Delio B1CSsmiw8ML Handheld Ultrasonic Inhaler Instruction Manual
- June 12, 2024
- Delio
Table of Contents
- Delio B1CSsmiw8ML Handheld Ultrasonic Inhaler
- Getting to know your instrument
- Safety information and warnings
- Useful information on the device
- Description of device and accessories
- Initial Use
- Assembly
- Replacing the new rechargeable battery image 1
- Operating the device with theA/CAdapter
- Operation
- Breathing technique
- Cleaning and disinfection
- Cleaning
- Disinfection
- Disinfection with hot water
- Drying
- Storage
- Troubleshooting
- Technical specifications
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Delio B1CSsmiw8ML Handheld Ultrasonic Inhaler
Getting to know your instrument
This nebulizer is a mesh type nebulizer for treating the upper and lower
respiratory tract.
The nebulization and inhalation of medication prescribed or recommended by a
physician can prevent against diseases of the respiratory tract, treat the
side effects of these disorders and accelerate healing.
For more information on possible applications, consult your physician or
pharmacist.
The device is suitable for inhalation at home. Medicines should only be
inhaled when instructed by a physician.
Inhalation should be performed in a calm and relaxed atmosphere. Inhale slowly
and deeply to enable the medicine to reach the small bronchi deep into the
lungs. Exhale normally.
Safety information and warnings
warning
-
Before use, ensure that there is no visible damage to the unit or accessories. When in doubt, do not use the unit and contact your dealer or the customer service address provided.
-
Using the unit is not a substitute for consultation with or treatment form a physician. Whenever you have any pain or illness, always contact your physician first.
-
lfyou have health concerns of and kind , contact your physician.
-
Only use medicines that have been prescribed or recommended by your physician or pharmacist.
-
lf the unit dose is not functioning correctly or if you begin to feel unwell or feel pain, stop using the unit immediately.
-
During use, hold the unit away from the eyes, as some nebulizer medicines may cause damage to the eyes.
-
Do not operate the unit in the presence of flammable gases.
-
Do not use the device near electromagnetic transmitters.
-
This device is not intended to be used by children or persons with restricted physical, sensory (e.g. insensitivity to pain) or mental abilities or persons lacking in the required experience or knowledge for safe operation of the device, unless supervised or instructed by a person responsible for their safety.
-
Do not dip the unit into water and not use it in wet rooms. Do not allow any liquid to penetrate the unit.
-
Remove the batteries from the device if you will not be using it for an extended period of time. Leaking batteries may damage the device.
Important -
Store the device in a location protected against environmental influences. The device must be stored in the specified ambient conditions.
-
Do not use any liquids that have a viscosity (thickness) greater than 3, as these may damage the mesh irreparably.
-
Do not use any medicines in powder form (even if dissolved.)
-
Do not shake the atomiser, as this may cause liquid to leak, and the device may be damaged.
-
Unplug from the socket after using the adapter.
-
Switch the unit off immediately if it is defective or malfunctioning.
-
The manufacturer shall not be held liable for damage or injuries caused by improper or incorrect use.
Useful information on the device
Accessories
The safe functioning of the device can only be guaranteed if the accessories
recommended by the manufacturer are used.
Important
Protecting against leaks
When filling the medicine container with medicine, ensure that you do not
exceed the maximum mark (8ml). The recommended filling amount is between 2ml
and 8ml.
During use, you can tilt the device up to a maximum angle of 45° in any
direction without impairing its nebulizing function or the success of the
treatment. However, please try to hold the device as upright as much as
possible.
Automatic switch-off
The device has an automatic switch-off function. To prevent damage to the
mesh, the device switches off automatically when the medicine/liquid has been
almost entirely used up.
Do not use the device if the medicine container or water tank is empty. The
device switches off automatically if the substance to be sprayed is no longer
in contact with the mesh.
Description of device and accessories
For convenience, the following “1. Medicine oF container” and “2.Mesh atomizer” will be called by a joint name: medicine cup module ( 12)
-
Medicine container
-
Mesh atomizer
-
On/Off button / Operation LEDs:
Blue: device ready
Red: change batteries -
Apparatus body
-
Battery compartment
-
Adapter connector
Initial Use
Notes
The atomizer and accessories must be cleaned and disinfected before they are
used for the first time. For information about this, see “Cleaning and
disinfection”.
Assembly
- Remove the unit from the packaging.
- Fitthe Medicine cup module(12) onto the apparatus body(4 ). (image 1)
When remove the medicine cup module(12), please press the PUSH button, and pull up the medicine cup module. (image 2)
Replacing the new rechargeable battery image 1
- Open the battery cover on the bottom of machine.
- Putt in Battery
- Close the battery cover.
Notes
- Make sure that the medicine cup is completely empfy and the adapter and machine are disconnected before replacing a reN rechargeable battery.
- Toe machine can not ‘Mlf”k when the ballBly is being charged.
- Toe machine can ‘Mlf”k Continuously for about 80-100 minutes after a full charging.
Operating the device with theA/CAdapter
Before using the adapter, remove the batteries.
- lnsert the USB cable into the connection socket.
- Plug the adapter the A/C outlet.
Notes
- Remove the batteries when operating the device with the Adapter.
Operation
Preparing the nebulizer
For reasons of hygiene, it is imperative that the nebulizer and accompanying
accessories are cleaned and disinfected after every use.
If the therapy requires that various different medicines be inhaled in
succession, ensure that the medicine cup module is rinsed after every use.
Filling the atomizer
-
Open the medicine container(1) by opening the clip, and fill the medicine container with an isotonic saline solution or medicine. Do not overfill! The maximum recommended fill amount is 8ml. (image 4, 5, 6)
-
Use medication only on the advice of a physician, and check the appropriate inhalation duration and quantity for your needs.
-
lf the recommended dose of the medicine is less than 2ml, add this quantity to at least
4ml using only isotonic saline solution. Viscous medicines may also require dilution. Always follow the instructions of your physician -
Close the lid of the medicine container and lock the clip in place. (image 7)
Fitting accessories
- Attach the accessories (mouthpiece, adult it mask or child mask) firmly to the assembled medicine container(1)
- Place the device to your mouth and close your lips firmly around the mouthpiece. When using either of the masks, place it over the nose and mouth.
- One-key operation: Press the button once to turn on the machine and enter the adult atomization mode. Press the button again to switch the adult atomization mode to the child atomization mode. Press the button again to turn off. After booting into adult atomization mode. Press the key twice to shut down. Press and hold the key for 2 seconds to enter the cleaning mode.
- The spray mist coming from the device and the illuminated blue LEDs(3) show that the device is working perfectly. Inhaling correctly
Breathing technique
- The right breathing technique is important to ensure that the particles are distributed as widely as possible into the air ways. So that they can enter the air ways and lungs, you must inhale slowly and deeply, hold your breath briefly (5 to 10 seconds) and then exhale quickly.
- You should only use nebulizers to treat respiratory disorders if advised by your physician.
- Certain medicines are only available on prescription.
Notes
You should hold the device as upright as possible. Tilting it slightly will not, however, affect its function, since the container is leak proof, To ensure that the device functions correctly, make sure that you do not tilt it more than 45° in any direction and that the medicine is in contact with the mesh while in use.
Important
Essential oils , cough medicines, solutions designed for gargling, and drops for application to the skin or for use in steam baths are wholly unsuitable for inhalation using the nebulizer.
End of inhalation
- After treatment, switch off the device using the on I off button (3)
- The LEDs (3) go out.
- Once the inhalant has been fully sprayed out, the device will switches off automatically. For technical reasons, a small amount may still remain in the medicine container.
- lf applicable, disconnect the adapter(10) from the A/C outlet.
Cleaning and disinfection
Awarning
Observe the following hygiene instructions in order to avoid any health risks.
The nebulizer and accessories are intended for multiple use. Please note that
cleaning and hygienic reconditioning requirements differ according to the
different areas of application.
Notes
- Do not put pressure on the mesh with brushes or similar, as this may easily puncture the mesh causing irreparable damage, and will void the warranty. The metal is very thin and can be punctured easily.
- Prior to cleaning the body or the medicine chamber, always remove first the battery or adapter before cleaning as the liquid may leak into the electronic components.Any liquid that penetrates into the unit can damage the electrical parts and impair the function of the device.
Cleaning
- Remove the mouthpiece and any accessories from the atomizer.
