EXTECH HD500 Heavy Duty Psychrometer Plus IR Thermometer User Guide
- June 13, 2024
- EXTECH
Table of Contents
USER GUIDE
Heavy Duty Psychrometer + IR Thermometer **Model HD500
Introduction
Congratulations on your purchase of the Extech HD500 Psychrometer. This handheld meter measures and displays Air Temperature, Relative Humidity, Dew Point, Wet Bulb and also Surface Temperature using the built-in IR thermometer. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Features
- Triple digital LCD display.
- Fast response, all data is calculated four times per second.
- Standard type K (NiCr-NiAl) Thermocouple input jack suitable for any style of type K probe.
- Infrared thermometer to measure surface temperature.
- Red laser pointer included.
- LCD with Backlight
- Automatic range selection
- USB interface.
- Low battery indication.
- Auto Power off.
Safety
- Use extreme caution when the laser pointer beam is on
- Do not point the beam toward anyone’s eye or allow the beam to strike the eye from a reflective surface
- Do not use the laser near explosive gases or in other potentially explosive areas
Meter Description
- Humidity & Air Temperature Sensor
- Type K Thermocouple input jack
- Laser pointer beam
- IR temperature sensor
- USB Interface
- LCD Display
- Upper display HOLD button
- Upper display Temp./Wet Bulb/Dew Point button
- Upper display record Max/Min button
- IR Measurement button
- °F/°C units button
- Lower display HOLD button
- Lower display record Max/Min button
- Backlight button
- Power button
NOTE: Battery Compartment, Tilt Stand and Tripod Mount are located on the reverse side of the meter
Display Description
- Low Battery icon
- PC communication icon
- Upper display function icons
- Upper display
- Laser pointer icon
- Lower display function icons
- Lower display
- Lower display function icons
Operation
Basic Measurements
-
Press the button to turn power on.
-
The upper display will indicate Air Temperature, Wet Bulb Temperature or Dew Point Temperature and % Relative Humidity.
Press the WB/DP button to toggle between Air, WB or DP. -
The lower display will indicate Type K temperature or IR Temperature.
Press and hold the IRT button to select the IR Thermometer.
The TYPE-K function will display “ “ if a type k probe is not inserted into the meter. -
Press the °C/°F button to change the temperature units from °C or °F.
Non-contact IR Surface Temperature Measurements
The built-in IR sensor can remotely measure the temperature of most surfaces.
The Laser pointer allows the user to aim accurately when taking non-contact
measurements.
- Turn ON the meter using the button.
- The IR sensor and laser pointer are located at the top of the meter.
- Point the sensor toward the surface to be measured.
- Press and hold the IRT button to begin measuring the surface temperature of a desired target. IR TEMP and will appear on the display. The laser pointer will switch on to help aim the meter.
- The measured IR surface temperature will appear on the lower display.
- When the IRT button is released, the laser pointer will switch off and the reading will freeze (data hold) on the display for approximately 7 seconds.
- After the 7 second hold time the meter returns to the type k mode.
WARNING: Do not directly view or direct the laser pointer at an eye. Low power visible lasers do not normally present a hazard, but may present some potential for hazard if viewed directly for extended periods of time.
Data Hold
- Press the HOLD buttons (one for upper display and one for lower display) to freeze the displayed value for the respective display. Press again to unlock the display.
- The HOLD icon will appear on the display when the Data Hold mode is active.
MIN-MAX Recording Mode
- Press the MAX/MIN button (one for upper display and one for lower display) to begin recording the Maximum and Minimum reading. The REC MAX icon will appear and only the maximum value measured will appear in the display. The display will update only if a value higher than the currently displayed value is measured.
- Press the MAX/MIN button again to display the minimum values. The REC MIN icon will appear and only the minimum recorded value will appear in the display.
- Press the MAX/MIN button again to display the currently measured values. The REC icon will appear in the display and the Max and Min values will be stored in memory.
- Press and Hold the MAX/MIN button for >2 seconds to exit the mode.
Backlight
Press the backlight button to turn the backlight ON or OFF.
Note: continuous use of the backlight function will reduce battery life.
Auto-Power Off
The meter will automatically turn off after 15 minutes of operation if no
buttons are pressed during this period. Auto-power off can be disabled by:
- Hold the IRT button and then press the button to turn the power on. When “disAPO” appears in the display, release the IRT button and the APO is disabled.
Low Battery
When the battery reaches the minimum operating voltage the battery icon will
appear in the display.
Replace the 9V battery when this happens.
Battery Replacement
When the battery icon appears on the LCD, the 9V battery must be replaced.
- The battery compartment is located on the rear of the meter.
- Press in and down on the arrow located above the tilt stand hinge.
- Replace the 9V battery
- Replace the battery cover.
** Never dispose of used batteries or rechargeable batteries in household
waste.
As consumers, users are legally required to take used batteries to appropriate
collection sites, the retail store where the batteries were purchased, or
wherever batteries are sold.
Disposal:** Do not dispose of this instrument in household waste. The user
is obligated to take end-oflife devices to a designated collection point for
the disposal of electrical and electronic equipment.
Other Battery Safety Reminders
-
* Never dispose of batteries in a fire. Batteries may explode or leak.
Never mix battery types. Always install new batteries of the same type
USB Interface and Software
The HD500 is equipped with a communication jack on its upper left side. The
supplied communications cable connects to this jack and to a USB port on a PC.
The supplied software allows the user to view and save readings to the PC.
Instructions for use and features are detailed in the supplied software HELP
utility.
InfraRed Measurement Considerations
- When taking IR measurements the meter automatically compensates for ambient temperature changes. Note that it may take up to 30 minutes to adjust to extremely wide ambient changes.
- Low temperature measurements quickly followed by high temperature measurements may require several minutes to stabilize as a result of the IR sensor cooling process.
- If the surface of the object under test is covered with frost, oil, grime, etc., clean before taking measurements.
- If an object’s surface is highly reflective apply masking tape or flat black paint before measuring.
- Steam, dust, smoke, etc. can obstruct measurements.
- To find a hot spot, aim the meter outside the area of interest then scan across (in an up and down motion) until the hot spot is located.
- IR measurements cannot be made through glass.
IR Theory
IR thermometers measure the surface temperature of an object. The meter’s
optics sense emitted, reflected, & transmitted energy that is collected and
focused onto the meter’s detector. The meter’s circuitry translates this
information into an LCD reading.
IR Field of View
Ensure that the desired target is larger than the spot size. As the distance
from an object increases, the spot size of the area measured by the meter
becomes larger. The meter’s field of view ratio is 30:1, meaning that if the
meter is 30 inches (cm) from the target, the diameter (spot) of the object
under test must be at least 1 inch (cm). Refer below to the field of view
diagram. Emissivity
Most organic materials and painted or oxidized surfaces have an emissivity of
0.95. Inaccurate readings will result when measuring shiny or polished
surfaces. To compensate, cover the surface under test with masking tape or
flat black paint. Allow time for the tape to reach the same temperature as
the material underneath then measure the temperature of the tape or the
painted surface.
Thermal Emissivity Table for Common Materials
Material | Emissivity | Material | Emissivity |
---|---|---|---|
Asphalt | 0.90 to 0.98 | Cloth (black) | 0.98 |
Concrete | 0.94 | Human skin | 0.98 |
Cement | 0.96 | Leather | 0.75 to 0.80 |
Sand | 0.90 | Charcoal (powder) | 0.96 |
Earth | 0.92 to 0.96 | Lacquer | 0.80 to 0.95 |
Water | 0.67 | Lacquer (matt) | 0.97 |
Ice | 0.96 to 0.98 | Rubber (black) | 0.94 |
Snow | 0.83 | Plastic | 0.85 to 0.95 |
Glass | 0.85 to 1.00 | Timber | 0.90 |
Ceramic | 0.90 to 0.94 | Paper | 0.70 to 0.94 |
Marble | 0.94 | Chromium oxides | 0.81 |
Plaster | 0.80 to 0.90 | Copper Oxides | 0.78 |
Mortar | 0.89 to 0.91 | Iron Oxides | 0.78 to 0.82 |
Brick | 0.93 to 0.96 | Textiles | 0.90 |
Specifications
General Specifications
Display | Dual Display Multi-function LCD with 9999 counts |
---|---|
Data Hold | Freezes displayed reading |
Sampling rate | 1 reading per second |
Sensors | Relative Humidity: Capacitance, Air Temp: Thermistor |
IR Distance to Spot ratio | 30:01:00 |
IR Spectral response | 6 to 14µm |
IR Emissivity | 0.95 fixed |
MIN-MAX | Record and Recall lowest and highest readings |
Auto Power OFF | Automatic shut off after 15 minutes (can be disabled) |
PC Interface | USB PC Communication with supplied software and cable for data |
acquisition
Over range indication| Dashes appears on the LCD
Low battery indication| Battery symbol appears on the LCD
Power supply| 9V Battery
Operating conditions| Meter: 0 to 50 o C (32 to 122 F); 80% RH max.
Dimensions / Weight| Main instrument: 257 x 76 x 53mm (10.1 x 3.0 x 2″)
Weight| 355g (12.5 oz.)
Range Specifications
Function | Range | Resolution | Accuracy |
---|---|---|---|
Temp (type-K) | -148°F to -20°F | 1°1000 | |
0.1°<1000 | ±(3.0% reading + 4°F) | ||
-20°F to 2501°F | ±(3.0% reading + 2°F) | ||
-100°C to -30°C | ±(3.0% reading + 2°C) | ||
-30°C to 1372°C | ±(3.0% reading + 1°C) | ||
IR Temp | -58 to -4°F -50 to -20°C | 0.1°F/°C | ±9.0°F / 5.0°C |
-4 to 932°F -20 to 500°C | 0.1°F/°C | ±2% reading or ±4°F/2°C | |
Air Temp. | -4 to 140°F -20 to 60°C | 0.1°F/°C | ±(2% reading + 2°F/1°C) |
%RH | 10% to 90% | 0.1%RH | ±2°/0 RH |
<10% and >90% | 0.1%RH | ±3% RH | |
Wet Bulb | -6.88 to 140°F -21.6 to 60°C | 0.1°F/°C | calculated |
Dew Point | -90.4 to 140°F -68 to 60°C | 0.1°F/°C | calculated |
Copyright © 2013‐2015 FLIR Systems, Inc.
All rights reserved including the right of reproduction in whole or in
part in any form
ISO‐9001 Certified
www.extech.com
HD500-en-GB_v3.4 4/15