INSTRUCTION MANUAL
715 Portable Portable Sludge Blanket Detector
Introduction
The 715 Portable Sludge Blanket Level Detector is designed for spot checking
sludge interface levels in settling tanks. The 715 works by measuring the
change in the level of suspended solids as the sensor is lowered into the
tank. When the sensor reaches the sludge blanket, an audible signal changes
from a fast bleeping to a slow bleeping. A red LED provides a visual
indication that the sludge blanket has been detected. To enable measurement of
the sludge interface position in a tank, the sensor cable is marked in one-
metre intervals from the sensor.
The 715 consists of two main parts, a cable reel and an IR sensor. The cable
reel houses the detector electronics and battery and is also used to store the
sensor cable when not in use. The 715 is supplied with either an IR8, IR15,
IR40 or IR100 sensor as required for the application.
The IR sensors use light attenuation across a single measurement gap to
determine the suspended solids in a solution. The length of the sensing gap
determines the measurement range of the sensor.
Operation
The 715 is ready for operation once the power switch is moved to the ON
position. Before using the 715 to determine the sludge interface position, it
is worthwhile performing the operation test below.
21 Operation Test
The operation of the 715 can be quickly checked. When the unit is switched on
and the sensor gap is clear (sensor in air or in clean water) the unit should
emit a fast bleep and the LED should not be illuminated. If the sensor gap is
then blocked, the unit will switch to a slow bleep and the LED will
illuminate. Unblocking the sensor should return the unit to a fast bleep with
the LED off.
2.2 Sludge Interface Detection
Release the sensor and allow it to hang below the cable reel. Slowly lower the
sensor into the settling tank, listening for a change in the sounder output.
When the sound changes, the sludge blanket has been detected, and the lowest
visible cable marking indicates the position of the sludge blanket below the
supernatant surface.
23 Cleaning
After use the sensor should be cleaned by rinsing with clean water. On a
regular basis the sensor and sensor cable should be cleaned by rinsing with a
solution of water and a mild detergent. The cable reel can be cleaned using a
damp cloth and mild detergent.
Note: The cable reel should not be immersed
2.4 Reeling Technique
When using the 715, the sensor cable should not be allowed to go slack when
reeling the sensor in or out. If the cable is allowed to become slack, it is
possible for the cable to jam inside the cable reel – if this does occur the
cable must be carefully freed — the cable reel must not be forced.
Technical Support
Technical Support is available by phone, fax, or email, the details of which are shown below.
- Phone: +44 (0) 1726 879800
- Fax: +44 (0) 1726 879801
- Email: techsupport@partech.co.uk
- Website: www.partech.co.uk
To enable us to provide quick and accurate technical support please have the following information ready when you contact us:
- Serial Number or original purchase details
- Sensor Type, and Serial Number
- Application details
- Description of fault
3.1 Returning Equipment for Repair
If equipment needs to be returned to Partech for repair or service the
following address should be used:
SERVICE DEPARTMENT
PARTECH INSTRUMENTS
ROCKHILL BUSINESS PARK
HIGHER BUGLE
ST AUSTELL
CORNWALL
PL26 8RA
UNITED KINGDOM
Please include the following information with the returned equipment. Also ensure that sensors are adequately protected for transportation (Advice on packing can be provided by our service department).
- Contact name and phone number
- Return address for equipment
- Description of fault or service required
- Any special safety precautions because of nature of application
Technical Specification
4.1 General
Battery Type: …………………………. IVDC (Alkaline Battery PP3)
Battery Life………………………. 2200 Hours
Operating Temperature Range…….0 to 50°C
Storage Temperature Range……….-20 to +60°C
Environmental Rating…………………. P65
Cable Type……………………….. 4 Core, polyurethane coated
Standard Cable Length……………….70 metres
Maximum Cable Length………………15 metres
Cable Marking………………………………. Every Metre
WIGHT… …………………………………… 1.2 KG
DIMENSIONS. ……………… 200 X 230 X130mmM (hxwxa)
4.2. Sensor
Measuring Principle…………………….Light absorption
Light Source………………………………….Infrared LED, 950nm
Environmental Rating…………………. P68
Material……………………………………. Hastelloy, PVC, Epoxy Potting Compound
Parameter | IR100 | IR40 | IRIS | IR8 |
---|---|---|---|---|
Maximum Supernatant (mg/I)* | 180 | 1350 | 9000 | 27000 |
Minimum Sludge (mg/I)** | 20 | 150 | 1000 | 3000 |
Weight (q) | 350 | 300 | 250 | 250 |
Width (mm) | 160 | 100 | 75 | 75 |
Height (mm) | 95 | 95 | 95 | 95 |
Depth (mm) | 25 | 25 | 25 | 25 |
- The maximum supernatant value is the point at which the 715 will not detect sludge.
The minimum sludge value is the starting point for detection of sludge.
Note:** The values quoted are based on the typical solids found on a municipal sewage treatment works.