ROMER 10007055 AgraStrip Pro Allergen Test Kit User Guide
- June 13, 2024
- ROMER
Table of Contents
AgraStrip® Pro Allergen Quick Guide
AgraStrip® Pro Allergen LFD Kits include:
- 15 AgraStrip® Pro lateral flow devices
- 1 bottle of ready-to-use extraction buffer
- 15 extraction tubes and caps
- 15 incubation vials
- 15 swabs with break-off tips
- 1 rack for incubation vials
Test procedures for:
- Raw materials/processed food
- Rinse waters
- Swab samples
Available kits:
- AgraStrip® Pro Almond 10007055
- AgraStrip® Pro -Lactoglobulin 10007056
- AgraStrip® Pro Brazil nut 10007057
- AgraStrip® Pro Casein 10007058
- AgraStrip® Pro Cashew/Pistachio 10007059
- AgraStrip® Pro Coconut 10007060
- AgraStrip® Pro Crustacea 10007061
- AgraStrip® Pro Gluten* 10007062
- AgraStrip® Pro Hazelnut 10007063
- AgraStrip® Pro Lupin 10007064
- AgraStrip® Pro Macademia 10007065
- AgraStrip® Pro Milk 10007066
- AgraStrip® Mustard 10007067
- AgraStrip® Pro Peanut 10007068
- AgraStrip® Pro Sesame 10007069
- AgraStrip® Pro Soy 10007070
- AgraStrip® Pro Walnut 10007071
- AgraStrip® Pro Whole Egg 10007072
Also Available:
-
AgraQuant® Allergen ELISA Test kits
-
AgraQuant® Plus Allergen ELISA Test kits
-
AgraStrip® Pro Gluten requires an additional dilution step prior to the transfer of the extract into the incubation vial (step 3). Please, refer to the product manual.
Making the world’s Food Safer®
FMCG Industry Solutions Pty Ltd
ABN 22 135 446 007
Factory 11A, 1 – 3 Endeavour Rd, Caringbah NSW 2229
www.fmcgis.com.au
sales@fmcgis.com.au
1300 628 104 or 02 9540 2288
Procedure for liquid and solid samples
Add 0.2 mL liquid sample (or rinse water), or weigh in 0.2 g of homogenized sample into the extraction tube. Fill with extraction buffer to the level shown in the picture.
Close the tube with the cap and vigorously shake by hand for 1 minute. You
can squeeze the tube to help suspend the sample.
Procedure for environmental swabs
Fill one extraction tube with extraction buffer up to the neck of the tube
(see step 1 for solid/liquid samples). Wet a swab by dipping it into the
buffer. Wipe an area of 5 x 5 cm using side-to-side movements, rotating the
swab tip as you go.
Place the swab into the extraction tube and carefully break off the end at the pre-scored point. Close the tube with a cap and shake vigorously for 1 minute.
Remove the smaller cap to open the dropper tip and transfer 6 drops (200
μL) into an incubation vial. Close the lid of the vial.
Shake the incubation vial vigorously by hand for 15 seconds. Make sure
that the liquid comes into contact with the lid of the vial.
Tap the base of the vial onto a surface to ensure that no liquid is remaining
in the lid. Let it incubate on the rack at room temperature for 5 minutes.
Place one test strip vertically (arrows pointing downwards) into the
incubation vial. Incubate for 5 minutes. Read the result immediately.
Result Interpretation
Negative Result:
Only the control band (C) appears in the results area of the test strip
Positive Result:
The control band (C) and the test band (T) appear in the results area of the
test strip. The sample could contain the target allergen in a concentration
higher than the cut-off value and further investigations should be performed
(e.g. using AgraQuant® Allergen ELISA kits).
Invalid Result:
No control band appears. Regardless of whether the test line appears, please
repeat the test with a fresh strip.
- Cut-off values must be validated for each matrix.