Dr Trust 1201 Portable ECG Smart Connect User Manual
- June 10, 2024
- DR Trust
Table of Contents
- 1201 Portable ECG Smart Connect
- GENERAL NOTES
- IMPORTANT SAFETY NOTES
- PRODUCT DESCRIPTION
- PREPARATION BEFORE TAKING MEASUREMENT
- TAKING MEASUREMENT
- RESULT DISPLAY
- CLEANING AND MAINTENANCE
- MESSAGES & TROUBLE SHOOTING
- SPECIFICATIONS
- EMC INFORMATION
- SYMBOL DEFINITIONS
- CONTACT US
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Dr Trust
Portable ECG Smart Connect -1201
QUICK START GUIDE
1201 Portable ECG Smart Connect
(There are three different methods -A “right index finger chest, B “left index finger chest”, and C left-hand right hand” of taking the measurement.)
Without Bluetooth
Press the On/Off button to turn the device on and the device shows the full
display.
Choose the method and put the Dr Trust Portable ECG Smart Connect on your body
part accordingly to start measurement.
The result will be shown on the LCD after 30 seconds automatically.
With Bluetooth
Download Dr Trust 360 App from the Play Store/App store to connect to your
smartphone or tablet with Bluetooth.
Enable Internet, GPS & Bluetooth for completing registration.
You rdata is automatically synchronized with the App for evaluation and easy
sharing with your doctor in real-time.
DR TRUST 360 APP DOWNLOAD
System Requirements: Smartphones or Tablets with Bluetooth Connectivity
Please scan the QR code to enter Dr Trust 360 App web page.
https://link.drtrust360.com/3wSZPux
Follow the instructions to register and use the app.
GENERAL NOTES
- This is an Electrocardiogram (ECG) device that provides referential data that could be taken into consideration by medical professionals.
- The applied part is ECG
- Self diagnosis or treatment is dangerous behavior. Please don’t use the cardiogram or data generated by the device to diagnose yourself.
- Consult with medical professionals if you notice any change in your condition, regardless of the measurement results.
- The measurement results produced by the device only reflect the measurement taken at the moment.
- The cardiogram cannot show all variations and symptoms of the heart disease, please consult with medical professionals when feeling discomfort with your heart.
IMPORTANT SAFETY NOTES
- It is not recommended to use the device with a cardiac pacemaker or other implanted devices. Be sure to follow your doctor’s instruction if necessary.
- Do not use this device during an MRI scan.
- Do not use this device with a defibrillator.
- Do not submerge (or immerse) the device in water or liquid. Do not clean the device with acetone, or other volatile solutions. Clean the unit with a cloth dampened with water or mild detergent. Then wipe it dry with a dry cloth.
- Do not put the device in autoclaves, or gas sterilization devices.
- Avoid hitting, dropping, stamping, shaking of the device.
- Do not disassemble the device that may result in damage, malfunction, or abnormal operation.
- Do not use this device on people with sensitive skin or allergies.
- This device is not approved for use on children weighing less than 10 kg.
- Do not store/expose the device in the following environments: Locations exposed to direct sunshine, high temperature or humidity, heavy dirt, near water or fire, and under high electro- magnetic influence.
- Store the device in safe place to avoid small parts swallowed by children under any circumstances.
*Static electricity may cause damage to the electronic components, to prevent electrostatic damage to the device, grounding yourself to discharge the static electricity from the body is required before handling the device. Increasing humidity, moisturizing skin can effectively reduce the accumulation of body surface electrostatic charge.
PRODUCT DESCRIPTION
3.1 Main Features
- Dr Trust Portable ECG Smart Connect- 1201 can complete your cardiogram measurement in about 30 seconds.
- Your Dr Trust Portable ECG Smart Connect- 1201 will automatically analyze and provide referential results using measurement data.
- Dr Trust Portable ECG Smart Connect- 1201 can take up to 500+ measurements with two new coin-typed batteries and keep records of 36 measurements.
- The APP can assist you to:
– Download ECG data.
– Configure your Dr Trust Portable ECG Smart Connect- 1201 .
