IDMED ToFscan NeuroMuscular Transmission Monitor User Manual
- June 13, 2024
- idmed
Table of Contents
- About this product
- Intended Use / Indication for use
- Expected technical performance
- Clinical performance
- Clinical benefits
- Important Information about the use of this device
- SAFETY MEASURES
- Explanation of the symbols
- I General information
- II ToFscan setup
- III Using the ToFscan
- IV Servicing, Disinfection cleaning
- V End-of-life disposal / Recycling
- VI Technical specifications and warranty
- VII Accessories
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
IDMED ToFscan NeuroMuscular Transmission Monitor User Manual
About this product
This operating manual provides the instructions on how to configure and use the ToFscan from IDMED. It also describes the specific procedures to clean and verify the device as part of the necessary maintenance. This manual is intended to be used only by qualified medical personnel.
Keep this operating manual with the ToFscan. A service manual is available for the technicians in charge of maintenance.
Before you start, please make sure you read carefully and understand the safety information contained in this manual.
Intended Use / Indication for use
The ToFscan is a neuromuscular transmission monitor for monitoring the neuromuscular block of a patient in the operating theatre, recovery room or intensive care unit. The effect of neuromuscular blocking agents (NMBAs) is monitored by measuring the acceleration of the muscle movement (acceleromyography) or by visually observing muscle contractions consequent to electrical stimulation. The ToFscan has a three-dimensional acceleration sensor (accelerometer) to detect and quantify a patient’s muscle movement. For the thumb (contracting adductor pollicis), the sensor is directly integrated into the finger’s splint, making it possible to obtain its optimal and reproducible positioning. For the eyebrow and big toe sensors, a correct sensor position allows for optimal and reproducible measurement.
Expected technical performance
The following features are essential performance of the device:
- Stimulation of an anesthetized patient with electrical pulse: repeated or single 200µs square pulse, 20mA to 60mA intensity. All values within +/- 10%
- Allow several stimulation sequences in accordance to usual practice, TOF, PTC, ATP, DBS, TET, ST.
- Measure the muscle response of the thumb, big toe of the foot or the corrugator muscle of the eyebrow following electrical stimulation (TOF). The result of this measurement is to detect a move- ment generated by the muscle concerned and the ratio of the amplitude of the first and last move- ment to the during the same stimulation
Clinical performance
The following characteristics are the clinical performances of the ToFscan:
- Intra-operative: Enable practitioners to monitor patients’ intra-operative muscle relaxation
- Post-operative: To diagnose residual blockade of the patient
Clinical benefits
NMT monitoring allows to:
- avoid complications during intubation or extubation related to the use of neuromuscular blocking agents
- adjust the dose of neuromuscular blocking agents required for each patient
Important Information about the use of this device
The compact ToFscan is intended for use by health professionals (anaesthetists, doctors or fully qualified nurse anaesthetists) specially trained in the use of this instrument. The device and all of the settings associated with it are designed for use on adult and paediatric patients in hospital or health institutions so that the patient’s neuromuscular block level can be monitored.
The ToFscan measurements of the patient’s muscular response can be used to monitor the effects of neuromuscular blocking agents.
The interpretation of ToFscan results must always be subjected to clinical assessment and compared against other observed clinical signs. Sole reliance on the results or values rendered by the ToFscan for the monitoring of curare- administered patients, is strongly discouraged. The values measured in patients with neurological disorders, nervous system disorders, Bell’s palsy, myasthenia or general neuromuscular disorders must be carefully interpreted.
The ToFscan is compliant with the European directive on medical devices and with the current regulatory requirements of the countries where it is distributed.
For further information, kindly contact the ToFscan manufacturer IDMED through his internet website (www.idmed.fr) or by mail at the following address:
IDMED
Hôtel Technoptic
2 rue M. DONADILLE
13013 MARSEILLE FRANCE
ToFscan® and IDMED® marks are property of IDMED company (France) in several countries.
SAFETY MEASURES
INTRODUCTION
Read this entire manual carefully before using the ToFscan.
WARNINGS, CAUTIONS, NOTES
The terms Warning, Caution and Note have specific meanings in this manual.
- A WARNING cautions against certain actions or situations likely to cause bodily harm or death.
- The word CAUTION warns against actions or situations likely to damage the equipment, produce inaccurate data or cancel a procedure, even if bodily harm is less than likely to occur.
- A NOTE provides relevant information about a function or procedure.
EXPLANATION OF THE SYMBOLS
The symbols which may be displayed on the ToFscan screen are recapitulated and
explained at the end of this section.
