Intended use
The Cadwell Sleepmate™ Reusable Airflow Sensor is to detect patient nasal and/or oral breathing during sleep research and clinical studies of sleep.
Instruction for use
The sensor should be installed so that the probes are exposed to the maximum
available airflow. The sensor probes are flexible to allow each probe to be
bent gently to insure that they are in the path of airflow. The probes should
not touch the skin as contact may attenuate the signal amplitude. All cannula
style airflow sensors are installed onto the patient in a manner similar to
that of an oxygen cannula. The sensors are delicate and must be handled
carefully. Proper handling and cleaning will increase durability and
longevity in the life of the sensor. When bending the prongs to fit the
patient, never bend at a sharp angle, but in a smooth round curve. The prong
should be bent a short distance away from the center bar. The assembly should
be carefully secured using both the sliding cinch tubing and surgical tape.
Connect the sensor to the appropriate channel input of your recorder. No
external power source is required.
Ensure the sensor is securely attached to the patient as described and you
have clear, reliable signals prior to lights out.
Cleaning
This sensor was designed to be reusable and can be cleaned by rubbing with any cleansing agent that is not corrosive toward plastic. The entire assembly must be completely dry before using again to insure patient safety. For sterilization implement a standard procedure for gas sterilization. Do not autoclave this product. Use the molded ends of the connectors when removing the connector from the junction box.
Technical Specifications
Description | Properties |
---|---|
Lead Wire | Two conductors white lead wire, approx. 200 cm (80″) long. |
Sensor | Flexible, rubber coated probes for oral (one probe) and nasal |
(two probes) airflow.
Connectors| Two 1.5 mm (0.1″) safety pins. Some recorders may use Lemo or
other connectors.
Filter Settings| Low Hz = 0.1 Hz; High Hz = 15 Hz;
Gain: Output range is +/-100 MV.
Sensitivity is dependent on variables such as probe placement, patient
airflow, ambient room temperature, CPAP pressure, etc.
Caution Statement :
US Federal Law restricts this device to sale by or on the order of a
physician. This product is for diagnostic purposes only and is not intended to
be used as an apnea monitor or in a life supporting or sustaining capacity.
Cadwell Industries Inc.
909 N. Kellogg Street, Kennewick, WA 99336 USA
1 800-245-3001 U.S.A. toll-free
+1 509-735-6481
+1 509-783-6503 Fax
www.cadwell.com
info@cadwell.com
©2021 Cadwell Industries, Inc. All rights reserved. PN 302993-639-5 Rev. 00
The information contained in this document is subject to change without
notice.
This document contains trademarks that belong to Cadwell Industries, Inc. and
other companies, respectively.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>