– Store measurement records in chronicle order.
– Display ECG waveforms for reference to medical professionals .
3.2 Package Contents
Your package should include the following items
Main Unit | Dr Trust Portable ECG Smart Connect- 1201 |
---|---|
Accessories | Instruction manual |
Two coin – typed batteries (trial batteries)
3.3 Components
3.4 LCD Display
-
Heartbeat Indicator: Flashes in sync with the heartbeat during measurement.
-
Heart Rate: Shows average heart rate of recording.
-
Battery Level Indicator: Shows current battery level.
-
Digits Display:
a. Shows the sequence number of the measurement results stored in memory.
b. Shows the 30 second count down during measurement. -
Time: Shows date and time.
-
Bluetooth Connection: Displays the status of Bluetooth connection.
PREPARATION BEFORE TAKING MEASUREMENT
4.1 Battery Installation
To charge your Dr Trust Portable ECG Smart Connect- 1201, please follow the steps below:
- Remove the battery cover by inserting the tip of your finger into the opening of the cover.
- Follow the +(positive) and -(negativep ol a rities indi cat edon the battery compartment.
- Slide the battery cover until it clicks into place.
4.2 Things to know Before taking Measurement
To assure accurate measurement, please observe the following:
- Make sure that electrodes are directly in contact with your skin.
- If your skin or hands are dry, moisten your skin with damp cloth before taking measurement.
- If the electrodes are dirty, use a soft cloth or a cotton swab moistened with disinfectant alcohol to clean dirt from the surface.
- During the measurement, do not touch your body with the hand with which you are taking the measurement.
- Make sure there is no skin contact between your right and left hand (measuring method C), or between your hand and chest (measuring methods A/B). Otherwise, the measurement cannot be taken correctly.
- Ideally perform the measurement either sitting down or lying down. Do not stand up.
- Do not move or talk during the measurement and hold the device still. Movements of any kind will falsify the measurements.
TAKING MEASUREMENT
5.1 Operation Procedure
There are three different methods of taking the measurement. Start with
measurement method A, “right index finger-chest.” If this method does not
deliver measurements, or delivers only unstable measurements (“EE” is
frequently displayed), switch to method B “left index finger-chest” and, if
necessary, to method C”left hand-right hand.”
The most suitable method/ procedure depends on the heart configuration (shape
of the heart) of each individual user. If it is not possible to take stable
measurements using a given measurement method, this could have a harmless
cause such as the shape of the
heart. However, the cause may also be an illness or disease.
Measurement method C offers maximum comfort, but offers a much lower
measurement stability than methods A or B.
Press the on/ off button for about three seconds to turn the device on. The
device shows the full display.
ATTENTION
If segments are missing, stop using the device and immediately contact
customer services.
To test whether the full display is completely displayed, hold the On/Off
button when switching on the device.
5.2 Measurement Methods
Measurement Method A| “right index finger-chest” (approximately corresponds to
“lead 2”)
---|---
| Place your right index finger on the two upper electrodes on the device and
hold the device.
| Place the bottom electrode of the Dr Trust Portable ECG Smart Connect-1201
on bare skin about 5 centimeters (2 inches) below your left nipple supported
by the ribcage, and press it GENTLY until you hear the click sound.
Lightly press the electrode against your chest until you hear or feel a click.
Important: Do not exert excessive pressure on the electrodes with your
fingers.
Measurement Method A| “left index finger-chest” (approximately corresponds to
“lead 3”)
---|---
| Place your left index finger on the two upper electrodes on the device and
hold the device vertically in your hand.
| Place the bottom electrode of the Dr Trust Portable ECG Smart Connect-1201
on bare skin about 5 centimeters (2 inches) below your left nipple supported
by the ribcage, and press it GENTLY until you hear the click sound.
Lightly press the electrode against your chest until you hear or feel a click.
Important: Do not exert excessive pressure on the electrodes with your
fingers.
Measurement Method C| “left hand – right hand” (approximately corresponds to
“lead 1)
---|---
| Place your right index finger on the two upper electrodes on the device.
Place a finger of your left hand on the lower
electrode.
