HOBO U20 Water Level Logger User Manual
- June 14, 2024
- Hobo
Table of Contents
HOBO U20 Water Level Logger
PRODUCT INFORMATION
Logger (U20-001-0x and U20-001-0x-Ti) Manual
The HOBO U20 Water Level Logger is used for monitoring changing water levels in a wide range of applications including streams, lakes, wetlands, tidal areas, and groundwater. The loggers are typically deployed in existing wells or stilling wells installed specifically for deploying the loggers. This logger features high accuracy at a great price and HOBO ease-of-use, with no cumbersome vent tubes or desiccants to maintain.
The logger uses a maintenance-free absolute pressure sensor and features a durable stainless steel or titanium housing (depending on model) and ceramic pressure sensor. The HOBO Water Level Titanium is recommended for saltwater deployment for recording water levels and temperatures in wetlands and tidal areas. The logger uses precision electronics to measure absolute pressure and temperature and has enough memory to record over 21,700 combined pressure and temperature measurements.
HOBO Water Level Logger
Models:
- U20-001-01 (30-foot depth) and U20-001-01-Ti (30-foot depth/Titanium)
- U20-001-02 (100-foot depth) and U20-001-02-Ti (100-foot depth/Titanium)
- U20-001-03 (250-foot depth) and U20-001-03-Ti (250-foot depth/Titanium)
- U20-001-04 (13-foot depth) and U20-001-04-Ti (13-foot depth/Titanium)
Required Items:
- Coupler (COUPLER2-B) with USB Optic Base Station (BASE-U-4) or HOBO Waterproof Shuttle (U-DTW-1)
- HOBOware® Pro
Accessories:
- Cable (CABLE-1-300 or CABLE-1-50) and Cable Crimp (CABLE-1-CRIMP)
- Replacement Coupler (COUPLER2-B)
Specifications
Pressure (Absolute) and Water Level Measurements U20-001-01 and U20-001-01-Ti
- Operation Range 0 to 207 kPa (0 to 30 psi); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude
- Factory Calibrated Range 69 to 207 kPa (10 to 30 psi), 0° to 40°C (32° to 104°F)
- Burst Pressure 310 kPa (45 psi) or 18 m (60 ft) depth
- Water Level Accuracy*
- Typical error: ±0.05% FS, 0.5 cm (0.015 ft) water
- Maximum error: ±0.1% FS, 1.0 cm (0.03 ft) water
- Raw Pressure Accuracy** ±0.3% FS, 0.62 kPa (0.09 psi) maximum error
- Resolution <0.02 kPa (0.003 psi), 0.21 cm (0.007 ft) water
- Pressure Response Time (90%)*** <1 second; measurement accuracy also depends on temperature response time
Pressure (Absolute) and Water Level Measurements U20-001-02 and U20-001-02-Ti
- Operation Range 0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) of water depth at sea level, or 0 to 33.6 m (0 to 111 ft) of water at 3,000 m (10,000 ft) of altitude
- Factory Calibrated Range 69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure 500 kPa (72.5 psia) or 40.8 m (134 ft) depth
- Water Level Accuracy*
- Typical error: ±0.05% FS, 1.5 cm (0.05 ft) water
- Maximum error: ±0.1% FS, 3 cm (0.1 ft) water
- Raw Pressure Accuracy** ±0.3% FS, 1.20 kPa (0.17 psi) maximum error
- Resolution <0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) water
- Pressure Response Time (90%)*** <1 second; measurement accuracy also depends on temperature response time
Pressure (Absolute) and Water Level Measurements U20-001-03 and U20-001-03-Ti
- Operation Range 0 to 850 kPa (0 to 123.3 psi); approximately 0 to 76.5 m (0 to 251 ft) of water depth at sea level, or 0 to 79.5 m (0 to 262 ft) of water at 3,000 m (10,000 ft) of altitude
- Factory Calibrated Range 69 to 850 kPa (10 to 123.3 psi), 0° to 40°C (32° to 104°F)
- Burst Pressure 1200 kPa (174 psi) or 112 m (368 ft) depth
- Water Level Accuracy*
- Typical error: ±0.05% FS, 3.8 cm (0.125 ft) water
- Maximum error: ±0.1% FS, 7.6 cm (0.25 ft) water
- Raw Pressure Accuracy** ±0.3% FS, 2.55 kPa (0.37 psi) maximum error
Pressure (Absolute) and Water Level Measurements U20-001-03 and U20-001-03-Ti (continued)
- Resolution <0.085 kPa (0.012 psi), 0.87 cm (0.028 ft) water
- Pressure Response Time (90%)*** <1 second; measurement accuracy also depends on temperature response time
