SPER SCIENTIFIC 850012 Sound Level Calibrator User Manual
- June 12, 2024
- SPER SCIENTIFIC
Table of Contents
850012 (ST-120)
2-Pt. Acoustical Calibrator
Class 1
USER MANUAL
INTRODUCTION
The ST-120, Sound Level Calibrator, is used for the calibration of the pressure sensitivity of the microphone and sound level measuring equipment. The design of ST-120 is based on a feedback arrangement to ensure a high stable sound pressure level. The performance is very stable; it does not required for the revision to the microphone equivalent volume. ST-120 conforms to the standard of IEC60942 (2003) Class 1.
EQUIVALENT FREE-FIELD
Usually we use the free-field microphone more than others. For example, when
use the ST-120 to calibrate the sound level meter and other environmental
noise measuring equipments need to revise the pressure-field to the equivalent
free-field. The corrected value is the difference between the pressure-field
response and free field response when at 1000Hz. The value for the 23.77mm (1
inch) microphone is –0.4dB, and for the 12.7mm (1/2 inch) microphone is
–0.2dB. Also the equivalent free-field for the 23.77mm
(1 inch) microphone is 93.6dB, and for the 12.7mm (1/2 inch) microphone is
93.8dB.
SPECIFICATIONS
- Standard : IEC 60942(2003) and ANSI S1.40. (1984)
- Sound Pressure Levels : 94 and 114Db. (reference of 2×10-5 Pa ).
- Accuracy : 94 ± 0.3dB and 114 ± 0.5dB.
- Frequency : 1000Hz ± 1%.
- Distortion : ≤ 3 %.
- Battery : 9V 006P or IEC6F22 or NEDA 1604.
- Dimension : φ62×134(mm ).
- Weight : about 350g (included 1/2″ adaptor and 9V battery ).
- Stable Times : 20s.
- Operating Temperature : 0 ℃ ~ 40 (℃ 32 ~ 1℉ 04 ).℉
- Storage Temperature : -10 ~ ℃ 60 (℃ 14 ~ 1℉ 40 ).℉
- Humidity : <90%.
- Atmospheric Pressure : 65KPa~100KPa
- Accessories : User’s manual x1 、Calibration certificate x1
OPERATING INSTRUCTIONS
- Insert the 1/2 inches microphone to the hole of ST120 Sound Calibrator until you hear a “click” sound. (refer to Fig. 1)
- Select output dB level and turn on power, LED will be lightened and the meter emits a sound.
- Read the value on the sound level meter’s display and perform the calibration
- Extract the microphone.
CALIBRATION AND ADJUSTMENT
If calibration out of specification, you can calibrate and adjustment it:
- Unscrew completely the metal silver battery cover.
- Uses a small screw driver trimming interior variable resistor(VR1) until attain a designated standard the accurate 94dB normal value.
CAUTION
When verification should use the pressure field response laboratory scale
microphone as well as the precision measurement amplifier carries on the
measuring. Does not need to 114.0dB to carry on the adjustment.
BATTERY REPLACEMENT
CAUTION
The battery voltage should maintain more than 7V,when the LED flash, please
change battery soon.
CAUTION
If you not use this product for a long time , please remove battery
- Unscrew completely the metal silver battery cover.
- Remove the battery.
- Replace the battery with a new one of the same type.
Description Of Calibrator
- Power & output level select switch .
- Low battery indicator .
- Battery cover.
- 1/2″ adapter.
CLEANING
To clean the instruments use a soft dry cloth. Never use a wet cloth, solvents or water.
END OF LIFE
Caution: this symbol indicates that equipment and its accessories shall be subject to a separate collection and correct disposal
ENVIRONMENT CONDITIONS
For inside use, max height: | 2000m |
---|---|
Reference temperature: | 23° ± 3 ℃ |
Operation temperature: | 5 ~ 40 ℃ |
Operation humidity: | <80% RH |
Storage temperature | -10 ~ 60 ℃ |
Storage humidity | <70% |
EMC
This instrument was designed in accordance with EMC Standards in force and its compatibility has been tested in accordance with EN61326-1 (2006).
ACCESSORIES
- Meter: ST-120.
- User’s manual.
- Carrying case.
- 1 battery 9V 006P or IEC6F22 or NEDA 1604.
WARRANTY
Sper Scientific warrants this product against defects in materials and
workmanship for a period of one (1) year from the date of purchase, and agrees
to repair or replace any defective unit without charge. If your model has
since been discontinued, an equivalent
Sper Scientific product will be substituted if available. This warranty does
not cover probes, batteries, battery leakage, or damage resulting from
accident, tampering, misuse, or abuse of the product. Opening the meter to
expose its electronics will break the waterproof seal and void the warranty.