seitron Americas S500 Portable Combustion Analyzer User Guide

June 13, 2024
seitron Americas

seitron Americas S500 Portable Combustion Analyzer

Product Information

The S500 analyzer is a combustion analysis device that offers various features and functions. It can perform combustion analysis in both Auto and Manual modes, calculate stack heat loss and efficiency, monitor ambient CO levels, measure differential pressure, draft, and pressure, and generate a QR Code for downloading the data of the analyses. The instrument also has the capability to print the analyses and performed measures on a ticket.

Product Usage Instructions

  1. Before starting the combustion analysis, select the utilized fuel.
  2. If you want to print the ticket of the average analyses and additional measures, enable the printer in the PRINT menu.
  3. If you want to print the complete analysis and performed measures, enter the MEMORY menu, select the related memory number, and push the PRINT interactive function button.
  4. To download the data of the single performed analyses, enter the MEMORY menu, select the memory number used to save the analyses and measures, and select one at a time the single rows.
  5. Scan the QR code using the Seitron APP SEITRON SMART ANALYSIS to download the acquired data related to the average analyses and additional measures if performed.
  6. Keep in mind that the magnets in the back of the instrument can damage credit cards, hard drives, mechanical watches, pacemakers, defibrillators, and other devices sensitive to magnetic fields. It is recommended to keep the instrument at least 10 inches away from any of these devices.
  7. Ensure all connections are tight to assure accurate sampling.
  8. When conducting measurements, make sure the water trap/filter assembly is in a VERTICAL position to prevent damage from moisture and particles to sensors.

For more detailed information on the use and maintenance of the S500 analyzer, refer to the complete manual available online at www.seitronamericas.com.

If you have any further questions or need assistance, please contact Seitron Americas Inc. at:
140 Terry Drive, Suite 101 – Newtown (PA) 18940 – USA
Tel.: 215-660-9777
Email: info@seitronamericas.com
Website: www.seitronamericas.com

FEATURES

  • Combustion analysis in Auto or Manual Mode
  • Calculation of stack heat loss and efficiency
  • Ambient CO monitor
  • Measuring differential pressure
  • Draft measurement
  • Pressure measurement
  • Generation and display of a QR Code for downloading the data of the analyses
  • Possibility to print on ticket the analyses and the performed measures with a Bluetooth® printer (optional)

The magnets in the back of the instrument can damage to credit cards, hard drives, mechanical watches, pacemakers, defibrillators and other devices proven sensitive to magnetic fields.
It is recommended to keep the instrument at a distance of at least 10” away from any of these devices.

KEYBOARD FUNCTIONS

KEYS FUNCTION
Activate the context keys shown on the display.
Turns on and off the instrument.
  • If pressed briefly, accesses the instrument menu.
  • If pressed for at least 2 seconds, turns off the instrument.

| Exits the current screen.
| Confirm settings.
| Select and/or Modify.

CONTEXT KEYS

CONTEXT KEY FUNCTION
Save Saves the data in the instrument memory.
Print Display the printing options screen and consequently execute it.
Keep Proceed with the combustion analysis.
Repeat Repeat the autozero phase.

The complete manual for use and maintenance of S500 is available online at www.seitronamericas.com.
Seitron Americas respects the nature and the environment, therefore provides this quick user guide of the S500 analyzer. However, complete documentation is available online. Respect your environment: think before printing the full manual on paper.

USING THE FLUE GAS ANALYZER

Be sure all connections are tight to assure accurate sampling.
When conducting measurements, the water trap/filter assembly MUST be in a VERTICAL position to prevent damage from moisture & particles to sensors.

  • When testing is completed, always drain the water trap with any condensation (after EACH test!).

On/Off

  • Before switching on the instrument, insert the Tc-K connector of the smoke probe to the instrument for primary air acquisition.
  • Perform the Zero phase of the instrument in fresh clean air.
  • When the Zero phase is over, push the button related to the interactive function “Keep” to store the acquired temperature value.

MANUALLY SET PRIMARY AIR TEMP

Menu→ Configuration→ Analysis→ Air temp
Procedure to manually set the primary air temperature value. Press OK to enter T air edit mode.
Press the arrows to select the digits you want to change and press OK when done.

DRAFT •MEASUREMENT

Menu→ Measurements→ Draft
To measure the draft proceed as follows:

  • Connect the probe pressure input hose to the instrument P+ input.
  • Before ZEROing the Pressure/Draft  sensor, please be sure to remove the gas probe from the stack first
  • Upon completion of the Zeroing of the Draft Sensor insert the probe in the stack to measure the Draft.

AMBIENT CO

Menu→ Measurements→ CO air

WARNING
It is compulsory to perform the Zero phase in clean air, so that the environment CO measurement results correct. It is advisable to turn on the instrument and wait for the Zero phase completion outside the area where the test is being performed.

PRINT

Menu→ Configuration→ Print

Ticket print
Set the parameter “Copy number” for how many printouts Set the parameter “printer” on BT or IR

Only if the printer is set on IR:
Set the parameter “Mode” on fast or slow

Only if the printer is set on BT:
Link the instrument to the BT printer through the parameter Pairing BT (only for the first time)

QR code generation
Set the parameter “QR code” on ON

COMBUSTION ANALYSIS

  • Before starting the combustion analysis, select the utilized fuel.
  • If it is desired to print the ticket of the average analyses and the additional measures enable the printer in the “PRINT” menu.
  • If it is desired to print the complete analysis and the performed measures, it will be necessary to enter the memory menu “MEMORY”, select the related memory number and push the “PRINT” interactive function button.
  • To download the data of the single performed analyses, it’s necessary to enter the “MEMORY” menu, select the memory number used to save the analyses and the measures and select one at a time the single rows.

Seitron Americas Inc.
140 Terry Drive, Suite 101 – Newtown (PA) 18940 – USA
Tel.: 215-660-9777
Email: info@seitronamericas.com
Website: www.seitronamericas.com