SO-LOW MV85 Series Ultra Low Freezer User Manual

October 30, 2023
SO-LOW

MV85 Series Ultra Low Freezer

Product Information

The product is a freezer with various safety features and
symbols to help users operate it safely. The manual includes
information on pre-installation, temperature and electrical
specifications, alarm system, custom settings, advanced settings,
service settings, advanced service settings, download, and upload.
The product comes in three models: MV85-1 with 0.25 cu. ft., MV85-2
with 2 cu. ft., and MV85-3 with 3 cu. ft. It has a 3/8 HP
refrigeration system and requires a dedicated line with 115 volts,
60 Hertz, and 15 Amps.

Product Usage Instructions

  1. Read all instructions and follow basic safety precautions
    before using the unit.

  2. Do not modify the plug provided with the freezer. If it will
    not fit the outlet, have a proper outlet installed by a qualified
    electrician.

  3. Position the freezer at least 6 inches away from any wall or
    object on any side.

  4. Do not attempt to repair or replace any part of the freezer for
    servicing without first contacting the So-Low Service Department
    while under warranty.

  5. Save these instructions for future reference.

  6. Refer to the illustrated symbols section of the manual to
    understand the warnings and cautions before operating the
    unit.

  7. Connect the appliance directly to a fixed installation without
    using extension leads or adaptors. If the appliance is connected to
    a long extension cable, there is a risk that the cable will become
    hot.

  8. Unplug the freezer before any technical service is performed on
    the unit.

  9. Only plug this unit into the proper outlet. Do not attempt to
    modify the plug in any way, as improper use of the electrical plug
    will void the warranty.

  10. Conform to the electrical codes to ground the equipment to
    avoid personal injury or damage to the equipment.

To reduce the risk of fire, electric shock or injury to persons using this freezer, read all instructions and follow basic safety precautions before using the unit, including the following:
Do not modify the plug provided with the freezer. If it will not fit the outlet, have a proper outlet installed by a qualified electrician.
Do not position equipment so it is difficult to disconnect from the power supply. freezer must be at least 6″ away from any wall or object on any side.
While under warranty, do not attempt to repair or replace any part of the freezer for servicing without first contacting the So-Low Service Department.
SAVE THESE INSTRUCTIONS

MEANING OF ILLUSTRATED SYMBOLS…………………………………… STARTING INSTRUCTIONS…………………………………………………..

PRE-INSTALLATION INFORMATION…………………………………………

TEMPERATURE SPECIFICATIONS…………………………………………. ELECTRICAL SPECIFICATIONS……………………………………………..

ALARM SYSTEM……………………………………………………………….. 8
· HOW TO MUTE AND ACCEPT ALARMS

CUSTOM SETTINGS……………………………………………………………. 10 ADVANCED SETTINGS………………………………………………………… 11 SERVICE SETTINGS…………………………………………………………….. 12 ADVANCD SERVICE SETTINGS……………………………………………… 12

DOWNLOAD…………………………………………………………………………. 13 UPLOAD………………………………………………………………………………. 13

MV85-1 DIAGRAM………………………………………………………………….. MV85-1 PARTS LIST……………………………………………………………….. MV85-2 DIAGRAM………………………………………………………………….. MV85-2 PARTS LIST……………………………………………………………….. MV85-3 DIAGRAM………………………………………………………………….. MV85-3 PARTS LIST………………………………………………………………..

Various symbols are used in this safety manual in order to use the unit without danger of injury and damage of the unit. A list of problems caused by ignoring the warnings and improper handling is divided as shown below. Be sure that you understand the warnings and cautions in this manual before operating the unit.
CAUTION
BLACK WITH YELLOW BACKGROUND LIGHTNING BOLT
CAUTION, RISK OF ELECTRICAL SHOCK
WARNING
BLACK WITH YELLOW BACKGROUND EXCLAMATION POINT
CAUTION, REFER TO ACCOMPANYING DOCUMENTS

WARNING
THIS FREEZER IS PROVIDED WITH AN INPUT CIRCUIT PROTECTIVE DEVICE WHICH SHALL BE MAINTAINED AND SERVICED BY QUALIFIED PERSONNEL ONLY.
CAUTION
UNDER NO CIRCUMSTANCES USE EXTENSION LEADS OR ADAPTORS. CONNECT THE APPLIANCE DIRECTLY TO A FIXED INSTALLATION. IF THE APPLIANCE IS CONNECTED TO A LONG EXTENSION CABLE THERE IS A RISK THAT THE CABLE WILL BECOME HOT.
CAUTION
UNPLUG FREEZER BEFORE ANY TECHNICAL SERVICE IS PERFORMED ON THE UNIT.
WARNING
DO NOT POSITION EQUIPMENT SO IT IS DIFFICULT TO DISCONNECT FROM THE POWER SUPPLY.

