4Healthcare 4H860 Chapel Trolley User Manual

June 13, 2024
4Healthcare

User Instruction Manual
4H860 Chapel Trolley 4Healthcare 4H860 Chapel
Trolley

PRODUCT DESCRIPTION

This product has been designed to be a multi-role trolley. It reduces the amount of manual handling required to move a casket and provides a platform for the casket to be viewed in a Chapel.
The High/Low and Casket brake are powered by a 24V electrical system with a rechargeable battery.
The High/Low and Casket braking features are operated via an InfraRed cordless handset and a cable connected handset.
Braking is provided by a foot operated floor brake.
The Chapel Trolley also comes with a two-piece removable skirt which surrounds the trolley.  This provides an aesthetically pleasing look when the trolley is used in the Chapel viewing role.

SPECIFICATIONS

Maximum Load: 250kg
Platform Height
High: 890mm
Low: 490mm
Overall Dimensions (low height setting)
Width: 1210mm
Height: 500mm
Depth: 520mm

OPERATING INSTRUCTIONS

  1. InfraRed handset
    When operating the electronic features with this handset make sure to point it towards the center of the trolley as best as possible. This will insure the trolley operates smoothly. The working range is up to 5m.
    The top two toggle switches operate the functions. Top switch operates the casket brake and the bottom switch operates the High/Low function. See figure 1
    The handset is powered by three ‘AAA’ sized batteries.

  2. Cable connected handset
    This is generally used as a back-up handset if the IR handset has been lost or if the batteries need replacing. This handset is also used when the unit needs to be re- activated after the main battery has been removed for charging. The top switches operate the casket brake and the bottom switches operate the High/Low. See figure 24Healthcare 4H860 Chapel Trolley -
Fig1

  3. High/Low function (Raising & Lowering the platform)
    Raising Platform: Press the UP button until desired height is reached. Lowering Platform: Press the down button until desired height is reached.

  4. Casket brake function
    To apply the brake, press the BRAKE ON button until a two-tone beep is heard. To turn off the brake press the BRAKE OFF button until the casket is lowered onto the rollers and can be moved along the platform.

  5. Floor brake
    The floor brake is positioned in the middle of the trolley, there is one on each side.
    To apply the brake use your foot and push down on the black and yellow pad until it locks down into position. See figure 3.
    To release the floor brake push down on one of the plates either side of the black and yellow pad. See figure 4.4Healthcare 4H860 Chapel Trolley -
Fig2

  6. Removing the main battery for charging.
    The battery needs to be removed from the trolley and placed on a wall charging unit. The battery is located at one end of the trolley, raise the skirt to locate it. Remove one section of the skirt that covers the battery to make it easier for removal. See figure 5.1
    Use one hand to support the battery and the other to locate the retaining clip. Release the clip and slide out the battery. See figure 5.2 & 5.34Healthcare 4H860 Chapel Trolley - Fig3

  7. Charging the main battery
    The trolley is supplied with a battery charger on a mounting bracket. This should be secured in a cool, dry place with access to a 240V power point. Mount the wall charger securely to a wall.
    Itis recommended to charge the battery when the trolley in not in use. The battery can be left on the charger overnight.
    Place the battery onto the wall charging unit and make sure the retainer clip locates onto the bracket. Check that the charger power supply and the charge light is on.
    When the battery is fully charged the charge light will turn green. This process from a flat battery generally takes 4Hrs. See figure 6.1 & 6.24Healthcare 4H860 Chapel Trolley - Fig4

  8. Fitting main battery to trolley
    Hold the battery in the middle with one hand and with the other hand hold the retaining clip end. Slide the battery onto the bracket and pull the retaining clip open, then engage the battery into the box and release the clip. See figure 7.1 & 7.2
    NOTE: THE ELECTRICS MUST BE RE-ACTIVATED BEFORE USE.
    To do this grab the cable connected handset and press the BRAKE ON button once. The IR handset will then be ready to use. See figure 7.34Healthcare
4H860 Chapel Trolley - Fig5

  9. Emergency stop button
    The red button is located on the box next to the main battery. When pressed it will shut down the electronics. To re-activate the system, turn the button a % clockwise and it will spring out. Wait 5 seconds and press any button on the cable connected handset, this will re-activate the system. See figure 8.1 & 8.2
    NOTE: YOU MUST WAIT 5 SECONDS BEFORE PRESSING THE BUTTONS ON THE CABLE CONNECTED HANDEST.4Healthcare 4H860 Chapel Trolley -
Fig6

  10. Fitting skirt to trolley
    The skirt comes in two pieces, find the center of the platform by using the center roller as a guide. Grab one skirt and start in the center and press the Velcro into the frame. See figure 9.1
    Pull the skirt tight as you move it along the frame so there are no creases. Work your way along to the other side. Repeat the same process with the second piece but this time overlap the start with the first one. See figure 9.2, 9.3, 9.4 & 9.54Healthcare 4H860 Chapel Trolley - Fig7

FAULT FINDING

FAULT
IR handset does not operate the electrics: check items below

  • Check that the main battery is fitted and the electrics have been activated as per instructions in section 8.
  • The emergency button has not been pressed in. Rotate button clockwise, if button spring out re-activate as per instructions in section 9.
  • Check that the blue light comes on at the end of the handset when a button is pressed. If no light comes on it indicates that the batteries in the handset are flat and need changing. Three AAA’s required. See image below
  • Check that the switch in the end of the handset is in the correct position. See image below.

Both handsets do not operate the electrics and when the buttons are pressed a continuous beep sound is made.

  • The electrical system needs to be reset. Remove the main battery and then follow the instructions in section 8.

CLEANING

The trolley can be cleaned with a mild detergent using a soft cloth or sponge.
Avoid spraying water directly into the electrical components.
The skirt can be machined washed on a delicate cycle and air dry, do not tumble dry.

MAINTENANCE

Daily

  • Visually check the trolley for any signs of damage.
  • Check that the handset and electrical features work correctly
  • Check castors roll freely
  • Check Floor Lock operates correctly

Weekly

  • Charge battery if not done daily as part of the routine

Annually

  • Check all fasteners, fixings and electrical components

SPARE PARTS

4Healthcare 4H860 Chapel Trolley - PARTS

Parts Supplied; Chapel Trolley IR Handset
Battery Wall Charging Unit Skirt
Revision: IR
Date: 14 August 2017 Distributor
EVOCARE AUSTRALIA PTY LIMITED
A.BN. 98 078 566 604
Trading as EVOCARE and L&M EQUIPMENT
P.0. Box 1144, Stafford Qld. 4053
Ph: 07 33558000 Fax: 07 3355 5043
Website: http://www.evocare.com.au
Emai: [email protected]
[email protected]
[email protected]
[email protected]
6 Welsh Place Lansdale WA 6065 Tel: (08) 9302 4757 Fax: (08) 9302 4524

Website:www.4healthcare.com.au
Email:[email protected]
ABN 51 156 666 161

Documents / Resources

| 4Healthcare 4H860 Chapel Trolley [pdf] User Manual
4H860 Chapel Trolley, 4H860, Chapel Trolley, Trolley
---|---

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>