Pressure (Absolute) and Water Level Measurements U20-001-04 and U20-001-04-Ti
- Operation Range 0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) of water depth at sea level, or 0 to 7 m (0 to 23 ft) of water at 3,000 m (10,000 ft) of altitude
- Factory Calibrated Range 69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure 310 kPa (45 psia) or 18 m (60 ft) depth
- Water Level Accuracy*
- Typical error: ±0.075% FS, 0.3 cm (0.01 ft) water
- Maximum error: ±0.15% FS, 0.6 cm (0.02 ft) water
- Raw Pressure Accuracy** ±0.3% FS, 0.43 kPa (0.063 psi) maximum error
- Resolution <0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) water
- Pressure Response Time (90%)*** <1 second; measurement accuracy also depends on temperature response time
Temperature Measurements (All Models)
- Operation Range -20° to 50°C (-4° to 122°F)
- Accuracy ±0.44°C from 0° to 50°C (±0.79°F from 32° to 122°F), see Plot A
- Resolution 0.10°C at 25°C (0.18°F at 77°F), see Plot A
- Response Time (90%) 5 minutes in water (typical)
- Stability (Drift) 0.1°C (0.18°F) per year
Logger
- Real-time Clock ± 1 minute per month 0° to 50°C (32° to 122°F)
- Battery 2/3 AA, 3.6 Volt lithium, factory-replaceable
- Battery Life (Typical Use) 5 years with 1 minute or greater logging interval
- Memory (Non-volatile) 64K bytes memory (approx. 21,700 pressure and temperature samples)
- Weight Stainless steel models: approximately 210 g (7.4 oz) Titanium models: approximately 140 g (4.8 oz)
- Dimensions 2.46 cm (0.97 inches) diameter, 15 cm (5.9 inches) length; mounting hole 6.3 mm (0.25 inches) diameter
- Wetted Materials Stainless Steel models: 316 stainless steel, Viton® O-rings, acetal cap, ceramic sensor Titanium models: Titanium, Viton O-rings, acetal cap, ceramic sensor
- Logging Interval Fixed-rate or multiple logging intervals, with up to 8 user-defined logging intervals and durations; logging intervals from 1 second to 18 hours. Refer to the HOBOware software manual.
- Launch Modes Immediate start and delayed start
- Offload Modes Offload while logging; stop and offload
- Battery Indication Battery voltage can be viewed in status screen and optionally logged in datafile. Low battery indication in datafile.
- Environmental Rating IP68
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).
- Water Level Accuracy: With accurate reference water level measurement, known water density, accurate Barometric Compensation Assistant data, and a stable temperature environment.
- Raw Pressure Accuracy: Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors.
- Changes in Temperature : Allow 10 minutes in water to achieve full temperature compensation of the pressure sensor. Maximum error due to rapid thermal changes is approximately 0.5%.
Software
HOBOware Pro software is required for logger operation. Using a reference
water level, HOBOware Pro automatically converts the pressure readings into
water level readings. The software also supports compensation for temperature,
fluid density, and barometric pressure.
Communication
For launching and reading out the Water Level logger in the field, you can use
a laptop computer with HOBOware Pro and an Onset Optic USB Base Station
(BASE-U-4), with a coupler (COUPLER2-B) or the HOBO Waterproof Shuttle
(U-DTW-1) with a coupler (COUPLER2-B). The optical interface allows the logger
to be offloaded without breaking the integrity of the seals. The USB
compatibility allows for easy setup and fast downloads.
Barometric Compensation
The HOBO Water Level Logger records absolute pressure, which is later
converted to water level readings by the software. In this application,
absolute pressure includes atmospheric pressure and water head. Atmospheric
pressure is nominally 100 kPa (14.5 psi) at sea level, but changes with
weather and altitude. Left uncompensated, barometric variations could result
in errors of 0.6 m (2 ft) or more.
