UNI-T UT139C True RMS Digital Multimeter Instruction Manual
- June 12, 2024
- UNI-T
Table of Contents
UNI-T UT139C True RMS Digital Multimeter
Overview
UT139AfB/C DMM is a small auto range, hand-held 3 1/2~3 5/6 bit true RMS
multimeter featuring complete function, novel structure, high reliability and
safety as well as large screen for display . It can be used for measuring
AC/DC voltage and current, variable frequency voltage (V.F.C), resistance,
diode, circuit on-off, capacitance, frequency ratio, NCV non-contact AC
voltage sensing and so on, is an ideal portable maintenance instrument for
users.
Accessory
Unpack and take out the instrument, please check following attachments
carefully for completeness or intactness.
In case of any shortage or damage, please contact with your supplier.
- An operating instruction manual
- A test pen (CAT III 600V) 10 A
- Point-type K( nickel chromiumcnnickel silicon) thermocouple (UT139C only)
- An optional current clamp (UT139C only)
Safety Information
UT139A/B/C is designed in compliance with standards such as IEC/EN61010-1,
pollution grade II, overvoltage (CATill600V) and double insulation standards.
Please comply with operation instruction specified in the Manual;
otherwise the protection provided by the instrument may be affected.
CE
This Meter complies with the standards EN 61010-1,EN 61010-2-030: in
pollution degree 2, overvoltage
category (CAT Ill 600V) and double insulation.
ETUCETL
CONFORMS TO UL STD 61010-1, 61010-2-030
CERTIFIED TO CSA STD C22.2 No. 61010-1, 61010-2-030
- It is forbidden to use the product without having rear cap in place, or otherwise there will be electric shocking.
- Prior to use, inspect the insulation layer of test pen for intactness, confirming no breakage and broken line.
- When LCD display shows the , it is required to replace the battery in time to ensure the measurement accuracy.
- Range switch shall be set at the correct measurement position.
- In case of electric shock and damage to the ins1rument, signals being measured shall not exceed rated limit value.
- To prevent any damage to the instrument, it is forbidden to change the gear of range switch in measurement.
- After each measurement, disconnect table pen and the circuit being measured; after the current measurement, especially the measurement of large current, it is necessary to power off before disconnecting table pen and the circuit being measured.
- In case of electric shock, it is required to be cautious when voltage being measured higher than DC 60V or AC 30Vrms.
- Do not use the product in high-temperature or high-humidity environment, particularly in the damp environment in where the instrument performance may be severely degraded
- Refrain from changing the internal wiring in the clamp ammeter to guard against damage to the meter and danger.
- Clean the meter case with damp cloth and mild detergent rather than the abrasive material and solvent.
Electric Symbol
Comprehensive Specification
-
Maximum voltage between input tenninal and grounding: see instruction about each input tenninal protection voltage.
-
10A terminal (CE) is equipped with:
F 10A H 600V fast-acting fuse (6×25) mm -
mNµA terminal (CE) is equipped with:
UT139A-FF 0.2A H 600V fast-acting fuse (6×32) mm
UT139B-FF 0.5A H 600V fast-acting fuse (6×32) mm
UT139C-FF 0.6A H 600V fast-acting fuse (6×32) mm -
Maximum display: (UT139A): 1999; (UT139B): 4000; (UT139C): 6000Refresh 2~3 times per second, display “OL” in case of overrange .
Capacitance and frequency (only applicable for UT139B/C): 9999 count.
Duty ratio (only applicable for UT139B/C): 1~99.9%Diode: about 2.1V (UT139A) and 3.2V(UT139B/C), displaying “OL” in case of overrange.
Range: auto/manual
Polarity: auto
Working temperature: o·c~40’C
Relative humidity: ?75% when o·c~30’C, and ?50% when 30’C~40’C
Storage temperature: -10’C~50’C -
Electromagnetic compatibility:
In 1V/m radio frequency (RF) field: Overall accuracy=specified accuracy+ 5% of range, and no specified index for RF over 1V/m. -
Operating ASL: 0~2000m
-
Internal battery: AA R6P 1.5Vx2
-
Low electricity: LCD displays””.