Any serious incident occurring in relation to the device must be notified to the manufacturer and the competent authority of the Member State in which the user and/or patient is established.
Warning
Risk of explosion: do not use the ToFscan in a flammable environment or in places where flammable anaesthetic products are accumulated.
The ToFscan is not designed to operate in an environment where there are SCANNERS, M.R.I.’s or any other device creating large magnetic fields. The same applies for short-wave or
medium-wave treatment devices.
The electrode cables, electrodes and connections must not come into contact with any other item, conductive or otherwise.
In order to reduce the risk of burns when using high-frequency surgical devices, do not place the ToFscan stimulation electrodes between the surgical site and the electrode leading back to the electrosurgical unit.
Simultaneously connecting a patient to a high-frequency surgical device may cause burns at the contact points of the ToFscan electrodes and cause damage to the device.
Never use the ToFscan at the same time that defibrillation devices are being used.
Like all neuromuscular transmission monitor, the ToFscan must be connected to electrical stimulation electrodes capable of supporting up to 300 volts with 60 mA current. The contact surface of the electrodes must be greater than 1.8cm² (0,28in²).
The output from electrical stimulation causes nociceptive stimulations and the intensity of these stimulations must be adapted to the patient’s analgesic level.
Do not use the ToFscan on patients wearing pacemakers without verifying and identifying the possible consequences. The user must take all the necessary precautions during the operation for this kind of patient.
Prior to use, check that no other equipment, device or material is in contact with the electrodes.
Sensors and electrodes should only be in contact with clean and healthy skin
Prior to each use, check that the device, the display and the cables (electrodes and sensors) are not damaged. Do not use if a part is damaged
Handle the device with care to prevent any fall.
The ToFscan should be used during a limited time on one patient at a time and cleaned between patients.
The ToFscan can be used fully or partially during the surgery. The duration should not exceed 24H.
After positioning one of the sensors in contact with the patient, check regularly at least every 2 to 3 hours that the sensor does not cause excessive pressure or stress on the patient’s skin. If skin appearance changes, change the site sensor.
Portable RF communication equipment (including peripherals such as antenna cables and external antennas) should not be used within 30 cm (1 foot) of any part of the ToFscan, including accessories specified by the manufacturer.
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
Use cables other than the one recommended by the manufacturer of this equipment could result in increased cyber security risks.
Caution
Read this entire manual carefully before using the ToFscan.
Never put the ToFscan or any of its parts or accessories into an autoclave.
Do not submerge or spray the appliance or any of its parts with liquids.
The ToFscan and its parts are not suitable for the processes of gas, radiation
(gamma or other), water bath, steam or heat sterilisation.
Follow the instructions given in the “Cleaning and disinfection” section for
cleaning and disinfecting the ToFscan.
The ToFscan carries an internal lithium-ion battery. The ToFscan battery
should under no circumstances be dissembled, modified or replaced. Any
tampering with the battery poses the risk of combustion or explosion. Only an
authorised technician or IDMED employee is qualified to perform such
operations.
After an extended period of disuse (storage), recharge the ToFscan battery for
at least 2 hours before use. If the ToFscan does not switch on when the wheel
selector is pressed, the battery needs to be replaced.
Only qualified technicians are authorised to carry out repairs or maintenance
procedures with the consent of IDMED.
The ToFscan user must take care not to come into contact with other electrical
devices when using the ToFscan.
Before performing electrical stimulation with the ToFscan, the practitioner
must assess the appropriateness and strength of the stimulation that can be
applied to the patient.
Never touch the electrodes during the stimulation phases. The electrodes are
only surface electrodes and are suited for the application of electrical
stimulation.
Do not use cables or accessories other than those supplied with the ToFscan.
The using of electrosurgery device in the same time as ToFscan can interfere
the measurement and the results of it.
In order to prevent electrostatic shock, the device must be used in an
electrostatic limited environment. (see Environment section)
ToFscan has been designed to stimulate patients via electric impulsions. As a
matter of fact, data recorder (EEG, ECG) can detect these electrophysiological
signal impulses. These perturbations are transitional and depend how these
devices are configured.
Notice on Electromagnetic Compatibility (EMC) : This device generates, uses,
and can radiate radio frequency energy. If not set up and used in accordance
with the instructions in this manual, electromagnetic interference may result.
The equipment has been tested and found to comply with the norm IEC60601-1-2
for medical electrical equipment. These limits provide reasonable protection
against electro-magnetic interference when operated in the intended use
environments (e.g. hospitals)
Contraindications:
None
Side effects:
None
The pictures in this manual are for illustration purposes.
