Hillrom Life2000 Ventilation System
Product Information
The Life2000 System is a ventilator designed for use in home and institutional
settings. It is not intended for ambulance or air transportation.
The system includes a ventilator, compressor, battery charger, belt clip,
external power supply, ventilator carry pouch, and various connections.
Package Content Overview
- The ventilator can be used with the Life2000 Compressor or an alternate 50-PSI pressure source.
- The compressor is an electropneumatic power unit that provides the ventilator with a continuous pressure source and also serves as a charging station.
- The belt clip is used to secure the ventilator when it is used in wearable configurations.
- The battery charger and AC power cord connect the ventilator to an AC power source.
- The external power supply and AC power cord connect the compressor to an AC power source.
- The ventilator carry pouch is an alternative to the belt clip for securing the ventilator when using it in wearable configurations.
System Overview
The Life2000 Ventilator has two released versions. The version can be
identified based on the REF number located on the label on the back of the
ventilator.
The functionality of the Communication Port, Battery Charge Icon, and System
Alarms may differ for each version.
- The ventilator version REF MS-01-0100, released prior to July x, 2019.
- The ventilator version REF MS-01-0118, released after July x, 2019.
- The ventilator features a touch screen, activity buttons, power button, power indicator light, alarm speaker, backup alarm buzzer, breath indicator light, interface connection, gas inlet connection, belt clip sockets, and optional Battery Charger Dongle.
- The compressor features a power button, battery charge status and indicator lights, interface connection, handle, air inlet and filter, cooling filter cover, cooling vents, condensation tray, power supply connection and cover, and internal alarm buzzer.
- The docking cradle for the ventilator includes a power source indicator light, power button for the compressor, locked and unlocked icons for ventilator charging, battery charge status button, and battery charge indicator lights.
Product Usage Instructions
- Before using the ventilation system, ensure that you have read and understood the product manual.
- Identify the REF number of the ventilator to determine its version and ensure proper use of the system.
- Connect the ventilator to the Life2000 Compressor or an alternate 50-PSI pressure source.
- If using wearable configurations, secure the ventilator using either the belt clip or the ventilator carry pouch.
- Connect the battery charger and AC power cord to the ventilator to provide power from an AC power source.
- Connect the external power supply and AC power cord to the compressor to provide power from an AC power source.
- If available, configure the Battery Charger Dongle with the ventilator for charging purposes.
- Use the touch screen and activity buttons on the ventilator to control its functions.
- Monitor the power indicator light, alarm speaker, backup alarm buzzer, breath indicator light, and system alarms for proper operation.
- If necessary, use the power button on the compressor to control its operation.
- Check the battery charge status and indicator lights on both the ventilator and compressor.
For more detailed information, please consult the full product manual
available for download at
https://www.hillrom.com/en/products/life2000-system/ or contact Breathe
Technologies at 877-698-1325 to request
a copy by email or mail.
Before using the Life2000® Ventilation System refer to the Life2000 ®
Ventilation System Instructions for Use for a full list of warnings and
cautions, detailed operating instructions, and additional information. This
guide is not a substitute for the Instructions for Use. A copy of the
Instructions for Use can be downloaded at
https://www.hillrom.com/en/products/life2000-system/
Indications for Use
- The Life2000®Ventilation System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
- The Life2000 Ventilation System consists of the Life2000 Ventilator and the Life2000 Compressor. The System is intended for use by qualified, trained personnel under the direction of a physician. Specifically, the System is applicable for adult patients who require the following types of ventilatory support:
- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask).
- Assist/Control mode of ventilation.
- The System is suitable for use in home and institutional settings and is not intended for ambulance or air transportation.
Safety Information
- Please consult the Patient Instructions for Use for a full list of Tips, Notes, Cautions and Warnings. The Instructions for Use can be downloaded at https://www.hillrom.com/en/products/life2000-system/ or contact Breathe Technologies at 877-698-1325 to request a copy by email or mail.
- The operator of the ventilation system is responsible for reading and understanding the product manual before use.
WARNINGS:
- The Life2000 Ventilation System is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician.
