Adam Equipment CBD Series Bench Counting Scales User Manual
- June 12, 2024
- Adam Equipment
Table of Contents
- Adam Equipment CBD Series Bench Counting Scales User Manual
- INTRODUCTION
- Details of your company
- Details of the unit purchased
- Brief description of the problem
- TROUBLE SHOOTING GUIDE
- ERROR CODES
- CBD DESCRIPTION
- SETTING UP THE REMOTE SCALE
- REMOTE SCALE CONNECTION
- REMOTE SCALE SET UP
- USER PARAMETERS
- USER CALIBRATION
- TECHNICAL PARAMETERS
- LOAD CELL DAMAGE
- PRINCIPLE OF OPERATION
- TESTING LOAD CELLS IN THE SCALE ****
- REPLACEMENT PARTS AND ACCESSORIES
- SCHEMATICS
Adam Equipment CBD Series Bench Counting Scales User Manual
INTRODUCTION
**** The CBD series of scales are general purpose counting scales. Refer to
the Operators Manual for details of operation.
This manual will cover details of calibration and troubleshooting.
If you have a problem with the scales that is not directly addressed by this
manual then contact your dealer or Adam Equipment for more assistance.
In order to provide further assistance, the supplier will need the following
information which should be kept ready:
Details of your company
-Name of your company:
-Contact person’s name:
-Contact telephone, e-mail, fax or any other methods:
Details of the unit purchased
(This part of information should always be available for any future correspondence. We suggest you to fill in this form as soon as the unit is received and keep a print-out in your record for ready reference.)
Model name of the scale: | CBD __ |
---|---|
Serial number of the unit: |
Software revision number
(Displayed when power is first turned on):
|
Date of Purchase:
|
Name of the supplier and place:
|
Brief description of the problem
Include any recent history of the unit. For example:
-Has it been working since it’s delivered
-Has it been in contact with water
-Damaged from a fire
-Electrical Storms in the area
-Dropped on the floor, etc.
TROUBLE SHOOTING GUIDE
PROBLEM POSSIBLE CAUSE
Display is blank Power Switch is
faulty or not turned on
No turn on test Battery is faulty
or not charged Power supply is incorrect
Display all zeros after turn on Pan not installed or E6 is
displayed Unstable weight Load cell damaged Electronics faulty
appears on display Maximum capacity
exceeded
Load Cell damaged Electronics is faulty
Display is unstable Drafts or air
currents Obstruction under pan Sample is moving (animal weighing) Vibrations
through table Temperature changed dramatically Electronics faultBattery low
Weight value incorrect Calibration error,
Recalibrate Unit calibrated with inaccurate weight Balance not level
Obstruction between sample and cover Wrong unit of weight displayed
Cannot use Full Capacity Over-load Stops hitting pan
support or hitting bottom of load cell Electronics faulty Parameters set
incorrectly Load Cell Damaged
Not Linear Overload stops
hitting too soon Faulty Factory Calibration
Off Centre Loading error Overload Stops not correct
Obstruction under pan
Battery will not charge Charging circuit failure
Battery failure Incorrect adapter (below 800 mA) Processor failure
ERROR CODES
ERROR CODE | DESCRIPTION | POSSIBLE CAUSES |
---|---|---|
Err 4 | Initial Zero is greater than the permissible value (typically 4% of |
maximum capacity) when power is turned on or when the [Zero] key is pressed,| Weight on the pan when turning the scale on.
Excessive weight on the pan when zeroing the scale.
Improper calibration of the scale.
Damaged load cell. Damaged Electronics.
Err 5| Keyboard error.| Improper operation of the scale.
Err 6| A/D count is not correct when turning the scale on.| Platform is not
installed. Load cell may be damaged.
Electronics may be damaged.
FAIL H or FAIL L| Calibration error| Improper calibration.
If the problem persists contact your dealer or Adam Equipment for assistance.
CBD DESCRIPTION
The CBD scales have an enclosure with all components mounted within it. To
gain access to the components remove the 4 screws securing the cover to the
base.
