ZEISS CT LUCIA 202 Hydrophobe Monofokale Instruction Manual

June 12, 2024
Zeiss

Hydrophobe MonofokaleZEISS CT LUCIA 202 Hydrophobe
Monofokale

Preparing the ZEISS CT LUCIA 202

ZEISS CT LUCIA 202 Hydrophobe Monofokale - LUCIARotate or pull the plunger all the way back from its initial position to provide space for the cartridge to be inserted.ZEISS CT LUCIA
202 Hydrophobe Monofokale - LUCIA 1.To set the screw injector mode, press the white plunger mechanism lock inside the finger flanges to the correct position.ZEISS CT LUCIA 202
Hydrophobe Monofokale - LUCIA 12Inject OVD into the back opening of the injector tip along the channels of the cartridge.ZEISS CT LUCIA 202 Hydrophobe Monofokale - LUCIA
12Place the IOL in a central position in the channels of the cartridge.
Point the leading haptic in towards the tip of the cartridge.ZEISS CT LUCIA
202 Hydrophobe Monofokale - LUCIA 12While closing the cartridge wings, place the trailing haptic over the groove at the back of the cartridge, so that it will not be caught by the plunger. Make sure that it is not squeezed between the wings.ZEISS CT LUCIA
202 Hydrophobe Monofokale - LUCIA 12Place the closed cartridge into the opening of the injector and slide all the way forward until you hear the click sound. Make sure that it is not squeezed between the cartridge and the injector body.ZEISS CT LUCIA 202
Hydrophobe Monofokale - LUCIA 12Ensure that the IOL is moving freely when advancing the plunger to inject the IOL. To commence implantation, when the tip is inside the incision, turn the injector one quarter rotation clockwise to ensure that the leading haptic is pointing left as it exits the tip.ZEISS CT LUCIA 202 Hydrophobe
Monofokale - LUCIA 12As the IOL optic leaves the injector and starts to unfold, slowly turn the injector counterclockwise back to the original position with the bevel edge pointing down.
Important:

  • When advancing the plunger to inject the IOL, it is critical to ensure that the injector has been changed into screw mode for implantation.
  • After the IOL is released, and the injector / cartridge has been removed, the trailing haptic may remain outside of the incision.
    Use a suitable instrument to manually move the haptic and IOL into position.

EN_32_022_0004III Printed in Germany CZ-VIII/2017
The contents of the brochure may differ from the current status of approval of the product or service offering in your country.
Please contact our regional representatives for more information. Subject to change in design and scope of delivery and due to ongoing technical development.
CT LUCIA 202 is either a trademark or registered trademark of Carl Zeiss Meditec AG or other companies of the ZEISS Group in Germany and / or other countries.
DualTec is a trademark of Ophtec B.V. ( **** 0344).
Printed on elemental chlorine-free bleached paper.
© Carl Zeiss Meditec AG, 2017. All rights reserved.

0297
CT LUCIA 202 DualTec™
OD655
0034Carl Zeiss Meditec AG
Goeschwitzer Strasse 51–52
07745 Jena
Germany
www.zeiss.com/ct-lucia
www.zeiss.com/med/contacts

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals