microlife FR1DQ1-B Infrared Forehead Thermometer User Manual
- June 4, 2024
- microlife
Table of Contents
FR1DQ1-B Infrared Forehead Thermometer
FR1DQ1-B
Infrared Forehead Thermometer
Distributed by: Microlife USA, Inc. 1617 Gulf to Bay Blvd.,Floor 2
Clearwater,FL 33755UNITED STATES Phone: (727)-442-5353 www.microlifeusa.com
Made in China
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
1 Measuring sensor 2 Tracking light 3 Display 4 START/IO button 5 MEMO button
6 MODE 7 Battery compartment cover 8 All segments displayed 9 Ready for
measuring
Start the measurement Measurement complete
12 Body mode 13 Object mode 14 Low battery indicator 15 Changing between
Celsius and Fahrenheit 16 Recall mode 17 Recall the last 30 readings
Measured temperature too high Measured temperature too low
20 Ambient temperature too high 21 Ambient temperature too low 22 Error
function display 23 Blank display 24 Flat battery 25 Date/Time 26 Beeper
function setting 27 Protective cap
Read the instructions carefully before using this device
Type BF applied part
Overview This Microlife thermometer is a high-quality product incorporating
the latest technology and tested in accordance with international standards.
With its unique technology, this device can provide a stable, heat-
interference-free reading with each measurement. The device performs a self-
test every time it is switched on to always guarantee the specified accuracy
of any measurement. This Microlife thermometer is intended for measurement and
monitoring not only the temperature of forehead of human body, but also the
temperature of object. This thermometer has been clinically tested and proven
to be safe and accurate when used in accordance to the operating instruction
manual. Please read through these instructions carefully in order for you to
understand all functions and safety information.
Indication For Use The Microlife Infrared Forehead Thermometer Model FR1DQ1-B
is intended for the intermittent measurement and monitoring of human body
temperature. The device is indicated for use by people of all ages in the
home. The device can be used in connection with a smart phone running the
«Microlife Connected Health +» APP. The memory data can be transferred to the
smart phone via Bluetooth.
Table of Contents 1. The Advantages of this Thermometer · Measures in a matter
of seconds · Measurement with distance control · Multiple uses (wide range of
measurement) · Accurate and reliable · Gentle and easy to use · Multiple
readings recall · Safe and hygienic · Fever alarm · Guidance system for self-
measurement · Bluetooth function 2. Important Safety Instructions 3. How this
Thermometer measures Temperature 4. Control Displays and Symbols 5. Setting
Date, Time and Beeper Functions 6. Changing measurement mode 7. Directions for
Use · Measuring in body mode with distance control
· Measuring in object mode without distance control
8. Changing between Celsius and Fahrenheit 9. How to recall & delete 30
readings in Memory Mode 10. Bluetooth® Function – App Quick start guide –
Downloading the «Microlife Connected Health+» App – Setup User Profile –
Transfer Data – Bluetooth® Icon Indicator
– Notes – Wireless communication interference 11. Error Messages 12. Cleaning
and Disinfecting 13. Battery Replacement 14. Guarantee 15. Technical
Specifications 16. www.microlife.com
1. The Advantages of this Thermometer Measures in a matter of seconds The
innovative infrared technology allows the measurement without even touching
the object. This guarantees safe and sanitary measurements within seconds.
Measurement with distance control The device will show good on display when
the device detects the distance is appropriate within 5 cm. Multiple uses
(wide range of measurement) This thermometer offers a wide range of
measurement from 0.1 99.9 °C / 32.2 – 211.8 °F, meaning the unit can be used
to measure body temperature or it also has a feature allowing it to be used to
measure surface temperature of the following examples: · Milk surface
temperature in a baby’s bottle · Surface temperature of a baby’s bath Accurate
and reliable The unique probe assembly construction incorporates an advanced
infrared sensor, ensuring that each measurement is accurate and reliable.