Explanation of the symbols
General symbols
Caution : USA federal law restricts this device to sale by or on the order of an anesthesiologist or other qualified practitioner.
Symbols displayed on the ToFscan screen
I General information
Overview of the ToFscan and its accessories
Main menu, Display screen
Menu selection
The menus, options and various tests are accessed using the selection wheel on
the front of the ToFscan. The user navigates through the various menus by
turning the wheel (clockwise or anti-clockwise).
To select a menu or function, press the wheel an release the wheel
button(holding it down for less than 1 second).
To start a test or an electrical stimulation, to access the settings or switch
off the device, press and hold the wheel button for 2 seconds. The following
icon is displayed on the screen together with the required time to hold down
the button in order to start the stimulation.
Example of the selection screen for starting the PTC mode:
Battery / AC power supply operation
The ToFscan carries a battery enabling it to function independently on battery
power for close to 1 month (for more information refer to the battery
section). This battery is recharged by the power supply provided with ToFscan.
The power supply unit can be used as a permanent mains power supply unit. This
means that the ToFscan will operate via its mains power supply unit without
necessarily running down the battery. In this operating mode, the ToFscan
displays the results and all information continuously. It will go into energy-
saving mode 2 hours after it is last used or the last measurement is taken.
When ToFscan operates on battery “ECO” mode can be activated by the user (for
more information on “ECO” mode refer to the section “Parameters Menu” “ECO”)
Note:
Position the ToFscan and the power supply in order to disconnect easily.
Fully charge the battery before the first use.
If the power supply unit should malfunction, never use power supply units
other than those supplied by IDMED
II ToFscan setup
Cable / Cable connection
The user connects the sensor+ electrode cable to the ToFscan prior to use. The
user will make sure sensor symbol is displayed in green on the right side of
the screen after this cable is connected to the ToFscan.
Electrodes
The ToFscan must be connected only to surface electrodes by “press-on”
connection. The electrodes must be electrodes used for electrical stimulation
of patients. They should be compatible with the stimulation values currently
used by a NeuroMuscular Transmission Monitor.
Positioning of the electrodes
NMT blockade may be monitored by stimulating various nerves and observing the
response of the particular muscles.
In the case of continuous monitoring, the most commonly used technique is
considered to be stimulation of the ulnar nerve and measurement of
acceleration in the adductor muscle of the thumb.
“Thumb” sensor:
In the case of monitoring the adductor muscle of the thumb (adult or
paediatric sensor), the electrodes will be positioned along the ulnar nerve on
the inner arm near the wrist. The electrodes will be spaced 2 to 5 cm (1 to 2
inches) apart when using single electrodes.
Note:
It is vital to position the electrodes properly in order to stimulate the
nerve and not the muscle.
Position of the sensor and the electrodes:
“Eyebrow” sensor
With the « eyebrow » sensor, it is possible to evaluate the level of blockade
of the patient while measuring the response level of the corrugator supercilli
muscle.
The stimulation electrodes are to be positioned on the root of the facial
nerve next to the tragus. The positioning will be done on each side of an
imaginary line going from the tragus extremity to the middle of the nose. The
distance between the electrodes will range from 2 to 5 cm (1 to 2 inches).
Note:
It is essential to correctly position the electrodes in order to stimulate a
nerve and not a muscle.
“Big Toe” sensor
The stimulation electrodes are to be positioned on the tibial nerve above the
ankle. The distance between the electrodes ranges from 2 to 5 cm (1 to 2
inches).
Note:
It is essential to correctly position the electrodes in order to stimulate a
nerve and not a muscle.
Placing of the sensor
When positioning the sensor, the sensor cable must not apply any pressure to
the sensor or the sensor clamp. It must allow the sensor to move freely
according to muscle contractions.
Positioning the sensor in contact with the patient must not cause excessive
pressure or stress that could injure the patient.
“Thumb” sensor
The splint must follow the shape of the hand as closely as possible and be
positioned so that it makes contact with the last phalanx of the thumb.
‘Eyebrow’ sensor
The positioning must allow the free movement of the sensor. The sensor is
positioned on the corrugator supercillii muscle. The sensor cable exert no
tension on the sensor. The sensor is fixed with a double sided tape on the
patient’s skin. This tape must be adapted to a medical usage and allow a
reliable fixing during the whole duration of the monitoring.
‘Big Toe’ sensor
The positioning must allow the free movement of the sensor. The sensor cable
exert no tension on the sensor. The sensor is fixed with a medical tape to the
patient’s big toe. It is essential to note that the toe and ankle of the
patient must be free to move.