The basic unit consists of:
- Base
- Load Cell frame
- Power switch
- Main PCB assembly
- Battery
- Display PCB assembly
- Keypad
All models of CBD are similar except the selection of load cells and the
program.
Normally if a problem is found with a circuit board the most cost effective
method of solving the problem is to replace the circuit board.
Schematics are attached in order to complete the documentation. However it is
rare that the problem due to a faulty component on the PCB is traced by a
user.
The main PCB assembly includes:
- A/D converter (AU1),
- Microprocessor (DU1) and
- Power supply/battery charger circuits (DU5, T2).
SETTING UP THE REMOTE SCALE
The CBD Series can be connected to any size of load cell type weighing base
via the Remote scale port on the right side of the scale case. Ensure you have
the correct base for the scale as each is matched for calibration.
Place the remote scale platform in the position where it is to be used. Level
the scale by adjusting the four feet. If fitted with a spirit level then it
should be adjusted such that the bubble is in the centre.
Press [Local/Rem] and test weighing performance.
REMOTE SCALE CONNECTION
The cable for the load cell goes to a 9 pin D-subminiature plug connector with the following connections:
Pin numbers | Connection |
---|---|
Pins 1,2 | – Excitation (0v) |
Pins 4,5 | + Excitation (+5v) |
Pin 7 | + Signal |
Pin 8 | – Signal |
(The sense wires connections of a six wire load cell are not used but can be connected to the respective Excitation pins).
REMOTE SCALE SET UP
The remote scale should set for a realistic resolution with respect to the
input provided by the load cell/s.
If a single 2mV/V load cell is fitted and more than 60% of the load cell is
used for full capacity then the high output of >6mV span makes it possible to
set a high resolution
If this criterion is met then the remote scale can be set to a high resolution
with a maximum of 1:30,000, i.e. 300kg x 10g.
It will also be possible to sample on the remote scale with the same accuracy
as the Local.
Where more than one load cell is fitted or the total load cell capacity is not
utilised then a reduced resolution should be selected in the remote scale
technical set up. For example, if a system uses four 2mV/V 1000kg load cells
for a scale of 1000kg capacity then the span output at full scale will be only
2.5mV.
In this situation the resolution should be reduced to give a good number of
ADC counts per displayed division. ie. Set to 1:5000 or 1000kg x 0.2kg.
Setting a high resolution without providing a good input to the remote scale
ADC will not give better accuracy and may make the scale difficult to meet
performance specification.
For best performance ensure a minimum of 0.1mV/d.
USER PARAMETERS
Enter the User Parameters section by pressing the [Pst] key during the
self-test when the power is turned on. This will allow the user to set the way
he wants the scale to work by choosing specific values from some options.
Press the [U.Wt./Units] key to scroll through the other parameters. To
enter any parameter, press the [Print] key.
Press [U.Wt./Units] to scroll through the sub-parameters. To see the
earlier setting, press the [Print] key. To change and scroll through the
other available settings, press [U.Wt.]. To choose the desired setting and
thereby go back to the sub-parameter, press the [Print] key.
To return to the parameter, press the [Tare/Zero] key.
PARA- METER | SUB- PARAMETER | DISPLAYS AND SETTINGS |
---|---|---|
F1 oFF | bEEP | “bEEP”“ “oFF”” |
“bEEP”” “on I n “” | Beeper is set to on between limits | |
“bEEP” “”on oUt” ” | Beeper is set to on outside limits (>0) | |
EL | “LitE”” “”oFF”” | Backlight is set to off |
“LitE”” “on”” | Set to on at all times | |
“LitE”” “AUt” | Set to work automatically when a weight is placed on the scale |
or a key is
pressed.
Un I t| “UnI t” ” KG/ Lb”| Kg/Lb both are enabled
”UnI t” “KiLo”| Kg only is enabled
”UnI t” “Lb” | Lb only is enabled | ||
---|---|---|---|
F2 Prt | P Mo dE | Au oFF Prints only when |
the Auto- Accumulation is set to off.
