.
INTRODUCTION
This centrifuge is suitable for 15 ml, 7,5 ml and 5 ml centrifuge, Blood
Collection Tubes and 100 mm tube for centrifuging.
Before starting up this centrifuge for the first time, please read the rest of
this operations manual.
DELIVERY PACKAGE
SERVOspin PLUS centrifuge | 1 unit |
---|---|
Rotor | 1 unit |
Centrifuge tube adapter | 1 unit |
Operations manual | 1 unit |
STRUCTURE DESCRIPTION
- Cover lock
- Mater rotor
- Safety switch
- Emergency door handle
- Time setting ( )
- Speed setting ( )
- Lid open button
- Short-Spin
- Start/Stop
- Power switch
- Power connector
INSTALLING THE DEVICE
Place the centrifuge onto a level , horizontal surface. Make sure that the
ventilation slits are not blocked.
Insert the column connector of the adapter to power connector of the device
,and insert another connector of the adapter to mains power supply.
A safety distance of 30 cm should be observed around the centrifuge during
operation.
No objects which could cause additional damage in the event of a centrifuge
crash should be positioned in this space.
Power on the main switch. The centrifuge is ready to operate when the display
becomes visible. Place the rotor onto the rotor axle and tighten using the
rotor nut.
Before starting up the centrifuge for the first time, make sure the rotor nut
is securely fastened.
TECHNICAL DATA
Model | SERVOspin PLUS |
---|---|
Power supply | 100 – 240 V ~ 50 – 60 Hz |
Power | 40 Watt |
Max. speed | 5000 U/min |
Max. centrifugal force | 3074 rcf |
rpm/rcf | Ja |
Time | 10 sec. ~ 99 min. |
Max. load | 8 x 15 ml |
Max. permitted density | 1,2 g/ml |
Acceleration time to max. speed | 20 – 80 seconds |
Braking time to max. speed | 20 – 80 seconds |
Ambient temperature | 5 °C – 35 °C |
Abmessungen (B xT x H) | 35 x 40 x 32 cm |
Gewicht mit Rotor | 5 kg |
SAFETY PRECAUTIONS
The rotor and the rotor lid must always be securely fastened. If the
centrifuge makes unusual noises when started , the rotor or rotor lid is not
fastened correctly.
Switch the device off immediately by pressing “START/STOP” key.
Do not use damaged rotors!
Do not move the centrifuge during the run!
PLACE THE TUBES
A liquid density of 1.2 g/ml must not be exceeded at the max. speed.
Damaged tubes can not be centrifuged. This is because broken tubes can, in
addition to sample loss, result in further damage to the centrifuge.
Close the test tube lids before centrifuging. Open lids can be ripped off
during centrifugation and damage the centrifuge. Must load the test tubes
symmetrically.
MAINTAIN OF ROTOR
Even slight scratches and tears of the rotor can lead to serious internal
material damage.
Avoid damage caused by aggressive chemicals, including among others: strong
and weak alkali, strong acids.
In the event of contamination caused by aggressive agents, The rotor must be
cleaned immediately using a neutral cleaning liquid.
OPERATION GUIDE
CONTROL ELEMENTS
- START/STOP for start or stop centrifugation
- SHORT-SPIN for reduced-time centrifugation
- OPEN to open the lid
- for setting the time and speed
ROTOR INSERTION AND REMOVAL
Fit the rotor onto the motor shaft, followed by the rotor nut. Tighten the rotor nut by turning clockwise. To release the rotor, turn the rotor nut counterclockwise. Before each start, check that the rotor is firmly tightened!
LOADING THE ROTOR
Rotor must always be loaded symmetrically. Minimize differences in weight between the filled sample tubes-taring with a scale is recommended. This will reduce wear on the drive and cut running noise.
CENTRIFUGATION WITH A PRESET TIME
Turn on the mains switch, pressing “OPEN” to open the lid, load the rotor
symmetrically.
Fasten the rotor lid and close the centrifuge lid.
Pressing Time’s change the run time, the time can be pre-selected between
20 seconds and 99 minutes, and pressing speed’s to change the run speed,
the max. speed is 5000 rpm.
Pressing the first “START/STOP” to start the run. Pressing the second
“START/STOP” to end the run prematurely. The remaining run time appears in the
display.
After the run, the lid lock of centrifuge opens automatically.
The time setting and the rotational speed may be changed during the run. The
remaining run time appears in the display.
SHORT-SPIN CENTRIFUGATION
Turn on the mains switch, pressing “OPEN” to open the lid, load the rotor
symmetrically. Fasten the rotor lid and close the centrifuge lid.
Pressing “SHORT-SPIN”. Short-spin centrifugation is possible for as long as
this key is held down. The maximum rpm is 5000.
SWITCHING TO THE RCF DISPLAY
Pressing speed’s simultaneously the display switches from rpm to rcf. Also
pressing speed’s simultaneously again, the display switches from rcf to
rpm.
You can calculate with the following formula:
rcf = 1118 10-5 n2 * rmax
n : rotational speed in 1/min
rmax = 6 cm, max. centrifuging radius in cm
OPEN THE CENTRIFUGE IN THE EVENT OF A POWER FAILURE
Disconnect the centrifuge from the mains supply. Wait until the rotor has come to a standstill, Then left up the device, remove the lock pole to right , the lid lock will be opened.
MAINTENANCE AND CLEANING
The rotor and the outside of centrifuge should be cleaned regularly with a moist cloth. Disconnect the centrifuge from the mains supply, remove the rotor and clean it separately. Only neutral agents may be used for cleaning.
Please check the rotor and especially the rotor bores regularly for deposits or damage, then reinsert the rotor and tighten the rotor nut.
TROUBLESHOOTING
Error | Cause | Solution |
---|---|---|
No display | No main power connection, Power failure | Plug in mains cable both |
sides, check mains fuse
Lid can´t be opened| Power failure
Rotor is still spinning. Lid lock failure| Emergency lid release (4.7) Stop
centrifuge, Contact service
Centrifuge shaked during acceleration| Rotor not loaded symmetrically| Stop
centrifuge and load centrifuge symmetrically
Er=01| Open lock failure| Contact service
Er=02| Close lock failure| Contact service
Er=08| Motor failure| Contact service
Customer Support
servoprax GmbH
Am Marienbusch 9 · D-46485 Wesel/Germany
Tel. +49 281 95283 0· Fax +49 281 56071
info@servoprax.de ·
www.servoprax.de
Edited at 07.12.2015