Gentle and easy to use · The ergonomic design enables simple and easy use of
the thermometer. · This thermometer can even be used on a sleeping child
without causing any interruption. · This thermometer is quick, therefore child
friendly. Multiple readings recall Users will be able to recall the last 30
readings with a record of both time and date when entering the recall mode,
enabling efficient tracking of temperature variations. Safe and hygienic · No
direct skin contact · No risk of broken glass or mercury ingestion. · Children
should be used under the care of adults to avoid danger caused by accidental
ingestion Fever alarm 10 short beeps with a red LCD backlight for 5secs and
time and date switch every 1 sec alert the patient that he/ she may have a
temperature equal to or higher than 37.5 °C.
Guidance system for self-measurement A blue tracking light shows the user that
the device is at the right distance and ready to take a measurement. Bluetooth
function
This device connects to the «Microlife Connected Health+» App by using
Bluetooth®4.2 and enables monitoring of
temperature
2. Important Safety Instructions · Follow instructions for use. This document
provides important product operation and safety information regarding
this device. Please read this document thoroughly before using the device and
keep for future reference. · This device may only be used for the purposes
described in these instructions. The manufacturer cannot be held liable
for damage caused by incorrect application. · Never immerse this device in
water or other liquids. For cleaning, please follow the instructions in the
«Cleaning
and Disinfecting» section. · Do not use this device if you think it is damaged
or notice anything unusual. · Never open this device. · A basic physiological
effect called vasoconstriction can occur in the early stages of fever,
resulting in a cool skin
effect. The recorded temperature using this thermometer can, therefore, be
unusually low. · Please note that before measurement, the thermometer and the
person whose temperature is to be measured
should be in the room temperature for at least 30 minutes · If the measurement
result is not consistent with the patient’s finding or unusually low, repeat
the measurement
every 15minutes or double check the result by another core body temperature
measurement. · This device comprises sensitive components and must be treated
with caution. Observe the storage and operating
conditions described in the «Technical Specifications»section. · Ensure that
children do not use this device unsupervised; some parts are small enough to
be swallowed. · Do not use this device close to strong electromagnetic fields
such as mobile telephones or radio installations. Keep a
minimum distance of 3.3 m from such devices when using this device. · Data
Transmission: This product emits radio frequencies (RF) in the 2.4 GHz band.
DO NOT use this product in
locations where RF is restricted, such as on an aircraft or in hospitals. Turn
off this thermometer; remove batteries when in RF restricted areas. For
further information on potential restrictions refer to documentation on the
Bluetooth usage by the FCC. · Protect it from: – Extreme temperatures – Impact
and dropping – Contamination and dust – Direct sunlight – Heat and cold
If the device is not going to be used for a prolonged period, the batteries
should be removed.
® WARNING: The measurement results given by this device is not a diagnosis. It
is not replacing the need for the
consultation of a physician, especially if not matching the patient’s
symptoms. Do not rely on the measurement result only, always consider other
potentially occurring symptoms and the patient’s feedback. Calling a doctor or
an ambulance is advised if needed.
3. How this Thermometer measures Temperature This thermometer measures
infrared energy radiated from the forehead as well as objects. This energy is
collected through the lens and converted to a temperature value.
4. Control Displays and Symbols
All segments displayed8 : Press the START/IO button 4 to turn on the unit; all
segments will be shown for 1 second,
default body mode.
Ready for measuring 9 : after all segments have been show for 1 second, the
unit is ready for measuring, the «°C» or
«°F» icon will keep flashing while the mode (3s body / 1s body / object), time
and beep icon will be displayed. Taking a measurement: for 3s body mode ,when
the device detect the distance is appropriate (within 5cm),the
screen shows “good” 10 with 1 short beep ,then press START/IO button 4 to
start measuring. For 1s mode, when the device detects the distance is
appropriate (within 5cm), the screen shows “good” 10 with 1short beep, then
press START/IO button 4 to start the measurement. Measurement complete 11 :
The reading will be shown on the display 3 with the «°C» or «°F» icon and the
mode
icon steady. The unit is ready for the next measurement as soon as the «°C» or
«°F» icon is flashing again.