To compensate for barometric pressure changes, you can use the HOBO U20 Water Level Logger as a barometric reference. The barometric reference is typically deployed in the same well or at the same location as the water level of interest, but rather than being placed in the water column, it is deployed above the water in air.
Barometric pressure readings are consistent across a region
(except during fast-moving weather events), so you can generally use
barometric pressure readings that are taken within 15 km (10 miles) of the
logger or more, without significantly degrading the accuracy of the
compensation. Therefore, one U20 or weather station (HOBO U30 or H21
recommended) can be used to compensate all of the water level loggers in an
area. The U20-001-01 model with its 0–9m (0–30 ft) range or the U20-001-04
with its 0–4 m (0–13 ft) range are both good barometric references due to
their smaller range, temperature-compensated accuracy, and rugged stainless
steel case. HOBOware Pro includes a Barometric Compensation Assistant for easy
and accurate barometric compensation.
LEDs
A light (LED) in the communications window of the logger confirms the logger
operation.
The following table explains when the logger blinks during logger operation:
When: | The Light: |
---|---|
The logger is logging | Blinks once every one to four seconds (the shorter the |
logging interval, the faster the light blinks); blinks when logging a sample
The logger is awaiting a start because it was launched in Start At Interval or
Delayed Start mode| Blinks once every eight seconds until logging begins
Calibration
The pressure sensor in each HOBO Water Logger is individually calibrated.
During calibration, raw pressure sensor data is collected at multiple
pressures and temperatures over the calibrated range of the logger (see the
specifications table). This data is used to generate calibration coefficients
that are stored in the logger’s non-volatile memory. The calibration
coefficients are then checked to be sure that the logger meets its stated
accuracy over the calibrated range. The pressure sensor can be used at
pressures and temperatures that are outside of the calibrated range, but the
accuracy cannot be guaranteed.
Important : Never exceed the burst pressure of the sensor!
Sleep Mode
The logger consumes significantly more power when it is “awake” and connected
to a base station or shuttle. To conserve power, the logger will go into a
low-power (sleep) mode if there has been no communication with your computer
for 30 minutes. To wake up the logger, remove the logger from the coupler,
wait a moment, then re-insert the logger.
Sample and Event Logging
The logger can record two types of data: samples and events. Samples are the
sensor measurements recorded at each logging interval (for example, the
pressure every minute). Events are independent occurrences triggered by a
logger activity, such as Bad Battery or Host Connected. Events help you
determine what was happening while the logger was logging. The logger stores
64K of data and can record over 21,700 samples of pressure and temperature.
Setup
Before you deploy the HOBO U20 Water Level Logger in the field, perform the
following steps in the office:
- Start HOBOware.
- Connect the logger to the computer. See the next section.
- Verify the status. Click Status on the toolbar and observe that the absolute pressure is near barometric pressure for the location and the temperature is near the actual temperature.
- Launch the logger. See the HOBOware User’s Guide for details.
- Make sure both Abs. Pressure and Temperature are selected (temperature is required for temperature compensation of pressure).
- Logging Battery Voltage is not essential since you can check the battery voltage using the Status screen at launch or readout of logger.
Connecting the Logger to a Computer
The HOBO Water Level Logger requires a coupler (COUPLER2-B) and USB Optic Base
Station (BASE-U-4) or HOBO Waterproof Shuttle (U-DTW-1) to connect to the
computer.
- Follow the instructions that came with your base station or shuttle to attach the base station or shuttle to a USB port on the computer.
- Unscrew the black plastic end cap from the logger by turning it counterclockwise.
- Attach the coupler to the base station or shuttle
- Insert the logger into the coupler with the flat edge on the logger aligned with the arrow on the coupler label. Gently twist the logger to be sure that it is properly seated in the coupler (it should not turn).
NOTE : If you are using the Waterproof Shuttle, briefly press the coupler lever to put the shuttle into base station mode.
If the logger has never been connected to the computer before, it may take a few seconds for the new hardware to be detected by the computer.
Important : USB communications may not function properly at temperatures below 0°C (32°F) or above 50°C (122°F).
Deploying the Logger
The HOBO Water Level Logger is designed to be easy to deploy in many
environments. The logger uses an absolute pressure sensor, so no vent tube is
required. The small size of the logger is convenient for use in small wells
and allows the logger to be mounted and/or hidden in the
field.
