molift Mover 180, 205 and 300 Electric Patient Lift User Guide
- June 4, 2024
- molift
Table of Contents
Mover 180, 205, and 300 Electric Patient Lift
User Guide
Periodic inspection checklist
In accordance with ISO: 10535
PI06101 Rev. D 2022-06-15
Product label, Mover 180
QR code URL: http://q.etac.com/004e
ID label, Mover 180
Product label, Mover 205
QR code URL: http://q.etac.com/004k
ID label, Mover 205
Product label, Mover 300
QR code URL: http://q.etac.com/004l
ID label, Mover 300
A copy of this document is available for download at
www.etac.com.
Scan the QR code on the product label for accessing the product home page. The
user manual can be found in the document section.
Article no. (see ID label): —————————–
Serial no. (see ID label): ——————————-
Production Year: —————————————-
Owner: —————————————————-
The sling shall be inspected separately and is not included in this
inspection.
Specify model:
Mover 180
Mover 205
Mover 300
Situation of use
Home
Hospital
Nursing home
Other
The periodic inspection is to be performed by a person who is suitable and
properly qualified and well acquainted with the design, use, and care of the
hoist
The hoist’s software Molift Service tool provides an option to enter the
date of PI and the name of the inspector along with additional information.
Molift recommends adding all observations/remarks concerning the hoist in the
Molift Service tool for a complete history log of the hoist
Service light indicator
Note color of service light indicator:
Status | Service light | Mode |
---|---|---|
(No light) | Power saving (Stand by) | |
Green | Ready for use | |
Yellow | Order service | |
Red | Perform service | |
Red (and | Perform service |
Visual examination
Visual examination of load-bearing structure to make sure there is no damage, cracks, frays, or deformation. All checkpoints must be checked to approve the hoist for further use
OK| Not OK|
---|---|---
| | Lifting column
| | Lifting arm
| | Legs
| | Product label
| | Plastic covers
| | Hand control
| | Actuator
| | Wheels/castors
| | Brakes on wheels/castors
| | Charger
| | Cables
| | Battery
| | Other electronics
| | Sling bar, incl. padding
| | Leg spreading mechanism
| | No corrosion
| | Weldings
| | Instruction for use/user manual is present/available
Functional examination
Test FUNCTION and inspect for wear and damage.
All checkpoints must be checked to approve the hoist for further use
OK| Not OK|
---|---|---
| | Wheels/castors
| | Rear wheel/castor brakes
| | With legs in inner position (parallel): Check that the distance between
legs can not decrease when pushed together by hand
| | For Mover 180 only: Lifting the column is completely fixed with hexagon
locking screws tightly secured
| | For Mover 205 and 300 only: The lifting column is completely fixed when
the locking handle is engaged
| | The emergency stop button is fully operational and interrupts power when
engaged.
Reset switch.
| | Electrical emergency lowering
| | Manual emergency lowering
| | Hand control (buttons, strain relief on hand control cable, hook, and
battery indicator)
| | Sling bar coupling (swivel joint moves freely and without noise)
| | Sling bar quick release
| | If all points so far are “OK”, the hoist shall be tested with SWL.
| | Perform one lifting cycle with weight equal to SWL (see product label for
value kg/lbs).
The actuator must not slip when loaded (causing the lifting arm to lower).
| | Leg spreading mechanism, outwards and inwards with SWL
| | Perform new visual control;
Damage, play, and deformations as described above. Any damaged parts must be
repaired
or replaced and the test is repeated once more until the hoist performs
correctly
| | Charger is operational
| | With a fully charged battery inserted, after the initial cycle, the
battery light is off, and the wrench/the service light is green
Accessories: ———————
Performed by
Full name: ———————–
Date/Place: ———————-
Signature: ————————-
Approved without faults
Next inspection (YYYY / MM): ………../…………
The hoist is marked with “Out of order” and sent for repair
The hoist is not eligible for repair and is taken out of service
If the periodic inspection reveals any defect, wear, or another damage that
jeopardizes the safety of the patient the hoist may not be used until the
deficiency has been eliminated
The owner is notified
Due to legal requirements, this document or a copy must be filed with the
owner’s service log book
If approved without faults, apply inspection label and mark with month and
year for next inspection. When ordering labels, use item no. 1100306
Notes and observations
Documents / Resources
|
molift Mover 180, 205 and 300 Electric Patient
Lift
[pdf] User Guide
Mover 180, 205 and 300 Electric Patient Lift, Mover 180, Mover 205, Mover 300,
Electric Patient Lift, Patient Lift, Electric Lift
---|---
References
- Molift Mover 180
- Molift Mover 205
- Molift Mover 300
- Etac - A world-leading developer of assistive devices
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>