EXTECH EA10 EasyView Dual K Thermometer User Guide

June 13, 2024
EXTECH

EXTECH EA10 EasyView Dual K Thermometer User Guide
EXTECH EA10 EasyView Dual K Thermometer

Introduction

Congratulations on your purchase of the EdTech Eastview Dual K Thermometer. This device offers dual thermocouple inputs with multifunction LCD display. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.

Meter Description

  1. LCD display
  2. Data HOLD button
  3. Power button
  4. Max/Min button
  5. OFFSET button
  6. Backlite button
  7. C/F/K units select button
  8. Elapsed TIME button
  9. T1 or T1‐T2 or T2 display select button
  10.  T1 Thermocouple input socket
  11.  T2 Thermocouple input socket
  12. Battery compartment (rear)
    Meter Description

Operation

Connecting thermocouples

  1. This meter accepts two type K thermocouples with spade plugs (sub‐miniature type with one spade wider than the other).
  2. Plug the thermocouple(s) into the meter’s T1 and/or T2 thermocouple input jack(s).

Turn Power ON

  1. Press the button to turn power on. The meter will perform a short self‐test.
  2. If a probe is not inserted to the meter, an “OL” indication will appear on the display.

Selecting °C, °F or °K units of measure
Press the C/F/K button to select the desired temperature unit of measure.

Selecting T1, T2, and T1‐T2 Displays
Press the T1 T2 T1‐T2 button to step through and select a display of:

  • a. T1 (thermocouple 1) displays in the upper large display and T2 (thermocouple 2) temp. displays in the lower small display.
  • b. T1 plus T2
  • c. T1‐T2 plus T1
  • d. T1‐T2 plus T2

MIN, MAX, AVG Recording Function with Timer

Press the MX/MN button to begin capturing the Minimum (MN), Maximum (MX), and Average (AVG) temperature values. The Elapsed Time clock will appear on the lower left of the display and the lower display will indicate the MAXIMUM (MX) value captured since the MX/MN button was pressed.

Press the MX/MN button to change the lower display to reflect the stored MX to MN and then AVG value of the input selected in the upper display (T1, T2 or T1‐T2). Each reading will be accompanied by a Timer value indicating the time when each reading was recorded. Press the Time button to change the elapsed time format from min:sec to hour:min. Press and Hold the MX/MN button until the display returns to standard operation to exit the MX/MN function.

Data Hold
Press the HOLD button to freeze the reading in the display. The ‘H’ hold icon will appear on the upper left‐hand side of the display. Press the HOLD button again to return to normal operation (the ‘H’ hold icon will disappear).

Offset
The Offset Mode permits the user to store the temperature reading as a reference value and compare this reference value to subsequent temperature readings. In Offset mode, the meter displays the actual measured temperature minus the stored reference value.

Press the OFFSET button to store the currently displayed temperature readings and enter the Offset mode. “OFFSET” will appear in the display. Press the OFFSET button to exit the mode

Backlight
Press the backlight button to turn on the LCD display backlight. The backlight will automatically turn off after one minute or press the button again to turn the light off.

Auto Power OFF
The meter will automatically turn off after 30 minutes if no keys are pressed during this period. Auto power off is disabled in the max/min recording mode.

Turning Power OFF
Press the button to turn power off

Specifications

Measurement Range 200 to 1999oF (‐200 to 1360oC)
Resolution 0.1oC/ oF <200o , 1 oC/ oF >200o
Accuracy   (@23 ± 5 oC) ±(0.3% reading + 2 oF) @ ‐200oF to 1831.9oF

±(0.5% reading + 2 oF) @ 1832oF to 1999oF
±(0.3% reading + 5 oC) @ ‐200oC to ‐93.1oC
±(0.3% reading + 1 oC) @ ‐93oC to 999.9oC
±(0.5% reading + 1 oC) @ 1000oC to 1360oC
Temperature Coefficient| 0.1 times the applicable accuracy specifications per oC from 0 oC to 18oC and 28oC to 50oC (32oF to 64oF and 82oF to 122oF)

Note: Temperature accuracy does not include the accuracy of the type K probe.

Note : The temperature scale is based on the international temperature scale of 1990 (ITS90).

General Specifications

Display s Display    Dual Display Multi‐function LCD with Backlight
Input Protection 60VDC; 24VAC rms
Display Update Rate 1 per second
Over range indication “OL” appears on the LCD
Open input indication “OL” appears on the LCD
Low battery indication “BT” appears on the LCD
Power supply 6 AAA Batteries
Battery life Approximately 200 hours with carbon zinc batteries
Operating Temperature 32 to 122oF (0 to 50oC)
Operating Humidity < 80% RH
Storage Temperature 14 to 140oF (‐10 to 60oC)
Storage Humidity 10 to 80% RH
Dimensions 5.91×2.8×1.4″ (150x72x35mm)
Weight Approx. 8.29 oz. (235g) with battery

Warranty

FLIR Systems, Inc. warrants this Extech Instruments brand device to be free of defects in parts and workmanship for three years from date of shipment (a six month limited warranty applies to sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department for authorization. Visit the website www.extech.com for contact information. A Return Authorization (RA) number must be issued before any product is returned. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. FLIR Systems, Inc. specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. FLIR’s total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied.

Calibration, Repair, and Customer Care Services

FLIR Systems, Inc. offers repair and calibration services for the Extech Instruments products we sell. We offer NIST traceable calibration for most of our products. Contact our Customer Service Department for information on calibration service availability. Annual calibrations should be performed to verify meter performance and accuracy. Technical support and general customer service is also provided, refer to the contact information provided below.

Support Lines: U.S. (877) 439‐8324; International: +1 (603) 324‐7800
Technical Support: Option 3; E‐mail: support@extech.com
Repair & Returns: Option 4; E‐mail: repair@extech.com
Product specifications are subject to change without notice
Please visit our website for the most up‐to‐date information
www.extech.com
FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA
ISO 9001 Certified
Copyright © 2014‐2016 FLIR Systems, Inc.

All rights reserved including the right of reproduction in whole or in part in any form
www.extech.com 

EXTECH Logo

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals