covetrus ProMAX IV Touch Digital Infusion Pump User Guide

June 12, 2024
covetrus

covetrus ProMAX IV Touch Digital Infusion Pump

This guide provides a quick reference for operating the Covetrus ProMAX IV Touch Digital Infusion Pump. Please refer to the User Manual for more detailed operations and settings.

PRODUCT INSTRUCTIONS FOR USE

  1. If using on a pole – Install Pole Clamp on the back of the pump with provided hex wrench and bolt and tighten on to the pole like, shown in Figure A.
  2. If hanging from a kennel – Install carabiners and hang from the front of the kennel, as shown in Figure B.
  3. If you would like to use the pump on a tabletop, please rotate it and set it on the table, as shown in Figure C.

Basic Operation

  1. Press and hold Power Button for 2 seconds to Power ON.
  2. Select Yes or No to clear the last treatment.
  3. Open door by pulling door handle; push free flow clamp (see Figures 1 and 2); install IV Admin Set following flow direction; secure line all the way down line guide by pressing line into both air sensor holders (see Figure 3); Close door by firmly pressing both sides of the door and confirm door is completely latched.
  4. Verify IV Admin Set Brand.
  5. Select Infusion mode, set Infusion Parameters, purge if necessary, and verify settings.
  6. Press Start.

Infusion Modes

To access Modes, press the bottom of the Home Screen or select Menu, then Modes.

Mode VTBI Range Rate Mode Time Mode
Rate Mode 0.01 – 9999.99 ml Set VTBI and Rate (ml/hr) Set VTBI and Time
Syringe Size Rate Range ml/hr
3 ml 0.1-100
5/6 ml 0.1-150
10/12 ml 0.1-300
20 ml 0.1-600
30/35 ml 0.1-900
60 ml 0.1-1500
Body-Weight Mode Select Unit, enter Body Weight, Concentration,

Dose, VTBI
Dose Rates| mcg/min, mcg/hr, mcg/kg/min, mcg/kg/hr, mg/min,
mg/h, mg/kg/min, mg/kg/h plus more available
Drug Library Mode| Select Drug, enter Concentration, Dose Rate,
Weight
Sequence Mode| Set Rate1, set Time1, then proceed to following
sequences as needed

Bolus Operation

  1. Purge: While the pump is idle, press Purge on the screen and select Yes or Press and Hold the Purge Softkey (see Figure 1).
  2. Programmed Bolus: During infusion Press Bolus on screen and set Bolus VTBI, Bolus rate, and Bolus time, and then select start (see Figure 2 and Figure 3).
  3. Manual Bolus: During infusion, press and hold Bolus softkey (see Figure 2) – Audible tone every 1 ml bolus infused. To change the default Manual Bolus rate: while the pump is idle, go to Menu, Settings, then select Bolus rate (See Figure 4).

Product Settings

To enter additional settings, select the Menu button on the touchscreen or softkey

  • Modes: Change the current Infusion Mode.
  • General: Rotate Screen ON/OFF, Change Date & Time, Screen Brightness, Sound Level, Screen Lock time, Night Mode, and Battery Capacity Indicator.
  • Records: Recall History of alarms and recent therapies.
  • System: Change Language, Factory default, Version, and Maintenance.
  • Home: Go back to the Home screen.
  • Standby Mode: To Access Standby Mode press the Power button, select Standby and enter standby time.

Service Information

All Warranty and Non-Warranty Repair Services will be performed by Triumph Medical Services. When sending your pump in for repair, please go to the following link and fill out the Veterinary RMA Request form: www.triumphmed.com/vet-rma-request. The shipping address will be provided when your RMA confirmation is emailed to you. For any questions, call 888.962.0662, ext 3.

This guide provides a quick reference for operating the Covetrus ProMAX IV Touch Digital Infusion Pump. Please refer to the User Manual for more detailed operations and settings.

INSTRUCTIONS FOR USE

  • If using on a pole – Install Pole Clamp on the back of the pump with provided hex wrench and bolt and tighten on to the pole like, shown in Figure A.
  • If hanging from a kennel – Install carabiners and hang from the front of the kennel, as shown in Figure B.
  • If you would like to use the pump on a tabletop, please rotate it and set it on the table, as shown in Figure C.