Low battery indicator 14 : When the unit is turned on, the «battery» icon will
keep flashing to remind the user to
replace the batteries.
5. Setting Date, Time, and Beeper Functions
Setting the date and time25
After the new batteries are fitted, device will enter time setting
automatically. Under any status, hold “MODE” button for 8 seconds, will enter
time setting mode. Under time setting mode, first will be year setting, then
Date setting, following by time setting, (then jump back to year setting again
as cycle, until “Start/IO” button be pressed to enter stand-by mode”) with
press MODE button.
1. The year number flashes in the display (1 time/per sec) to indicate year
setting25 . Press “MEMO” 5 button to
select year (21~40), then Press “MODE” 6 button to confirm the years’ setting,
then move to month setting. 2. Press “MEMO” 5 button to select month (1~12),
then press “MODE” 6 button to confirm the month setting. then
move to day setting
3. Press “MEMO” 5 button to select day (1~31), Press “MODE” 6 button to
confirm the day setting. then move to
next setting 4. Follow the previously mentioned instructions to set the day,
hours, and minutes.
5. Once you have set the minutes and pressed the START/IO button4 , the date
and time are set and the time is
displayed.
If no button is pressed for 20 seconds, the device will go back to stand-by
for measurement with no date/time
information display and then switches to ready for measuring.
Press “Start/IO” button at any time-setting status, will allow to enter
measuring stand-by mode directly. (If there is any
item of time-setting have not been completed, they will be show with no-
information–)
Setting the beeper26
At any status, press “MODE” 6 button for 3 seconds, will enter beep on/off
setting
1. In “beeper on/off” setting mode press and hold the “MODE” button6 for
3seconds to set the beeper26 . 2. Press the MEMO5 to either turn the beeper on
or off. The beeper is activated when the beeper icon26 is
shown without across. When the beeper setting has been chosen, press the
START/IO button4 to enter the
stand-by measurement mode; otherwise, the device automatically switches to
ready for measuring after 10 seconds.
6. Changing measurement mode
1. Press the START/IO button4 . The display3 is activated to show all
segments for 1second. 2. The default mode is 3S body mode. Press the MODE
button 6 to switch between 3S, 1S body mode 12 and
object mode13 .
7. Directions for Use
Always remove the protective cap27 before use.
Measuring in 3S and 1S body mode with distance control
1. Press the START/IO button4 . The display3 is activated to show all
segments for 1second. 2. A flashing «°C»/«°F» icon, the blinking blue tracking
light 2 and a beep indicate that the device is ready to take
a measurement 10 .
3. Press the MODE button6 to switch between 3S and 1S body mode.
4. Remove any hair, sweat or dirt from the forehead before measuring to
ensure the accuracy of the readings. 5. 3S body mode: Aiming the thermometer
at the center of the forehead with a distance no more than 5 cm. Once
the display shows “good”, when the measuring sensor1 detects the distance is
appropriate within 5 cm. Then
press the START/IO button and slowly scanning from forehead to temple, a long
beep will verify the completion of measurement. 6. 1S body mode: Aiming the
thermometer at the center of the forehead with a distance no more than 5 cm.
Once
the display shows “good”, when the measuring sensor1 detects the distance is
appropriate within 5 cm. Then
press the START/IO button, a long beep will verify the completion of
measurement. 7. Read the recorded temperature from the LCD display. 8. For the
next measurement remove the thermometer from the forehead and wait until the
«°C»/«°F» icon is flashing.
Follow steps 4-5 above.
9. Press and hold the START/IO button 4 for 3seconds to enter stand by
ambient mode or 6 seconds to turn off the
device totally; otherwise, the device will automatically switch off after
approx. 30seconds
Measuring in object mode without auto measurement
1. Press the START/IO button4 . The display3 is activated to show all
segments for 1second.
2. Press the MODE button6 to switch to object mode. 3. A flashing «°C»/«°F»
icon, the blinking blue tracking light2 and a beep indicate that the device is
ready for
measurement9 .