Deployment Guidelines
Full Temperature Equilibrium
The pressure sensor is temperature compensated over the range of 0° to 40°C
(32° to 104°F). To obtain the highest level of accuracy, the logger should be
allowed to come to full temperature equilibrium (approximately 20 minutes)
before the reference level is recorded.
Sudden Temperature Changes
Sudden temperature changes should be avoided. When deploying a HOBO Water
Level Logger for barometric pressure reference, some considerations should be
made to minimize the rate of temperature fluctuations. Ideally, the barometric
pressure reference logger should be hung several feet below ground level in an
observation well where ground temperatures are stable (while making sure the
logger remains above the water level). If that is not possible (or if a well
is not used), try to put the logger in a location where it will not be subject
to rapid daily temperature cycles.
Venting
When deploying a HOBO Water Level logger in a well, make sure the well is
vented to the atmosphere. Typically, a small hole can be drilled in the well
cap to ensure that the pressure inside and outside the well is at equilibrium.
If this is not possible, the barometric pressure reference logger should be
used inside the same well.
Wire
Use a no-stretch wire to hang the water level logger. Any change in length of
the wire will result in a 1-to-1 corresponding error in the depth measurement.
Always pull-test a cable prior to deploying a logger in a well to make sure it
does not stretch.
Stilling Well
If you are deploying the logger in a lake, river, or stream, you must first
build a stilling well to protect the logger from vibration, shock, and
movement. A simple stilling well can be constructed with PVC or ABS pipe. A
properly constructed stilling well helps to protect the logger from currents,
wave action, and debris. Suspend the logger in the stilling well so it is
always underwater, but not on the bottom to be buried by silt.
For more information, see the Technical Application Note for Constructing a Stilling Well at: http://www.onsetcomp.com/water_level_stilling_well.html
Burst Pressure
Be very careful not to exceed the burst pressure for the logger. The pressure
sensor will burst if the maximum depth is exceeded (see specifications table).
The logger should be positioned at a depth where the logger will remain in the
water for the duration of the deployment, but not exceed the rated bursting
depth.
Deployment Procedure
- Cut wire to suspend logger.
- Measure the physical depth to the surface of the water from the suspension point.
- Cut a piece of stranded, stainless steel wire (Teflon coated is best) so that the logger will be deep enough to always be in the water. Estimate the low water level and make the cable length such that the logger will be about 2 feet below that level.
- Attach the wire to the suspension point and to the logger cap.
- Relaunch the logger if desired (if a PC or a HOBO U-Shuttle is available).
- Lower the logger into the well or stilling well.
- Measure the water depth from the desired reference point (top of pipe, ground level, or sea level).
- To maximize accuracy, allow 20 minutes after deploying the logger before measuring water depth to allow the logger to reach temperature equilibrium with the water.
- If the well is too small in diameter to measure the water depth after deployment, measure the water depth before deployment, then deploy the logger immediately and record deployment time.
- For good deployments: If the water level surface is below the reference point (such as referencing groundwater measurements to the top of the well), record the water level as a negative number. If the water level surface is above the reference point (such as height above sea level), record the water level as a positive number.
- For lake, stream, and river deployments: If the water level is being referenced to some point above the logger (such as the top of the stilling well), record the water level as a negative number. If the water depth is being referenced to a point below the water surface such as the bottom of the stream, record the water level as a positive number.
- Record the reference measurement date and time.
Deploying a U20 Logger for Barometric
Pressure Data (Optional)
If you are using a U20 logger to record barometric pressure data, install one
logger in one of the wells as follows:
- Cut wire for suspending the logger.
- Measure the physical depth to the surface of the water from the suspension point.
- Cut a piece of stranded, stainless steel wire (Teflon coated is best) so that the logger will hang about 2 feet below the ground surface but always above the water surface.
- Attach the wire to the suspension point and to the logger cap.
- Relaunch the logger if desired (if a PC or a HOBO U-Shuttle is available)
- Lower the logger into the well or stilling well. Make sure the logger does not go below the water surface.
- Record the deployment time.
Collecting Data
For reading out the Water Level logger in the field, you can use either of the
following:
- Laptop computer with HOBOware Pro and an Optic USB Base Station (BASE-U-4), with a coupler (COUPLER2-B)
- HOBO Waterproof Shuttle (U-DTW-1) with a coupler (COUPLER2-B)
- Measure the water depth using the original reference point with the correct sign.
