SKC 690-105 ULTRA Passive Samplers Instruction Manual
- June 13, 2024
- SKC
Table of Contents
SKC 690-105 ULTRA Passive Samplers
Product Information
Product Name: ULTRA Passive Samplers with Anasorb CSC Sorbent
Product Model: Cat. No. 690-105
Manufacturer: SKC
Applications: SKC-patented ULTRA Passive Samplers are designed to detect
organic vapors in ambient and indoor air at low ppb to ppt levels.
Performance Profile:
- Sampling Rates: Refer to the SKC Passive Sampling Guide at www.skcinc.com
- Analysis: Solvent extraction; gas chromatography. See Analysis Instructions at www.skcinc.com/knowledgecenter
- Shelf-life: Six months from the date of manufacture. Store at 39.2°F (4°C)
- Patent: U.S. Patent No. 6,607,581
Product Usage Instructions
-
Open the pouch and remove the sampler. Do not discard the pouch, instead store it in a clean, organic solvent-free environment.
-
Record the following information:
- Sample ID on the pouch and sampler ID label
- Sampling location on the pouch
- Date the sample was taken on the pouch
- Indoors: Clip the sampler(s) to the stand accessory 3a (Cat. No. 690-302) and position it at the appropriate height.
- Outdoors: Clip the sampler(s) inside a shelter that is open on two ends and the bottom to allow air to flow freely. Hang the shelter securely at the appropriate height. Clip the sampler(s) in the appropriate location:
Ensure that the diffusion holes in the sampler are not obstructed in any way.
-
Start sampling by sliding the sampler cover to the open position (diffusion holes visible). Write the sample start time on the pouch label.
-
Stop sampling by sliding the sampler cover to the closed position (arrow and SKC visible). Write the sample stop time on the pouch label.
Important: Measure and record the ambient temperature and atmospheric pressure. Include this information when sending the samples to the laboratory. -
Immediately after sampling, place the sampler in its resealable pouch. Make sure that the sample ID on the sampler matches the sample ID on the pouch.
-
Seal the pouch(es) and place them in a container with a cool pack.
-
Use expedited shipping to send the sample(s), pertinent information, and Chain of Custody form(s) to a laboratory for analysis.
Storage:
- Before Sampling: Up to 6 months at 39.2°F (4°C) in a clean, organic solvent-free environment
- After Sampling: 21 days at 39.2°F (4°C) in a clean, organic solvent-free environment
Ordering:
Sampling Accessories:
- Stand for Indoor Sampling – Cat. No. 690-302
- Shelter for Outdoor Sampling – Cat. No. 690-303
Warranty and Return Policy:
SKC products are subject to the SKC Limited Warranty and Return Policy. For
more information, refer to the complete SKC Limited Warranty and Return Policy
at www.skcinc.com/warranty.
Form 40126 Rev 230227
Operating Instructions
Applications
- SKC patented* ULTRA Passive Samplers provide for low ppb to ppt detection of organic vapors in ambient and air.
Performance Profile
-
Sampling Rates : Go to the SKC Passive Sampling Guide at www.skcinc.com.
-
Analysis : Solvent extraction; gas chromatography
See Analysis Instructions at www.skcinc.com/knowledgecenter. -
Shelf-life : Six months from the date of manufacture. Store at 39.2 F (4 C).
The US. Patent No.6,607581
Sampling
-
Open pouch. Remove sampler.
a. Do not discard the pouch.
b. Store the pouch in a clean, organic solvent-free environment. -
Record the following:
a. Sample ID on pouch and sampler ID label
b. Sampling location on pouch
c. The date sample was taken on the pouch!
Do not use Sharpie markers or equivalent to mark labels or pouches as these can off-gas VOCs. -
Clip sampler(s) in appropriate location.
a. Indoors : Clip sampler(s) to stand accessory (Cat. No. 690-302) and locate at the appropriate height. b. Outdoors : Clip sampler(s) inside the shelter (open on two ends and bottom to allow air to flow freely). Hang shelter securely at the appropriate height.
Ensure diffusion holes in the sampler are not obstructed in any way. -
Start sampling: Slide sampler cover to open position (diffusion holes visible). Write sample start time on pouch label.
-
Stop sampling: Slide the sampler cover to the closed position (arrow and SKC visible). Write sample stop time on pouch label.
Important : Measure and record ambient temperature and atmospheric pressure. Include information sent to the laboratory. -
Place the sampler in its resealable pouch immediately after sampling.
Ensure the sample ID on the sampler matches the sample ID on the pouch. -
Seal pouch(es). Place pouch(es) in a container with a cool pack.
-
Use expedited shipping to send sample(s), pertinent information, and Chain of Custody form(s) to a laboratory for analysis.
Storage Before
- Sampling : Up to 6 months at S 39.2 F (4 C) in a clean, organic solvent-free environment
- After Sampling: 21 days at 39.2 F (4 C) in a clean, organic solvent-free environment
Ordering
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC’s sole liability and the buyer’s exclusive remedy. To view the
complete SKC Limited Warranty and Return Policy, go to
skcinc.com/warranty.
863 valley View Road, Eighty Four, PA 15330
724-941-9701 •
skcinc.com