4. Aim the thermometer at the center of the object you want to measure with a
distance no more than 5 cm. Press
the START/IO button4 . After 1seconds a long beep will verify the completion
of measurement.
5. Read the recorded temperature from the LCD display. 6. For the next
measurement wait until the «°C»/«°F» icon is flashing and follow steps 4-5
above.
7. Press and hold the START/IO button4 for 3seconds to turn off the device
totally, otherwise the device will
automatically switch off after approx. 30seconds. NOTE: Patients and
thermometer should stay in similar room condition for at least 30 minutes. ·
Don’t take a measurement while or immediately after nursing a baby. · Don’t
use the thermometer in high humidity environments. · Patients should not
drink, eat or exercise before/while taking the measurement. · Don’t move the
measurement device from the measuring area before hearing the termination
beep. · 10 short beeps and a red LCD backlight alert the patient that he/ she
may have a temperature equal to or higher than 37.5°C.
· Always take the temperature in the same location since temperature readings
may vary according to locations.
· Doctors recommend rectal measurement for newborn infants with in the first 6
months, as all other measuring methods might lead to ambiguous results. If
using a Infrared Forehead Thermometer on those infants, we always recommend
verifying the readings with a rectal measurement.
· In the following situations it is recommended that three temperatures are
taken with the highest one taken as the reading:
1. Children under three years of age with a compromised immune system and for
whom the presence or absence of fever is critical.
2. When the user is learning how to use the thermometer for the first time
until he/she has familiarized himself/herself with the device and obtains
consistent readings.
3. If the measurement is surprisingly low. · Readings from different
measuring sites should not be compared as the normal body temperature varies
by
measuring site and time of day, being highest in the evening and lowest about
one hour before waking up. Normal body temperature ranges: – Axillar: 34.7 –
37.3 °C / 94.5 – 99.1 °F – Oral: 35.5 – 37.5 °C / 95.9 – 99.5 °F – Rectal:
36.6 – 38.0 °C / 97.9 – 100.4 °F – Microlife FR1DQ1: 35.4 – 37.4 °C / 95.7 –
99.3 °F
8. Changing between Celsius and Fahrenheit This thermometer can display
temperature readings in either Fahrenheit or Celsius. To switch the display
between
°C and °F, at any status , press and hold the MODE button 6 for 3 seconds will
enter °C/°F setting , In “°C/°F” setting
mode the °C/°F icon is shown on the display, change the measurement scale
between °C and °F by pressing the
MEMO button5 . When the measurement scale has been chosen, press the START/IO
button 4 to enter the
«ready for measuring» mode otherwise, the device automatically switches to
ready for measuring after 10 seconds.
9. How to recall readings in Memory Mode
This thermometer can recall the last 30 readings with a record of both time
and date.
· Recall mode16: Press the MEMO button5 to enter recall mode when the power is
off or in measuring mode.
The memory icon «M» will display.
· Reading the last reading 17 : Press and release the MEMO button 5 to recall
the last reading. Number «1» and «M» are
displayed.
· Pressing and releasing the MEMO button 5 after the last 30 readings have
been recalled will resume the above sequence from
reading. After memory recall 1. Press “Start/ IO” button to go back to ready
for measuring 2. Press “Start/ IO ” button for 3 seconds to switch off. 3. If
no button be press for 30 seconds, the device will auto off.
10. BluetoothR Function
This device can be used in conjunction with a smart phone running the
«Microlife Connected Health+» App. The measurement results will be
automatically transferred via Bluetooth®. App Quick start guide Please refer
to APP quick start guide for more details Downloading the «Microlife Connected
Health+» App Download «Microlife Connected Health+» App for free from Google
PlayTM (Android) or App Store (iOS) and install it on your smartphone. Setup
User Profile 1. Open the App and select “Start” to begin using the App. 2.
Select “Settings” on the navigation menu (bottom) to enter “My profile”. 3.
Edit profile information, then select “Save” to complete Profile setup.