Au on Prints only when
the Auto- Accumulation is set to on.
P Cont| Sets the RS-232 interface to print continuously and the accumulation function
is disabled.
SEr r E| Sets the RS-232 to print continuously the weight only.
P bAU d| b 600
b 1200
b 2400
b 4800
b 9600
| Sets the required baud rate (speed for the RS- 232 communications). Default
rate is 4800.
PAritY| 8 n 1
7 E 1
7 o 1
| 8 data bits, no parity
7 data bits, even parity 7 data bits, odd parity
U id| “U id”
“ Abc234” “ “
| Shows the current user ID (if any).
Enter a new User ID as described in the Description under the PLU section. The ID can be alpha- numeric but is limited to 6 characters.
SC id| “Sc id”
“ Abc234” “ “
| Shows the current scale ID (if any). Enter a new Scale ID as described in the
Description under the PLU section. The ID can be alpha-numeric but is limited to 6 characters.
tECH| | Allows access to the Technical parameters using a password. Not normally accessed by user.
USER CALIBRATION
ACTION | DISPLAYS |
---|
Press the [Tare/Zero] key during the self-test at power on. The scale will ask to enter the password.| ****
“ Pi n”
Default Password is 0000. Enter “0” four times.
Password can be changed in technical parameters.
Press [Print].
| ****
“ Pi n” “—- ”
Select the scale to be set up by using the [Local/Remote] key.
Press the [Print] key to enter the technical section.
| “tEch” “ LocAL” “ “
“tEch” “rEmo tE” “ “
Use the [U.Wt.] to select the weighing unit to be used for setting of the
calibration. The arrow in the “ Weight ” window will indicate the unit
selected. Press the [Print] key to continue.| “tEch” “ Uni t” “ “
You will enter the first parameter – Calibration. Display will ask to unload
any weight on the platform.
Press the [Print] key to continue.
| “Un L o Ad”
Enter the calibration weight to be loaded and press [Print].
Load the calibration weight onto the scale and press [Print] again.
The Calibration is complete, the scale will run the self-test during which the weight should be removed.
| “SEL” “ 0010”
“Lo Ad”
TECHNICAL PARAMETERS
The technical parameters are accessed via the “tEch” prompt at the end of user parameters and are password controlled to prevent unauthorised access. These parameters set the metrology for the scales. Each scale is set independently. The parameters will set capacity, division, decimal point position, initial zero range, auto and manual zero range as well as factory calibration.
ACTION | DISPLAYS |
---|
From the “tech” user parameter, Press the [Print] key. The scale will ask you to enter the password.|
“ Pi n”
Default Password is 0000. Enter “0” four times.
9999 will override any other user password. Press [Print].
|
“ Pi n” “—- ”
Select the scale to be set up by using the [Local/Rem] key.
Press the [Print] to enter the technical section for that scale.
|
“t Ech” “ Lo c A L” “ “
“t Ech” “r E mo t E” “ “
Use the [U.Wt./Units] to select the weighing unit to be used for setting up the parameters for the scale. The arrow in the “ Weight ” window will indicate the unit selected.
Press the [Print] key to continue.
|
“t Ech” “ Uni t” “ “
Press the [U.Wt./Units] key to scroll through the menus for the chosen scale. User [Print] to select and set parameters and the
[Tare/Zero] to escape as without changing data as with the user parameter section.
|
“Cn t”
Both scales have the same menus except that the capacity of the local scale is fixed and cannot be entered.
Press the [U.Wt/Unit] to go to next menu. Press [Print] to enter.|
Use [U.Wt/Unit] to select the decimal point position and press [Print].
To change the remote scale cap press CE to clear and then enter new value followed by [Print].
| “CAp”. Scale capacity setting.
“dESC” “ 0.00”
“SEL” “0060”
---|---
Press the [U.Wt/Unit] to go to next menu. Press [Print] to enter.
Use [U.Wt/Unit] to select the division which you would like the display to increment in followed by [Print].
|
“div” weighing division.
