Dr Trust 9003 Glucometer Strips 9003 Instruction Manual

June 12, 2024
DR Trust

9003 Glucometer Strips 9003
Instruction Manual

Warnings

Warnings

  • For in vitro diagnostic use (for use outside of the body only).
  • For single use only.
  • Healthcare professionals and other users testing multiple patients with this system should handle everything that comes into contact with human blood carefully to prevent transmitting infectious diseases, including sanitized objects.
  • Please read this sheet and your Blood Glucose Monitoring System Owner’s Manual before you use this test strip. Use only Dr Trust Gold Standard Test Strips with Dr Trust Gold Standard Blood Glucose Monitoring System to obtain accurate results, and be covered by the manufacturer’s
  • Results may be inaccurate when testing on patients with abnormally low blood pressure or those who are in shock.
  • For patients with impaired peripheral circulation, collection of capillary blood from the approved sample sites is not advised as the results might not be a true reflection of the physiological blood glucose level. It may apply under the following circumstances: severe dehydration as a result of diabetic ketoacidosis or due to stress hyperglycemia, hyperosmolar non-kenotic coma, shock, decompensated heart failure NYHA Class IV or peripheral arterial occlusive disease.
  • Keep test strips and lancets away from small children. If swallowed, consult a doctor immediately.

Intended Use

Dr Trust Gold Standard test strips, when used together with Dr Trust Gold Standard Blood Glucose Monitoring System, allow your blood glucose levels to be measured by yourself at home or by healthcare professionals. It uses fresh whole blood samples from the finger, palm, upper arm and forearm. This system is not intended for use in the diagnosis or screening of diabetes mellitus. Professionals may use test strips to test capillary and venous blood sample; home use is limited to capillary whole blood testing.

About Alternative Site Testing (AST)

IMPORTANT: There are limitations for performing AST. Please read the Dr Trust Gold Standard Blood Glucose Monitoring System Owners Manual and consult the doctor before you perform AST.

Alternative site testing (AST) is when individuals check their blood glucose levels using areas of the body other than the fingertip. The Dr Trust Gold Standard test strips allow AST to be performed on sites other than the fingertip. We strongly recommend that you perform AST ONLY at the following times:

  • During a pre-meal or fasting state (more than 2 hours since the last meal).
  • Two hours or more after taking insulin.
  • Two hours or more after exercise.

Do NOT use AST if:

  • You think your blood glucose is low.
  • You may not notice if you are hypoglycemic.
  • Your AST results are inconsistent with the way you feel.
  • You are testing for hyperglycemia.
  • Your routine glucose results often fluctuate.

To obtain a blood sample from the alternative sites, please rub the puncture site for approximately 20 seconds before following the procedures of “Testing Your Blood Glucose”.

Limitations

  • Xylose: Do not test blood glucose during or soon after a xylose absorption test. Xylose in the blood can give falsely elevated results.
  • Hematocrit: The hematocrit level is limited to between 35% and 60%. Please ask your healthcare professional if you do not know your hematocrit level.
  • Metabolites: Dopamine, L-Dopa, methyldopa, uric acid, ascorbic acid, and acetaminophen at normal blood concentration do not significantly affect blood glucose test results.
  • There are no significant interference in the presence of galactose, maltose, or fructose observed in blood glucose tests.
  • Lipemic Effects: Blood triglycerides up to 3000 mg/dt. (33.9 mmol/L) do not affect the results significantly, but may affect results at higher levels.
  • Altitude Effects: Altitudes up to 10.742 feet (3,275m) do not affect test results.
  • Use only heparin for anticoagulation of fresh capillary or venous whole blood.

The following compounds when determined to be in excess of their limitation and tested with the Dr Trust Gold Standard glucose meter may produce elevated glucose results:

Summary of substances and concentrations in excess of limitation with interference

Substance| Limiting Concentration
(mg/dL)| Therapeutic / Physiologic Concentration  Range (or Upper Limit ) (mg/dL)
---|---|---
Acetaminophen (Paracetamol)| > 6.25| 0.45 – 3
Ascorbic Acid| > 5| 2
Pralidoxime Iodide| > 5| — 10 (IV Dose 500 mg)
Uric Acid| > 10| 2 – 8

Storage and Handling

IMPORTANT: Do not use the test strips if they have expired.