Single-use hand sensor
The user will position the single-use sensor on the patient’s hand according
to the instructions on the packaging. Ensure that the sensor does not exert
excessive pressure on the patient’s skin or impede the blood circulation of
the fingers (avoid excessive tightening of the fingers, hand or any other
patient limbs by the adhesives of the sensor). The integrated electrodes will
be positioned on the path of the ulnar nerve at the wrist area inside the arm.
After positioning the single-use sensor on the patient’s hand, the sensor will
be connected to the ToFscan cable (TOF-CS1) itself connected to the ToFscan.
Notes:
While the device is being operated, the user should check that the sensor must
keep the same position as the one from the initial set-up. The same applies
for the patient’s arm, leg or head which should not change position for the
duration of the monitoring process.
If the ‘Thumb’ sensor does not fit nicely the hand, it can be fixed with
medical adhesive tape to be maintained in an ideal position. The user can
immobilize the last three fingers with an adhesive strap to improve the thumb
range of motion and obtain more precise measure during the monitoring of the
thumb.
In the case of the thumb sensor, check that the splint part of the sensor or
the ring around the index finger does not cause pressure or excessive stress,
an adhesive positioning (see image «Positioning by adhesive») can then be put
in place.
After a certain period of use of the sensor, a slight mark or redness of the
skin in the contact area with the sensor may appear. This mark or redness is
due to the presence of the sensor in contact with the skin. This must remain
limited, harmless and not look like an injury.
Skin impedance
The ToFscan is an electrical stimulator with a constant current. Therefore,
irrespective of skin impedance it will stimulate the patient with an identical
current. It will function in this way so long as the needed voltage is below
300 V. Because of this limit you need to have a good skin impedance.
For example, to get a 60 mA current through a resistive charge the maximum
impedance should be equal to 5 Kohms. The skin impedance is more complex than
a simple resistive charge and the ToFscan will help you getting a good
impedance with a colored electrodes symbol.
Only the green symbol allows to use in good condition the ToFscan. With yellow
symbol the intensity of electric stimulation may be lower than expected.
If the symbol is red the ToFscan doesn’t provide electric stimulations.
If you get the screen below it is necessary to check or modify the patient’s
connection to the electrodes.
Note:
Cleaning the patient’s skin prior to positioning the electrodes significantly
lowers skin resistance. The user should therefore ensure that the patient’s
skin has been cleaned before attaching the electrodes. The electrodes quality
and condition are essential in the measured impedance value.
Connecting the cable to the ToFscan
After putting the electrodes on the patient, the user must connect them to the
ToFscan with the electrodes cable. Before connecting the cable, he/she should
check that the ToFscan displays the main menu and is not in stimulation phase
or programmed in automatic stimulation mode.
The proximal electrode (nearest to the heart) will be connected to the red-
coloured positive electrode clamp. The distal electrode (further away from the
heart) will be connected to the black-coloured negative electrode clamp.
Once the cable has been connected to the electrodes, the ToFscan displays the
electrodes and the sensor’s connection symbol (icon in green if present and
functional).
Reference or “REF”
The “REFERENCE” mode enables the user to measure the patient’s motor response
to TOF electrical stimulation when the patient is anaesthetised but not under
the effect of a neuromuscular blocking agent. This measurement is used to
display the calculation comparisons between the range of the patient’s muscle
response under curare and without curare during TOF stimulations.
For further information about this test, refer to the “TOF mode” section,
under the “REFERENCE” menu.
III Using the ToFscan
General principle
The ToFscan is used to perform 5 modes of electrical stimulations. Some of
these modes can be configured or programmed by the user.
All modes are selected by a brief push on the wheel button. Once in the
submenu, press and hold the wheel button to start the electrical stimulation.
The ToFscan will give a “beep” simultaneously to the start of the electrical
stimulation.
It is important to observe a waiting time between each stimulation to avoid
distorting the results. The ToFscan memorises the elapsed time since the
previous stimulation and displays it at the bottom of the screen. If this time
is shorter than the waiting time to be observed between each stimulation, the
waiting time before the next stimulation is displayed in the centre of the
screen together with the following symbol .
The symbol in front of a result indicates a possible presence of a impaired measurement. The user can do the test again (while respecting the waiting time) or wait for the next test scheduled in the case of “AUTO TOF” mode.
Note:
The recommended waiting time between each stimulation are shown at the end of
each description of the various stimulations (or tests).
Only ‘TOF’ tests are commonly used for the NMT monitoring for the eyebrow
muscle (stimulation of the facial nerve).