- Off-label use of equipment is prohibited by law.
- Always have an alternate means of ventilation or oxygen therapy available.
- Any modifications to the Life2000 Ventilation System equipment and interfaces is prohibited.
- Breathe Pillows Entrainment Interface and Universal Circuit® connector are for single patient use only.
- Always inspect the interface equipment for cracks, breaks, kinks, or for any particular odor.
- Patient must secure interface equipment properly to avoid breathing distress and possible strangulation.
Package Content Overview
- Life2000® Ventilator (ventilator)
- The ventilator can be used with the Life2000 Compressor or an alternate 50-PSI pressure source.
- NOTE: Ventilators may not be configured with the Battery Charger Dongle.
- Life2000® Compressor (compressor)
- The compressor is an electropneumatic power unit that provides the ventilator with a continuous pressure source and is a charging station for the ventilator
- Belt clip for ventilator
- The belt clip is used to secure the ventilator when it is used in wearable configurations.
- Battery charger and AC power cord for the ventilator
- The battery charger and AC power cord connect the ventilator to an AC power source.
- External power supply and AC power cord for the compressor
- The external power supply and AC power cord connect the compressor to an AC power source.
- Ventilator Carry Pouch
- Alternative to belt clip for securing the ventilatorwhen using in wearable configurations.
System Overview
Life2000® Ventilator Versions
- There are two released ventilator versions of the Life2000 Ventilator. You will be able to identify the version of the ventilator based on the REF number.
- The functionality of the ventilator’s Communication Port, Battery Charge Icon, and System Alarms differ for each version of the venitlator. Please make sure to identify the REF number of the ventilator to ensure propper use of your system.
Identifying the REF number
- The REF number is located on the label on the back of the ventilator, see examples below.
Life2000® Ventilator
- Battery charger connection
- Silence Alarm button
- Ventilator REF MS-01-0100
- For manufacturer’s use only Ventilator REF MS-01-0118 Communication port
- Touch screen
- Activity buttons
- Power button for ventilator
- Power indicator light
- Alarm speaker
- Backup alarm buzzer
- Breath indicator light
- Interface connection
- Gas inlet connection
- Belt clip sockets
- Battery Charger Dongle*
Ventilators may not be configured with the Battery Charger Dongle.*Dongle.
Life2000® Compressor (Dockless)
- Power button for compressor
- Battery charge status and indicator lights
- Interface connection
- Handle
- Air inlet and filter
- Cooling filter cover
- Cooling vents
- Condensation tray
- Power supply connection and cover
- Alarm buzzer (internal)
Life2000® Compressor (with Dock)
- Ventilator docking cradle
- Power source indicator light
- Power button for compressor
- Locked icon (ventilator charging indicator light)
- Locking knob
- Unlocked icon
- Battery charge status button
- Battery charge indicator lights W and battery charge scale
- Interface connection
- Outlet fitting
- NOTE: Docking function is no longer applicable Docking function is no longer applicable..
- Handle
- Air inlet and filter
- Cooling filter cover
- Cooling vents
- Condensation tray
- Power supply connection and cover
- Alarm buzzer (internal)
- Silence Alarm button
Setting Up & Charging Your Battery Life2000® Compressor Setup
- Position the compressor upright on a flat, level surface. Make sure that the cooling vents, cooling filter cover, and air inlet on the back of the compressor are not blocked, and there is sufficient clearance from surrounding objects.
- Protect the compressor from falling. The compressor should not be used adjacent to or stacked with other equipment.
- Open the power supply connection cover.
- Insert the AC power cord connector into the power supply connection on the back of the compressor until it clicks into place.
- Connect the AC power cord to the compressor’s external power supply and cord.
- Connect the pronged end of the compressor’s AC power cord to an AC power source.
- Verify that the green LED indicator on the external power supply lights up to indicate the AC connection.
Charging the Ventilator
- Plug the ventilator AC power cord into the ventilator battery charger.
- Connect the ventilator AC plug into an AC power source.
- The ventilator battery charger indicator light turns on when connected to AC power.
- Connect the battery charger cord to the battery charger dongle on the ventilator. For ventilators configured without battery charger dongle, connect the battery charger cord directly to the battery charger connector located on the ventilator.
NOTE: The ventilator can be used while the battery is charging.
Life2000 Interfaces
The Breathe Pillows Entrainment Interface
- Nasal pillows (patient side)
- Entrainment ports
- Tube fit adjuster (cinch)
- Interface tubing
- Compressor or ventilator connector
- Oxygen tubing connector
Wearing the Breathe Pillows Entrainment Interface
- Place the interface in front of the patient with the curve of the interface towards the patient’s face and the entrainment ports facing up.
- Loop the interface tubing over the ears and position the nasal pillows snugly inside the nostrils. Using the tube fit adjuster (cinch), adjust the tubing length under the chin so the interface is secured that snugly and comfortably
Checking the Breathe Pillows Interface Positioning
The interface is placed correctly when:
- The interface pillows rest snugly inside the nostrils, as shown. The fit is comfortable
- The interface does not make breathing difficult.
- Air does not flow to the eyes, cheeks, or lips.
- Entrainment ports are not obstructed.
- If any one of these conditions is not met, reposition the interface. If problems persist, try a different interface size.
Connecting The Life2000 Universal Circuit® Connector
- The Universal Circuit® connector is used to connect any commercially available non-invasive mask (full face, nasal, or pillows) to a ventilator or compressor.
- The Universal Circuit® connector is only compatible with the Life2000 Ventilation System. When connecting the Universal Circuit® Connector, and oxygen therapy is required a third party Oxygen Adapter and Oxygen tubing will need to be connected.
WARNING:
- Interfaces are designed for single-patient use. To prevent risk of cross-contamination use a new Universal Circuit® connector for each new patient. For the third-party mask or tube, refer to the user guide provided by the manufacturer.
- The interface, source gas supply hose, and power cords should be positioned to avoid restricting movement, causing a tripping hazard, or posing a strangulation risk.
CAUTION:
- Hillrom recommends a 90-day replacement schedule for the Universal Circuit® connector.
- Compressor or ventilator connector
- Interface tubing
- Entrainment ports
- Universal Circuit® Connector (patient side)
- Sense Tube
- Oxygen adapter (third party)
- Barb connection
- Oxygen tubing (third party)
NOTE: The interface assembly is packaged clean but not sterile. The Universal Circuit® connector does not need to be cleaned or sterilized prior to first use.
System Setup
Extended Range Configuration
- Ensure the ventilator is powered off.
- Attach the outlet fitting connector on the CombO2® hose to the outlet fitting on the compressor.
- Connect the CombO2® hose oxygen concentrator tubing to the oxygen concentrator (if available).
- Connect the Breathe Pillows Entrainment Interface to the Life2000® Ventilator.
- Connect the quick connect end on the CombO2® hose to the ventilator by pushing the small quick connect end onto the gas inlet connection on the ventilator; when connected, the quick connect end will click into place.
- Attach the CombO2® oxygen tubing to the Breathe Pillows Entrainment Interface.
WARNING:
The interface, source gas supply hose, and power cords should be positioned to avoid restricting movement, causing a tripping hazard, or posing a strangulation risk.
CAUTION:
Use only a Hillrom source gas supply hose with the ventilation system. If an unauthorized source gas supply hose is used with the ventilation system, the system may be damaged.
Stand Alone Configuration
- Attach the oxygen regulator (per the manufacturer’s instructions for use) onto the oxygen cylinder.
- Connect the oxygen hose to the Diameter Index Safety System (DISS) connector end of the regulator by turning it clockwise.
- Connect the quick connect end on the oxygen hose to the ventilator by pushing the small quick connect end onto the gas inlet connection on the ventilator; when connected, the quick connect end will click into place.
- Connect the Breathe Pillows Interface to the Life2000® Ventilator.
Starting Therapy
Activity Buttons
- Each Activity Icon on a button in the Prescription Settings screen represents a prescription that can be programmed by a clinician and made available to the patient.
- Each button on the Prescription Settings screen corresponds to an Activity Button on the ventilator.
