NATIONAL INSTRUMENTS SHC68-C68-S Shielded Cable User Guide
- June 12, 2024
- NATIONAL INSTRUMENTS
Table of Contents
NATIONAL INSTRUMENTS SHC68-C68-S Shielded Cable User Guide
This document lists specifications for the NI PXI-2536 544-crosspoint, FET matrix (NI 2536).
All specifications are subject to change without notice. Visit ni.com/manuals for the most current specifications.
Topology 1-wire 8 × 68 matrix
Caution To ensure the specified EMC performance, operate this product only with shielded cables and accessories.
Caution Refer to the Read Me First: Safety and Electromagnetic Compatibility document for important safety and electromagnetic compatibility information. To obtain a copy of this document online, visit ni.com/manuals and search for the document title.
Related Information
Refer to the NI Switches Help for detailed topology information.
About These Specifications
Specifications characterize the warranted performance of the instrument under
the stated operating conditions.
Typical Specifications are specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 °C ambient temperature. Typical specifications are not warranted.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
Caution The protection provided by the NI 2536 can be impaired if it is used in a manner not described in this document.
Input Characteristics
Maximum switching voltage (channel-to-ground) | ±12 VDC, 8 VAC |
---|---|
Maximum switching power (per channel, resistive) | 1.2 W |
Maximum switching current | 100 mA |
DC isolation resistance (between open terminals) | >1 GΩ, typical at 23° C >334 |
MΩ, typical at 55° C
Current leakage between column and ground (closed path)| 10 nA, typical (12
VDC applied at 25 °C)
Offset voltage| 10 µV, typical
Overvoltage protection|
Powered on| ±36 VDC
Powered off| ±40 VDC
Total path resistance, row-to-column
Typical| 10 Ω
Maximum| 15.5 Ω
RF Performance Characteristics
Test setups for RF characteristics used two 1-meter cables and two NI TBX-68 connector blocks.
Single crosspoint bandwidth (50 Ω system, one row to one column)| >1 MHz,
typical
---|---
Crosstalk (50 Ω system)
10 kHz| <-53 dB, typical
100 kHz| <-33 dB, typical
1 MHz| <-30 dB , typical
Dynamic Characteristics
FET operate time1
Typical| 12 µs
Maximum| 16 ms
Note Certain applications may require additional time for proper settling.
Maximum scan rate | 50,000 crosspoints/s |
---|---|
Simultaneous drive limit | 544 switches |
Expected relay life | Unlimited, when operated within specified limits |
Caution During chassis power up, the row and column connections may
produce a
charge injection. Refer to the following figures for information about how
this might affect loads that are connected to the front panel I/O connectors
and referenced to earth ground.
Chassis power-up charge injection
Row | 7.5 μC |
---|---|
Column | 1.5 μC |
Figure 1. Impact of Charge Injection at Power Up: Typical Voltage
Developed vs. Resistive Load (Using Test Setup in Figure 2)
Figure 2. Test Setup for Row and Column Connections
Related Information
Refer to the NI Switches Help for information about including additional
settling time.
Trigger Characteristics
Input trigger|
---|---
Sources| PXI trigger lines 0 to 7
Minimum pulse width| 70 ns
Output trigger|
Destinations| PXI trigger lines 0 to 7
Pulse width| Programmable (1 µs to 62 µs)
Physical Characteristics
Relay type | FET switch |
---|---|
I/O connector | Four 68-pin receptacle VHDCI |
Power requirement | 1.7 W at 3.3 V, typical, 1.3 W at 12 V, typical |
Dimensions (L × W × H) | 3U, one slot, PXI/cPCI module, 21.6 × 2.0 × 13.0 cm |
(8.5 × 0.8 × 5.1 in.)
Weight| 159 g (5.6 oz)
Pollution Degree| 2
Maximum altitude| 2,000 m
Shock and Vibration
Operational Shock| 30 g peak, half-sine, 11 ms pulse (Tested in accordance
with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-
28800F.)
---|---
Random Vibration|
Operating| 5 Hz to 500 Hz, 0.3 grms
Nonoperating| 5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC
60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-
28800F, Class 3.)
Diagrams
Figure 3. NI 2536 Hardware Diagram
Figure 4. NI 2536 Connector Pinout
Related Information
For topology-specific connection information, refer to your device in the NI
Switches Help.
Accessories
Visit ni.com for more information about the following accessories.
Caution NI products typically must be operated with shielded cables and accessories to ensure compliance with Electromagnetic Compatibility (EMC) requirements. To determine if shielded cables or accessories are required for this product, refer to the EMC specifications in the Electromagnetic Compatibility section of this document. If shielded cables or accessories are required for EMC compliance, do not use unshielded cables or accessories unless they are installed in a shielded enclosure with properly designed and shielded input/output ports, and are connected to the NI product using a shielded cable. If unshielded cables or accessories are not properly installed and shielded, the EMC specifications for the product are no longer guaranteed.
Table 1. NI Accessories for the NI 2536
Accessory | Part Number |
---|---|
CB-68LP/R unshielded, 68-pin I/O connector block | 777145-01, 777145-02 |
NI TBX-68 unshielded, I/O connector block with DIN-rail mounting | 777141-01 |
SH68-68, 68 pin VHDCI to 68 pin SCSI, twisted pair cable with basic shielding
for use with connector blocks (1 m, 2 m)| 191945-01, 191945-02
SHC68-C68-S, 68 pin VHDCI to 68 pin VHDCI, shielded cable (0.5 m, 2 m)|
186380-0R5, 186380-02
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical
equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for
sensitive electrical equipment for measurement, control, and laboratory use:
- EN 61326-2-1 (IEC 61326-2-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- AS/NZS CISPR 11: Group 1, Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
Note For EMC declarations and certifications, and additional information, refer to the Online Product Certification section.
CE Compliance
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory
compliance information. To obtain product certifications and the DoC for this
product, visit ni.com/ certification, search by model number or product line,
and click the appropriate link in the Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally
responsible manner. NI recognizes that eliminating certain hazardous
substances from our products is beneficial to the environment and to NI
customers.
For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle, all NI products must be
disposed of according to local laws and regulations. For more information
about how to recycle NI products in your region, visit
ni.com/environment/weee.
ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2007—2015 National Instruments. All rights reserved.
COMPREHENSIVE SERVICES
We offer competitive repair and calibration services, as well as easily accessible documentation and free downloadable resources.
SELL YOUR SURPLUS
We buy new, used, decommissioned, and surplus parts from every NI series. We
work out the best solution to suit your individual needs. Sell For Cash Get
Credit Receive a Trade-In Deal
OBSOLETE NI HARDWARE IN STOCK & READY TO SHIP
We stock New, New Surplus, Refurbished, and Reconditioned NI Hardware.
Bridging the gap between the manufacturer and your legacy test system.
1-800-915-6216
www.apexwaves.com
sales@apexwaves.com
References
- Test and Measurement Systems, a part of Emerson - NI
- Test and Measurement Systems, a part of Emerson - NI
- Engineering a Healthy Planet - NI
- Product Take-Back Program and Recycling - NI
- Product Documentation - NI
- NI Trademarks and Logo Guidelines - NI
- Test and Measurement Systems, a part of Emerson - NI
- Product Certifications - NI
- Engineering a Healthy Planet - NI
- Product Take-Back Program and Recycling - NI
- Product Documentation - NI
- Product Documentation - NI
- National Instruments SHC68-C68-S Shielded Cable | Apex Waves
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>