Basic Operation

  • Press and hold Power Button for 2 seconds to Power ON
  • Select “Yes” or “No” to clear the last treatment
  • Open door by pulling door handle; push free flow clamp (see Figures 1 and 2); install IV Admin Set following flow direction; secure line all the way down line guide by pressing line into both air sensor holders (see Figure 3); Close door by firmly pressing both sides of the door and confirm door is completely latched
  • Verify IV Admin Set Brand
  • Select Infusion mode, set Infusion Parameters, purge if necessary, and verify settings
  • Press Start

Infusion Interface

During Infusion Interface

Infusion Modes

To access Modes, press the bottom of the Home Screen or select Menu, then Modes

Syringe Size Rate Range ml/hr
3 ml 0.1-100
5/6 ml 0.1-150
10/12 ml 0.1-300
20 ml 0.1-600
30/35 ml 0.1-900
60 ml 0.1-1500
  • VTBI Range : 0.01 – 9999.99 ml
    • Rate Mode: Set VTBI and Rate (ml/hr)
  • Time Mode: Set VTBI and Time
    • VTBI Range: 0.01 – 9999.9 ml
    • Time Range: 0 – 99 hours 59 minutes
  • Body-Weight Mode: Select Unit, enter Body Weight, Concentration, Dose, VTBI
    • Dose Rates: mcg/min, mcg/hr, mcg/kg/min, mcg/kg/hr, mg/min, mg/h, mg/kg/min, mg/kg/h plus more available
  • Drug Library Mode: Select Drug, enter Concentration, Dose Rate, Weight
    • Drug library is editable by going to settings – Drug Library – DrugLib maintenance – enter password 8888 – options to edit drugs and change password
  • Sequence Mode: Set Rate1, set Time1, then proceed to following sequences as needed
Bolus Operation
  • Purge: While the pump is idle, press Purge on the screen and select Yes or Press and Hold the Purge Softkey (see Figure 1)
  • Programmed Bolus: During infusion Press Bolus on screen and set Bolus VTBI, Bolus rate, and Bolus time, and then select start (see Figure 2 and Figure 3)
  • Manual Bolus: During infusion, press and hold Bolus softkey (see Figure 2) – Audible tone every 1 ml bolus infused
  • To change the default Manual Bolus rate: while the pump is idle, go to Menu, Settings, then select Bolus rate (See Figure 4)

Settings

To enter additional settings, select the Menu button on the touchscreen or softkey

  • Modes: Change the current Infusion Mode
  • Settings: Change current IV admin Brand, drug library, default KVO(Keep Vein Open) Rate (0.01 ml/hr – 5.0 ml/hr), default Bolus rate, Occlusion Pressure (4 Levels 50 mmHg – 900 mmHg), Bubble sensor size (8 Levels 10 ul – 800 ul), Pump idle alert time, Almost empty alarm, Micro Mode
  • General: Rotate Screen ON/OFF, Change Date & Time, Screen Brightness, Sound Level, Screen Lock time, Night Mode, and Battery Capacity Indicator
  • Records: Recall History of alarms and recent therapies
  • System: Change Language, Factory default, Version, and Maintenance
  • Home: Go back to the Home screen
  • Standby Mode: To Access Standby Mode press the Power button, select Standby and enter standby time

Service Information

All Warranty and Non-Warranty Repair Services will be performed by Triumph Medical Services. When sending your pump in for repair, please go to the following link and fill out the Veterinary RMA Request form: www.triumphmed.com/vet-rma-request. The shipping address will be provided when your RMA confirmation is emailed to you. For any questions, call 888.962.0662, ext 3

Questions?

Distributed by

  • Covetrus® North America
  • 400 Metro Place North Dublin, OH 43017
  • covetrus.com

Reorder: 077737

Made in China

© 2023 Covetrus, Inc. All rights reserved. No copying without consent.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

covetrus User Manuals

Related Manuals