LYCEBELL LC-38C Digital Pen Type Multimeter Instruction Manual
- June 14, 2024
- LYCEBELL
Table of Contents
LYCEBELL LC-38C Digital Pen Type Multimeter
Overview
The pen type digital multimeter is characterized at slim size, portable,
stable performance and anti-dropping capacity. Using 6000 counts digit LCD
monitor with character 12mm high, they offer clear readings. With overall
circuitry design centering on large-scale IC A/D converters in conjunction and
overload protection circuit, the meters give excellent performance and
exquisite making as a handy utility instrument.
The meters can be used to measure DC & AC voltage, resistance, capacitor,
positive diode voltage fall and audible continuity. If the equipment is used
in a manner not specified by the manufacturer, the protection provided by the
equipment may be impaired.
Panel Layout
-
VΩ: V/ Input test lead.
-
Lamp light.
-
Test lead fixture: Fix the COM test lead.
-
LCD display: 6000 counts digit, full function symbol display.
-
Toggle Switch: Use this switch to select functions and ranges.
-
HOLD key: Press the “HOLD” key to lock display value, and the “DH” sign will appear on the display, press it again to exit. Press the “HOLD” key over 2 seconds to light the back light, press it again for more than 2 seconds to lights off it, it can lights off automatically after 15 seconds too.
-
REL Key: Press the “REL” key, the meter enters relative measuring mode, “REL” is displayed on the LCD and the present reading becomes the reference value and displayed on the display.
Relative measurement REL△ =measurement value-Reference value. Press it again to exit. -
SELECT key: This key work on the ” ” range, press the key to choose resistance, diode, continuity test or capacitance test, on the voltage range, change to DC or AC.
-
Protective casing.
-
Battery cover.
-
COM: COM input test lead.
Safety Information
The meter is designed according to IEC-1010 concerning electronic measuring
instruments with an over-voltage category 600V (CAT Ⅲ) and pollution 2.
Follow all safety and operating instructions to ensure that the meter is used
safely and is kept in good operating condition.
safety symbols
Important safety information, refer to the operating manual.
Dangerous voltage may be presence.
Double insulation (protection Class II)
Special Cautions for Operation
- To avid risk of electric shock, do not use the meters before the cover is in place.
- The toggle switch should be right position for the testing.
- To avoid electric shock and damaging the instruments, the input signals are forbidden to exceed the specified limits.
- When measuring TV set or switched power, attention should be paid to the possible pulses that may bring destruction to the circuit.
- Toggle switch position is forbidden to be changed at random during measurement.
- Take caution against shock in the course of measuring voltage higher than DC 60V & AC 30V.
- After operation is finished, set toggle switch at OFF to save battery power.
- If the meter is without usage for long time, take out battery to avoid damage by battery leakage.
GENERAL SPECIFICATIONS
- Max Voltage between input terminal and Earth Ground: CAT Ⅲ 600V.
- Over-range Indication: display “OL” for the significant digit.
- Automatic display of negative polarity “-”.
- Low Battery Indication : “ ” displayed.
- Max LCD display: 6000 counts digit.
- Auto range control.
- Auto Power Off: When measurement exceeds 15 minutes without switching mode and pressing key, the meter will switch to standby mode. Press any key to exit standby mode. When restart the system, press and hold SELECT key to disable auto power off.
- Power supply: 1.5V×2 “AAA” R03P battery
- Operating Temp.: 0℃ to 40℃ (relative humidity <85%)
- Storage Temp. : -10℃ to 50℃ (relative humidity <85%)
- Guaranteed precision Temp.: 23±5℃ (relative humidity <70%)
- Dimension: 235 x 54 x 30mm
- Weight: approx. 200g (including battery)
Testing Specifications
Accuracy is specified for a period of year after calibration and at 18℃ to 28℃ (64℉ to 82℉) with relative humidity to 70%.
DC Voltage
Range | Resolution | Accuracy |
---|---|---|
600.0mV | 0.1mV | ±(0.5% of rdg + 2 digits) |
6.000V | 1mV | |
60.00V | 10mV | |
600.0V | 100mV | |
600V | 1V | ±(0.8% of rdg + 2 digits) |
- Impedance: 10MΩ, More than 100MΩ on 600mV range
- Overload protection: 600V DC or AC rms
AC Voltage (True RMS)
Range | Resolution | Accuracy |
---|---|---|
6.000V | 1mV | ±(1.0% of rdg + 3 digits) |
60.00V | 10mV | |
600.0V | 100mV | |
600V | 1V | ±(1.5% of rdg + 3 digits) |
- Impedance: 10MΩ
- Overload protection: 600V DC or AC rms
- Frequency Range: 40 to 2kHz
Resistance
Range | Resolution | Accuracy |
---|---|---|
600Ω | 0.1Ω | ±(1.0% of rdg + 3 digits) |
6kΩ | 1Ω | ±(1.0% of rdg + 2 digits) |
60kΩ | 10Ω | |
600kΩ | 100Ω | |
6MΩ | 1kΩ | |
60MΩ | 10kΩ | ±(1.5% of rdg + 3 digits) |
- Overload protection: 600V DC or AC rms
Capacitance
Range | Accuracy | Resolution |
---|---|---|
6nF | ±(3.0% of rdg + 10 digits) | 1pF |
60nF | ±(2.5% of rdg + 5 digits) | 10pF |
600nF | 100pF | |
6µF | 1nF | |
60µF | ±(5.0% of rdg + 10 digits) | 10nF |
600µF | ±(10.0% of rdg + 20 digits) | 100nF |
6mF | 1µF | |
60mF | 10µF |
- Overload protection: 600V DC or AC rms
Diode and Audible continuity test
Range | Description | Test Condition |
---|
| Display read approximately forward voltage of diode| Forward DC current approx. 1.5mA Reversed DC voltage approx. 4V
| Built-in buzzer sounds if resistance is less than 50Ω| Open circuit voltage approx. 2V
Overload protection: 600V DC or AC rms
OPERATING INSTRUCTIONS
Attention before operation
Check battery. When the battery voltage drop below proper operation range, the
“ ” symbol will appear on the LCD display and the battery need to changed.
Pay attention to the “ ” besides the input jack which shows that the input
voltage should be within the specified value.
The toggle switch should be positioned to desired range for measurement before
operation.
Measuring DC & AC Voltage
Set the toggle switch at the desired “ ” range position, it shows symbol for
testing DC voltage, if you want to test AC voltage, push “SELECT” button
switch.
Connect test leads across the source or load under measurement.
You can get reading from LCD. The polarity of the red lead connection will be
indicated along with the DC voltage value.
NOTE:
- “ ” means you can’t input the voltage more than 600V, it’s possible to show higher voltage, but it may destroy the inner circuit or pose a shock.
- Be cautious against shock when measuring high Voltage.
Measuring Resistance
- Set the toggle switch at the desired “ ” range position.
- Connect test leads across the resistance under measurement.
- You can get reading from LCD.
NOTE: Max. input overload: 600V rms<10sec
- For measuring resistance above 1MΩ, the mete may take a few seconds to get stable reading.
- When the input is not connected, i.e. at open circuit, the figure ‘OL’ will be displayed for the over-range condition.
- When checking in-circuit resistance, be sure the circuit under test has all power removed and that all capacitors have been discharged fully.
Measuring Capacitance
Set the toggle switch at the desired “ ” range position, push “SELECT” to
choose Capacitance measurement.
Connect test leads across the capacitance under measurement.
You can get reading from LCD.
NOTE: Max. input overload: 600V rms<10sec
- Capacitors should be discharged before being tested.
- When testing large capacitance, it will take longer time before the final indication ( For 100uF~60mF range, it will take about 10 seconds ).
- When testing small capacitance (≤1uF), to assure the measurement accuracy, first press “REL”, then go on measuring.
Diode & Audible continuity Testing
- Set the toggle switch at the “ ” range position, push “SELECT” to choose Diode or Aud i ble continuity measurement.
- On diode range, connect the test leads across the diode under measurement, display shows the approx. forward voltage of this diode.
- On Audible continuity range, connect the test leads to two point of circuit, if the resistance is lower than approx. 50Ω, the buzzer sounds.
NOTE: Make sure the power is cut off and all capacitors need to be discharged under this measurement.
Battery replacement
- When the battery voltage drop below proper operation range the ” ” symbol will appear on the LCD display and the battery need to changed.
- Before changing the battery, set the toggle switch to “OFF” position. Open the battery cover by rotary the battery cover.
- Replace the old battery with the same type battery (AAA R03P 1.5V×2).
- Close the battery cover and rotary the battery cover to lock it.
Maintenance
- Use only moist fabric or small amount of detergent but not chemical solution for cleaning.
- Do not use the meter before the back cover is properly closed and screw secured. Upon any abnormality, stop operation immediately and send the meter for maintenance.
- Please take out the battery when not using for a long time.
Above picture and content just for your reference. Please be subject to the
actual products if anything different or updated.
Please pardon for not informing in advance.
CONTACT US
For any problem or concern, welcome to email us for prompt response.
AFTERSALES1010@HOTMAIL.COM
P.S.
To make sure you can receive immediate solution and your requests processed
quickly, please email us with these information:
- Order Number
- Platform of Your Purchase
- Full Model Number
- Description of the Problem(Attaching videos or photos can help us troubleshoot the problems even faster)
Documents / Resources
|
LYCEBELL LC-38C Digital Pen Type
Multimeter
[pdf] Instruction Manual
LC-38C Digital Pen Type Multimeter, LC-38C, Digital Pen Type Multimeter, Pen
Type Multimeter, Type Multimeter, Multimeter
---|---
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>