- When the ventilator is first powered on, you must select an Activity Button before a breath will be delivered.
- One, two, or three Activity buttons may be available, as directed by a physician. Choose an Activity Button appropriate for your activity.
- Ensure that the ventilator is powered on.
- Ensure a pressure source (the compressor, an oxygen cylinder, or an oxygen wall source) is connected to the ventilator and turned on.
- Press and hold an Activity Button until you hear a tone that indicates it is active. The touch screen will display the Home Screen and the ventilator will begin delivering therapy.
- Confirm the selected Activity Icon is displayed at the bottom of the touch screen and the icon and volume are displayed at the top of the screen. The ventilator will begin ventilating using the chosen prescription parameters for the next breath.
Home Screen & Menu Overview
Home Screen
- When the ventilator is powered on, it completes a self test and then displays the Home Screen. This screen indicates that the ventilator is ready for use.
- When an Activity Button is selected on the ventilator, the Home Screen will display the breath rate per minute (Breath/min or BPM), Peak Inspiratory Pressure in centimeters of water pressure (PIP cmH2O), and gas flow rate in liters per minute (Air LPM or O2 LPM).
- The Wrench Button is used to access the Menu screen
- The current Activity Icon and Output Volume (displayed on the Home Screen during ventilation)
- The Flip Button flips the screen 180°.
- Peak Inspiratory Pressure (PIP cmH2O) indicator (displayed on the Home Screen during ventilation)
- Current breath rate (Breath/min or BPM) (displayed on 1 2 3 Home Screen during ventilation)
- Average gas flow in liters per minute (Air LPM or O2 LPM) based on prescription and patient’s current breath rate (displayed on the Home Screen during ventilation).
- Battery Charge Icon
- Current Activity Icon (displayed during ventilation)
- The Vibration Icon indicates that the ventilator is set for vibration.
- Time and date
NOTE: The Home Screen will initially display LPM when the ventilator is first powered on. After an Activity Button is selected and ventilation begins, the Home Screen will display Air LPM or O2 LPM based on the option prescribed and selected.
Menu Screen
Use the Menu screen to access the Settings screen or Information screen
(including software version and total operation time). To get to the Menu
screen, touch the Wrench button from any screen.
- Screen title.
- Touch to go to the Home Screen
- Touch to go to the Settings Screen for Trigger Sensitivity, Clinician’s Settings, and Utilities.
- Touch to go to the Information Screen.
NOTE: To toggle between the two screens, touch the Wrench button for the Menu Screen and the Home Screen button for the Home Screen.
Battery Information
If you intend to use the ventilator on battery power, make sure the battery is sufficiently charged for your duration of use. It takes approximately 3 to 4 hours to fully recharge the battery whether the ventilation is off or on. A fully charged battery should last up to four (4) hours.
- Press the Power button to turn on the ventilator.
- The green power indicator light shows that the ventilator is on.
- Check the Ventilator Battery Charge Icon on the touch screen to see the current battery charge level for the ventilator. Refer to the chart below to determine the appro ximate amount of ventilator battery charge and the approximate time remaining.
NOTE: The Ventilator Battery Charge Icon is always displayed on the touch screen during normal operation.
- Ventilators may not be configured with the Battery Charger Dongle.*Dongle.
- Ventilator Battery Charge Icons for Ventilator REF MS-01-0100
- Ventilator Battery Charge Icons for Ventilator REF MS-01-0118
- The charging icon may still appear when the ventilator is 100% charged.
- Very low battery alarm will sound with less than 15% charge.
- Low battery alarm will sound with less than 25% charge.
- It may take up to 20 seconds, for the charging sign to appear/disappear after the charger is connected/disconnected from the ventilator.
Cleaning & Maintenance
Cleaning
Life2000® Ventilation System
- Once a week (or more often if necessary): clean and disinfect the ventilation system using a clean cloth, clean and disinfect the external surfaces of the ventilation system with 70% isopropyl alcohol or Clorox Disposable Wipes. Wipe the surface of the ventilation system with clean dry cloth to remove any residual cleaner.
