APEX WAVES NI PXIe-2525 Multi-Bank Configurable 2-Wire Multiplexer
Product Information
The NI PXIe-2525 is a multi-bank configurable 2-wire multiplexer relay module. It is designed to provide reliable switching capabilities for various applications. The module’s specifications are subject to change without notice.
Input Characteristics
The NI PXIe-2525 is suitable for voltage measurements in Measurement
Categories CAT I and CAT O. These categories are intended for measurements on
circuits that are not directly connected to the electrical distribution
system, such as signal levels, limited-energy parts of hardware, and circuits
powered by regulated low-voltage sources.
It is important to note that when hazardous voltages (>42.4 Vpk/60 VDC) are present on any channel, the safety low-voltage (42.4 Vpk/60 VDC) cannot be connected to any other channel.
Switching Power
The maximum switching power of the NI PXIe-2525 is 60 W or 62.5 VA. It is
essential to ensure that the switching power does not exceed these limits, as
it is limited by the maximum switching current and voltage.
Maximum Current
The maximum switching current for each channel is 2 A. The maximum current for
carrying purposes is also 2 A per channel.
Minimum Switch Load
The minimum switch load for the NI PXIe-2525 is 20 mV/1 mA. However, it is not
recommended for 2-wire resistance measurements.
DC Path Resistance
The initial DC path resistance for the NI PXIe-2525 is not specified in the
provided text extract.
Product Usage Instructions
- Ensure that the NI PXIe-2525 is properly installed in your system according to the manufacturer’s instructions.
- Familiarize yourself with the input characteristics and limitations of the module to ensure safe and accurate measurements.
- Avoid connecting the safety low-voltage (42.4 Vpk/60 VDC) to any other channel when hazardous voltages (>42.4 Vpk/60 VDC) are present on a channel.
- Ensure that the switching power does not exceed the maximum specified limits of 60 W or 62.5 VA.
- Take note of the maximum switching current of 2 A per channel for both switching and carrying purposes.
- Avoid using the NI PXIe-2525 for 2-wire resistance measurements as it is not recommended.
- For switching inductive loads, such as motors and solenoids, consider implementing additional protection to prevent high voltage transients that can interfere with module operation and impact relay life.
- For more information on transient suppression and other related topics, visit ni.com/info and enter the Info Code relayflyback.
DEVICE SPECIFICATIONS
NI PXIe-2525
Multi-bank Configurable 2-wire Multiplexer
This document lists specifications for the NI PXIe-2525 (NI 2525) multiplexer
relay module. All specifications are subject to change without notice. Visit
ni.com/manuals for the most current specifications.
About These Specifications
- Specifications characterize the warranted performance of the instrument under the stated operating conditions.
- Typical Specifications are specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 °C ambient temperature. Typical specifications are not warranted.
- All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
- Caution The protection provided by the NI 2525 can be impaired if it is used in a manner not described in this document.
Input Characteristics
Maximum switching voltage
- Channel-to-channel…………………………………………………………..150 V
- Channel-to-ground…………………………………………………………..150 V, CAT O
Note
This module is rated for Measurement Category I. It is intended to carry
signal voltages no greater than 150 V. This module can withstand up to 800 V
impulse voltage. Do not use this module for connection to signals or for
measurements within Categories II, III, or IV. Do not connect to MAINS supply
circuits (for example, wall outlets) of 115 VAC or 230 VAC. Refer to the Read
Me First: Safety and Electromagnetic Compatibility document for more
information about measurement categories.
Measurement Categories CAT I and CAT O (Other) are equivalent and are for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. This category is for measurement of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special hardware, limited-energy parts of hardware, circuits powered by regulated low-voltage sources, and electronics.
Caution
When hazardous voltages (>42.4 Vpk/60 VDC) are present on any channel, safety
low-voltage (≤42.4 Vpk/60 VDC) cannot be connected to any other channel.
-
Maximum switching power………………………………………………………………….60 W, 62.5 VA
Caution The switching power is limited by the maximum switching current, the maximum voltage, and must not exceed 60 W, 62.5 VA. -
CH-COM DC isolation resistance………………………………………………………………….>1 GΩ, typical at 25 °C
Maximum current
- Switching…………………………………………………………..2 A (per channel)
- Carry…………………………………………………………..2 A (per channel)
- Minimum switch load………………………………………………………………….20 mV/1 mA
Note The NI 2525 is not recommended for 2-wire resistance measurements.
Note
Switching inductive loads (for example, motors and solenoids) can produce
high-voltage transients in excess of the module’s rated voltage. Without
additional protection, these transients can interfere with module operation
and impact relay life. For more information about transient suppression, visit
ni.com/info and enter the Info Code relayflyback.