When disconnecting the ToFscan from a patient at the end of the operation and
before connecting a new patient for new operation, the display of the ToFscan
must be reset. For this, press the wheel an release the wheel button.
TOF mode
The “TOF” mode includes various options or sub-menus. Each of these options is
detailed below. With this mode, the TOF stimulation can be done two ways,
either directly or manually by the operator, or automatically at a repeated
interval selected by the user.
“TOF” sub-menu
After selecting the “TOF” menu and then the “TOF” sub-menu, the user can start
a “TOF” stimulation (or test) by pressing and holding the selection wheel.
Prior to this, he must verify that the output (current in mA) of the selected
stimulation is appropriate for the level of anaesthesia, the level of
neuromuscular blockade and the patient’s profile. For further information on
stimulation output, refer to the “Parameters” section.
TOF stimulation is one of the most commonly used forms of stimulation,
comprising 4 stimulations (of 200 µs) at 0.5 second intervals.
If the ToFscan is connected to a cable fitted with an accelerometer sensor,
after the electrical stimulation it will display a calculation of the
percentage of the range of the 4th response against the first (ratio T4/T1 TOF
as a %) in yellow in the middle of the screen. The ToFscan also displays a bar
graph allowing the visualization of the range of the different responses.
If a “Reference” test (or stimulation) was performed, it will be symbolized on
the top of the bar graph by a horizontal yellow line. ToFscan automatically
displays the T4/Tref ratio.
Example of a screen with a result of 100% following a TOF stimulation:
The number of responses detected (TOF count) is displayed as an X/4 ratio (X
being the number of muscle response detected).
When ToFscan detects artefact movements or electric noises in the measurement
process, a symbol
will be displayed in front of results. This symbol informs users about a
possible non reliable measurements/results.
The waiting time required by the ToFscan between two “TOF” stimulations is 12
seconds.
Note:
The user should validate the results read out by pressing the selection wheel
in order to perform other stimulations. The percentage calculations are
limited to 100% (only for T4/T1) so that values which are not pertinent will
not be displayed.
“AUTO TOF” sub-menu
The “TOF AUTO” mode allows you to program TOF stimuli at regular intervals.
The available intervals are every 15s., 30s., 1min., 2min., 5min. and 15min.
After selecting the stimulation interval, the user starts the stimulation
cycle by pressing and holding the selection wheel (for at least 2 seconds).
The first stimulation is delivered 4 seconds after the selection wheel has
been pressed.
To stop a programme, press the selection wheel; the ToFscan then returns to
the “AUTO TOF” menu.
The results displayed are the same as in the “TOF” menu.
The stimulation interval can be changed after starting TOF AUTO mode.
The frequency of intervals can be changed by rotating and pressing the knob
without holding it down.
“REFERENCE” or “REF” sub-menu
The reference mode enables the user to store the patient’s response to a TOF
electrical stimulation when the patient is anaesthetised but not under the
effect of neuromuscular blocking agent. This value can help the user to
evaluate the recovery of neuromuscular function and the efficiency of agent
depolarizing neuromuscular blocking.
The ToFscan delivers the TOF stimulation in order to calculate the average
amplitude of the four muscle responses; this value will be recorded as Tref.
This average amplitude will be used to calculate T4/Tref and will be displayed
for subsequent TOF stimulations.
The waiting time between two “REFERENCE” stimulations is 12 seconds.
Reference value can be erased by holding down the wheel button in the sub-menu
“Reference”.
Example of a screen for deleting a reference.
Note:
The reference value is used only to calculate T4/Tref during a TOF electrical
stimulation and only if the ToFscan is connected to a cable with an integrated
sensor (accelerometer sensor).
Like all electrical stimulations, the stimulation used for the reference must
only be performed on anaesthetised patients. Stimulations can be very painful
for non-anaesthetised patients.
After the results are displayed, the selection wheel must be pressed to return
to the selection menu.
TET mode
Tetanic stimulation or “TETANUS” stimulation is used to stimulate a patient
for 5s at 50 Hz. As the ToFscan does not display a measurement at the end of
this test, no user validation is expected at the end of the stimulation in
order to reactivate the selection wheel navigation function.
The patient’s motor response is not measured by the ToFscan sensor, but is
visually gauged by the user.
Note:
The waiting time required by the ToFscan between two “TET” stimulations is 3
minutes. The “TET” stimulation is absolutely not recommended in case of the
eyebrow muscle monitoring.