Transfer Data 1. Turn on Bluetooth of Smartphone. 2. Open the app on the
Smartphone, and select the Temperature Overview interface. 3. Make sure to
FR1DQ1-B is within 5 meters of your Smartphone, and urn off other Bluetooth
device if possible in case it interfere the data transfer. 4. Turn on FR1DQ1-B
in Body Temperature mode. Bluetooth of FR1DQ1-B is activated (when not taking
measurement) and automatically searches for Smartphone to connect to. 5. When
FR1DQ1-B is connected to the App on a Smartphone, FR1DQ1-B will transfer data
automatically if new data (hasn’t been transferred previously) is available,
while transferring. The LCD of FR1DQ1-B will show new data with “-” and 3
digits thermometer designated ID for 3 seconds, no new measurement can be
start within this 3 seconds. 6. Upon data transfer, the App will display a
message with “-” and 3 digits thermometer ID (MAC address) to confirm which
Patient Log in the App the data should be assigned to. Be sure to check”-“and
3 digits thermometer ID is identical on FR1DQ1-B and APP then select the
Patient Log and Confirm to complete the data transfer.
If”-“and 3 digits thermometer is not identical of both App and FR1DQ1-B,
please discard the message show on the app to stop the data transfer, and
record the temperature data into the App manually. Or take a new measurement
and start Transfer Data from step 1 to 6 again.
If no message showed on App to confirm which patient log the data should be
assigned to, please record the temperature data showed on FR1DQ1-B into the
App manually or takes a new measurement and start Transfer Data step from 1 to
6 again. Bluetooth® Icon Indicator The Bluetooth® icon indicator on your
device, located in the middle right portion of the display, is designed to
provide information about the connection between your device and smartphone.
The Bluetooth® function is enabled: The Bluetooth® icon indicator14 appears
steady on the display. Connecting with a device or uploading the data: The
Bluetooth® icon indicator14 flashes.
The Bluetooth® function is automatically switched off during measurement;
after measurement the device will switch it on and upload the data to the
«Microlife Connected Health+» App. NOTES The Bluetooth® of this thermometer is
a medical device data system (MDDS) as its only function is for records and no
additional functions. It does not modify the data or modify the display of the
data, and it does not by itself control the functions or parameters of any
other medical device. It is not active when the thermometer is recording data
or during measurement. The thermometer will not sound any alarm with or
without Bluetooth®. The Bluetooth® is used only to transfer data from point A
to point B. The App on your smart devices cannot be used to start or stop the
thermometer, nor update the firmware of thermometer via Bluetooth®. Wireless
communication interference This product operates in an unlicensed ISM band at
2.4 GHz. In the event this product is used near other wireless devices such as
microwave and wireless LAN, which operate on the same frequency band as this
product, there is a possibility that interference may occur. If interference
occurs, stop the operation of the other devices or relocate this product away
from other wireless devices before attempting to use it.
11. Error Messages
Measured temperature too high 18 : Displays «ErrH» when measured temperature
is higher than 43°C/109.4°F in body mode or
Displays «H» when measured temperature is higher than 99.9°C /211.9°F in
object mode.
Measured temperature too low 19 : Displays «ErrL» when measured temperature is
lower than 34.0 °C / 93.2 °F in body mode
or Displays «L» when measured temperature is lower than 0.1 °C / 32.2 °F in
object mode.
Ambient temperature too high 20 : Displays «AH» when ambient temperature is
higher than 40.0 °C / 104.0°F in body model
and object model.
Ambient temperature too low 21 : Displays «AL» when ambient temperature is
lower than 15.0 °C / 59.0 °F in body mode and
5.0 °C / 41.0 °F in object mode.
Error function display 22 :
«Err»: The system has a malfunction. «Errd»: Device is over 5cm from forehead
/object. «Er2»: Keep the measuring distance of 1-5 cm. Do not touch the bottom
side (sensing area) of the measuring sensor.
Blank display 23 : Check if the batteries have been inserted correctly. Also
check polarity (<+> and <->) of the
batteries.
