NATIONAL INSTRUMENTS PXIe-4163 Precision PXI Source Measure Unit Owner’s Manual

June 12, 2024
NATIONAL INSTRUMENTS

NATIONAL INSTRUMENTS - logo

SPECIFICATIONS
PXIe-4163
PXIe, 24-Channel, ±24 V, 50 mA Precision PXI Source Measure Unit

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Nominal unless otherwise noted.

Conditions

Specifications are valid under the following conditions unless otherwise noted.

  • Ambient temperature of 23 °C ± 5 ºC 1

  • Chassis with slot cooling capacity ≥38 W 2
    – For chassis with slot cooling capacity = 38 W, fan speed set to HIGH

  • Calibration interval of 1 year

  • 30 minutes warm-up time

  • Self-calibration performed within the last 24 hours

  • niDCPower Aperture Time property or NIDCPOWER_ATTR_APERTURE_TIME attribute set to 2 power-line cycles (PLC)

Instrument Capabilities

Channels 0 through 23
DC voltage range ±24 V

The following table and figure illustrate the voltage and the current source and sink ranges of the PXIe-4163.

Table 1. PXIe-4163 DC Current Source and Sink Ranges, Warranted

Chassis Slot Cooling Capacity

≥58 W| 38 W
10 μA
100 μA
1 mA
10 mA
50 mA| 30 mA

1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).
2 For increased capability, NI recommends installing the PXIe-4163 in a chassis with slot cooling capacity ≥58 W.

Figure 1. PXIe-4163 Quadrant Diagram, Any Channel

NATIONAL INSTRUMENTS PXIe 4163 Precision PXI Source Measure Unit -
Instrument Capabilities 1

Valid on any channel in chassis with slot cooling capacity ≥ 58 W.
Valid on any channel in all other compatible chassis. 1
1 Maximum 480 mA per module.

SMU Specifications

Voltage
Table 2. Voltage Programming and Measurement Accuracy/Resolution, Warranted

Range| Resolution and Noise
(0.1 Hz to 10 Hz)| Accuracy (23 °C ± 5 °C)
± (% of Voltage +
Offset)| Tempco3 a (% of Voltage +
Offsety°C, 0 °C to 55 °C
---|---|---|---
Tad ± 5 °C
24 V| 200µV| 0.05%+ 5 mV| 0.0005%+ 1µV

3 Temperature coefficient applies beyond 23 °C ± 5 °C within 5 °C of Tcal.

Current

Table 3. Current Programming and Measurement Accuracy/Resolution, Warranted

Range| Resolution and Noise (0.1 Hz to
10 Hz)| Accuracy (23 °C t 5 °C) ± (% of  Current + Offset)| Tempco4 s (% of Current + Offset)/°C, 0 °C to 55 °C
---|---|---|---
Teal ± 5 °C
10µA| 100 pA| 0.10%+ 5 nA| 0.004% + 10 pA
100 RA| 1 nA| 0.10% + 50 nA| 0.004% + 100 pA
I mA| 10 nA| 0.10% + 500 nA| 0.004% + 1 nA
10 mA| 100 nA| 0.10%+ 5µA| 0.004%+ 10 nA
30 mA or 50 mA5| 500 nA| 0.10%+ 25 RA| 0.004% + 50 nA

Available DC Output Power

Chassis Slot Cooling Capacity| Per Channel Maximum| Absolute Maximum
---|---|---
≥58 W| 1.2 W| 28.8 W
38 W| 0.7 W| 11.5 W

Additional Specifications

Settling time 6 <500 μs, typical 7
Transient response 8 <100 μs, typical 9
Wideband source noise 10 15 mV RMS, typical <100 mVpk-pk, typical

4 Temperature coefficient applies beyond 23 °C ± 5 °C within 5 °C of Tcal.
5 50 mA range available only when installed in chassis with slot cooling capacity ≥58 W. 30 mA range available in all other chassis.
6 Current limit set to ≥1 mA and ≥10% of the selected current limit range. PXIe-4163 configured for fast transient response.
7 To settle to 0.1% of voltage step.
8 PXIe-4163 configured for fast transient response.
9 To recover within ±20 mV after a load current change from 10% to 90% of range.
10 20 Hz to 20 MHz bandwidth. PXIe-4163 configured for normal transient response. Measured at the end of the 1 m SHDB62M-DB62M-LL cable.

Remote sense|
---|---
Voltage| No additional error due to lead drop
Current| No additional error due to lead drop
Maximum lead drop| 1 V drop/lead
Load regulation|
Voltage 11| 50 μV/mA, typical
Current| (30 pA + 20 ppm of range)/volt, typical
Functional isolation voltage, any pin to
earth ground| 60 V DC
Absolute maximum voltage to Output LO|
From Sense HI 12|
When VOutput HI > 0 V| -0.5 V to (VOutput HI + 0.5 V)
When VOutput HI ≤ 0 V| (VOutput HI – 0.5 V) to 0.5 V
From all other pins| ±25 V

The following figures illustrate noise as a function of measurement aperture for the PXIe-4163.

Figure 2. Voltage RMS Noise Versus Aperture Time,13 Typical

NATIONAL INSTRUMENTS PXIe 4163 Precision PXI Source Measure Unit -
Additional Specifications 1

11 At connector pins when using local sense.
12 Where VOutput HI is the voltage at the Output HI pin in the same channel as a Sense HI pin.
13 All channels averaged. Channels 9 and 22 have degraded performance.

Figure 3. Current RMS Noise Versus Aperture Time, 14 Typical

NATIONAL INSTRUMENTS PXIe 4163 Precision PXI Source Measure Unit -
Additional Specifications 2

14 All channels averaged. Channels 7, 9, and 11 have degraded performance.

Supplemental Specifications

Measurement and Update Timing

Available sample rates
15| (600 kS/s)/N
---|---
where
N = 6, 7, 8, … 2 20
S is samples Sample rate accuracy| ±50 ppm
Maximum measure rate to host 16| 100,000 S/s per channel, continuous
Maximum source update rate 17|
Single channel| 100,000 updates/s
All channels simultaneously| 40,000 updates/s per channel
Input trigger to|
Source event delay| 8.5 µs
Source event jitter| 1.7 μs
Measure event jitter| 1.7 μs

Triggers

Input triggers|
---|---
Types| Start
Source
Sequence Advance
Measure
Sources (PXI trigger lines 0 to 7) 18|
Polarity| Active high (not configurable)
Minimum pulse width| 100 ns

15 When source-measuring, both the NI-DCPower Source Delay and Aperture Time properties affect the sampling rate. When taking a measure record, only the Aperture Time property affects the sampling rate.
16 Load dependent settling time is not included. Normal DC noise rejection is used.
17 As the source delay is adjusted or if advanced sequencing is used, maximum source update rates may vary.
18 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1.

19 Destinations  (PXI trigger lines 0 to 7) 18|
---|---
Polarity| Active high (not configurable)
Minimum pulse width| >200 ns
Output triggers (events)|
Types| Source Complete
Sequence Iteration Complete
Sequence Engine Done
Measure Complete
Destinations (PXI trigger lines 0 to 7) 18|
Polarity| Active high (not configurable)
Pulse width| 230 ns

Calibration Interval

Recommended calibration interval 1 year

Physical

Dimensions| 3U, one-slot, PXI Express/CompactPCI Express module
2.0 cm × 13.0 cm × 21.6 cm
(0.8 in. × 5.1 in. × 8.5 in.)
---|---
Weight| 394 g (13.9 oz)
Front panel connectors| Custom 62-position D-SUB, female

Power Requirements

Chassis Slot Cooling Capacity| +3.3 V Current Draw, Typical| +12 V Current Draw, Typical
---|---|---
Idle| Full Output Load| Idle| Full Output Load
38W| 1A| 1 A| 1.5 A| 3A

58W| 1A| 4.5 A

19 Input triggers can come from any source (PXI trigger or software trigger) and be exported to any PXI trigger line. This allows for easier multi-board synchronization regardless of the trigger source.

Environmental Characteristics

Temperature and Humidity|
---|---
Temperature|
Operating|
Chassis with slot cooling capacity ≥58 W 20| 0 °C to 55 °C
All other compatible chassis| 0 °C to 40 °C
Storage| -40 °C to 71 °C
Humidity|
Operating| 10% to 90%, noncondensing
Storage| 5% to 95%, noncondensing
Pollution Degree| 2
Maximum altitude| 2,000 m (800 mbar) (at 25 °C ambient temperature)
Shock and Vibration|
Random vibration|
Operating| 5 Hz to 500 Hz, 0.3 g RMS
Non-operating| 5 Hz to 500 Hz, 2.4 g RMS
Operating shock| 30 g, half-sine, 11 ms pulse

20 Not all chassis with slot cooling capacity ≥58 W can achieve this ambient temperature range. Refer to PXI chassis specifications to determine the ambient temperature ranges your chassis can achieve.

Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end- user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2018 National Instruments. All rights reserved.
376735B-01 December 13, 2018

COMPREHENSIVE SERVICES
We offer competitive repair and calibration services, as well as easily accessible documentation and free downloadable resources.

SELL YOUR SURPLUS
We buy new, used, decommissioned, and surplus parts from every NI series.
We work out the best solution to suit your individual needs.
My Sell For Cash Get Credit Receive a Trade-In Deal

OBSOLETE NI HARDWARE IN STOCK & READY TO SHIP
We stock New, New Surplus, Refurbished, and Reconditioned NI Hardware.

Bridging the gap between the manufacturer and your legacy test system.

1-800-915-6216
www.apexwaves.com
sales@apexwaves.com

All trademarks, brands, and brand names are the property of their respective owners.

Request a Quote PXIe-4163

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

NATIONAL INSTRUMENTS User Manuals

Related Manuals