CEFALY Enhanced Quick
Start Guide
Welcome to the CEFALY community! Use this as a guide for your first treatment
session, but remember, these are just highlights. Be sure to read the full
user manual before your first treatment as well.
First Things First
1. Fully charge device until indicator light turns green (yellow = low
battery)
2. Wash & dry forehead just above
eyebrows
3. Position electrode & attach
device
Starting Your Session
Select & confirm a treatment program as follows:| ACUTE
Program| PREVENT
Program
---|---|---
When To Use| Use at the earliest sign of a migraine attack| Use daily to
prevent future migraine attacks
How To Start Program| Tap button 1X| Tap button 2X (with a slight pause in
between each press)
What You Will Hear| 1 beep| 2 beeps
What You Will See| LED indicator flashing single turquoise light
intermittently| LED indicator flashing two purple lights intermittently
Session Time| 60 minutes| 20 minutes
Find Your Sweet Spot Follow these instructions to set your
preferred intensity level:
Intensity for both programs ramps up gradually before reaching full intensity
at 14 min.
During the first 14 minutes of either treatment as intensity increases, press the device button 1X to stabilize the intensitylevel for the remainder of the session. To ramp up intensity faster during the first 14 min, press & hold button. This is more applicable for experienced CEFALY users. To end treatment early, simply pull device off electrode.
During treatment, you may feel a tingling sensation. This is caused by tiny
electrical impulses designed to stimulate the trigeminal nerve. Using your
CEFALY device should never be painful.
When new to CEFALY, aim to use the device at the highest intensity level
that’s comfortable for you. You do not need to treat at maximum intensity and
you can increase intensity during future sessions when you feel ready.
Electrode Care
Proper care of your CEFALY electrodes is key to maximizing the number of treatments each can help perform. Follow these guidelines:
- Because our fingers have natural oils that reduce the effectiveness of the gel pad, we suggest holding electrode pads by the edge.=
- Be sure you’re thoroughly washing and drying your forehead before each treatment session. Use basic soap, avoiding alcohol wipes and facial cleansers.
- During active treatment, avoid excessive movement if possible.
- When you remove an electrode after a treatment, immediately replace it on the plastic backing, adhesive side down. Then seal it in the Resealable Electrode Bag.