CALIBRATION
- Immerse the meter in the proper calibration solution: 1382 ppm (M10032) for T75 and 6.44 ppt (M10038) for T76.
- Unscrew the battery compartment cap on the top of the meter.
- Allow the reading to stabilize and with a small screwdriver turn the calibration trimmer inside the battery compartment until the display shows “1382” for T75 or “6440” for T76.
- Replace the battery compartment cap.
The calibration is now complete.
OPERATION
- Remove the protective cap.
- Turn the meter on by pressing the ON key.
- Immerse the probe in the solution to be tested.
- Stir gently and wait for the reading to stabilize.
- After use, rinse the electrode with water to minimize contamination.
- Always replace the protective cap after use.
Note: if the reading is overrange, T75 will display “1” while T76 shows “. . .”.
PROBE REPLACEMENT
The probe can be easily replaced in the following way:
- Remove the protective cap.
- Unscrew the plastic ring on the top of the probe.
- Pull out the MA73075 (or MA73076) probe and replace it with a new one (aligning the colored pin towards the LCD side).
- Make sure the gaskets are in place before screwing back the plastic ring.
BATTERY REPLACEMENT
The meter is supplied with BEPS (Battery Error Preventing System) which avoids
any erroneous reading due to low battery level, by switching the meter off. To
replace the batteries unscrew the battery compartment cap and replace all
three 1.5V batteries while paying attention to their polarity. Make sure the
gasket is in place before screwing back the cap.
Batteries should only be replaced in a nonhazardous area using the battery
type specified in this instruction manual.
SPECIFICATIONS
Range | T75 0 to 1999 ppm (mg/L) |
---|
T76 0 to 9990 ppm (mg/L)
Resolution| T75 1 ppm (mg/L)
T76 10 ppm (mg/L)
Accuracy| (@20°C) ±2% F.S.
Typical EMC Deviation| ±2% F.S.
Temperature Compensation| Automatic, 0 to 50°C with b=2%/°C
TDS factor| 0.5
Calibration| Manual, at 1 point through trimmer
Probe| T75
MA73075
(replaceable)
T76
MA73076
(replaceable)
Environment| 0 to 50°C; 100% RH
Battery type| 3 x 1.5V, alkaline
Battery life| approx. 250 hours of continuous use
Dimensions| 165 x 30 x 30 mm
Weight| 80 g
ACCESSORIES
MA73075 | Replaceable Conductivity probe, LR |
---|---|
MA73076 | Replaceable Conductivity probe, HR |
M10032B | 1382 ppm (mg/L) calibration solution, 20 mL (25 pcs) |
M10038B | 6.44 ppt (g/L) calibration solution, 20 mL (25 pcs) |
CERTIFICATION
Milwaukee Instruments conform to the CE European Directives.
Disposal of Electrical & Electronic Equipment. Do not treat this product as
household waste. Hand it over to the appropriate collection point for the
recycling of electrical and electronic equipment.
Disposal of waste batteries. This product contains batteries. Do not dispose
of them with other household waste. Hand them over to the appropriate
collection point for recycling.
Please note: proper product and battery disposal prevents potential negative
consequences for human health and the environment. For detailed information,
contact your local household waste disposal service or go to
www.milwaukeeinstruments.com (USA &
CAN) or www.milwaukeeinst.com.
RECOMMENDATION
Before using this product, make sure it is entirely suitable for your specific
application and for the environment in which it is used. Any modification
introduced by the user to the supplied equipment may compromise the meter’s
performance. For your and the meter’s safety do not use or store the meter in
hazardous environment. To avoid damage or burn, do not perform any measurement
in microwave ovens.
WARRANTY
These instruments are warranted against defects in materials and manufacturing
for a period of 2 years from the date of purchase. Probes are warranted for 6
months. This warranty is limited to repair or free of charge replacement if
the instrument cannot be repaired. Damage due to accidents, misuse, tampering
or lack of prescribed maintenance is not covered by warranty. If service is
required, contact your local Milwaukee Instruments Technical Service. If the
repair is not covered by the warranty, you will be notified of the charges
incurred. When shipping any meter, make sure it is properly packaged for
complete protection.
Milwaukee Instruments reserves the right to make improvements in design,
construction and appearance of its products without advance notice.
ISTRT75 07/20
SATISFACTION GUARANTEED
Documents / Resources
|
milwaukee T75 TDS Waterproof
Meters
[pdf] User Manual
T75, T76, T75 TDS Waterproof Meters, TDS Waterproof Meters, Waterproof Meters,
Meters
---|---
References
- Shop Milwaukee Instruments - the Official EU Site
- Shop Milwaukee Instruments - Official US Site
- Shop Milwaukee Instruments - the Official EU Site
- Shop Milwaukee Instruments - Official US Site
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>