“inc 5”
Press the [U.Wt/Unit] to go to next menu, use [Print] key to enter.
Use [U.Wt/Unit] to increment the values then [Print].
| “Azt” Autozero tracking range.
Select from 0.5d, 1d, 2, 4,d
Press the [U.Wt/Unit] to go to next menu, use [Print] key to enter.
Use [U.Wt/Unit] to increment the values then [Print].
| “0 Auto” Zero auto range at power on.
Select form 0%, 2%, 5%, 10% or 20%
Press the [U.Wt/Unit] to go to next menu, use [Print] key to enter.
Use [U.Wt/Unit] to increment the values then [Print].
| “0 manl ” Zero manual range.
Select form 0%, 2%, 4%, 10%, 50% or
100%
Press the [U.Wt/Unit] to go to next menu, use [Print] key to enter.
Enter the new PIN number and press
[Print].
Confirm the new PIN number and press [Print].
| “Pin” Password number for “tEch”.
“Pin1”
“Pin2”
“donE”
Press [U.Wt./Units] to continue or escape the parameters with the
[zero/Tare]
|
LOAD CELL DAMAGE
The most common reason a scale fails is that the load cell has been damaged. The damage can be from 2 primary causes. The first is physical damage due to an overload or an impact from the side and the second is damage due to the environment, such as moisture, extreme heat or a cut cable if the cables are exposed.
PRINCIPLE OF OPERATION
The stain gauge load cell is a method of measuring the amount of stress put onto a metal structure due to a weight being supported by the structure. The stress is measured using 4 strain gauge sensors mounted in a particular way on the metal structure.
These 4 strain gauges are connected in such a way that they form a wheatstone bridge. See figure below. In the simplest load cells there are only the 4 resistances of the strain gauges to consider. Normally without a load on the load cell all the resistances are the same. However when the load cell has a weight on it, two of the strain gauges will be in compression (A and D) and the resistance will decrease and two will be in tension (B and C) and their resistance will increase.
This will cause the bridge to become unbalanced. When the bridge has a voltage across it from +E to -E then the signal output at +S and -S will show a voltage of zero volts with no load and a small voltage proportional to the load as the load is increased. Typical load cells show 20-30mv of signal if the excitation voltage is 10volts and the load cell is fully loaded.
Most load cells used in scales are not as simple as this example. They have additional resistance elements added to compensate for temperature variations and to set the outputs to correct voltages. The circuit of a typical load cell is shown below.
A method to check the basic function and integrity of a load cell can be done
using an ohmmeter and voltmeter with up to a 10volt power supply.
Before the load cell is connected to the power supply use the ohmmeter to
measure the resistance between the wires. Typical resistance values are:
+E to -E 410 ohms ±30 ohms
+S to -S 350 ohms ± 2 ohms
+E to +S or any other combination similar approximately 270 ohms.
If the load cell has Sense connections in addition they are connected to the
excitation internal to the load cell.
Connect the load cell +E and -E to a power supply, typically 10 volts. Never
use more than 12 volts as it may damage the load cell. Measure the voltage
between +S and -S.
With no load the voltage should be approximate 0mv ±3mv.
With a mass on the load cell the output voltage should increase. The amount of
increase will depend upon the sensitivity of the load cell, capacity of the
load cell, the excitation voltage and the amount of mass placed on the load
cell. Most load cells have a sensitivity of either 2mv/V or 3mv/V.
The expected change to the output is:
Vout = (2mv/V) Vext Mass capacity of load cell
For example a 2mv/V load cell of 30Kg capacity with 20Kg placed on it. Vext is
5 volts would give a output of approximately:
Vout = (2mv/V) Vext Mass = 2mv/V 5V 20Kg = 6.7mv capacity of
load cell 30kg
If the load cell has been damaged the no load voltage will likely be greater
than 3mv or the loaded voltage will be grossly different from the expected
value. Remember the loaded voltage will be offset by the amount of zero load
voltage measured.
The load cell can be tested while it is connected to the A/D converter circuit
board, using the scale power supply for excitation voltage.