-40°C TO -85°C

· ·

MV85-1 MV85-2 MV85-3

115 VOLTS 60 HERTZ 1 PHASE

15 AMP
DEDICATED LINE

NEMA 5-15

WARNING
ONLY PLUG THIS UNIT INTO THE PROPER OUTLET. DO NOT ATTEMPT TO MODIFY PLUG IN ANY WAY. IMPROPER USE OF THE ELECTRICAL PLUG WILL VOID WARRANTY
CAUTION
FAILURE TO GROUND THE EQUIPMENT MAY CAUSE PERSONAL INJURY OR DAMAGE TO THE EQUIPMENT. ALWAYS CONFORM TO THE ELECTRICAL CODES.

MODEL

CU. FT.

MV85-1 0.25

MV85-2 2

MV85-3 3

REFRIGERATION SYSTEM 3/8 HP REFRIGERATION SYSTEM 3/8 HP REFRIGERATION SYSTEM 3/8 HP REFRIGERATION SYSTEM

REFRIGERANT EP88 ­ 6.10 OZ. EP88 ­ 6.88 OZ. EP88 ­ 6.90 OZ.

This appliance does not offer automatic defrosting as such function would increase the chamber temperature (and stored samples) to a value that could destroy stored samples that are temperature sensitive. Therefore, you have
to perform defrosting manually by following these instructions:
Defrost the appliance at least twice a year. 1. If the appliance is loaded with samples or products, move these to another
appliance to maintain the temperature of the products. 2. Turn off the appliance on the display by pressing the ON/OFF button.
3. Unplug the main power cable from the power socket. 4. Open the door and inner doors of the appliance.
5. Place cloths or towels in the bottom of the appliance to collect any defrost water. 6. Leave the appliance open for a minimum of 24 hours allowing the ice to melt.

CAUTION
DISCONNECT THIS UNIT FROM THE POWER SUPPLY BEFORE PERFORMING MAINTENANCE ON THE UNIT.
· · · ·
WARNING
IT IS RECOMMENDED TO SLOWLY RE-ADD YOUR PRODUCT INTO THE FREEZER TO PREVENT AN EXTREME LOAD ON THE COMPRESSORS, WHICH COULD SHORTEN FREEZER LIFE EXPECTANCY.

·
·
·
WARNING
WHEN FILLING THE UNIT WITH RACKS/BOXES, MAKE SURE TO PRE-COOL THEM IN ORDER NOT TO EFFECT THE INSIDE TEMPERATURE OF THE UNIT.
WARNING
THE INNER DOORS SHOULD BE USED AT ALL TIMES. MAKE SURE TO CLOSE THEM TIGHTLY. THE MANUFACTURER IS NOT RESPONSIBLE FOR DAMAGE TO THE FREEZER CAUSED BY FAILURE TO DO SO.
WARNING
CLEAN THE FREEZER BOTH INSIDE AND OUT USING A DAMP, WRUNG OUT CLOTH. DRY WITH A DRY CLOTH. THE SMELL FROM THE PLASTIC PARTS IN THE FREEZER WILL DISAPPEAR ONCE THE FREEZER HAS BEEN COOLED DOWN.

CONTROL DISPLAY

Mute an alarm
Press the Alarm mute button to mute an alarm. This procedure will silence the alarm for 5 min, where after it will start sounding again until being muted or accepted. * Please note that muting an alarm will only silence the alarm, the alarm is still active.
Accept an alarm
Press Enter to show an alarm (if an alarm is active). – Select Back to return to main screen – Select Next to see other active alarms – Select OK to accept the shown alarm If multiple alarms are active, the next alarm will be shown after accepting the previous alarm. Please note that accepting an alarm will definitively cancel the current alarm. (if the alarm conditions remain unchanged, a new alarm will be activated according to the alarm settings).
Remote alarm contact
The alarm relay will not be switched off when muting an alarm. If all alarms are accepted, the remote alarm will be turned off.

Read User Manual Online (PDF format)