- Daily (or more often if necessary): check and empty the compressor’s condensation tray and replace if necessary.
- Every three to six months (or more often if necessary): check the compressor’s air inlet filter and replace if necessary.
- Every three to six months (or more often if necessary): check the compressor’s cooling filter assembly and replace if necessary.
CombO2® Hose
- If dirt is visible on the outside of the hose, use a clean cloth and mild detergent such as dish-washing soap to remove it. Replace every six months. WARNING:
- Do not subject Breathe source gas supply hoses to heat sterilization, hot water pasteurization, autoclaving, radiation sterilization, ethylene oxide gas sterilization, or attempt to clean them in a dishwasher or microwave oven. Doing any of these may damage the hoses and impair gas delivery.
Breathe Pillows Entrainment Interface
- Breathe Technologies recommends a 90-day replacement schedule for the Breathe Pillows Entrainment Interface. Once a week (or more often if necessary): Clean and purge the Breathe Pillows Entrainment Interface.
Preventive Maintenance
- Contact your service representative to make arrangements for preventive maintenance, service, and component replacement per the chart below. The ventilation system can only be serviced or repaired by an authorized service center.
REVISION | VENTILATOR | COMPRESSOR |
---|---|---|
Before Revision C | 2.5 Years from ship date | 1 Year from ship date |
Revision C and after | 2.5 Years from ship date | 2.5 Years from ship date |
Environmental Specifications
- Do not use the ventilation system if the ambient temperature is greater than 40°C (104°F) or less than 5°C (41°F). Store the ventilation system in ambient temperatures less than 60°C (140°F) and greater than -20°C (-4°F).
WARNING:
- The rear housing of the ventilator enclosure may reach 49°C in a 40°C environment.
Cleaning the Breathe Pillows Entrainment Interface & Universal Circuit® Connector
- Place the patient on an alternate means of ventilation, if necessary.
- With the Interface fully disconnected from the ventilator, submerge the patient side of the Interface or the Universal Circuit® connector in a clean container of mixed warm water suitable for drinking and a mild detergent (e.g., dish-washing soap) and agitate the patient side of the interface to clean it.
- Rinse the patient side of the interface thoroughly with warm water.
Purging Interfaces using the Compressor
After cleaning and completely drying the interface or when you suspect dust or debris has entered the airflow passage, purge the interface with the purge tube connector and purge tube.
- Connect the Purge Tube Connector to the gas inlet.
- Connect the Purge Tube to the Purge Tube Connector.
- Turn on the Compressor
- Proceed to purge each port on the Interface
Alarms and Troubleshooting
Ventilator Alarms
When an alarm notification occurs, there is a distinct sound and a display message corresponding to the priority level of the alarm. The priority level of an alarm is indicated by the color and the rate at which the message flashes.
High-Priority Alarm
- A red, rapidly flashing alarm message is an alarm that indicates a situation that requires immediate attention.
- Sound: Sequence of two sets of five tones.
Medium-Priority Alarm
- A yellow, steadily flashing alarm message is an alarm that indicates a potentially hazardous situation that must be resolved in a timely manner.
- Sound: Sequence of three tones.
Low-Priority Alarm
- A blue, non-flashing alarm message is an alarm that indicates a problem that is not hazardous but should be resolved.
- Sound: Single tone.
- Please see next page for assistance in solving ventilator alarms.
- NOTE: When attempting to resolve alarm conditions, ensure that the patient receives adequate ventilation therapy; place the patient on an alternate means of ventilation if necessary.
How to Silence and Clear On-Screen Alarms
- Alarm notifications that appear on the touch screen originate from the ventilator.
- Silencing and cleaning on-screen alarms depends on alarm priority and number of active alarms.
- Press the Silence Alarm button to temporarily silence the on-screen alarm for 60 seconds. Pressing the Silence Alarm button silences only one alarm at a time. If more than one on-screen alarm occurs, press the Silence Alarm button once for each alarm. If the on-screen alarm is a medium or high priority alarm and is not silenced after 60 seconds, the alarm will continue with an additional buzzer.