-
Contour dimension: about (175x80x48.5) mm
-
Weight: about 350g (inclusive of battery)
-
Safety Standard: IEC/EN 61010-1: CATIII600V; Pollution grade II
-
Accreditation: CE3
Appearance Structure (See Figure 1)
- Case
- LCD Display
- /4/7.: Selection buttons
- . Range Switch
- Measuring input terminal
LCD Display (See Figure 2)
Knobs and Keys for Range selection
Button
RANGE button: it can be used for selecting auto/manual range. After pressing,
it will switch one gear of switch, when reading the highest gear, jump to the
lowest gear range and in tum. When the time of pressing button is
~2s or switching a range, you’ll exit the manual range mode. (Only applicable
for V~ /WI~ )
MAX/MIN button
it can be used to automatically enter the manual range mode. In such case,
auto shutdown function is disabled and maximum value is displayed, after
another pressing on the button, the minimum value will be displayed and
values are displayed in tum (maximum value-minimum value). When the time of
pressing button is ~2s or switching a range, you’ll exit data recording mode
(only applicable for v~. C, I~ and ‘C/”F)
REL button
it can be used to automatically enter the manual range mode. The current
displayed value will be taken as the reference value and then the difference
between the measured value and reference value will be displayed,
after another press, you’ll exit the relative measurement mode. (Only
applicable for V~ . C, I~ . ‘C/”F and~~} The backlight will be illuminated
when the time of pressing button is ~2s, after about 15s, the backlight will
be automatically turned off; the backlight will be turned off if pressing the
key ~2s when the backlight is illuminated, (Only applicable for UT139A full
range: REL/LIGHT button)
NCV/mVrv button (Only appllcable for UT139A):
It can be used to switch NCV/mVrv. For detailed information, see the operation
instruction for non-contact AC voltage sensing.
Hz/% button (only applicable for UT139B/C):
It can be used to select the mode Hz/%, only applicable for the selection of
frequency, AC voltage/ current measurement modes.
SELECT button:
it can be used to select range (only applicable for multi-range). Under AC
mode, press the button ~2s. display “UFC”, enter V.F.C measurement mode and
measure the variable frequency voltage. After another ~2s
pressing on the button, display “End” and exit the V.F.C measurement mode.
HOLD button : (Applicable for full range)
It can be used to lock and hold the displayed value, in such case, LCD
displays the prompt “Cl”, after another press, it is unlocked and enter the
normal measurement mode.
The backlight will be illuminated when the time of pressing button is ;;:,2s,
after about 15s, the backlight will be automatically turned off; and the
backlight will be turned off if pressing the key ~2s when the backlight is
illuminated, (Only applicable for UT139A full range: HOLD /LIGHT button) 8
Measurement Instruction
Check the built-in AA 1.5Vx2 battery, display will show the symbol “C:.” when
lack of power, and then replace battery in time. It is required to pay
attention to the symbol beside the test pen socket, which reminds one of the
fact that in case of measurement safety, testing voltage or current shall not
exceed the specified value
AC and DC Voltage Measurement (See Figure 3)
- Connect the instrument with the load in parallel for measurement.
- When the input impedance of the instrument is about 1 OM Q , the load may cause measurement error in the circuit with high impedance. In most cases, the error can be neglected (0.1 %or lower) if the circuit impedance is under 1 Ok
Notes
- It is forbidden to input voltage higher than 600Vrms, despite of the possibility of measuring higher voltage, as it may damage the instrument.
- It is required to avoid the electric shock in measuring high voltage.
Resistance Measurement (See Figure 4a)
Connect the instrument with the load in parallel for measurement.
Notes
- The display will show “OL” when the measured resistance open-circuit or resistance value exceeds the maximum range of the instrument.
- Prior to measuring online resistance, it is necessary to switch off all powers in the circuits to be measured, and release all residual charges to ensure the measurement accuracy.
- In measuring low resistance, a measurement error in resistance of about 0.1 Q ~0.2 Q will be resulted by the test pen. In order to acquire accurate reading, it is required to short circuit the test pen, take REL relative measurement mode to ensure the measurement accuracy.
- Check the test pen for any loosening or other reasons in case there is a resistance value no less than 0.50 when test pen is short circuited.
- Several seconds may be required for the reading stability when measuring high resistance, which is normal for high resistance measurement.
- By using the resistance measurement function, it is allowed to make self-checking of the built-in fuse, see (Figuer 4b).
- No input higher than DC 60V or AC 30V is allowed.
Circuit on-off measurement (See Figure 5)
If the resistance of two terminals to be measured is bigger than 150 Q, there
will be a circuit break and buzzer will make no sound; if the resistance is
::::; 10 Q, the circuit is deemed with good conductivity and buzzer will
continuously sound.
Notes
- Prior to measuring online circuit on-off, it is necessary to switch off all power supplies in the circuits to be measured and release all residual charges to ensure the measurement accuracy.
- To prevent personal injury, it is not allowed to input the voltage higher than DC 60V or AC 30V.
Diode measurement (Figure 6)
“OL” will be displayed when the diode to be measured is an open circuit or
polarity is reversely connected. For Silicon PN junction, the normal value is
normally 500~800mV.
Notes
-
Prior to measuring online diode, it is necessary to o switch off all power supplies in the circuits to be measured and release all residual charges to ensure the measurement accuracy.
-
Test voltage for diode is about:
2.1V(UT139A) ,
3.2V(UT139B/C). -
To prevent personal injury, it is not allowed to input the voltage higher than DC 60V or AC 30V
Capacitance measurement (Only applicable for UT139B/C)
(See Figure 7) The instrument, when without any input, will display a fixed
reading which is the internal fixed capacitance value. When measuring small
range gear capacitance, the above value shall be subtracted from the value to
be measured to ensure the accuracy. The relative measurement REL function can
be used to automatically subtract the value to facilitate the measurement.
Notes
- The display will show “OL” when the capacitor to be becomes short-circuited or the capacitance value exceeds the maximum range of the instrument.
- Generally, several seconds will be taken to measure high-capacity capacitor.
- To prevent damage to the instrument and personal injury, it is required before testing to release all residual charges, which is particularly important for capacitor with high voltage.
Frequency/Duty Ratio Measurement (Only applicable
(See Figure 8) At the frequency measurement gear, press the button Hz/% to
select frequency/duty ratio measurement mode.
Notes
- To prevent personal injury, it is not allowed to input the voltage higher than DC 60V or AC 30V
Temperature measurement (Only applicable for UT139C)
(See Figure 9) Temperature sensor: It is only applicable for K (Ni-Cr and
Ni- Si) thermocouple. After startup, “OL” is displayed, complete Celsius or
Fahrenheit temperature measurement by connecting the product with K-type
temperature sensor. °F=1.8 C+32
Notes
The point type K (Ni-Cr and Ni- Si) thermocouple is only applicable for the
measurement of temperature under 230 C/446°F.
AC and DC current measurement (See Figure 10)
Connect the instrument with the load in serial for measurement. AC measurement
value will be true RMS.
Notes
-
Before connecting instrument in serial with the loop to be measured, switch off the power supply.
-
In measurement, it is required to use proper input terminal and function gear; if unable to estimate the current, the · Fuses are provided inside the 10A, mA7 µ I input jacks. It is forbidden to connect the table pen test pin in paralfel with any circuit, especially the power supply terminal, which may cause damage to the instrument and personal injury.
For security purposes, when measuring current higher than 5A, the time of each measurement should be controlled less than 10s and an interval of at least 15min should be maintained. · When measuring AC current online, it is allowed to press the button Hz/% to display online AC frequency/ duty ratio. -
60A AC and DC current clamp measurement (See Figure 11).
