SKC 500-300 Passive Sampler for Ammonia Instruction Manual
- June 15, 2024
- SKC
Table of Contents
- SKC 500-300 Passive Sampler for Ammonia
- Product Information
- Instruction
- Performance
- Analysis Instructions for Ammonia
- SKC Limited Warranty and Return Policy
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
- UMEx passive samplers are designed for single use only. Do NOT reuse UME x samplers.
SKC 500-300 Passive Sampler for Ammonia
Product Information
- Product Name: Passive Sampler for Ammonia
- Catalog Number: 500-300
- Sampling Rate : 39.92 ml/min with a relative standard deviation of 11.4%
- Validated Concentration Range: 0.1 to 100 ppm
- Analysis Method: Solvent extraction and analysis of ammonium ion by IC with conductivity detection or visible absorption spectrometry
Instruction
The SKC UMEx 300 Passive Sampler eff actively collects ammonia at 39.92 ml/min.
The UMEx 300 Passive Sampler contains a tape treated with sulfuric acid. Each
sampler incorporates a “blank/correction” section of tape
in addition to the sample section of tape. Analysis is by ion chromatography
(IC) with conductivity detection or visible absorption spectrometry for
identic cation of ammonia that may be present in the sample. Designed for
single use, the SKC UMEx 300 Passive Sampler is packaged in an aluminized
pouch for easy sample transport to a laboratory.
Performance
- Sampling Rate: 39.92 ml/min with a relative standard deviation of 11.4% Validated
- Concentration Range : 0.1 to 100 ppm
- Analysis Method: Solvent extraction and analysis of ammonium ion by IC with conductivity detection or visible absorption spectometry
- Lower Detection 15 min: 2.4 ppm
- Limits: 8 hrs: 0.075 ppm
- 24 hrs: 0.025 ppm
- Shelf-life: 9 from date of manufacture at ≤ 39.2 F (4 C)
- Storage: Before use: ≤ 39.2 F (4 C)
- After use: Ambient temperature (72 F [22 C]) or ≤ 39.2 F (4 C) and analyze within 3 weeks
- Accuracy: ± 23.4%
- Temperature Effects: No effect on sampling rate from 22 to 40 C
- Humidity Eff ects : No eff ect from 20 to 80% relative humidity (RH)
- Wind Velocity Effects: No effect from 5 to 100 cm/sec
- Interferences: None identified
- Dimensions: 3.4 x 1.1 x 0.35 in (8.6 x 2.8 x 0.89 cm)
- Weight: 0.38 oz (10.9 gm)
- Slide Cover: Pink
Cautions
- Do not store with food.
- Before sampling, check the expiration date on the label on the outside of the pouch. Do not use after the last day of the month indicated.
- SKC recommends using gloves when handling chemically treated media.
- UME x samplers are designed for single use only. Do NOT reuse UMEx samplers.
- Open the aluminized pouch and remove the sampler. Do not discard the pouch; use it to send sampler to the laboratory. Store the pouch away from potential ammonia sources.
- Enter date, start time, and location in the space provided on the back of the sampler.
- Position the sampler on a worker’s collar for personal sampling or in an appropriate location for area sampling.
- Slide the sampler cover to the “on” position to begin sampling.
- After sampling for the desired time, up to 24 hours, slide the sampler cover to the “off ” position to stop sampling. Enter the sample stop time in the space provided on the back of the sampler.
- Place the sampler in the original pouch immediately after sampling.
- Seal the pouch. Send pertinent information and sample to an accredited laboratory for analysis.
Analysis Instructions for Ammonia
Purchase commercially available ammonium standards appropriate for your application. Prepare a range of ammonium standards in de-ionized ultrafiltered (DIUF) water as outlined in Sample Preparation.
Note : The ammonium curve is not linear. A graphing program will be needed to calculate results.
Sample Preparation for Ammonia
- Remove the sampler from the pouch and the sliding cover from the sampler.
- Use clean forceps to lift out the treated tape from each section. Place each piece of tape in a sealed vial. This provides a sample and a blank/correction.
- Note : It is also acceptable to use a “lot” blank to correct for background. One lot blank is included in each box of badges.
- Note: The blank/correction section of tape has an indentation for easy identic cation.
- Ammonium is desorbed from both the sample and the blank/correction tapes by adding 3 ml of DIUF water to each vial. Shake the sample by hand and then let stand for 60 minutes. Excessive shaking causes the fiter to break apart in the vial.
- Immediately transfer 1 ml of extract to an auto-sampler vial for analysis of ammonia.
Ammonia Sample Analysis
- The sample extracts are analyzed for the ammonium ion by ion chromatography with conductivity detection or visible absorption spectrometry.
- A 25-microliter portion of the extract is injected onto a Dione 3 x 250-mm CS16 column and with a 36-mM methane sulfonic acid eluent.
- Calculate the ammonium results by comparing against a standard calibration curve.
- Convert the results from ammonium to ammonia using the following formula: Concentration μg/ml ammonia = Concentration μg/ml ammonium x 0.944
- Total mass of ammonia is calculated below: Concentration ammonia (μg/ml) x Desorption volume (3 ml)
- The ammonia from the blank/correction tape must always be subtracted from the sample tape when calculating air concentrations.
- Calculate the air concentration in ppm using the following equations
- Volume of air (liters) = Time (minutes) x Sampling rate (39.92 ml/min) 1000
- Concentration (ppm) = Mass (mg) x 24450
- Air volume (L) x Molecular weight (17.0 gm/mole)
UME x Passive Samplers # | Cat. No. |
---|---|
UME x **300,* †** for ammonia, pk/10 | 500-300 |
UME x **100,* †‡** for formaldehyde and other aldehydes, pk/10 |
500-100
UME x **200,* † for sulfur dioxide and/or nitrogen dioxide,
pk/10| 500-200
Accessory
Treated Tape, for QC purposes only, pk/25| P20083
Stand for Area Sampling| 690-302**
-
- Limited shelf-life
- † Do not store with food.
- ‡ Storage at ≤39.2 F (4 C) required
-
UMEx passive samplers are designed for single use only. Do NOT reuse UME x samplers.
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to skcinc.com/warranty .
References
- SKC, Inc. | Industrial Hygiene and Occupational Safety Products and Technology
- Warranty Information | Learn More About Industrial Hygiene & Occupational Safety Products at SKC, Inc.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>