DBS mode
The ToFscan is used to perform “Double Burst Stimulations” or “DBS”. It offers
the user 2 types of DBS mode under the “DBS MODE” menu. The DBS mode can help
to detect a possible residual blockade. DBS stimulations consist of two bursts
of 50 hertz stimulations spaced 750 ms. Depending on the selected DBS mode,
each burst will have 2 or 3 impulses (impulse duration: 200 µs). After DBS
stimulation applied, the number of responses measured is displayed with their
relative amplitudes by 2 white bars. The percentage of the ratio between the
amplitude of the second response and the first is displayed at left side on
the screen.
“DBS” sub-menu
The ToFscan provides “DBS 3.3” stimulation by default.
The user can activate this stimulation by pressing and holding the selection
wheel, or the “DBS 3.2” stimulation after having selected it under the “DBS
Mode” menu.
“DBS MODE” sub-menu
This menu is used to select the various types of DBS stimulations. The ToFscan allows DBS 3.3 and DBS 3.2.
Note:
The waiting time after a “DBS” stimulation is 20 seconds. The « DBS »
stimulation is absolutely not recommended in case of the eyebrow muscle
monitoring.
PTC Mode
“PTC” or “Post Tetanic Count” stimulation is generally used for deep
neuromuscular block or when there is no response to TOF stimulation. “PTC”
stimulation comprises a 5-second “TETANUS” stimulation at 50 Hz followed by a
3-second pause and then 10 “SINGLE TWITCH” stimulations.
“PTC” sub-menu
When selecting this sub-menu, the user starts PTC stimulation by pressing and
holding the selection wheel. Upon completion of the stimulation (17 seconds’
duration) the ToFscan displays the number of muscle responses detected. It
maps out each of these in the form of a bar chart to compare their respective
ranges.
When measurements are disrupted (parasitised), a symbol will be displayed
instead of the measurement (bar).
“ATP” sub-menu
ATP (Automated TOF PTC) is an automatic mode to measure deep, average and
light neuromuscular blockade. ATP mode uses TOF and PTC stimulations in
appropriate ways. Stimu- lations are repeated every 30 seconds or every 5
minutes depending on the number of responses measured after each TOF or PTC
stimulation.
Example of an ATP mode screen during a TOF stimulation:
If you get the screen below, it means that the measurement conditions are disrupted. Please check the position of the sensor.
Principle
The ATP mode is an automatic mode. ATP can be stopped at any time by the user
by pushing the wheel button. ATP mode uses TOF and PTC stimulations depending
on the depth of blockade. Results are displayed as per the stimulation
applied. ATP delivers TOF stimulation, in case of no response from the TOF
stimulation, a PTC stimulation will be applied. After all stimulations (TOF or
PTC) the ToFscan displays the results. If the patient has had at least one
response to TOF stimulation, the ToFscan will then display the measured
results and there will be a pause of 30s before the next set of TOF
stimulation. If the patient doesn’t have a response to TOF stimulation, the
ToFscan will stimulate the patient with a PTC stimulation and display the
results. After a PTC stimulation there is a pause of 5 minutes before the next
set of stimulation.
ATP Process
Note:
The ATP mode must only be used with the thumb sensor and only with non-
depolarizing neuromuscular blocking agents. It can be stopped at any time by
pushing the wheel button. In the absence of movements following several
stimulations, the ATP mode will stop.
The waiting time required by the ToFscan following “PTC” stimulation or “ATP”
mode is 3 minutes. It is important to remember that “PTC” stimulations is only
normally used when no responses to “Single
Twitch” or “TOF” stimulations are detected. ”PTC” stimulation and “ATP” mode
are absolutely not recommended for eyebrow muscle monitoring.
ST Mode
The “Single Twitch” stimulation is a stimulation 0.2 ms pulse duration. It
produces a single muscle contraction. The patient’s motor response is not
measured by the ToFscan sensor but is visually gauged by the user.
“Twitch” sub-menu
Stimulation is activated by pressing (2s) the selection wheel.
“0.1HZ” sub-menu
The ToFscan produces one Single Twitch every 10 seconds when the selection
wheel is pressed and held. The user stops the 0.1 HZ stimulation by simply
pressing the selection wheel.
“1HZ” sub-menu
The ToFscan produces one Single Twitch every second when the selection wheel
is pressed and held. The user stops the 1 HZ stimulation by simply pressing
the selection wheel.
Note:
The “0.1 HZ” or “1 HZ” stimulations are automatically repeated during 10
minutes, after which the ToFscan stops the stimulation. No waiting time is
required by the ToFscan after this category of stimulations. The practitioner
will gauge the length of the waiting period required depending on the number
of stimulations delivered.