DC path resistance
-
Initial…………………………………………………………..<0.8 Ω
-
End-of-life…………………………………………………………..≥1.8 Ω
DC path resistance is a combination of relay contact resistance and trace resistance. Measure path resistance by combining the resistance of the high and low signal paths from one row to one column. Contact resistance typically remains low for the life of a relay. At the end of relay life, the contact resistance rises rapidly above ≥1.8 Ω. -
Thermal EMF………………………………………………………………….<10 µV
RF Performance Characteristics
- Bandwidth (-3 dB, typical at ………………………………………………………………….23 ºC) ≤10 MHz 50 Ω termination
- Open channel isolation (typical at 23 ºC) 50 Ω termination
- 10 kHz…………………………………………………………..≥ 76 dB
- 100 kHz…………………………………………………………..≥ 63 dB
- 1 MHz…………………………………………………………..≥ 36 dB
- Channel-to-channel crosstalk (typical at 23 ºC) 50 Ω termination
- 10 kHz…………………………………………………………..≤ -73 dB
- 100 kHz…………………………………………………………..≤ -52 dB
- 1 MHz…………………………………………………………..≤ -32 dB
Dynamic Characteristics
Relay operate time
- Typical…………………………………………………………..1 ms
- Maximum…………………………………………………………..3.1 ms
- Simultaneous drive limit…………………………………………………………..79 relays
Expected relay life
- Mechanical…………………………………………………………..1 × 108 cycles
- Electrical
- ≤30 mV, ≤10 mA resistive……………………………………………………2.5 × 106 cycles
- 30 V, 1 A……………………………………………………5 × 105 cycles
- 30 V, 2 A……………………………………………………1 × 105 cycles
- 60 VDC, 1 ADC resistive……………………………………………………1 × 105 cycles
Note
Relays are field replaceable. Refer to the NI Switches Help at
ni.com/manuals for more information about replacing a
failed relay.
Physical Characteristics
- Relay type………………………………………………………………….Electromechanical, non-latching
- Relay contact material………………………………………………………………….Palladium-ruthenium, gold covered
- Front panel connector………………………………………………………………….160 DIN 41612, 160 positions, male
Power requirement
- PXI Express
- 12 V……………………………………………………20 W
- 3.3 V……………………………………………………3.4 W
- Dimensions (L × W × H)………………………………………………………………….3U, one slot, PXI/cPCI module, PXI Express compatible 21.6 cm × 2.0 cm × 13.0 cm (8.5 in. × 0.8 in. × 5.1 in.)
- Weight………………………………………………………………….230 g (8.1 oz)
Environment
- Maximum altitude………………………………………………………………….2,000 m (800 mbar) (at 25 °C ambient temperature)
- Pollution Degree………………………………………………………………….2
Indoor use only.
Operating Environment
- Ambient temperature range………………………………………………………………….0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 low-temperature limit and MIL-PRF-28800F Class 2 high-temperature limit.)
- Relative humidity range………………………………………………………………….10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.)
Storage Environment
- Ambient temperature range………………………………………………………………….-40 °C to 71 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 limits.)
- Relative humidity range………………………………………………………………….5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.)
Shock and Vibration
- Operating shock………………………………………………………………….30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Meets MIL-PRF-28800F Class 2 limits.)
- Random vibration
- Operating…………………………………………………………..5 Hz to 500 Hz, 0.3 grms
- Nonoperating…………………………………………………………..5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)
Diagrams
The following figure shows the NI 2525 power-on state.
Figure 1. NI 2525 Power-On State
Figure 2. NI 2525 Connector Pinout
Accessories
Table 1. NI Accessories for the NI 2525
Accessory | Part Number |
---|---|
DIN160 to 50-pin D-SUB switch cable, 1 m | 782417-03 |
DIN160 to DIN160 switch cable, 1 m | 782417-02 |
Table 1. NI Accessories for the NI 2525 (Continued)
Accessory | Part Number |
---|---|
DIN160 to bare wire switch cable, 1 m | 782417-01 |
Relay replacement kit for IME02TS Relays | 782461-10 |
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
Online Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for
electrical equipment for measurement, control, and laboratory use:
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Note
- Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.
- Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
- Note For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.
CE Compliance
This product meets the essential requirements of applicable European
Directives, as follows:
- 2006/95/EC; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory
compliance information. To obtain product certifications and the DoC for this
product, visit ni.com/certification, search by
model number or product line, and click the appropriate link in the
Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally
responsible manner. NI recognizes that eliminating certain hazardous
substances from our products is beneficial to the environment and to NI
customers.
For additional environmental information, refer to the Minimize Our
Environmental Impact web page at
ni.com/environment. This page contains the
environmental regulations and directives with which NI complies, as well as
other environmental information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers
At the end of the product life cycle, all NI products must be disposed of
according to local laws and regulations. For more information about how to
recycle NI products in your region, visit
ni.com/environment/weee.
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help» Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third- party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export- compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED
WARRANTIES
AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. The U.S.
Government Customers:
The data contained in this manual was developed at private expense and is
subject to the applicable limited rights and restricted data rights as set
forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
COMPREHENSIVE SERVICES
We offer competitive repair and calibration services, as well as easily
accessible documentation and free downloadable resources.
SELL YOUR SURPLUS
We buy new, used, decommissioned, and surplus parts from every NI series We
work out the best solution to suit your individual needs.
- Sell For Cash
- Get Credit
- Receive a Trade-In Deal
OBSOLETE NI HARDWARE IN STOCK & READY TO SHIP
We stock New, New Surplus, Refurbished, and Reconditioned NI Hardware.
Request a Quote CLICK HERE PXI-2523
Bridging the gap between the manufacturer and your legacy Test system.
Contact
- tell: 1-800-915-6216
- web: www.apexwaves.com
- email: sales@apexwaves.com
All trademarks, brands, and brand names are the property of their respective owners
© 2014 National Instruments. All rights reserved.
References
- Test and Measurement Systems, a part of Emerson - NI
- Test and Measurement Systems, a part of Emerson - NI
- Engineering a Healthy Planet - NI
- Product Take-Back Program and Recycling - NI
- Using Info Codes - NI
- Product Documentation - NI
- NI Trademarks and Logo Guidelines - NI
- Product Certifications - NI
- Engineering a Healthy Planet - NI
- Product Take-Back Program and Recycling - NI
- Using Info Codes - NI
- Product Documentation - NI
- PXIe-2525 National Instruments Multiplexer Switch Module | Apex Waves
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>