ZEISS CT SPHERIS 204 ACCUJECT 3.0-1P Injector Set Instruction Manual
- June 12, 2024
- Zeiss
Table of Contents
ZEISS CT SPHERIS 204 ACCUJECT 3.0-1P Injector Set Instruction Manual
ACCUJECTTM 3.0-1P injector set is compatible with ZEISS CT SPHERIS® 204 +31.0 to +45.0 D and ZEISS CT ASPHINA® 404 +31.0 to +42.0 D
Inject OVD into the cartridge tip and along the channels.
Place the IOL in the center of the open cartridge. Use forceps to push down on
the edges of the optic to ensure that the optic and haptics are positioned
under the ledges of the cartridge.
Maintain the IOL in place with the folding forceps while closing the
cartridge. Take care that the haptics do not slip out from the cartridge
grooves between the wings.
Advance the plunger until it begins to move the IOL forward and ensure that it is moving freely.
To control the release of the IOL from the tip, slow down the movement of the plunger when the optic portion of the lens reaches the bevelled end of the tip.
EN_32_022_0010I
The contents of the brochure may differ from the current status of approval of
the product in your country. Please contact our regional representative for
more information. Subject to change in design and scope of delivery and as a
result of ongoing technical development. CT ASPHINA and CT SPHERIS are
registered trademarks of Carl Zeiss Meditec. ACCUJECT is a trademark of
Medicel (0482). Printed on elemental chlorine-free bleached paper.