TESTING LOAD CELLS IN THE SCALE ****
The signal from the load cells is amplified by the circuits on the A/D circuit
board.
The signal needs to be amplified to make it acceptable for the A/D converter
used. The CBD/CBW scales amplify the input by 200, and change the no load
voltage to about +2.5V.
The amplified and offset voltage is measured at pin 6 of IC7, the INA118P
amplifier IC. This voltage is sent to the filter and then to the A/D
converter.
REPLACEMENT PARTS AND ACCESSORIES
If you need to order any spare parts and accessories, contact your supplier or Adam Equipment. A partial list of such items is as follows-
SCHEMATICS
**** The following schematics are included as a guide for the experienced
Service Technician. If you are not experienced we do not advice you to change
parts on the circuit boards or otherwise disturb the electronics.
If you require assistance please contact your dealer or Adam Equipment. We
will be happy to help you resolve problems or determine the cause of a problem
if you send us a full description of the scale and the problem.
CBD Parts List
Part Number | Revision | Diagram Number | Description | Qty Used |
---|---|---|---|---|
1.0001 | 1 | Stainless Steel Pan | 1 | |
1.0002 | 2 | ABS Pan | 1 | |
1.0044 | 3 | Front Overlay | 1 | |
1.0045 | 4 | Keyboard | 1 | |
1.0004 | 5 | Upper case | 1 | |
0.0026 | 6 | Screw M6 x 20 | 2 | |
1.0025 | 7 | Upper Bracket of Load Cell | 1 | |
1.0026 | 8 | Load Cell | 1 | |
0.0013 | 9 | Screw M4 x 20 | 4 | |
1.0046 | 10 | Washer | 4 | |
1.0027 | 11 | Bottom Bracket of Load Cell | 1 | |
0.0008 | 12 / 28 | Screw M4 x 12 | 4 | |
1.0047 | 13 | Washer | 4 | |
1.0048 | 14 | Main PCB | 1 | |
1.0022 | 15 | Display PCB | ||
1.0011 | 16 | Bottom Cover | 1 | |
0.0011 | 17 | Case screw M4 x 15 | 5 | |
1.0049 | 18 | Serial Port RS-232 | 1 | |
1.0050 | 19 | Duel Scale connector | 1 | |
1.0032 | 20 | Power Switch | 1 | |
1.0012 | 21 | Foot | 4 | |
0.0003 | 22 | Screw M3 x 10 | 4 | |
0.0013 | 23 | Screw M4 x 20 | 4 | |
1.0014 | 24 | Battery Cover | 1 | |
1.0051 | 25 | Internal Transformer (SA only) | 1 | |
1.0052 | 26 | Nut | 1 | |
1.0057 | 27 | Serial PCB board | 1 | |
0.0008 | 28 / 12 | Screw M4 x 12 | 4 | |
1.0010 | 29 | Battery (6V / 4MH) | 1 | |
1.0024 | 30 / 31 | Washer | 2 | |
1.0003 | 32 | Rear Overlay | 1 | |
0.1003 | AC Adapter 9VDC 800mA-1A | 1 | ||
0.5006 | In use wet cover |
WARRANTY INFORMATION
Adam Equipment offers one year Limited Warranty (Parts and Labour) for the
components failed due to defects in materials or workmanship. Warranty starts
from the date of delivery.
During the warranty period, should any repairs be necessary, the
purchaser must inform its supplier or Adam Equipment Company. The company or
its authorised Technician reserves the right to repair or replace the
components at the purchaser’s site or any of its workshops depending on the
severity of the problems at no additional cost. However, any freight involved
in sending the faulty units or parts to the service centre should be borne by
the purchaser.
The warranty will cease to operate if the equipment is not returned in
the original packaging and with correct documentation for a claim to be
processed. All claims are at the sole discretion of Adam Equipment.
This warranty does not cover equipment where defects or poor performance
is due to misuse, accidental damage, exposure to radioactive or corrosive
materials, negligence, faulty installation, unauthorised modifications or
attempted repair or failure to observe the requirements and recommendations as
given in this User Manual.
