ACC-SCOPE EXM-150 Darkfield Annulus Instruction Manual

June 13, 2024
ACC-SCOPE

ACC-SCOPE EXM-150 Darkfield Annulus

ACC-SCOPE-EXM-150-Darkfield-Annulus-PRODUCT

Product Information

The EXM-150 Darkfield Annulus (CAT# 150-3220) is an accessory designed for use with models of the EXM-150 series microscopes with Abbe condensers. It is effective with 4x and 10x objectives, but not recommended for use with 40x or 100x objectives. The darkfield annulus is included with the microscope when purchased, and it is compatible with digital cameras.

Product Usage Instructions

  1. Turn on the microscope and adjust the light intensity to ensure bright enough illumination for comfortable observation.
  2. Ensure that the darkfield annulus is in the out position, away from the light path. It should move freely in one direction and bump against a stop in the other direction.
  3. Place a specimen on the stage and focus using the 4x objective.
  4. Swing the darkfield annulus into the in position. The field of the specimen should appear dark, while the specimen and any debris or particles should appear bright against the dark background.
  5. Adjust the iris position to optimize the view. With practice, you can make these adjustments while continuously looking into the microscope.
  6. Switch to the 10x objective and adjust the iris position as necessary to optimize the image.
  7. To use brightfield observation, move the darkfield annulus to the out position and adjust the iris to achieve the desired contrast in the image.

Installation and Operation

ACC-SCOPE-EXM-150-Darkfield-Annulus-FIG-1
\(1\)

  • NOTE: If purchased with the microscope, your EXM-150 microscope should arrive with the darkfield annulus (stop) already installed.
  • NOTE: The EXM-150 darkfield annulus is designed for use only with models of the EXM-150 series microscopes with Abbe condensers (CAT# EXM-151, EXM-151-T, EXM-151-VT, and any of these models with digital cameras).
  • IMPORTANT: The darkfield annulus is effective with 4x and 10x objectives on the EXM-150 models specified above. It is not recommended for use with 40x or 100x objectives.

EXM-150 Darkfield Annulus Instructions v230711
The Principle of Darkfield

Darkfield illumination generates a dark background to the specimen whereas brightfield illumination generates the familiar bright (some call it white) background or field around the specimen. In brightfield (BF) observation, light passes through the specimen and is collected by the objective for observation. Color and contrast information in the specimen is transmitted with the light, producing the colorful specimen image against a light background (no color or contrast). In darkfield, light from the light source is presented to the specimen at a very high angle. In the absence of any specimen, the light would not ever be captured by the objective, passing right by it, and resulting in a dark (black) background. With the presence of any substance or specimen, light is scattered in many directions, some of which will make it into the objective and can therefore be observed. Some color information may be seen, but darkfield observation often appears monochromatic. Darkfield observation is useful  with unstained specimens or specimens with little to no coloration.

Installation

When ordered with the microscope, the darkfield annulus is already installed. No installation is required.

EXM-150 Darkfield Annulus Instructions v230711

Operation

  1. Turn on the microscope. Adjust light intensity at any time until illumination is bright enough and observation is comfortable.
  2. Start with the darkfield annulus in the “out” position. Swing or rotate the darkfield annulus out of the light path. It should move freely in one direction, but bump against a stop in the other direction.ACC-SCOPE-EXM-150-Darkfield-Annulus-FIG-1 \(4\)
  3. Place a specimen on the stage and focus on the 4x objective.
  4. Open the condenser iris using the iris lever. While looking into the microscope, move the iris lever to open and close the iris – you will see the image change in brightness, and you will also see the “texture” change in the image (texture/contrast increase when the iris is closed down).
  5. Swing the darkfield annulus into the “in” position. Observe the specimen. The field of the specimen should appear dark, and the specimen (and any debris, dirt, particles, etc.) should appear bright against the dark background.
  6. Adjust the iris position to optimize the view. With some practice, you will be able to make these adjustments while continuously looking into the microscope.
  7. Switch to the 10x objective and adjust the iris position as necessary to optimize the image.
  8. Move the darkfield annulus to the “out” position to use brightfield observation. Adjust the iris to achieve the desired contrast in the image.

EXM-150 Darkfield Annulus Instructions v230711