- Add 6ml of clean water to the medicine container (1) and switch on the device to first remove any medicine residue from the mesh. (Please use up the entire liquid).
- Remove the batteries.
- If any traces of medical solutions or impurities remain on the gold-colored contacts on the device and atomizer, use cotton buds to remove them.
Disinfection
Follow the steps listed below carefully in order to disinfect your atomizer
and the accessories. It is advisable to disinfect the individual components
after the last time they are used each day at the latest.
Fist , clean the atomiser and the accessories as described under ‘Cleaning’.
You can then continue whit disinfection.
Disinfection with hot water
- Place 5ml of hot water inside the medicine cup module.
- Spray until the water inside the medicine module runs out.
Drying
- Carefully dry all components using a soft cloth.
- Shake the mesh atomiser lightly backward and forwards(5 to 10 times), so that the water inside is removed.
- Place the individual components on a dry, clean and absorbent surface and allow to dry completely(for at least 4 hours).
Notes
Ensure that all parts are thoroughly dried after cleaning, otherwise, there is
an increased risk of bacterial growth.
Once all components are fully dry, reassemble the nebulizer and place the
components in a dry, sealed container.
Ensure thatthe medicine container has dried completely by shaking it.
Otherwise, nebulization may not function properly after reassembling the
device. If this is the case, disassemble the medicine container again and
shake it so thatthe water can escape. Once it has been reassembled, the
nebulizer should function properly again as usual.
Storage
- Do not store the nebulizer in a damp atmosphere(e.g. in the bathroom) or transport it together with damp objects.
- Store and transport the nebulizer away from direct sunlight.
Troubleshooting
Problem Question | Possible Cause/Remedy |
---|
The nebulizer produces little or no aerosol.
| 1. lnsufficient medicine in the Nebulizer.
2. The nebulizer is not held in an upright position. 3.The medicine used is unsuitable for spraying
(e.g. too thick, viscosity must be less than 3 cP) 4.The metal contacts between the medicine cup
module and body is dirty.
Output is too low.
| 1. The batteries are flat. Replace the batteries or connect the power supply unit and try again.
2. There are air bubbles inside the medicine container which are preventing the medicine from coming into continuous contact with the mesh. Please check and remove any air bubbles.
3. Particles on the mesh are impeding output. To remove the particles, fill the nebulizerwith 2 to 3 drops of vinegar and 3 to6ml of water, then nebulizer this solution completely. Do not inhale this spray and ensure that youclean and disinfect the medicine container afterwards.
4. Mesh is worn out.
Which medicines are suitable for inhalation?| Only the doctor can advise you which medicine to use to treat your condition. Consult your physician. With the mesh nebulizer, you can atomize medicines with a viscosity lower than 3cP.
Some inhalation solution remains in the nebulizer.
| This is normal and occurs for technical reasons. Stop inhaling as soon as you can hear a marked difference in tl’le sound made by the nebulizer, or when the device switches off automatically due to insufficient inhalant.
What special steps should be taken for babies and children.
| 1. ln babies, the mask should cover the mouth and nose to guarantee effective inhalation.
2. ln children, the mask should also cove the mouth and nose.
Note: Children should only use the device with help and under supervision of an adult. Never leave a child alone with the nebulizer.
Inhalation with the mask takes longer.
| This is for technical reasons. Less medicine is inhaled per breath through the mask holes than using the mouthpiece.
The aerosol is mixed with ambient air through the holes in the mask.
---|---
Does each user need their own accessories?|
This is absolutely necessary for hygiene reasons.
Technical specifications
- Power consumption: 1.5 W
- Input: 5V/1A(USB) or DC 3V (2 AA batteries);
- Oscillation frequency: 1 OS kHz± 10%
- Condensation rate: ;,,0.2 ml/min.
- Atomized particles:~ 4μm
- Volume: max. 8ml
- Product weight: 9±10%
- Size (LxWxH):
- Housing material:
- Operating conditions: Temperature:+5″C-+40″C;
- Relative humidity:<93% RHnon-con-dancing
- Storage and transport conditions:
- Operating conditions: Temperature: -2o·c-+5o·c
- Relative humidity:<93% non-condensing
Please follow the doctor’s prescription.
Specifications are subject to change without prior notice
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>