  • Test strips expire 6 months after first opening. Write the first opening date on the test strip vial when you first open it. (For strip vial only)

  • Store the test strips in a cool, dry place between 2°C and 30°C (35.6°F and 86°F) and below 85% relative humidity.

  • Keep the test strips away from direct sunlight. Do not store the test strips in high humidity.

  • Store the test strips in their original vial ONLY. Do not transfer them to a new vial or any other (For strip vial only)

  • Do not touch the test strips with wet hands.

  • Use each test strip immediately after taking it out of the vial or individual foil packet. Close the vial immediately after taking out a strip. (For strip vial only)

  • Keep the vial closed at all times. (For strip vial only)

  • Do not bend, cut, or alter the test strip.

Strip Appearance

  1. Absorbent Hole
    Apply a drop of blood here. The blood will automatically be absorbed.

  2. Confirmation Window
    This is where you confirm if enough blood has been drawn into the absorbent hole of the strip.

  3. Test Strip Handle
    Hold this part to insert the test strip into the slot.

  4. Contact Bars
    Insert this end of the test strip into the meter. Push it in firmly until it will go no further.

ATTENTION

The front side of the test strip should face up when inserting the test strip. Test results might be wrong if the contact bar is not fully inserted into the test slot.

Testing Your Blood Glucose

PLEASE WASH AND DRY YOUR HANDS BEFORE PERFORMING ANY TESTING.

STEP 1
Insert the test strip fully into the slot of the meter until it will go no further. When the strip is fully inserted, the meter will do several self- checks.
STEP 2
Collect a blood sample for about 0.5pL with the test strip. A sufficient quantity of blood is required for the test to provide accurate results. Touch the blood drop with the absorbent hole of the test strip, and wait until the confirmation window is fully covered. Do NOT apply a smeared blood sample. The meter will start counting down.
STEP 3
After a few seconds, the meter will display your blood glucose level. The last reading will be automatically saved in the meter. Turn it off by removing the test strip and throw away the used test strip.
Please refer to your Owner’s Manual for more information.
The used lancet and test strip are potentially biohazardous. Please dispose of them carefully according to your local regulations.

Reading Your Result

Your blood glucose readings deliver plasma equivalent results and are displayed either in milligrams of glucose per deciliter of blood (mg/dL) or in millimoles of glucose per liter of blood (mmol/L). The measurement range of this meter is 20-600 mg/dL (1.1-33.3 mmol/L).

Reference values

Time of day Normal plasma glucose range for people without diabetes
Fasting and before meal < 100 mg/dL (5.6 mmol/L)
2 hours after meals < 140 mg/dL (7.8 mmol/L)

Source: American Diabetes Association (2014). Clinical Practice Recommendations. Diabetes Care, 37 (Supplement 1): S16.

Please consult your doctor to determine a target range that works best for you.

Questionable or inconsistent results
If your test results are unusual or inconsistent with how you are feeling:

  • Make sure the confirmation window of the test strip is completely filled with blood.
  • Check the expiry date of the test strips
  • Check the performance of your meter and test strip with the control solutions.

PLEASE NOTE: Unusually high or low blood glucose levels may be symptoms of a serious medical condition. If most of your results are unusually high or low, please contact your healthcare professional.

Quality Control Testing

Our control solutions contain a known amount of glucose that can react with test strips. If you suspect the meter or test strips are not working properly, you can check the performance of the meter, test strip and your technique by comparing the control solution results with the range printed on the label of test strip vial or on the package. Please refer to the Owner’s Manual for complete testing instructions.
IMPORTANT: The reference range of the control solutions may vary with each new vial or package of test strips. Make sure you check the range on the label of your current vial or on the current package.

