microlife IB MT 100 Medical Thermometer Instructions
- June 12, 2024
- microlife
Table of Contents
- microlife IB MT 100 Medical Thermometer
- Product Information
- Description of this Thermometer
- Important Safety Instructions
- Turning on the Thermometer
- Storage of Measured Values
- Measuring methods / Normal body temperature
- Approx. measuring time: 1 minute!
- Approx. measuring time: 1 minute!
- Technical Specifications
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
microlife IB MT 100 Medical Thermometer
Product Information
Product Name | MT 100 |
---|---|
Model Number | IB MT 100 Var1 3023 |
Revision Date | 2023-07-10 |
Description | This thermometer is designed for measuring human bodytemperature |
through oral, rectal, or axillary methods. It featuresan ON/OFF button, a
display, a battery compartment cover, and ameasuring sensor/measuring tip. The
entire thermometer surface canbe cleaned and disinfected.
Type| Type BF applied part
Measurement Range| N/A
Measurement Accuracy| N/A
Operating Conditions| N/A
Storage Conditions| N/A
Battery| LR41 (1.5V) / SR41 (1.55V)
Battery Lifetime| N/A
IP Class| IP67
Reference to Standards| ASTM E1112; IEC 60601-1; IEC 60601-1-2 (EMC);
IEC60601-1-11
Expected Service Life| N/A
Description of this Thermometer
- ON/OFF button
- Display
- Battery compartment cover
- Measuring sensor / measuring tip
- Cleaning and disinfecting area (entire thermometer surface)
Important Safety Instructions
- Follow the instructions for use provided in the usermanual.
- This device is intended for measuring human body temperaturethrough oral, rectal, or axillary methods. Do not attempt to taketemperatures at other sites, such as in the ear.
- Do not use this device if it is damaged or shows any signs ofunusual behavior.
- Before first use, it is recommended to clean the deviceaccording to the cleaning instructions for personal hygiene.
- Rectal measurement in children younger than 3 years should onlybe performed by trained healthcare personnel. Use an alternativemeasurement method instead.
- Do not attempt rectal measurements on persons with rectaldisorders.
- Ensure that children do not use this device unsupervised, assome parts may pose a choking hazard.
- Avoid using this device close to strong electromagneticfields.
- The measurement result given by this device is not a diagnosis.Do not rely solely on the measurement result.
- Dispose of batteries and electronic devices in accordance withlocally applicable regulations.
- Follow instructions for use. This document provides important product operation and safety information regarding this device.
- Please read this document thoroughly before using the device and keep for future reference.
- This device is only to be used for measuring human body temperature through oral, rectal or axillary. Do not attempt to take temperatures at other sites, such as in the ear, as it may result in false readings and may lead to injury.
- Do not use this device if you think it is damaged or notice anything unusual.
- We recommend cleaning this device according to the cleaning instructions before first use for personal hygiene.
- The minimum measurement time until the beep is heard must be maintained without exception!
- Consider that different measurement locations may require continued measuring even after the beep, see section «Measuring methods / Normal body temperature».
- For safety reasons (risk of rectal perforation) rectal measurement in children younger than 3 years must be performed only by trained health care personnel (professional user). Use another measurement method instead. For rectal fever measurement in small children younger than 3 years, thermometers with a flexible tip are available.
- Do not attempt rectal measurements on persons with rectal disorders.
- Doing so may aggravate or worsen the disorder.
- Ensure that children do not use this device unsupervised; some parts are small enough to be swallowed. Be aware of the risk of strangulation in case this device is supplied with cables or tubes.
- Do not use this device close to strong electromagnetic fields such as mobile telephones or radio installations. Keep a minimum distance of 3.3 m from such devices when using this device.
- Protect the device from impact and dropping!
- Avoid ambient temperatures above 60 °C. NEVER boil this device!
- Use only the commercial disinfectants listed in the section «Cleaning and disinfecting» to clean the device to avoid damage to the device.
- We recommend this device is tested for accuracy every two years or after mechanical impact (e.g. being dropped). Please contact your local Microlife-Service to arrange the test.
- WARNING: The measurement result given by this device is not a diagnosis! Do not rely on the measurement result only.
- Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with domestic waste.
- Read the instructions carefully before using this device.
- Type BF applied part
Turning on the Thermometer
- To turn on the thermometer, press the ON/OFF button 1; a short beep signals «thermometer ON». A display test is performed. All segments should be displayed.
- Then at an ambient temperature of less than 32 °C, an «– -» and a flashing «°C» appear at the display field 2. The thermometer is now ready for use.
Function Test
Correct functioning of the thermometer is tested automatically each time it is
turned on. If a malfunction is detected (measurement inaccuracy), this is
indicated by «ERR» on the display, and a measurement becomes impossible. In
this case, the thermometer must be replaced.