Connect as shown in the figure with the attached current clamp. (Only applicable for UT139C)
NCV Non-contact AC voltage sensing (See Figure 12)
If need to detect whether there is a AC voltage or electromagnetic field,
allow the front end of the instrument be close to the object to be sensed. The
analog quantity of sensed AC voltage is about: “EF” is displayed when
,;;;critical voltage.” EF” is displayed when UT139A> critical voltage, a five-
level sound effect (buzzing sound) is generated according to the voltage Vd to
mark the difference of sensed voltage.
“-” is displayed when UT139B/C>critical voltage,”- – – -” is designed
according to the size of voltage Vd and accompanied by different buzzing
sounds to mark the difference of sensed voltage.
By pressing NCV/mV..,,(only applicable for UT139A), display the analog
quantity of sensed mV” ‘to make the difference of AC voltage .
Notes
Test pen is not required for measurement when the range knob switch is set at
“NCV”.
Battery voltage measurement (only applicable for UT139A)
(See Figure 13) The range gear 1.5V is only applicable for the measurement of
battery ,;;;2v, with load resistance of about 51 Q, while the range gear 9V is
only applicable for the measurement of battery ,;;;1 sv, with load resistance
of about 1 k Q
To prevent built-in fuse F1 inside the instrument from burning out due to
over-load, it is forbidden to measure the battery or power supply beyond the
rated range.
The measuring time for battery voltage should be as short as possible since
there will be built-in analog load power consumption which may shorten the
service life of battery.
Other functions
- After 2s of full display following startup, enter the normal measurement mode. “ErrE” will be displayed in case of any fault to the internal EEPROM.
- Auto power-off : the instrument will be “automatically powered off’ to save energy in case of no operation to the knob switch within 15min. Under auto power-off statutes, any press on the SELECT of UT139A or any button of UT139B/C will “automatically arouse” the instrument, or restart the instrument by turning the knob switch to OFF. Under power-off status, press on SELECT and power on, the buzzer will make five sounds to remind that the auto power-off function has been cancelled. Restarting the instrument after power-off will recover the auto power-off function.
- Buzzer: A “Beep” sound (about 0.25s) from buzzer means the function button is valid when pressing any button or turning the function button. When measuring voltage or current: AC and DC voltage > about 600V mA gear AC/DC current > 190mA(UT139A), 390mA(UT139B), 590mA(UT139C)When A gear AC/DC current is >10A, buzzer will make continuous sounds to remain the over-range. In addition, the buzzer make make 5 continuous sounds about 1 min prior to auto power-off and then make one long sound prior to power-off. When the auto power-off function is cancelled, 5 sounds will be made by the buzzer every 15min.
- Low-voltage detection : it is used to detect the internal VDD. When the voltage is lower than 2.4V, the battery under-voltage symbol “” will be displayed, and the normal operation is available; while when lower than 2.2V,no operation is allowed and only the symbol is displayed after startup.
- When the battery supply voltage is reduced to 2.6V, the LCD backlight is weak or can not start; measurement functions can still be used normally.
Technical Index
Accuracy: ± (a% reading +b word number), one year of guarantee period. Ambient
temperature: 23°C ±5°C C73.4°F ±9r) Relative humidity: <75%.
DC voltage measurement
Range | Resolution | Accuracy |
---|---|---|
UT139A | UT139B | UT139C |
20.00mV* | 40.00mV* | 60.00mV* |
200.0mV** | 400.0mV** | 600.0mV** |
2.000V | 4.000V | 6.000V |
± C0.7%+3)
20.00V| 40.00V| 60.00V| 10mV
200.0V| 400.0V| 600.0V| 0.1V
600V| 600V| 600V| 1V
OMO. (There will be unstable digital display in case of */** range open- circuit; after connecting with the load, it can be controlled s ±1)
- Range : 20mV range for UT139A, 40mV range for UT139B, 60mV range for UT139C.
- Range : 200mV range for UT139A400mV range for UT139B 600mV range for UT139C.
- Maximum input voltage: ±600V
AC voltage measurement
Range | Accuracy | Resolution |
---|---|---|
UT139A | UT139B | UT139C |
20.00mV | 40.00mV | 60.00mV |
± (1.0%+3)
| ± (1.0%+3)| 10µV
200.0mV| 400.0mV| 600.0mV| 0.1mV
2.000V| 4.000V| 6.000V|
± C0.8%+3)
| 1mV
20.00V| 40.00V| 60.00V| 10mV
200.0V| 400.0V| 600.0V| 0.1V
600V| 600V| 600V| ± (1.2%+3)| ± (1.0%+3)| 1V
V.F.C 200.0V-600V
&.
| 0.1/1V| ± (4.0% +3)
Input impedance: about 10MO.
Display the true RMS. Frequency response: UT139A 45~400Hz, UT139B/C 45~1
kHz.(VFC: 45?400Hz)
Assurance of accuracy: 5~ 100% range, an allowance of <10 words of
residual reading for short-circuit.
It will be up to 3.0 when AC crest factor reaches full value (with except for
600V range, which is up to 1.5 when the range reaches the full value). Maximum
input voltage: 600Vrms
Resistance measurement
Range | Accuracy | Resolution |
---|---|---|
UT139A | UT139B | UT139C |
200.00* | 400.00* | 600.00* |
± (1.0%+2)
| ± (1.0%+2)| 0.10
2.000kO| 4.000kO| 6.000kO|
± C0.8%+2)
| 10
20.00kO| 40.00kO| 60.00kO| 100
200.0kO| 400.0kO| 600.0kO| 1000
2.000MO| 4.000MO| 6.000MO| ± (1.2%+3)| 1k0
20.00MO| 40.00MO| 60.00MO| ± (1.2%+3)| ± (1.5%+5)| 10k0
Range: measured value = displayed value – short-circuit value of test pen.
- Open-circuit: about *1 V
- Overload protection: 600V-PTC
*Circuit on-off, diode measurement**
Range | Resolution | Remark |
---|---|---|
**** | 0.10 | Circuit breakage resistance value is set as: > 150 Q, buzzer is |
soundless.
Good conductivity is set as: ,s;10 Q buzzer sounds.
| 1mV| Open circuit voltage is about 2.1V(UT139A), 3.2V{UT139B/C) Normal voltage value of silicon PN junction is about 0.5~0.BV.
Overload protection: 600V-PTC
Capacitance Measurement (Only applicable for UT1398/C)
Range | Resolution | Accuracy |
---|---|---|
9.999nF | 1pF | Under REL status: ±(4%+10) |
99.99nF~999.9µF | 10pF~0.1µF | ±(4%+5) |
9.999mF~99.99mF | 1µF~1OµF | ± 10%(,s;2mF) |
Overload protection: 600V-PTC
For capacitor ?1 µF, it is recommended to adopt REL measurement mode to ensure
measurement accuracy.
Frequency/duty ratio measurement (only applicable for UT139B/C)
Range | Resolution | Accuracy |
---|---|---|
9.999Hz~9.999MHz | 0.001Hz~0.001MHz | ± (0.1%+4) |
1%~99.9% | 0.1% | Not defined |
- Over-load protection : 600V-PTC Input range a: (D C level is zero) ?1OO kHz:100mVrms?a?20Vrms
- 100 kHz~1MHz : 200mVrms?a?20Vrms 1MHz: 500mVrms?a?20Vrms 5MHz~10MHz: 900mVrms?a?20Vrms
Duty ratio %: only applicable for measurement ?100kHz Remark
- During measurement of AC voltage or AC current, if need to read online frequency value or duty ratio, following input should be met: frequency response: ?1 kHz;
- AC voltage: mV range input ;?: 1 00mV; V range input ;?: range x6% *. AC current: input range a4000/6000µA, 400/600mA, 1 0A range: a;?: range x6% 400/600µA, 40/60mA, 4/6A range: a;?: range x60% 7. Temperature measurement (only applicable for UT139C)
Range | Resolution | Accuracy |
---|---|---|
oc | **** |
-40~1000°C
| -40~0°C| ****
1°C
| ±3
0~100°C| ±(1.0%+3)
100~ 1000°C| ±(2.0%+3)
OF| ****
-40~1832″F
| -40~32°F| ****
1°F
| ±5
32~212°F| ± (1.5%+5)
212~1832°F| ± (2.5%+5)
Over-load protection: 600V-PTC
Remark: The point type K (Ni-Cr and Ni- Si) thermocouple is only applicable
for the measurement of temperature under 230″C/446″F 8. DC current measurement
Range | Accuracy | Resolution |
---|---|---|
UT139A | UT139B | UT139C |
µA | 200.0µA | 400.0µA |
± (0.7%+2)
| ****
± (0.7%+2)
| 0.1µA
2000µA| 4000µA| 6000µA| 1µA
mA| 20.00mA| 40.00mA| 60.00mA| 10µA
200.0mA| 400.0mA| 600.0mA| 0.1mA
A| 2.000A| 4.000A| 6.000A| ± (1.0%+3)| ± (1.0%+3)| 1mA
10.00A| 10.00A| 10.00A| 10mA
µA mArange
F1 fuse: (cp6x32)mm FF0.2A H 600V (CE) UT139A FF0.5A H 600V (CE) UT139B FF0.6A
H 600V (CE) UT139C 10 A range: F2 fuse: (cp6x25)mm F 10A H 600V (CE)
AC current measurement
Range | Resolution | Accuracy |
---|---|---|
UT139A | UT139B | UT139C |
200.0µA | 400.0µA | 600.0µA |
±(1.0+3)
2000µA| 4000µA| 6000µA| 1µA
mA| 20.00mA| 40.00mA| 60.00mA| 10µA
200.0mA| 400.0mA| 600.0mA| 0.1mA
A| 2.000A| 4.000A| 6.000A| 1mA| ±(1.2%+3)
10.00A| 10.00A| 10.00A| 10mA
- Frequency response : UT139A 45~400Hz, UT139B/C 45~1kHz
- Display : true RMS.
- Assurance of accuracy : 5~100% range, an allowance of <2 words of residual reading for short-circuit. It will be up to 3.0 when AC crest factor reaches full value.
Over-load protection: (the same as the DC current over-load protection)
(60A) current clamp measurement (only applicable for UT139C)
Range | Resolution | Accuracy |
---|---|---|
60A de | 0.01A | ±(1.0+3) |
60A ac | ±(1.2+3) |
Over-load protection: 600V-PTC
Range | Resolution | Load current | Accuracy |
---|---|---|---|
1.500V | 1mV | *30mA | ±5% |
9.00V | 10mV | *10mA |
Upkeep and Maintenance
Warning
Power shall be switched off before opening the rear cover of the instrument;
and the test pen shall be away from the input terminal and circuit to be
measured.
Conventional upkeep and maintenance
- For upkeep and maintenance, wet cloth and mild cleanser rather than abrasive or solution shall be used to clean the meter housing.
- Please stop using and send for maintenance in case of any abnormal condition about the instrument.
- The inspection or maintenance for instrument, if necessary, shall be performed by the qualified professional maintenance personnel or designated maintenance department.
Battery or fuse tube replacement (See Figure 14)
Built-in battery shall be replaced in time when LCD displays the under-
voltage prompt” ? “, otherwise measurement accuracy may be affected. Battery
specification: AA 1.5Vx2
Operating steps
-
Set the power switch at “Off”, take the test pen away from the input jack and remove the protective sleeve .
-
Screw off one screw securing batter rear cover with screw driver, remove the cover and replace:
The under-voltage used battery and burnout fuse:
F1 fuse: Ccp6x32) mm FF0.2A H 600V (CE)(UT139A) FF0.5A H 600V (CE)(UT139B)
FF0.6A H 600V (CE) (UT139C) -
Screw off the second screw securing the rear cover with screw driver and remove the cover to replace the burnout F2 fuse Ccp6x25) mm F 10A H 600V (CE).
UNI-T
UNI-TREND TECHNDLDGV (CHINA) CD., LTD.
No6, Gong Ye Bei 1st Road,
Songshan Lake National High-Tech Industrial
Development Zone, Dongguan City, Guangdong Province, China
Tel: (86-769) 8572 3888
http://www.uni-trend.com