Parameters Menu
The “Parameters” menu will enable the user to modify the standard parameters
of the ToFscan. This menu is represented on the screen by the following icon:
To go to the “Configuration” menu, the selection wheel must be pressed and held (2s).
“STIM” sub-menu
When this sub-menu is selected by pressing the selection wheel, the user can adjust the current of the stimulations. The ToFscan is configured to 50 mA by default. It is generally acknowledged that in order to obtain supra-maximal stimulation, the current required for the stimulation of the Ulnar or tibial nerve for the adult is 50mA. This value is 30mA for a paediatric usage on the same nerves. The value for the stimulation in the case of the eyebrow muscle (Corrugator Supercilli) is 30mA. In special cases where considered appropriate by the user, he/she may adjust this stimulation value. The user should consider the potential risks of an unsuitable stimulation current for a patient.
“SND” sub-menu
The SND sub-menu allows activating or deactivating the sound “beep” emitted by the ToFscan during measurements, selections and electrical stimulation.
“ECO” sub-menu
The ECO sub-menu allows selecting the eco mode. “ECO” mode reduces the length of display to increase the time life of battery if the ToFscan is not connected to the power supply. In this case, the display becomes inactive 40s after the last measurement or action of the user (5s in case of “TOF-AUTO” automatic mode). Otherwise, the measurements display lasts 16 minutes.
“T4/T2” sub-menu
“T4/T2” sub-menu allows inactivating display of ratio T4/T2 instead of ratio T4/T1 when the amplitude of T2 response is higher than the amplitude of T1 response. T2 bar is yellow when ratio T4/T2 is displayed instead of ratio T4/T1.
IV Servicing, Disinfection cleaning
Preventive servicing, Maintenance
In order to maintain its performance, it is strongly recommended that the
following tests be carried out on the device at least once every two years:
- Checking of the integrity of the casing, screen and labelling
- Checking of the battery charging process
- Checking of the condition of the electrode cable, its electrode clamp ends and its sensor clamp
- Checking for the value of current of electrical stimulations.
- Checking the sensor measures.
The lifetime of the ToFscan, under the required operating and maintenance conditions, is 5 years (2 years for accessories).
Caution:
Only qualified technicians are authorised to carry out some repairs with
the consent of IDMED.
Battery / Battery charge
Battery
The ToFscan includes a Lithium-Ion rechargeable battery. The battery is
equipped with thermal protection and short-circuits protection. At full
charge, battery power lasts for approximately 1 month at a rate of 10 “TOF”
stimulations per day (“Eco” mode activated).
Level and colour gauges indicates the level of charge of the battery
The battery has a one-year warranty (its battery power at one year should be 50% more than its estimated battery power). The normal service life of the battery is 2 years.
Note :
Only qualified technicians are authorised to carry out repairs or maintenance
operations with the consent of IDMED.
Battery Charge
The battery is charged with the charger supplied by IDMED. It can charge a
flat battery in less than 8 hours.
The battery can be charged regardless of its level. Whenever the ToFscan
displays the flat battery symbol (coloured red) , the battery must be charged
as soon as possible.
Charging is automatic, consequently when the charge is complete the ToFscan
stops the process.
Note :
Only qualified technicians or IDMED employees are authorized to carry out
repairs or maintenance operations on the battery.
Maintenance work on the battery is restricted to checking its charge cycle
once every two years. By doing this, it can be verified that the charge cycle
does not exceed 8 hours (a change from red to green for the battery charge
gauge).
Cleaning and disinfection
Caution:
Do not place the ToFscan or any of its parts or accessories inside an autoclave.
Under no circumstances must the ToFscan or any of its parts or accessories come into direct contact with, be immersed in or filled with liquid.
The ToFscan and its parts and accessories are non-sterile devices. Under no circumstances must the ToFscan or any of its accessories be sterilized.
The ToFscan must be cleaned and disinfected between each patient. Low-level disinfection is generally sufficient.
The surfaces of the ToFscan and its accessories have to be cleaned with a lint-free cloth moistened with a Quaternary Ammonium Compound (QACs), isopropyl alcohol 70%. Before using any of these solutions, refer to the manufacturer’s documentation and test on a reduced surface.
Examples of recommended products:
– mikrozid® sensitive liquid from the manufacturer Schülke & Mayr GmbH.
– Incidin OxyWipe from the manufacturer Ecolab.
Please check with your local authorized distributor or with the manufacturer
which products are available and approved in your country.
The ToFscan cable (electrode and/or sensor) must not come into direct contact with, be immersed in, splashed or filled with liquid and is to be cleaned in the same manner as the ToFscan.