- A visible message will be displayed for High Priority alarms (High Temperature, High Circuit Pressure, and High PEEP alarms) When the alarm is resolved, touch Ok in message box to clear the message.
Ventilator Alarms
Ventilator Alarms | What to do about it |
---|---|
High Circuit Pressure | Check the interface tubing for pinches and kinks. |
Replace the interface if alarm persists.
High PEEP Pressure| Interface may be blocked. Inspect and clean the interface
per IFU.
High PIP Pressure| Check the interface and tubing and connectors for possible
obstruction or damage.
High Temperature| Is the ventilator well aired & away from heat source? Call
your provider if alarm persists.
Very Low Battery| Battery charge is below 15%. Recharge the battery. Call your
provider if alarm persists.
Battery Low| Battery charge is below 25%. Recharge the battery. Call your
provider if alarm persists.
Breath Timeout| Check if patient is breathing through the nose. Ensure
interface is not leaking or clogged.
High Breath Rate| Check patient’s breathing rate. Check for leakage or
occlusion. Adjust trigger sensitivity.
High Delivery Pressure| Check the interface tubing for pinches and kinks.
Replace the interface if alarm persists.
High Gas Pressure| Check gas source pressure. Discontinue ventilator use; call
your provider if alarm persists.
Low Breath Rate| Check if patient is breathing through the nose. Ensure
interface is not leaking or clogged.
Low Delivery Pressure| Check interface connection. Check interface tubing for
possible leakage or damage.
Low Gas Pressure| Check gas supply and pressure. Check connectioins for leak.
Check tubing for kinks.
Low PIP Pressure| Check for leakage. Switch activity button. Call your
provider if alarm persists.
System Fault| Reboot the ventilator. If fault persists, use alternate
ventilation & call your provider.
POST System Fault| Restart the ventilator. If fault persists, use alternate
ventilation & call your provider.
Compressor Alerts
The compressor has alerts that are independent of the ventilator. Compressor alerts must be resolved in order for the compressor alerts to be silenced; there is no Silence Alarm button for alerts originating from the compressor. See below for compressor alerts and their resolutions.
Troubleshooting – Compressor Alerts
Low Battery Alert (Intermittent buzzer)| Connect the compressor to an AC power source.
If the battery does not recharge, place the patient on an alternate means of ventilation (if necessary) and contact your healthcare provider.
Constant Audible Alert (Compressor stops operating)| Discontinue use of the compressor. Place the patient on an alternate means of ventilation (if necessary), or connect the ventilator to an alternate pressure source (oxygen cylinder or wall source) in Stand Alone Configuration.
If running on battery, check the compressor’s battery charge status. If the status is less than two indicator lights, connect the compressor to an AC power source.
Power off the compressor and power it on again. If the alert persists, power off the compressor and contact your healthcare provider.
- NOTE: When attempting to resolve alert conditions, ensure that the patient receives adequate ventilation therapy; place the patient on an alternate means of ventilation if necessary.
- NOTE: For additional troubleshooting tips, refer to the IFU online or contact your provider.
Patient Setup Checklist
Setup By (Company Name):
Company Phone Number:
Clinician Name:
Setup Date:
Required for the Life2000 Ventilation System Patient Setup
- Life2000® Ventilator
- Life2000® Compressor
- Breathe Pillows Entrainment Interface
- Ventilator Battery Charger
- Compressor Battey
- Charger CombO2® 50 ft. Hose
- 6 ft. Oxygen Hose
- Ventilator Carry Case
- Ventilator Belt Clip
- Purge Tube
- Purge Tube Connector
- Diss Regulator (*not provided) Quick Reference
- Technical Support and Customer Service
- In the USA, call Hill-Rom at 800-426-4224.
- Outside of the USA, contact your distributor, local Hill-Rom representative, or go to respiratorycare.hill-rom.com.
- 80029589, April 2023
- ©2023 Breathe Technologies, A Hillrom Company. Breathe, Breathe Pillows Interface, Breathe Technologies, Life2000, and Universal Circuit are trademarks of Breathe Technologies, a HIllrom Company. All Rights Reserved.