Flat battery indicator 24 : If only «battery» icon is shown on the display,
the batteries should be replaced
immediately.
12. Cleaning and Disinfecting Use an alcohol swab or cotton tissue moistened
with alcohol (70% Isopropyl) to clean the thermometer casing and the measuring
sensor. Ensure that no liquid enters the interior of the device. Never use
abrasive cleaning agents, thinners or benzene for cleaning and never immerse
the device in water or other cleaning liquids. Take care not to scratch the
surface of the sensor lens and the display.
13. Battery Replacement This device is supplied with 2 new, long-life 1.5V,
size AAA batteries. Batteries need replacing when this icon
«battery» 14 is the only symbol shown on the display.
Remove the battery cover7 by sliding it in the direction shown. Replace the
batteries ensure correct polarity as shown
by the symbols in the compartment. Batteries and electronic devices must be
disposed of in accordance with the locally applicable regulations, not with
domestic waste.
14. Guarantee This device is covered by 5 years guarantee from the date of purchase. The guarantee is valid only on presentation of the guaranteed card completed by the dealer (see back) confirming date of purchase or the receipt. The guarantee covers the device. Batteries and packaging are not included. Opening or altering the device invalidates the guarantee. The guarantee does not cover damage caused by improper handling, discharged batteries, accidents or non-compliance with the operating instructions. Please contact Microlife- service.
15. Technical Specifications
Type: Measurement Range:
Infrared Forehead Thermometer FR1DQ1-B Body mode: 34.0 °C -43.0 °C / 93.2 °F -109.4 °F,
Object mode: 0.1-99.9 °C / 32.2 °F-211.8.0 °F
Resolution: 0.1 °C / °F
Measurement accuracy (Laboratory): Body mode: ± 0.2°C (35.0°C~42.0°C) / ± 0.4°F (95.0°F~107.6 °F)
± 0.3°C (34.0°C~34.9 °C and 42.1°C~43.0°C) / ± 0.5°F(93.2°F~94.8 °F and 107.8°F~109.4°F)
Object mode: ± 1.0°C (0.1°C~99.9°C) / ± 2 °F (32.2°F~211.8 °F)
Display: Liquid Crystal Display, 4 digits plus special icons
Acoustic: The unit is turned on and ready for the measurement: 1 short beep.
Complete the measurement: 1 long beep (1 sec.) if the reading is less than
37.5 °C / 99.5 °F, 10 short «beep» sounds, if the reading is equal to or
greater than 37.5 °C / 99.5 °F. System error or malfunction: 3 short «bi»
sounds.
Memory: 30 readings recall in the memory mode with a record of both time and
date. Backlight: The display light will be GREEN for 5 seconds, when a
measurement is completed with reading less
than 37.5°C/99.5°F but equal to or higher than 34.0°C/93.2°F. The display
light will be RED for 5
seconds, when a measurement is completed with a reading equal to or higher
than37.5°C/99.5°F but
equal or lower 43.0°C/109.4°F .
Operating conditions: Body mode: 15-40.0 °C / 59-104.0 °F, Object mode: 5-40.0
°C / 41-104.0 °F 15-95% relative maximum humidity
Storing conditions: -25 – +55 °C / -13 – +131 °F / 15-95% relative maximum
Automatic Switch-off: Approx. 30 sec when power on without measurement & after
measurement Battery: 2 x 1.5 V alkaline batteries; size AAA Battery Life:
Approx. 2000 measurements (using new batteries) Dimensions: 140 x 42.2 x 50 mm
Weight: 95.8 g (with batteries), 73.5 g (w/o batteries) IP Class: IP22
Reference to standard: ASTM E1965; IEC 60601-1; IEC 60601-1-2 (EMC); IEC
60601-1-11 Expected service life: 5 years or 12000 measurements This device
complies with the requirements of the Medical Device Directive 93/42/EEC.
Technical alterations reserved. According to the Medical Product User Act a
biennial technical inspection is recommended for professional users. Please
observe the applicable disposal regulations.
16. www.microlife.com Detailed user information about our thermometers and
blood pressure monitors as well as services can be found at www.microlife.com
FCC This device complies with part 15 of the FCC Rules. Operation is subject
to the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation. Changes or
modifications to the product are not approved by Microlife USA and could void
the user’s authority to operate the equipment under FCC jurisdiction. This
equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses, and can radiate
radio frequency energy, and if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures: reorient or relocate the receiving antenna,
increase the separation between the equipment and receiver, connect the
equipment into an outlet on a circuit different from that to which the
receiver is connected, consult the dealer or an experienced radio/TV
technician for help. This device complies with FCC radiation exposure limits
set forth for an uncontrolled environment. End users must follow the specific
operating instructions for satisfying RF exposure compliance.
Trademark Usage: Apple, the Apple logo, iPad, and iPhone are trademarks of
Apple Inc., registered in the U.S. and other countries. App Store is a service
mark of Apple Inc. Android and Google Play are both trademarks of Google Inc.
The Bluetooth® word mark and logos are registered trademarks owned by
Bluetooth SIG, Inc. and any use of such marks in this thermometer is under
license. Other trademarks and trade names are those of their respective
owners.
Appendix Guidance and manufacturer’s declaration electromagnetic emissions
The model FR1DQ1-B is intended for use in the electromagnetic environment specified below. The customer or the user of the model FR1DQ1-B should assure that it is used in such an environment.
Emissions
RF emissions CISPR 11
Compliance
Electromagnetic environment– guidance
Group 1
The model FR1DQ1-B uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions IEC 61000-3-3
Class B N/A N/A
The model FR1DQ1-B is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power
Guidance and manufacturer’s declaration electromagnetic immunity
The model FR1DQ1-B is intended for use in the electromagnetic environment
specified below. The customer or the user of the model FR1DQ1-B should assure
that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment –guidance
Electrostatic ±8 kV contact
discharge (ESD) IEC ±15 kV air
61000-4-2
±8 kV contact ±15 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
±2 kV for power
Electrical fast
supply lines and
transient/burst IEC
N/A
patient coupled
61000-4-4
lines
Surge IEC 61000-4-5
±1 kV line(s) and N/A
neutral
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
Voltage dips, short 40 % UT
interruptions and (60 % dip in UT) for
voltage variations 5 cycles N/A
on power supply 70 % UT
input lines
(30 % dip in UT) for
IEC 61000-4-11 25 cycles
<5 % UT
(>95 % dip in UT)
for 5s
Mains power quality should be that of a typical commercial or hospital environment. If a dips or an interruption of mains power occurs, the current of the model FR1DQ1-B may be dropped off from normal level, it may be necessary to use uninterruptible power supply or a battery.
Power frequency (50/60 Hz)
3 A/m magnetic field IEC 61000-4-8
3A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level
Guidance and manufacturer’s declaration electromagnetic immunity
The model FR1DQ1-B is intended for use in the electromagnetic environment specified below. The customer or the user of the model FR1DQ1-B should assure that it is used in such an environment.
Immunity test IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Portable and mobile RF communications equipment should be used no closer to
any part of the model FR1DQ1-B, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency
of the transmitter.
Conducted RF3 Vrms
N/A
IEC 61000-4-6 150 kHz to
80 MHz
Recommended separation distance
Radiated RF IEC3 V/m
61000-4-3
80 MHz to
2.5 GHz
3 V/m
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
Distance in metres (m).b
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,a should be less than the compliance level in each
frequency range. b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by
absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the model FR1DQ1-B is used
exceeds the applicable RF compliance level above, the model FR1DQ1-B should be
observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the
model FR1DQ1-B.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Recommended separation distances between
portable and mobile RF communications equipment and the model FR1DQ1-B
The model FR1DQ1-B is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
model FR1DQ1-B can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the model FR1DQ1-B as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
power
of transmitter
0.01
/
0.12
0.23
0.1
/
0.38
0.73
1
/
1.2
2.3
10
/
3.8
7.3
100
/
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
References