When cleaning the cables of the ToFscan, be careful not to create excessive traction on the splint that could cause premature breakage of the wires inside the sheath.
Diagnostics / Possible dysfunctions
The table below summarises a list of possible dysfunctions and the solutions for resolving them.
Note:
Should the problem persists or is not solved by the actions listed in the
table above, you must contact the distributor. .
V End-of-life disposal / Recycling
In the interest of environmental safety, you are required to pass your used
system on to a collection body with the capability to treat devices containing
electronic components and Lithium-ion storage batteries.
To dispose of or recycle device components, contact a company specialising in
the recycling of electronic devices.
Unsorted electronic waste products are potentially hazardous to the
environment.
Packaging materials must be disposed of or recycled in accordance with the
regulations in force.
VI Technical specifications and warranty
The ToFscan comes with a microcontroller and a colour LCD screen for optimal legibility and ease of use.
Safety
- Biocompatible material sensors (part in contact with the patient). Latex free
- Compliant with European regulation 2017/745. Class 2a device (CE 0459 LNE/G-MED)
- Compliant with standards IEC 60601-1. Class II equipment.
- Compliant with standards IEC 60601-2-10.
- EMC: IEC 60601-1-2
EMC Emission
EMC Immunity
Stimulations
- TOF (Train Of Four), T4/T1 and T4/Tref calculations.
- AUTO TOF (ToF programmed from 15s to 15min).
- TET (Tetanus 50 Hz)
- DBS (Double Burst Stimulation) modes 3.3 and 3.2
- PTC (Post Tetanic Count)
- ATP (Automatic TOF PTC)
- TWITCH (Single Twitch) 0.1 Hz and 1 Hz.
Acceleration sensor
- Three-dimensional accelerometer (+/- 8 G at 10 bits, Fq:200 Hz, Resolution 0.016G)
Electrical stimulations
- Constant output current of 0 to 60mA (accuracy +/- 10%) (on an resistive load of 4 Kohms)
- Monophasic, duration of impulse 200 µs, frequency 50 Hz
- Stimulation or ECG electrodes:
– Capable of supporting up to 300 volts with 60 mA current.
– Contact surface must be greater than 1.8cm² (0,28in²).
Examples of recommended electrodes:
– RED DOT electrodes ref.2560 from the company 3M
– F9047 electrodes from the company FIAB
Please check with your authorized distributor or with the manufacturer which products are available and approved in your country.
Data transfer
- Optical output for fiber optic connection
- TOF-RS1 and TOF-RS2 are the only recommended Optic-Serial (RS232) cable to connect ToFscan to other monitors.
Power supply
- 2900 mAh (minimum)/ 3.7V Lithium-Ion battery (comes with thermal protection and protection against short-circuits)
- Battery power for about one month with normal use (10 TOF measurements a day).
- Charger / External power supply (continuous 5V, 1 A minimum).
Dimensions / Weight
- 70x150x55 mm (monitor only).
- 320 g (approximately) with battery and cable and electrode. (190g excluding cable).
Warranty
- Length of warranty: 2 years, 6 months for accessories and sensor
Power consumption
- With power supply connected: 1 Watt while in use and 0,1 Watt while in sleep mode.
Environment
Shipping and storage conditions
The ToFscan and its accessories must be stored or transported under the following restrictions and conditions. These conditions apply to non- operational storage and transport conditions.
Temperature 10°C to +50°C
Humidity 15% to 95% (without condensation)
Pressure 500 hPa to 1060 hPa
For storage and transport, the original packaging must be used.
Protect the ToFscan from pikes in temperature which can cause condensation.
Operating environment
Reminders:
Risk of explosion: do not use the ToFscan in a flammable environment or in
locations where flammable anaesthetics may be accumulated.
The ToFscan is not designed to operate in an environment where there are SCANNERS, M.R.I.’s or any other devices creating large magnetic fields. In order to limit electrostatic discharge, the humidity must be maintaining above 35% and an antistatic floor is recommended.
The ToFscan is designed to operate safely under the following conditions. Situations other than those described are likely to compromise the reliability of the device.
Temperature 10°C to +40°C
Humidity 35% to 90% (without condensation)
Pressure 700 hPa to 1060 hPa
VII Accessories
The ToFscan (reference : TOF-MU) is delivered with a number of accessories. Here is a list of the main accessories with their designations and IDMED references. The complete accessories list is available from the ToFscan distributor.
Accessories of medical device ToFscan
Other Accessories
Check the availability of the accessories in your country with the authorized distributor or the manufacturer.