molift Mover 180, 205 and 300 Electric Patient Lift User Guide

June 4, 2024
molift

molift Logo Mover 180, 205, and 300 Electric Patient Lift
User Guide

Periodic inspection checklist

In accordance with ISO: 10535
PI06101 Rev. D 2022-06-15

molift Mover 180 205 and 300 Electric Patient Lift -
Overview

Product label, Mover 180

molift Mover 180 205 and 300 Electric Patient Lift - Fig
1 QR code URL: http://q.etac.com/004e

ID label, Mover 180

molift Mover 180 205 and 300 Electric Patient Lift - Bar
Code

molift Mover 180 205 and 300 Electric Patient Lift - Overview
1

Product label, Mover 205

molift Mover 180 205 and 300 Electric Patient Lift - Fig
2 QR code URL: http://q.etac.com/004k

ID label, Mover 205

molift Mover 180 205 and 300 Electric Patient Lift - Bar Code
1

molift Mover 180 205 and 300 Electric Patient Lift - Overview
2

Product label, Mover 300

QR code URL: http://q.etac.com/004l

ID label, Mover 300

A copy of this document is available for download at www.etac.com.
Scan the QR code on the product label for accessing the product home page. The user manual can be found in the document section.
Article no. (see ID label): —————————–
Serial no. (see ID label): ——————————-
Production Year: —————————————-
Owner: —————————————————-
The sling shall be inspected separately and is not included in this inspection.
Specify model:
Mover 180
Mover 205
Mover 300

Situation of use

Home
Hospital
Nursing home
Other
The periodic inspection is to be performed by a person who is suitable and properly qualified and well acquainted with the design, use, and care of the hoist
The hoist’s software Molift Service tool provides an option to enter the date of PI and the name of the inspector along with additional information. Molift recommends adding all observations/remarks concerning the hoist in the Molift Service tool for a complete history log of the hoist

Service light indicator
Note color of service light indicator:

Status Service light Mode
(No light) Power saving (Stand by)
Green Ready for use
Yellow Order service
Red Perform service
Red (and Perform service

Visual examination

Visual examination of load-bearing structure to make sure there is no damage, cracks, frays, or deformation. All checkpoints must be checked to approve the hoist for further use

OK| Not OK|
---|---|---
| | Lifting column
| | Lifting arm
| |  Legs
| | Product label
| |  Plastic covers
| |  Hand control
| | Actuator
| | Wheels/castors
| | Brakes on wheels/castors
| | Charger
| |  Cables
| | Battery
| | Other electronics
| | Sling bar, incl. padding
| | Leg spreading mechanism
| | No corrosion
| | Weldings
| | Instruction for use/user manual is present/available

Functional examination
Test FUNCTION and inspect for wear and damage.
All checkpoints must be checked to approve the hoist for further use

OK| Not OK|
---|---|---
| | Wheels/castors
| | Rear wheel/castor brakes
| | With legs in inner position (parallel): Check that the distance between legs can not decrease when pushed together by hand
| | For Mover 180 only: Lifting the column is completely fixed with hexagon locking screws tightly secured
| | For Mover 205 and 300 only: The lifting column is completely fixed when the locking handle is engaged
| | The emergency stop button is fully operational and interrupts power when engaged.
Reset switch.
| | Electrical emergency lowering
| | Manual emergency lowering
| | Hand control (buttons, strain relief on hand control cable, hook, and battery indicator)
| | Sling bar coupling (swivel joint moves freely and without noise)
| | Sling bar quick release
| | If all points so far are “OK”, the hoist shall be tested with SWL.
| | Perform one lifting cycle with weight equal to SWL (see product label for value kg/lbs).
The actuator must not slip when loaded (causing the lifting arm to lower).
| | Leg spreading mechanism, outwards and inwards with SWL
| | Perform new visual control;
Damage, play, and deformations as described above. Any damaged parts must be repaired
or replaced and the test is repeated once more until the hoist performs correctly
| | Charger is operational
| | With a fully charged battery inserted, after the initial cycle, the battery light is off, and the wrench/the service light is green

Accessories: ———————

Performed by
Full name: ———————–
Date/Place: ———————-
Signature: ————————-
Approved without faults
Next inspection (YYYY / MM): ………../…………
The hoist is marked with “Out of order” and sent for repair
The hoist is not eligible for repair and is taken out of service
If the periodic inspection reveals any defect, wear, or another damage that jeopardizes the safety of the patient the hoist may not be used until the deficiency has been eliminated
The owner is notified

Due to legal requirements, this document or a copy must be filed with the owner’s service log book
If approved without faults, apply inspection label and mark with month and year for next inspection. When ordering labels, use item no. 1100306

molift Mover 180 205 and 300 Electric Patient Lift - Fig
4

Notes and observations

molift Logo