Storage of Measured Values
If the ON/OFF button 1 is pressed for more than 3 seconds when turning on the thermometer, the automatically stored maximum temperature during the last measurement will be displayed. At the same time, a «M» for memory will appear on the display. About 2 seconds after the button is released, the temperature value disappears and the thermometer is ready for measurement.
Measuring methods / Normal body temperature
**In the armpit (axillary) / 34.7 – 37.3 °C**
Wipe the underarm with a dry towel. Place the measuring sensor 4 under the arm into the center of the armpit so the tip is touching the skin and position the patient’s arm next to the patient’s body. This ensures that the room air does not affect the reading.
Because the axillary takes more time to reach its stable tempera- ture wait at least 5 minutes , regardless of the beep sound.
In the mouth (oral) / 35.5 – 37.5 °C
Do not eat or drink anything hot or cold 10 minutes before the measurement. The mouth should remain closed up to 2 minutes before starting a reading.
Position the thermometer in one of the two pockets under the tongue, to the left or right of the root of the tongue. The measuring sensor 4 must be in good contact with the tissue. Close your mouth and breathe evenly through the nose to prevent the measurement from being influenced by inhaled/exhaled air.
If this is not possible due to blocked airways, another method for measuring should be used.
Approx. measuring time: 1 minute!
In the anus (rectal) / 36.6 – 38.0 °C
Attention: For the prevention of rectal perforation in children (younger than 3 years), we recommend using another measuring method, or the use of a thermometer with flexible tip.
Carefully insert the measuring sensor 4 of the thermometer 2 to 3 cm into the anal aperture.
The use of a probe cover and the use of a lubricant is recom- mended.
If you are unsure of this measurement method, you should consult a professional for guidance/training.
Approx. measuring time: 1 minute!
Cleaning and disinfecting
For disinfection in home use environment, use a 70% Isopropyl alcohol swab or a cotton tissue moistened with 70% Isopropyl alcohol to wipe surface pollutants off the thermometer probe. Always start wiping from the end of the thermometer probe (approx. at the middle of the ther- mometer) towards the thermometer tip. Afterwards the entire thermom- eter (see number 5 in the drawing) should be immersed in 70% Isopropyl alcohol for at least 5 minutes (max. 24 hours). After immer- sion, let the disinfectant dry off for 1 minute before next use.
The thermometer is not intended for professional use.
Battery Replacement
When the symbol (upside-down triangle) appears in the display
the battery is flat and needs replacing. To replace the battery remove the battery compartment cover 3 from the thermometer. Insert the new battery with the + at the top. Make sure you have a battery of the same type to hand. Batteries can be purchased at any electrical store.
Technical Specifications
- Type: Maximum thermometer
- Measurement range: 32.0° C to 43.9° C
- Temp. < 32.0 °C: display «– -» for low (too low)
- Temp. > 43.9 °C: display «H» for high (too high)
- Measurement accuracy:
- ± 0.1° C; 34° C – 42° C
- ± 0.2° C; 32.0 – 33.9° C and 42.1 – 43.9° C
- Operating conditions:
- 10 – 40 °C; 15 – 95 % relative maximum humidity
- Storage conditions: -25 – +60 °C; 15 – 95 % relative maximum humidity
- Battery: LR41 (1.5V) / SR41 (1.55V)
- Battery lifetime: approx. 4500 measurements (using a new battery)
- IP Class: IP67
- Reference to standards:
- EN 12470-3, clinical thermometers;
- ASTM E1112; IEC 60601-1; IEC 60601-1-2 (EMC);
- IEC 60601-1-11
- Expected service life: 5 years or 10000 measurements
This device complies with the requirements of the Medical Device Directive
93/42/EEC.
Technical alterations reserved
Guarantee
This device is covered by a lifetime guarantee from the date of purchase. During this guarantee period, Microlife will repair or replace the defective product free of charge. Opening or altering the device will invalidate the guarantee. The following items are excluded from the guarantee:
- Transport costs and risks of transport.
- Damage caused by incorrect application or non-compliance withinstructions.
This device is covered by a lifetime guarantee from the date of purchase. During this guarantee period, at our discretion, Microlife will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee. The following items are excluded from the guarantee:
- Transport costs and risks of
- Damage caused by incorrect application or non-compliance with the instructions for use.
- Damage caused by leaking batteries.
- Damage caused by accident or
- Packaging/storage material and instructions for
- Regular checks and maintenance (calibration).
- Accessories and wearing parts:
Should guarantee service be required, please contact the dealer from where the product was purchased, or your local Microlife service. You may contact your local Microlife service through our website: www.microlife.com/support
Compensation is limited to the value of the product. The guarantee will be granted if the complete product is returned with the original invoice. Repair or replacement within guarantee does not prolong or renew the guarantee period. The legal claims and rights of consumers are not limited by this guarantee.
Contact Information:
Microlife UAB, P. Luksio g. 32, 08222 Vilnius, Lithuania