Repairs carried out under the warranty does not extend the warranty
period. Components removed during the warranty repairs become the company
property.
The statutory right of the purchaser is not affected by this warranty.
The terms of this warranty is governed by the UK law. For complete details on
Warranty Information, see the terms and conditions of sale available on our
web-site.
Manufacturer’s Declaration of Conformity
This product has been manufactured in accordance with the harmonised European
standards, following the provisions of the below stated directives:
Electro Magnetic Compatibility Directive 89/336/EEC Low Voltage Directive
73/23/EEC
Adam Equipment Co. Ltd. Bond Avenue, Denbigh East Milton Keynes, MK1 1SW
United Kingdom
FCC COMPLIANCE
****This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. The equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
Shielded interconnect cables must be employed with this equipment to insure
compliance with the pertinent RF emission limits governing this device.
Changes or modifications not expressly approved by Adam Equipment could void
the user’s authority to operate the equipment.
Any Electrical or Electronic Equipment (EEE) component or assembly of parts intended to be incorporated into EEE devices as defined by European Directive 2002/95/EEC must be recycled or disposed using techniques that do not introduce hazardous substances harmful to our health or the environment as listed in Directive 2002/95/EC or amending legislation. Battery disposal in Landfill Sites is more regulated since July 2002 by regulation 9 of the Landfill (England and Wales) Regulations 2002 and Hazardous Waste Regulations 2005.
ADAM EQUIPMENT is an ISO 9001:2000 certified global organisation with more than 30 years experience in the production and sale of electronic weighing equipment. Products are sold through a world wide distributor network supported from our company locations in the UK, USA, SOUTH AFRICA and AUSTRALIA.
ADAM ’s products are predominantly designed for the Laboratory,
Educational, Medical and Industrial Segments. The product range is as follows:
-Analytical and Precision Laboratory Balances
-Counting Scales for Industrial and Warehouse applications
-Digital Weighing/Check-weighing Scales
-High performance Platform Scales with extensive software features including parts counting, percent weighing etc.
-Crane scales for heavy-duty industrial weighing
-Digital Electronic Scales for Medical use
-Retail Scales for Price computing
Adam Equipment Co. Ltd.
Bond Avenue Milton Keynes MK1 1SW
UK
|
Adam Equipment Inc. 26, Commerce Drive Danbury, CT
06810
USA
|
Adam Equipment S.A. (Pty) Ltd.
7 Megawatt Road, Spartan EXT 22, Kempton Park,
Johannesburg Republic of South Africa
|
Adam Equipment (S.E. ASIA) Pty Ltd.
2/71 Tacoma Circuit Canning Vale, Perth WA 6155, Australia
---|---|---|---
Tel:+44 (0)1908 274545
Fax: +44 (0)1908 641339
| Tel: +1 203 790 4774
Fax: +1 203 792 3406
| Tel: +27 (0)11 974 9745
Fax: +27 (0)11 392 2587
| Tel: +61 (0) 8 6461 6236
Fax: +61 (0) 8 9456 4462
E-mail: sales@adamequipment.co.uk| E-mail: sales@adamequipment.com| E-mail: sales@adamequipment.co.za| E-mail: sales@adamequipment.au.com
© Copyright by Adam Equipment Co. Ltd. All rights reserved. No part of this
publication may be reprinted or translated in any form or by any means without
the prior permission of Adam Equipment.
Adam Equipment reserves the right to make changes to the technology, features,
specifications and design of the equipment without notice.
All information contained within this publication is to the best of our
knowledge timely, complete and accurate when issued. However, we are not
responsible for misinterpretations which may result from the reading of this
material.
The latest version of this publication can be found on our Website.
Visit us at www.adamequipment.com
Read More About This Manual & Download PDF:
References
- All-guides – Innovative Search Service of Online Manuals
- au
- Laboratory & Industrial Weighing Scale Manufacturer - Adam Equipment USA
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>