**Criacr KA25 Professional Digital Scale User Manual

**

INSTRUCTION

Thank you for purchasing our scale. Normal care and proper treatment will provide for years of reliable service. Please read all operating instructions carefully before use. It may be necessary to calibrate the scale before the first use.

Standard Features Include

  • Unique hinged cover design protects the scale
  • Rugged plastic construction with stainless steel weighing pan
  • High contrast LCD
  • Weighs in g. oz, ozt, dwt, ct, gn
  • Low battery indicator
  • Energy-saving auto shut-off feature

SPECIFICATIONS

Capacity: 20g/100g/200g
Readability: 0.0019/0.01g
Units: g, oz, ozt, dwt, ct, gn
Calibration: External Calibration Tare Range: Full Capacity
Auto off(default) 60/120/180 seconds Operating Temperature Range : 10-30 C
Display: LCD Power Requirements: 2xAAA Batteries(included

LCD DISPLAY

  1. Current Weight
  2. Weight unit
  3. Low battery
  4. Negative weight value

OPERATION

  1. Open the battery compartment cover located on the backside of the scale.

  2. Insert 2xAAA batteries into the battery compartment (note: please follow the polarity +/- diagram in the battery compartment) The LCD will be on as below

  3. Replace the battery cover and place the scale on a horizontal flat surface, then wait until “0.00” or “0.000” is displayed,

  4. Press the MODE key to switch between weight units.
    DISPLAY

  5. Put the object or sample on the weighing platform and wait until the reading is stable.

TARE WEIGHING

  1. Press the ON/OFF button to turn on the scale.
  2. Place the “tare item (weighing container) on the weighing platform.
  3. Press TARE and wait until “0.00 ” or “0,000” is displayed.
  4. Add the “sample” into or on the weighing container.

CALIBRATION

1 Press the ON/OFF button to turn on the scale, then wait until “0.00” or “0.000” is displayed.
2.Press and hold the MODE key until the display reads “CAL
3.Press the MODE key again and wait for the LCD display to flash “50.00”
4 Place the 50g weight included) on the center of the platform.
5. Wat until the display displays “PASS” and then return to normal weighing mode.
6. Remove the 50g weight and complete the calibration

Auto off(default)

If there is not any operation and the weight is steady in weighing mode, ther scale will turn off automatically after about 180 seconde default)

How to change the auto-off time

  1. In weighing mode, press and hold the TARE key until the display reade the present auto-off time
  2. Press the MODE key to switch between 0, 60, 120, 180 seconds.(0 means to turn off the auto shut-off function.)
  3. Press the ON/OFF key to confirm your choice and complete the setup, the scale is turned off at the same time

Indication Message

Low battery indication:””,”Lo” or “88888”
Please replace the new batteries
DISPLAY

OUT2″: Please recalibrate the scale per the instructions above.

Overload indication: “0-Ld”

Please remove the object from the weighing tray immediately
DISPLAY

“LLLLL”: The internal components have been damaged. (i.e. it was shocked, dropped, or excessively overloaded). Try to recalibrate the scale. If the scale was not damaged too much, recalibration may fix the problem.

Unstable Indication: “UNST”

Please use the scale in a more stable (i.e. flat) position.
DISPLAY

TROUBLESHOOTING

The primary reason for inaccuracy or malfunction of the scale is due to low batteries or incorrect installation of batteries, incorrect or infrequent calibration, overloading the scale, or operating the scale on an unstable surface. Please keep in mind to maintain and operate your scale properly. The scale is a precision instrument and must be handled with the utmost care and caution.

PROPER OPERATION AND ENVIRONMENT

  • Avoid any exposure to extreme heat or cold, your scale works better when operated at normal room temperature. It is recommended to allow the scale to acclimate to normal room temperature for at least one hour before use.
  • Allow your scale sufficient time to warm up. Usually 30-60 seconds to give the internal components a chance to stabilize before calibration.
  • Keep your scale in a clean environment. Dust, dirt, moisture, vibration, air currents, and a close distance to other electronic equipment can cause an adverse effect on the reliability and accuracy of your scale.
  • Handle your scale with care. Gently apply all items or samples to be weighed onto the weighing surface. Although this scale is designed to be quite durable, try to avoid rough treatment as this may permanently damage the internal components.
  • Avoid shaking, dropping, or otherwise shocking the scale. This is a precision instrument and must be handled with extreme care.
  • Only operate the scale on a stable, vibration free surface.