APERA LabSen 833 pH Electrode User Manual
- June 12, 2024
- APERA
Table of Contents
LabSen 833 HF pH Electrode
User Manual
LabSen 833 pH Electrode
LabSen electrochemical sensors are premium pH electrode with manufacturing
technology and key components imported from Switzerland. LabSen833 pH
electrode adopts special HF glass membrane, suitable for strong acid solution,
or solution containing HF ( ≥3pH)
This probe has following features:
- Impact-resist membrane (see the right picture), there is no danger of electrode breakage during normal use.
- Extremely long service life, able to measure solution containing HF for about 1000 times. (test condition: 25 °C, 1% HF concentration, test period 1 minute )
- Blue gel inner solution, does not flow and will not cause bubble.
- Silver ion trap reference prevents the contamination of the junction when testing sample contains sulfides and proteins.
Technical Data
Measuring Range | (0 ~ 12) pH | Electrolyte | 3M KCl |
---|---|---|---|
Temperature Range | (0 ~ 100) °C | Soaking Solution | 3M KCl |
Shaft Material | Lead-free Glass | Temperatursensor | NTC 30KΩ |
Membrane Typ | HF | Electrode Dimension | (Ø12×120) mm |
Membrane Resistance | <400MΩ | Connector | BNC/RCA |
Reference | Silver Ion Trap | Cable | Ø5×1m |
Junction | Ceramic |
Usage and Maintenance
2.1 When measuring, please unscrew the bottle cap, pull out the electrode and rinse it with deionized water. After using, please put the electrode back into the bottle and screw tight the cap.
2.2 Prior to measurement, remove the rubber plug to maintain pressure of the
reference solution,
keep consistent flow rate of reference solution and stable potentials of
junction.
2.3 Measurement suggestions:
(a) The electrode can measure HF solution lighter than 0.1mol/L (pH≈2.1),
within 3 minutes for each time. After long time test, the electrode appearance
is not obviously dissolved or damaged, with good repeatability and stability.
(b) Measure HF solution lighter than 0.5mol/L (pH≈1.8), after long time test,
if the membrane is obviously dissolved, it will shorten the service life. Our
recommend is avoid such test or test faster, less than 2 minutes for each
time.
(c) The electrode is not suitable to measure HF solution heavier than 0.5ml/L
(pH <1.8), it will cause the electrode obviously to underperform or damage.
2.4 After a period of usage, the reference solution will be running low.
Whenever the level falls to 1/2 height of the electrode, add 3M KCL solution
to the refilling hole by using syringe or pipette.
2.5 The connector of the electrode should keep clean and dry. If being
contaminated, please clean it with medical cotton and absolute alcohol and
blow dry to prevent the short circuit of the electrode and slow reaction of
electrode.
2.6 The electrode’s measuring tip should be soaked in the soaking bottle
containing certain amount of storage solution to keep the membrane hydrated
and junction unblocked. Clean the bottle and replace the storage solution if
the storage solution gets turbid and mildewed. The electrode should never be
sinked in pure water or buffer solution for long.
2.7 After 1-year of use, we recommend replacing the electrode to ensure the
best accuracy.
Warranty
We warrant this electrode free from defects in material and workmanship and
agrees to repair or replace free of charge, at option of APERA INSTRUMENTS any
malfunctioned or damaged product attributable to responsibility of APERA
INSTRUMENTS for a period of six months. Warranty period is the time limit to
provide free service for the products purchased by customers, not the service
life of the tester or electrodes.
This limited warranty does not cover any damages due to:
transportation; storage; improper use; failure to follow the product
instructions or to perform any preventive maintenance; modifications;
combination or use with any products, materials, processes, systems or other
matter not provided or authorized in writing by us; unauthorized repair;
normal wear and tear; or external causes such as accidents, abuse, or other
actions or events beyond our reasonable control.
APERA INSTRUMENTS (Europe) GmbH
Address: Wilhelm-Muthmann-Str.18,
42329 Wuppertal, Germany
Phone: +49 202 51988998
Website: www.aperainst.de
Email: info@aperainst.de
References
- Apera Instruments® Meters and Sensors for pH, Conductivity, TDS, Salinity, ORP, DO, Turbidity, ISE
- Apera Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>