- Record depth date and time.
- Pull the logger out of the well.
- Remove the logger from its cap, leaving the suspension undisturbed.
- Read out the data using one of the options listed above.
- Save the data in a test folder location.
- Redeploy the logger (optional). See below.
Barometric Pressure Data
To read out a U20 logger used for barometric pressure data:
- Remove the logger from the well.
- Read out the data using one of the options listed above.
- Save the data in a test folder location.
- Redeploy the logger (optional). See the next section.
Redeploying the Logger
If you are redeploying the logger, you must first make sure that it is
launched. If you use the HOBO Waterproof Shuttle to offload data, the shuttle
automatically performs a synchronized relaunch of the logger so that data is
logged on the same measurement intervals. If you wish to change the launch
settings, you must launch the logger using HOBOware Pro. The existing
suspension can be reused as long as the water level logger remains in the
water and the barometric logger remains out of the water for the entire test
interval. Take a new reference reading with the date and time as described in
Collecting Data. Record this information in your field notebook to use later
to calibrate your data, which will zero out any drift error.
Processing Data using
Barometric Pressure Data
To determine water level using barometric pressure data, use the Barometric
Compensation Assistant in HOBOware Pro, as described below. If you are using
barometric pressure data from a HOBO weather station, you can use the data
file as if it were U20 barometric data. For data from sources other than Onset
products, see Barometric Data from Other Sources below.
- In HOBOware Pro, open the water depth data file. The Plot Setup window appears.
- Uncheck all boxes except Abs. Pressure.
- Run the Barometric Compensation Assistant.
- Click the Process button.
- Select the water density box that best describes the water that you are measuring or enter the actual water density.
- Check the Use a Reference Water Level box and enter the reference water level that you measured at the beginning of the deployment.
- Enter the water level as a positive number if it is measured upward from a reference point below the water’s surface, such as the water’s height above sea level. This illustration shows an example of when to enter the water level as a positive number.
- Enter the water level as a negative number if it is measured downward from a reference point above the water’s surface, such as the top of the well. This illustration shows an example of when to enter the water level as a negative number.
- Select the date and time from the pull-down menu that is closest to the recorded date/time for the measurement. If you measured the depth before deployment because of pipe size, then select a date/time after the start of the deployment.
- Check Use Barometric Data file.
- Click the Choose button. This will allow you to select the data file to use for barometric pressure compensation.
- Select and open the data file.
- Click the Create New Series button. A new Plot Setup window appears.
- Select the Water Level box and any other series that you want plotted. Click the Plot button to obtain a plot of the resulting water level data.
Measurement Error
Measurement error can be caused by manual measurement error, sensor drift, or
a change in the suspension cable length. To quantify measurement error (which
is ideally zero), compare the calculated water level at the end of the plot
with the water level measured just before you removed the water level logger.
Barometric Data from Other Sources
Third-Party Weather Station or Barometric LoggerIf you choose to use
barometric pressure from a third-party weather station or barometric logger,
you need to convert the date, time, and pressure data to a text file with
special header requirements. For information on how to set up the text file,
see the HOBOware Help or User Guide. It is easiest to do this work in EXCEL
and then save it as a text file.
Online Weather Station
If you choose to use barometric pressure from an online weather station, such
as the National Weather Service, the measured barometric pressure is modified
to be at sea level. This sea level pressure is useable since all pressure
offsets are zeroed when you enter the reference measurement.
In the Barometric Compensation Assistant, when you select the Barometric Data File, select the text file that you generated. HOBOware Pro will ask for the data format and data separation characters (tab or comma) and then import the barometric data.
Maintenance
Protecting the Logger
Important : Do not attempt to open the logger housing! Unscrewing the metal nose cone of the logger will cause serious damage to the pressure sensor and logger electronics. There are no user serviceable parts inside the case. Contact Onset technical support if your logger requires servicing.
This logger can be damaged by shock. Always handle the logger with care. The logger may lose its calibrated accuracy or be damaged if it is dropped. Use proper packaging when transporting or shipping the logger.
Biofouling
Periodically inspect the logger for fouling. Biological growth on the face of
the pressure sensor will throw off the pressure sensor’s accuracy. Organisms
that grow inside the sensor nose cone and on the sensor itself can interfere
with the sensor’s operation and eventually make the sensor unusable. If the
deployment area is prone to biofouling, check the logger periodically for
marine growth.
