Newtons4th LEM-6 Power Analyzers User Guide

October 31, 2023
Newtons4th

Newtons4th LEM-6 Power Analyzers User Guide
Newtons4th LEM-6 Power Analyzers

LEM current transducers are recognized as leaders in the field of high current AC+DC sensors and used in conjunction with PPA series power analyzers, they allow N4L to provide world leading high power measurement systems.

The performance of any system involving external current sensors is largely dependent on the interfacing and connection technique. This is more complex than commonly imagined and often represents the difference between a good and bad system.

Power supply stability, constant and peak current rating, burden resistor selection, connection cable volt drop, noise rejection and total burden limits for a CT are some of the issues easily overlooked despite being critical to achieving an effective solution.

N4L current transducer interface accessories provide a solution to almost any high current power measurement environment with hardware specifically designed for ease of use, reliability, and optimum performance.

Module

LEM-6 Interface
Module

LEM-1 Interface
Module

LEM-1 50W Power Module
Module

LEM-1 120W Power Module
Module

The ideal solution will depend upon the number of phases, the LEM CT’s being used for each phase plus the distance between the power analyzer and the current sensor(s).

Interconnecting cables used in each system follow the key below:

Connection cable key

D type – Multi core cable – 2m standard – up to 6m optional

Safety BNC to BNC Std 1m

± 15V DC Power

Touchproof 4mm – Coax – Safety BNC – up to 18m

Option 1:
LEM-6 Interface Unit 

The LEM-6 interface together with connection cables supplied as standard provides the most convenient and flexible solution for systems with up to 707 Arms on 3 to 6 phases.

Each phase has an individual DC-DC converter to optimise stability and 4 selectable burden resistor values. The 1 Ohm burden is used for full CT range measurements while the higher burden values provide additional measurement sensitivity in applications where a CT will continuously work in a low current region of its specified range.

Number of phases on one interface – 3 to 6
Calibrated Burden resistor values – 4 per phase (1, 2.5, 5 and 10 Ohm)
Max CT secondary current – 1A dc or ac peak per phase
Distance from Interface to CT(s) – Up to 6m
nterface Unit 

Where multiple burden resistors are not required, the CT current is above 707 Arms or the distance between the power analyzers and CT’s is above 6m, the LEM-1 Interface powered by one of our LEM-1 Power Modules provides the solution.

Option 2:
LEM-1 Interface + LEM-1 Power Module 

Number of phases on one interface – 1
Calibrated Burden resistor value – 1 (1 Ohm)
Max CT secondary current – 1.6Arms per phase
Distance from Interface to CT – Up to 6m
Connection

Option 3:
LEM-1 Interface + LEM-1 Power Module + BDR01 

Number of phases on one interface – 1
Calibrated Burden resistor value – 1 (1 Ohm)
Max CT secondary current – 1.6Arms per phase
Distance from Analyzer to Interface – Up to 18m
Connection

APP GUIDE – Current transducer cabling principles.pdf

Multi-phase LEM-1 systems

Featuring a universal AC power input and touchproof 4mm safety outputs for convenient connection to multiple LEM-1 interfaces, the LEM-1 Power Modules maintain a stable ±15V DC supply over their complete power range plus a x2 full power capacity for 100ms to support overload capability available on most LEM CT’s.
Multi-phase LEM-1 systems

CT Model| Max AC Amps RMS| LEM-6Interface| LEM-1Interface| LEM-1 50WMax LEM-1 / CT’s| LEM-1 120WMax LEM-1 / CT’s
---|---|---|---|---|---
IT 60-S| 42| Yes| Yes| 6| 6
IT 65-S | 60| Yes| Yes| 6| 6
IN 100-S| 100| Yes| Yes| 5| 6
IT 200-S| 141| Yes| Yes| 6| 6
IN 200-S| 200| Yes| Yes| 5| 6
IT 205-S
| 200| Yes| Yes| 6| 6
IT 400-S| 282| Yes| Yes| 6| 6
IN 400-S| 400| Yes| Yes| 4| 6
IT 405-S | 400| Yes| Yes| 4| 6
IN 500-S| 500| | Yes| 1| 3
IT 600-S| 425| Yes| Yes| 3| 6
ITN600-S| 424| Yes| Yes| 3| 6
IT 605-S
| 600| Yes| Yes| 3| 6
IT 700-S| 495| Yes| Yes| 3| 6
ITN 900-S| 636| Yes| Yes| 2| 5
ITN 1000-S| 707| Yes| Yes| 2| 4
IT 1000-S| 707| | Yes| 1| 3
IN 1000-S| 1000| | Yes| 1| 3
IN 1200-S| 1200| | Yes| 1| 3
IN 2000-S| 2000| | Yes| 1| 3

LEM-1 Power Module selection chart

LEM-1 Power Module selection chart

LEM-1 Power Module electrical specification

Parameter LEM-1 50W LEM-1 120W
Input Voltage 90-264 Vrms
Input Frequency 47-63Hz
Output Voltage DC ± 15V
Output Current 1.67A Continuous 4A Continuous
Load Regulation ±1% @ 10-100% full load
Overload capability Continuous rating x2 for 100ms – 10% duty cycle
Output protection Short circuit protection with foldback – Auto recovery
Operating Temp -30ºC to 70ºC
Typical Efficiency > 85 % @ full load

LEM-1 Interface and Power Module dimensions

nterface and Power Module dimensions

Product names and Options

OPTION 1

N4L-LEM6-3 Three Channel Interface| LEM6-3 – 19″ rack mountable Three Channel System Interface Unit + 0.5m Safety BNC-BNC Leads + 4mm shorting connectors
N4L-LEM6-4 Four Channel Interface| LEM6-4 – 19″ rack mountable Four Channel System Interface Unit + 0.5m Safety BNC-BNC Leads + 4mm shorting connectors
N4L-LEM6-5 Five Channel Interface| LEM6-5 – 19″ rack mountable Five Channel System Interface Unit + 0.5m Safety BNC-BNC Leads + 4mm shorting connectors
N4L-LEM6-6 Six Channel Interface| LEM6-6 – 19″ rack mountable Six Channel System Interface Unit + 0.5m Safety BNC-BNC Leads + 4mm shorting connectors
LEM Interface Cable Assembly-2m| LEM-1 & LEM6 CT to interface cable 2m length – Supplied standard
LEM Interface Cable Assembly-3m| LEM-1 & LEM6 CT to interface cable 3m length – Optional
LEM Interface Cable Assembly-4m| LEM-1 & LEM6 CT to interface cable 4m length – Optional
LEM Interface Cable Assembly-5m| LEM-1 & LEM6 CT to interface cable 5m length – Optional
LEM Interface Cable Assembly-6m| LEM-1 & LEM6 CT to interface cable 6m length – Optional
OPTION 2
LEM-1 Interface| LEM-1 Interface + 4mm & BNC shorting connectors + colour coded power leads + 0.5m Safety BNC-BNC Lead        Unit Only – NO PSU
LEM-1 50W         Power Module| Universal Input to ±15V @ 1.67A Output Power module + IEC input lead
LEM-1 120W       Power Module| Universal Input to ±15V @ 4A Output Power module + IEC input lead
LEM Interface Cable Assembly| As in Option 1 shown above
OPTION 3
LEM-1 Interface| As in Option 2 shown above
LEM-1 50W         Power Module| As in Option 2 shown above
LEM-1 120W       Power Module| As in Option 2 shown above
LEM Interface Cable Assembly| As in Option 1 shown above
LEM-1 Current Breakout Cable| 4mm – Coax – Safety BNC – Connection cable – 4m
Current Breakout Cable Extension| Breakout cable extension – charged per meter up to 14m (18m total)
BDR01 – 1 Ohm Burden Resistor| 1 Ohm Burden Resistor – 0.1% – BNC touchproof connectors

Newtons4th Ltd
office@newtons4th.com
www.newtons4th.com

Newtons4th Ltd 1 Bede Island Road Leicester LE2 7EA UK Tel +44 (0)116 2301066

Newtons4th logo

Documents / Resources

| Newtons4th LEM-6 Power Analyzers [pdf] User Guide
LEM-6 Interface, LEM-1 Interface, LEM-1 50W Power Module, LEM-1 120W Power Module, LEM-6, LEM-6 Power Analyzers, Power Analyzers, Analyzers
---|---

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals