NEUATION iFUGE M24PR Refrigerated High Speed Micro Centrifuge User Manual
- June 12, 2024
- NEUATION
Table of Contents
- NEUATION iFUGE M24PR Refrigerated High Speed Micro Centrifuge
- INTRODUCTION
- FEATURES
- STANDARD ACCESSORIES
- TECHNICAL SPECIFICATIONS
- SAFETY PRECAUTIONS
- INSTALLATION
- STANDARD PARTS LISTING
- USER INTERFACE & DISPLAY
- ROTOR INSTALLATION
- OPERATING THE CENTRIFUGE
- MAINTENANCE AND CLEANING
- TROUBLESHOOTING
- WARR ANTY STATEMENT
- PRODUCT DISPOSAL
- TRANSPORTATION & STORAGE
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
NEUATION iFUGE M24PR Refrigerated High Speed Micro Centrifuge
INTRODUCTION
This manual provides important safety information for this centrifuge. lt should be kept near the centrifuge for quick & easy reference. This centrifuge is equipped with a maintenance-free brushless DC motor drive with a microprocessor controlled that can speed up to 15,000 RPM and a safety feature of imbalance detection. This centrifuge is equipped with refrigeration which can be maintained as low as up to -20°C up to 4o•c.
INTENDED USE
This microcentrifuge is designed to separate, sediment, spin down aqueous
solutions & solvent suspensions of differing densities in compatible sample
containers.
NOTE:
Before using the centrifuge, please read this user manual carefully. This user
manual is intended to assist with the operation and care of the unit & is not
a document that aids in repair. For repair please contact the supplier.
SYMBOL
FEATURES
- Maximum volume of 44 x 1.5/2 ml*
- Temperature ranges from -20°C to 40°C
- The temperature of 4 ·c maintained even at high speed
- Imbalance detector with auto cutoff for safe operation
- Brushless DC motor for maintenance-free performance
- Large LCD display with speed and time indicator
- Speed setting from 500-15000 RPM
- One press changeover from RPM to RCF functionality
- Digital countdown Timer
- Reduction adaptors for 0.1/0.2 & 0.5ml tu bes
- Emergency lid release during power failure
- Automatic internal diagnosis & error display
- Silent operation
(*Available with 4 optional rotors)
STANDARD ACCESSORIES
- T- Allen wrench
- Adaptors for 0.4/0.5 ml tu bes
- Adaptors for 0.1/0.2 ml tubes
- Product user manual and warranty card
- Power cord
NOTE: Rotors has tobe ordered separately. (optional accessory)
TECHNICAL SPECIFICATIONS
Motor Type | Brushless DC Motor |
---|---|
Rotor Capacity | 24 x l.5/2ml tube, 44 x l.5/2ml tube, |
8 x 4 x 0.1/0.2ml PCR tube/strips, 8 x 5ml tube
Speed Setting| Variable 500 – 15000 RPM
22388 x g (Max RCF)*
Speed Step| ± 100 RPM
Speed Accuracy| ± 100 RPM
Temperature Accuracy| ± 1 ·c
Temperature Range| -2o·c to 4o•c.
Run Time| 30s to 999min & 59s
Acceleration Time| < 72 seconds
Deceleration Time| < 74 seconds
Noise Level
Ambient Temperature
| <60 dB
580%
Permissible Relative Humidity| 20 to 80%
Size (W X D X H)| 325 x 720 x 305 mm
Weight| 43 kgs
Refrigerant| R134a
The total mass of Refrigerant| 251 g
Maximum allowable pressure (PS) under normal conditions| 250 psi
Altitude| Use upto an altitude of 2000 m above MSL
Pollution Degree| 2
---|---
Environment
Power Supply
| For indoor use only
230 VAC, S0Hz
Power Consumption| 650W
Safety Fuse Rating| 10Amp
(*RCF varies for each rotor congregation -As mentioned in point no 11 (rotor Installation).
SAFETY PRECAUTIONS
Read all safety & usage information provided in this manual carefully before using the device.
- Never use the centrifuge in any manner not specified in this manual.
- Using equipment in any manner not specified in this manual or by the manufacturer, will void the warranty.
- Never move the centrifuge while the rotor is spinning.
- The rotor and the rotor lid must always be securely fastened. If the centrifuge makes an unusual noise during operation, the rotor or rotor lid fit needs to be checked. Switch OFF the device immediately by pressing STOP, check the rotor fit & fasten it well.
- The rotors must be loaded symmetrically. Each tube should be counterbalanced by another tube of the same weight.
- Do not use the centrifuge or rotor that has not been correctly installed or damaged.
- Repairs must only be performed by an authorized service technician.
- Using incorrect rotors & wrong spare parts will void the warranty.
- The centrifuge may be used for specified applications only. lt must not be operated in a hazardous or flammable environment and must not be used to centrifuge explosive or highly reactive substances.
- If liquids are spilled on the rotor or rotor chamber, the centrifuge must be cleaned carefully and properly before being used again.
- Prior to centrifugation, the tubes should be visually inspected for material damage. Damaged tubes may not be centrifuged. This is because broken tubes can, in addition to sample loss, create an imbalance which can result in further damage to the centrifuge and accessories.
- The maximum capacity of tubes must not exceed its corresponding rotor capacity. Do not use liquid with a density higher than l.2g/ml for full load operation.
- Do not lean on the equipment. it may damage the equipment or harm the operator.
- When moving the centrifuge from a cold room to a normal room, run the centrifuge for 30 minutes beforehand in the cold room to avoid condensation. Alternately, allow it to warm up in the lab for at least 3 hours before use, but do not plug in the centrifuge in order te prevent possible damage by condensation.
- Be sure to close the tu bes lid tightly prior to centrifugation. Open tu bes lid can be tarn off du ring centrifugation and can damage the rotor lid or centrifuge.
- Rotors and rotor lids are high-graded components which are subject to extreme mechanical strain. Even slight scratches and tears can lead to serious internal material damage. Ensure to check the rotor for any signs of damage before use. Rotor & rotor lid showing visible signs of corrosion or mechanical damage should not be used, contact your local supplier or mfg for replacement rotor & parts.
- Do not fill tu bes while they are in the rotor. Liquid spillage may harm the device.
- In the event of contamination caused by aggressive agents, the rotor must be cleaned immediately using a natural cleaning liquid (like water). This is particularly important for the bares of the tubes. lf any damage is seen, contact the service technician.
- Before using cleaning or decontamination methods, other than those stipulate by the manufacturer, contact the manufacturer to ensure that the intended method will not damage the centrifuge.
- The power cord given with centrifuge unit is designed to use for that particular centrifuge. Do not use any other power cord, using any other power cord /adaptor may damage the centrifuge and will void the warranty.
- Safety Fuse is provided of 10Amp configuration which can be replace by the operator. The same will protect the machine circuit during an electrical fault or overload.
INSTALLATION
Open the box, then remove the packaging and gently take the centrifuge out of the box. Before using this centrifuge open the centrifuge & remove all packaging from the rotor chamber & ensure rotor is firmly tightened. The user manual and accessories should be kept near the centrifuge. Please keep all packaging in safe storage for at least 2 years for warranty purposes.
MOUNTING
Place the centrifuge on a flat and leveled surface, ensuring that the four
feet of this centrifuge stand on the surface firmly. Avoid installing on a
slippery surface or surface prone to vibration.
- The ideal ambient temperature is 20°C ± S°C, avoid placing the centrifuge in direct sunlight.
- Keep a clearance of at least 30-35 cm on both sides and at least 20 cm behind it to guarantee cooling efficiency.
- Keep away from heat or water to avoid sample temperature issues or centrifuge failures.
- Do not place the centrifuge in any area where operating the unit may be difficult.
CONNECTING POWER CORD
Connect one side of the power cord to the rear side of the centrifuge and the other to the supply as shown in the figure below.
STANDARD PARTS LISTING
USER INTERFACE & DISPLAY
ltem | Name | Function |
---|---|---|
1 |
RPM/RCF
| Press button to select SPEED value in rpm mode. And
press+/- key to set desired run RPM. Press and Hold to switch to RCF mode. Press+/- key to set desired RCF.
2
|
TEMP
| Press button to set the temperature value in a range from 4°C to 40°C. And
press +/- key to set desired temperature.
3
4
| PROG
LINE
| Press button to enter in Program Mode and to select
any Program out of the 99 available programs.
Press button to select line of any specific program out of 4 available lines in each program.
5| TIME| Press button to select TIME mode. And press+/- key to
set desired run TIME.
6| START| Press button to START to start the centrifuge.
7
|
STOP/OPEN
| Single press Stop/Open button, will stop motor. Second press Stop/Open
button, will open the lid once the motor stop. Also used for lid open if
centrifuge motor is not running.
8| +/- Keys| Use to set speed, Temperature and time values. Press “+” to
increase and “-” to decrease values.
9| QUICK COOL| To start a temperature control run at optimized (S000)RPM to
achievemaximum cooling
10| QUICK SPIN| Ta set short spin an a set speed. Press continues press for
Quick spin operation.| short| spin
---|---|---|---|---
11| ROTOR| Ta select among multiple rotor listed below.
ROTOR INSTALLATION
ROTORS AND ACCESSORIES
The table below shows rotors compatible with the centrifuge and the Max. RCF
with different tu bes and adaptors.
NOTE:
Always use recommended accessories for best results and product safety.
Optional rotors & additional accessories must be ordered separotely.
ROTOR REMOVAL AND REPLACEMENT PROCESS
The rotor comes pre-installed with the centrifuge. lf you want to remove or
replace the rotor, follow the instructions below.
REMOVING THE ROTOR
- Do not remove or loosen the rotor lid before attempting to remove the rotor.
- Using the T – Allen Key, loosen the rotor nut by turning it eo unter clockwise. Do not tryto pull the rotor, the rotor will come up automatically.
- Once the rotor nut is loosen completely, pull up the rotor vertically.
REPLACING THE ROTOR
- To replace or install the rotor, take the rotor and load vertically onto the shaft.
- Place the rotor nut in the center hole of the rotor onto the motor shaft.
- Put T-Allen Key into the rotor nut and tur n clockwise to tighten and counter clockwise to loosen the rotor.
- After properly fastening the rotor, place the rotor lid on the rotor lid nut by hand and rotate the rotor lid nutclockwise.
NOTE:
- Check the rotor is firmly tightened before running the next program.
- Do not remove or loosen the rotor lid before removing the rotor.
BALANCING THE ROTOR
- Always balance the rotor before centrifugation. Above are examples of properly balanced rotors.
- The samples in the tu bes should be of equal volume.
- If the tu bes are not loaded correctly – vibration or imbalance can occur which can cause serious damage to the centrifuge.
- If the tu bes are not loaded symmetrically then the imbalance detector will cut off the running centrifuge for device & user safety. This will stop the centrifuge and Err 55 will be seen indicating tubes are not loaded symmetrically. To resume operation, load tu bes symmetrically & restart the centrifuge.
- Incorrect method of loading tubes in centrifuge rotor:
OPERATING THE CENTRIFUGE
STARTING THE CENTRIFUGE
After connecting the power cord, switch ON the power at the rear side of the centrifuge. Make sure to check the rotor fit before use. Centrifuge will not operate with an open lid.
NOTE:
Maintain a gap of 3 seconds between turning off and turning on the centrifuge.
00 NOT turn off and turn on the centrifuge instantly.
SWITCH ON THE CENTRIFUGE
After connecting the power cord. Switch ON the main power supply & then Switch
on the power switch located on the right side of the instrument. Make sure to
check the rotor fitment before use. Centrifuge will not operate with open lid.
NOTE:
Maintain a gap of 3 seconds between switch OFF and switch ON again. 00 NOT
switch OFF and ON again instantly.
SPEED SETTING
After closing the centrifuge lid, press “RPM/ RCF” to select speed setting in
RPM / RCF mode. Press “+” to increase the speed value and press “-” to
decrease the speed value. The minimum and maximum speed (RPM) of the
centrifuge is 500 rpm and 15000 rpm respectively. You can quickly change the
increment / decrement value by rotating setting knob quickly. Value is
accepted when left idle indicated by multiple blinks of the set value. The
speed can also be changed while the centrifuge is under operation. Press the
RPM / RCF button & use setting knob to change speed as instructed above.
Changing the speed between the ongoing centrifugation will run the centrifuge
at updated speed for the remaining time indicated by the tim er.
TIMER SETTING
Press “TIME” button to set the run time. After selecting press “+” to increase
the timer value and press”-” to decrease the timer value. Value is accepted
when left idle indicated by multiple blinks of the set value. The centrifuge
timer can be set from 30 s to 999mins and 59s. The timer in the centrifuge is
countdown timer and time in the display will be in “Min/Sec”. The same will
showed on the display.
- Toset the timer in minutes, pressing the time button “ONCE”. For example: if time is 005:00 (5 Mins0 Sec), then next time will be 006:00 (6 Mins0 Sec).
- To set the tim er in seconds, pressing the time button “TWICE”. For example: if time is 005:00 (5 Mins 0 Sec), then next time will be 005:01 (5 Mins 1 Sec).
- Pressing time button more than 2 tim es will start the process from point 1 again.
- The input will be accepted if we leave the setting knob idle for 3 seconds. The value be saved automatically after five blinks.
TEMPERATURE SETTING
Press temperature button for setting the temperature from -20 °C up to 40 •c.
The parameters can be set when the Temperature value blinks by pressing “+”
key for increasing the temperature & “-” to reduce the temperature. The values
will automatically saves after screen stops blinking.
QUICK COOL SETTING
PRESS Quick Cool Key, indicated by (cubic rotation) on top right
(lndication will be in dotted rotation in place of rotor digit indication). Du
ring quick cool operation rpm:FIX 5000 & TEMP:-On set value. Press START
button for centrifuge to start at FIX RPM & set TEMP. After reaching set TEMP,
motor stops thus RPM becomes zero but compressor continues to keep the chamber
temperature at the set value. The quick cool works for 1 hour only. (for
indication of refrigeration running process, TEMP display will blink) lf
timers stops and TEMP does not reach to set value, TIME will switch-over from
beginning. “RPM/RCF, Tim er, Rotor, Line, Prog”, “Quick Spin” button not work
du ring “Quick Cool” operation.
SHORT SPIN OPERATION
Short Spin Centrifugation is the feature for short/pulse/quick run. lt will
run as long as the button is pressed. Set rotational speed prior to short spin
as required. During short spin the tim er will be in incremental mode. After
releasing short spin button the time in the display will show duration of
short spin.
ROTOR SELECTION
Different rotors has different radius which means that RCF for that rotor
would also be different. Proper selection of rotor is mandatory to achieve
proper RCF. Press “Rotor” button and then rotate setting knob to select the
rotor from available 4 rotors. Press “Rotor” button . See below chart to
identify rotor no. from its volume:
Rotors
Rotor No.| Name| rpm| RCF at max rpm
1| 24 slot x 2ml rotor| 15000| 21420 g
2| 4 x 8 PCR rotor| 15000| 15397 g
3| 8 slot x 5 ml rotor| 15000| 18461 g
4| 44 slot x 2ml rotor| 15000| 22388 g
seconds of the adjustment. Then so the max value of speed and acc/dcc to be changed as per the rotor selection. (Max. Speed (RPM/RCF) and Acc/Dcc vaule for all the rotors have been added in the software) Now set the desired speed (RPM/RCF), time & temp and press START button to store and run the selection.
START AND STOP OPERATION
Press “START BUTTON” to start operation and press “STOP/OPEN BUTTON” to stop
the ongoing operation. When the centrifuge is running the symbol “ ” will
rotate. Pressing the “STOP/OPEN BUTTON” will stop the operation. Once
operation is stopped, press Button again to open the centrifuge lid. lf the
time gets over, centrifuge will stop automatically. When the centrifuge is not
running the symbol “ ” will be idle. To open the lid in non-operating stage,
press the “STOP/OPEN BUTTON”.
After completing run, before any another operation, it is mandatory to open the lid once before starting new operation. Centrifuge will not do 2nd operation if lid is not opened and closed for at least one time after completion of 1 st operation.
NOTE:
It is mandatory to open the lid once after completion of operation for 2nd
operation. Centrifuge will not start if lid is not opened and c/osed after
completion of a operation.
PROGRAM MODE
Long press “PROG” button to enter into program mode. Short press select
program and rotate setting knob (clockwise & anticlockwise) to select desire
program out of 99 available programs. User can select any program from 1 to 99
programs and can have user specific setting parameter for all programs. Each
program has 4 lines indicated by Ll, L2, L3, L4. Press “LINE” button to enter
in line selection mode and rotate setting knob to select line out of 4 lines.
Every line of the program can have different values for all setting
parameters. User can set and save different speed, different time, different
temperature for any line of program. Only Rotor will not remain save for any
program. Rotor selection is not Program specific. Rotor selection is universal
and it should be set before setting any other parameters.
On first time usage, all programs will have zero (0) values and once the “PROG” button is pressed below display will appear indicating selection of program mode.
As per above image, the circle O left of “P0l Ll” is blank indicating no values in any line of specific program. Once any line of Program in filled or set, below change of display will appear.
The circle left to “P0l L3” will have a tick mark indicating that any one line of the specific program is filled. lf any line leading to selected line is set, then the arrow below the line number will appear.
lf no leading line is filled or last line L4 is selected then the arrow below Line will not appear. lt indicates there is no further operation left as leading lines of the specific program are empty. program and selected rotor remains active for all programs until new rotor number is not selected. Select the rotor, Press “PROG” button to enter in Program mode, select specific program using setting knob, Press “UNE” button to select specific line and set different parameter for that particular line of program. Values for any parameter gets save after 3 blinks. Once Program is set, press “START” button to start program.
After completion of any operation, it is mandatory to open the lid at least once for another operation. Open the lid, close it again and press STARTtor another operation.
NOTE:
- Rotor is not program specific. Selected rotar will be active for any new operation.
- It is mandatory to open the lid once after completion of operation for l”d operation. Centrifuge will not start if lid is not opened and closed after completion of ongoing operation.
IMBALANCE DETECTION
The centrifuge is equipped with an imbalance detection safety feature. When
the rotor is not loaded symmetrically, the imbalance detector gets activated
and will cut off the centrifugation. The error “Err 55” will be shown on the
display. First correct the imbalance load then switch OFF & switch it ON
again. The values will be same as set before imbalance. The imbalance
detection feature cannot be deactivated, as it is factory fitted.
OPENING CENTRIFUGE LID IN POWER FAILURE
Disconnect the centrifuge from the main power supply. Wait until the rotor has
come to a standstill (this may take longer time). Once the rotor has stopped,
Pull the Emergency Lid release lever by hand or by a small Allen Key located
on the left side of the machine as show in image below. This will open the
centrifuge lid.
NOTE:
This method of opening the lid shou/d be used only in case of Emergency or
power failure.
MAINTENANCE AND CLEANING
-
The rotor and the outside of the centrifuge should be cleaned regularly with a mild wet (with water) cloth.
-
Ensure that while cleaning the unit is not plugging in.
-
Wear protective glove & safety glass while operating & cleaning the device.
-
The brushless motor in the centrifuge requires no routine maintenance. Any required service should be performed by authorized, qualified personnel only. Repairs performed by unauthorized personnel may void the warranty.
-
Always keep the centrifuge housing, rotor chamber and rotor clean. All parts should be wiped down periodically with a soft cloth. For more thorough cleaning, use a neutral cleaning agent (Ph between 6 and 8) and clean with a soft cloth. Exclusive amounts of liquid should be avoided.
NOTE: Liquid shou/d not come into contact with the motor. -
After cleaning, ensure that all parts are dry before re-use.
-
Regularly cleaning of the rotor is important.
-
If the rotor chamber needs cleaning, clean with cloth or sponge moistened with a neutral detergent solution.
-
Do not place the rotor into the cleaning solution.
-
If corrosive, toxic or pathogenic bacteria are accidentally spilled in the rotor or rotor chamber the centrifuge must be decontaminated throughly.
TROUBLESHOOTING
This centrifuge has a seif- diagnostic function. lf a problem occurs, an error/warning code will be displayed on the display screen and the operator can determine the malfunction with the warning code below.
ERROR | PROBLEM | SOLUTION |
---|---|---|
No display | No main power connection. |
Power failure.
| Power check & proper plug-in of main cable at both ends.
Check the mains fuse.
lmproper connection.| Connect adaptor properly.
Error 1
| Lid not closed correctly.| Close lid correctly and restat unit.
Error with lid closing and opening mechanism. Latch jammed or any limit switch
of latch got damaged.| Contact service.
Contact service representative.
Error 41
| Selected temp value not reachable for selected set speed.| Set temp value will be taken
automatically that it can be archived after 1 hour, Error indication will be displayed every 15 second. restart unit.
Error 42| Temperature is out of control.| Error indication will be displayed every 15 second.
lf it persist in every operation consistently, contact service representative. restart unit.
Error 43| Over temperature inside chamber.| Turn off the centrifuge and wait
until temperature goes down. restart unit.
Error 44| Temperature sensor failure.| Contact service representative.
---|---|---
Error 52| Rotor stuck or incorrect operating voltage.| Turn off the centifuge,
check roter fitment or apply correct 230 VAC ± lOVAC operating voltage.
restart unit
Error 55| Rotor not loaded symmetrically.| Load roter symmetrically & restart
centrifuge. restart unit.
Centrifuge lid cannot be opened.
| Rotor is still spinning.| Wait for the roter to come to a stop.