Solvents
Check a materials compatibility chart before deploying the logger in locations
where untested solvents are present. The logger is shipped with Viton O-rings
installed. Viton has an excellent resistance to most solvents and is suitable
for deployments in water that contain a mixture of most fuels, solvents and
lubricants. However, the Viton O-rings are sensitive to polar solvents
(acetone, ketone), ammonia, and brake fluids. The black acetal cap is provided
to help protect the communications window. Acetal is resistant to most
solvents, fuels, and lubricants. The polycarbonate communications window is
sealed as an additional barrier to water and dirt entering the logger housing.
Compensating for Drift
All pressure sensors drift over time. The drift for the pressure sensor and
electronics in the HOBO Water Level logger is less than 0.3% FS (worst case)
per year. In most applications, drift is not a significant source of error,
because the offset created by any drift is zeroed out when you take a manual
reference level measurement and use the logger software to automatically
calculate the level readings relative to the reference measurement. In effect,
you are re-zeroing the sensor each time you apply a reference reading to the
data file.
Pressure sensor drift matters only when absolute pressure values are needed, or if there are no recent reference level or depth measurements available. For example, if the logger is deployed for one year and no new reference level readings are taken during the deployment, it is possible that the sensor could have drifted as much as 0.3% FS by the end of the deployment. It is possible to determine the actual amount of drift during a deployment if a reference level is taken at the beginning and the end of a long-term deployment. The results of applying the two different reference levels (once at the beginning of the data file, and again at the end of the data file) can be compared. Any difference between the files indicates the amount of sensor drift (assuming accurate reference levels).
Verifying Accuracy
You can check the differential accuracy of your loggers for water level
measurements by deploying the loggers at two depths and comparing the
difference in level readings. When verifying the accuracy this way, be sure to
allow the loggers’ temperature to stabilize at each depth. Use the logger
software to convert the readings from pressure to level. The level readings
should be taken close enough together that the barometric pressure does not
change.
You can check the absolute pressure accuracy of your HOBO Water Level Logger by comparing its ambient pressure readings to a second HOBO logger. Their readings should be within each other’s specified accuracy. Alternatively, you can check the pressure reading against an accurate local barometer. If you use a non-local source of barometric information, such as the NOAA website, adjust for altitude.
Recalibration
If you would like to have your logger’s absolute accuracy verified against a
NIST standard, or to have your logger recalibrated, contact Onset or your
place of purchase for pricing and return arrangements.
The Battery
The battery in the HOBO Water Level Logger is a 3.6 Volt lithium battery.
Battery Life
The battery life of the logger should be about five years or more. Actual
battery life is a function of the number of deployments, logging interval, and
operation/storage temperature of the logger. Frequent deployments with logging
intervals of less than one minute, and continuous storage/operation at
temperatures above 35°C will result in significantly lower battery life. For
example, continuous logging at a one-second logging interval will result in a
battery life of approximately one month. To obtain a five-year battery life, a
logging interval of one minute or greater should be used and the logger should
be operated and stored at temperatures between 0° and 25°C (32° and 77°F).
Voltage
The logger can report and log its battery voltage. If the battery falls below
3.1 V, the logger will record a “bad battery” event in the data file. If the
data file contains “bad battery” events, or if logged battery voltage
repeatedly falls below 3.3 V, the battery is failing and the logger should be
returned to Onset for battery replacement.
Replacing the Battery
To have your logger’s battery replaced, contact Onset or your place of
purchase for return arrangements. Do not attempt to replace the battery
yourself. Severe damage to the logger will result if the case is opened
without special tools, and the warranty will be voided.
WARNING : Do not cut open, incinerate, heat above 100°C (212°F), or recharge the lithium battery. The battery may explode if the logger is exposed to extreme heat or conditions that could damage or destroy the battery case. Do not dispose of the logger or battery in fire. Do not expose the contents of the battery to water. Dispose of the battery according to local regulations for lithium batteries.
© 2008–2018 Onset Computer Corporation. All rights reserved. Onset, HOBO, and HOBOware are trademarks or registered trademarks of Onset Computer Corporation. All other trademarks are the property of their respective companies.
Patent #: 6,826,664