Chemical Components
Glucose dehydrogenase (E. coli) 8%
Electron shuttle 55%
Enzyme protector 8%
Non-reactive ingredients 29%

Additional Information for Healthcare Professionals

Always wear gloves and follow your facility’s biohazard control policy and procedures when performing tests involving patient blood samples. Use fresh whole blood samples only. Professionals may use test strips to test capillary and venous blood sample.

Sample Size: 0.5 ∪ L
Reaction Time: 5 seconds
System Measurement Range: 20 mg/dL to 600 mg/dL (1.1 to 33.3 mmol/L)
Hematocrit Range: 35% to 60%

Accuracy
Diabetes experts have suggested that glucose meters should be within ±15 mg/dL (0.83 mmol/L) of the reference method when the glucose concentration is lower than 100 mg/dL (5.55 mmol/L), and be within ±15% of the reference method when the glucose concentration is 100 mg/dL (5.55 mmol/L) or higher. The tables below display how often Dr Trust Gold Standard achieves this target. The chart is based on a study carried out on 160 patients (each patient was tested six times which had 960 test results) to see how well Dr Trust Gold Standard performed compared to YSI-2300 reference method
**Table 1 Accuracy results for glucose concentration < 100 mg/dL (5.55 mmol/L)
*Acceptance criteria in EN ISO 15197: 2015, 95% of all differences in glucose values (i.e., YSI-2300 reference values minus Dr Trust Gold Standard’s glucose values) should be within ±15 mg/dL (0.83 mmol/L) for glucose concentration < 100 mg/dL (5.55 mmol/L), and within ±15% for glucose concentration > 100 mg/dL (5.55 mmol/L).
Note:** When Dr Trust Gold Standard Test Strips results are compared to the reference values, difference values below 100 mg/dL (5.55 mmol/L) are expressed in mg/dL or mmol/L, while those above 100 mg/dL (5.55 mmol/L) are in percent.

User performance

154 subjects tested on the fingertip and the alternative sites, the palm, the forearm and the upper arm. The tables show how well Dr Trust Gold Standard performed compared to YSI-2300 reference method results.

Table 1 Difference distribution for glucose concentration < 100 mg/dL (5.55 mmol/L)

Tested sites| Difference within ±5mg/dL| Difference within ±10mg/dL| Difference within ±15mg/dL
---|---|---|---
Fingertip| 26/46| (56.5%)| 40/46| (87.0%)| 46/46| (100%)
Palm| 19/57| (33.3%)| 50/57| (87.7%)| 57/57| (100%)
Forearm| 16/57| (28.1%)| 44/57| (77.2%)| 57/57| (100%)
I Upper arm| 22/57| (38.6%)| 47/57| (82.5%)| 57/57| (100%)

Table 2 Difference distribution for glucose concentration 100 mg/dL (5.55 mmol/L)

Tested sites| Difference within ±5%| Difference within ±1%| Difference within ±15%
---|---|---|---
Fingertip| 54/108 (50.0%)| 93/108 (86.1%)| 105/108 (97.2%)
Palm| 30/97 (30.9%)| 68/97 (70.1%)| 97/97 (100%)
Forearm| 28/97 (28.9%)| 68/97 (70.1%)| 96/97 (99.0%)
Upper arm| 25/97 (25.8%)| 61,/97 (62.9%)| 96/97 (99.0%)

Precision
The CV (%) is less than 5% both in intermediate precision and repeatability precision.

Symbol Information

Symbol Referent
In vitro diagnostic medical device
Consult instructions for use
Temperature limitation
Use by
Batch code
Manufacturer
Authorized representative in the European Community
Do not reuse
CE mark
Humidity limitation

Imported and Marketed in India by:

Nureca Limited
No. 2, Part C, 3rd Floor, 3rd Cross, Aziz Sait Industrial Estate, Nayandahalli, Bangalore, Bangaluru – 560039.,
Bengaluru(Bangalore) Urban, Karnataka (India) – 560039
Customer Care Exceutive Email : customercare@nureca.com
Website : www.drtrustusa.com; www.drtrust.in
MedNet EC-REP GmbH
Borkstrape 10 48163, Munster, Germany
Use Only With Dr Trust Gold Standard Blood Glucose Monitoring System