Power Failure.| Emergency lid release after roter stop.
Centrifuge shakes during acceleration & exceptional running noise.| Rotor not
loaded symmetrically.| Load roter symmetrically & restart operation.
Either a broken tube, damage to the rotor or motor is cause for run noise.|
Replace broken tube. For damaged rotor/motor contact service representative.
Rotor damaged.| Remove & change roter.
Display Error.| Loose connection of display.| Contact service representative.
Power tripping.| Cable not fit properly.| Remove cable and connect properly.
System gets hang.| Electronic error.| Switch OFF centrifuge and then switch it
ON again. lf the error still shows, contact service representative.
IM PORTA NT NOTE
- Maintain a 3 second gap between restarting the centrifuge. Instant ON-OFF can lead to a reset, erasing last run memory.
- If the motor gets hot due to which there will be a fluctuation in speed value then allow the centrifuge to cool for atleast 30 minutes. Do not do any operation for 30 minutes.
- Do not use liquids with density higher than 1.2g/ml for fu/1 load operation.
WARR ANTY STATEMENT
This product is warranted tobe free from defects in material and workmanship
for a period of two (2) years from date of purchase. Your product will be duly
repaired upon prompt notification in compliance with the following conditions:
This warranty is valid only if the product is used for its intended purpose
and within the guidelines specified in this instruction manual. This warranty
does not cover damage caused by accident, neglect, misuse, improper service,
natural forces or other causes not arising from defects in original material
or workmanship. This warranty does not cover any incidental or consequential
damages, commercial lass or any other damages from the use of this product.
The warranty is invalidated by any non-factory modification, which will immediately terminate all liabilities on us for the products or damages caused by its use. The customer shall be responsible for the product or use of products as weil as any supervision required for safety. lf requested the products must be returned to the distributor in weil packed and insured man ner and all shipping charges must be paid.
NOTE:
Some states da not allow limitation on the length of implied waranties or the
exclusion or limitation of incidental or consequential damages. This warranty
gives you specific legal rights. This warranty is given expressly in lieu of
all other warranties, expressed or implied.
The purchaser agrees that there is no warranty of merchantability or of fitness for any intended purpose and there are no other remedies or warranties, implied, which extend beyond the description on the face of the agreement. This warranty is only applicable to the original purchaser. Products received without proper authorization will not be processed for warranty or service. All items returned for service should be sent with postage prepaid in the original packaging or other suitable packaging, padded to avoid damage. We will not be responsible for damage incurred by improper packaging.
NOTE:
This warranty is valid only if the warranty is registered with the supplier
within 30 days from the date of purchase.
For you reference, make a note of seria/ number, date of purchase and supp/ier here.
Serial No.: Purchase
Date:
Supplier:
PRODUCT DISPOSAL
In case the product is to be disposed of, the relevant legal regulations are to be observed. Information on the disposal of electrical and electronic devices in the European Community.
The disposal of electrical devices is regulated within the European Community by national regulations based an EU Directive 2012/19/EU an waste electrical and electronic equipment (WEEE). According to these regulations, any devices supplied after 13.06.05 in the business to business sphere, to which this product is assigned, may no langer be disposed off in municipal or domestic waste. They are marked with the following symbol to indicate this.
As disposal regulations within the EU may vary from country to country, please contact your supplier if necessary.
TRANSPORTATION & STORAGE
This devices is heavy & weighs approximately 43kgs. Care must be taken while lifting up. Always take few people help to lift this instrument. The refrigerated centrifuge must be kept always in upright position while storing/transporting from place to place.
- Use only original packaging du ring transportation
- For langer distance take transportation aid like hard trucks
- Avoid knocking, harsh shaking or jolting the device
- Always retain the packaging material & transportation protections for langer storage or transportation
- The transportation conditions for the instrument is -25° C to 60°C with a Relative Humidity up to 80% & max pressure of 106kPa for both general & air transportation
- For storage the instrument is recommended to be stored in original package. the instrument is -s•c to 45° C with a Relative Humidity up to 70% & max pressure of 106kPa
Accumax Lab Devices Private Limited
Plot No. 1S, 16 & 32 GIDC Electronic Park SEZ, Seetor 26, Gandhinagar,
Gujarat, lndia – 382026.
- Website: www.neuation.com.
References
- Leading Laboratory Equipment Manufacturers and Suppliers | Neuation
- iFuge M24PR Refrigerated Centrifuge | Lab Devices
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>