Press the lower electrode until you hear or feel a click.
Important: Do not press the device too firmly against your skin.
CAUTION
Please note that there must be no skin contact between your right and left
hand (measurement method C) or hand and chest (measurement methods A/B).
Otherwise, the measurement cannot be taken correctly. Do not move during the
measurement, do not speak and hold the device still. Movements of any kind
will falsify the measurements.
- The device will start beeping (if the beep is enabled) and the number indicator will flash for a few seconds. This indicates that the measurement is about to begin.
- The measurement is completed, and the results are displayed at the end of the 30-second countdown.
Note: If the bottom electrode is removed within the space of three
seconds after the measurement starts, the measurement will be displayed as EE.
5.3 Common Wrong Operations
Please avoid the following operations:
Finger not touching the electrodes tight enough.
Measuring over/through clothing.
Holding Dr Trust Portable ECG Smart Connect in the wrong direction.
Holding Dr Trust Portable ECG Smart Connect with left hand.
RESULT DISPLAY
6.1 Measurement Results
The possible results are shown below.
| : Normal Result
---|---
| : Slow Heart Rate (<60 bpm)
| : Fast Heart Rate (>100 bpm)
| : Paused Heart beat (pause>2 s)
| (Disrupted Rhythm) : Irregular Rhythm
| : Abnormal Waveform (Altered Waveform)
Please note that if the displayed heart rate is flashing, it indicates ECG
signals are unstable or weak. Please retake the measurement.
6.2 Reviewing results on Dr Trust Portable ECG Smart Connect- 1201
-
Press the button for 3 seconds to turn on your Dr Trust Portable ECG Smart Connect- 1201. The LCD shows the total number of records stored.
-
The button again, and the LCD shows the average heart rate of the latest measurement and the Digits Display shows the sequence number of stored records.
-
By pressing the button repeatedly, the result will be shown in a cyclic order.
CLEANING AND MAINTENANCE
Please clean the surface of the Dr Trust Portable ECG Smart Connect – 1201
with a soft cloth dampened with water.
If the electrodes are dirty, please use a soft cloth or a cotton swab
moistened with disinfectant alcohol to clean dirt from the surface.
MESSAGES & TROUBLE SHOOTING
Message Shown | Possible Cause | Suggestion |
---|
Measurement result shows flashing Heart
Rate.| Unstable or weak ECG signals.| Please retake the measurement following
the manual’s instruction.
Measurement result shows
EE.| Failure analysis resulted from either low contact pressure or high noise
level during measurement.| Please retake the measurement following the
manual’s
instruction.
Problem| Possible Cause| Suggestion
---|---|---
Power does not come on after pressing the
power button for 3 seconds.| 1. Battery power is low.
2. Batteries are installed with wrong direction.| 1. Replace with new
batteries.
2. Make sure the polarities of batteries are matched with the ones indicated
on the battery compartment.
Electrode touching skin, but measurement
does not start.| Pressing too lightly.| Make sure to press the bottom
electrode down enough.
Measurement records disappear from the
device.| Old records are overwritten by new ones due to storage overload.|
Download data periodically to your computer for record keeping.
Bluetooth connection failed.| Basic communication problem between the device
and Smartphone/Pad.| Please turn off the device, “Dr Trust iBP Monitor” APP
and Bluetooth function of Smartphone/Pad, then follow the instructions
mentioned above to reconnect again.
Device ID is not displayed on the box in the “Setting” function of “iBP
Monitor APP.| Problems of data exchanging for the first time connection.| If
the device is discovered successfully, there will be a blank on the box, then
click on the blank to manually link the device with APP and display the device
ID.
SPECIFICATIONS
Dr Trust Portable ECG Smart Connect- 1201
Model| Smart Portable ECG Pen
Induction System| single channel, bipolar
Bandwidth| 0.05 40 Hz
Sampling Rate| 256 Hz
Heart Rate Range| 5 199 beats/min
Memory| 36 measurements
Power Supply| two coin-typed batteries
Battery life| 500+ measurements with two new coin-typed batteries
Transportation Environment Storage Environment| -20 60°C, 10 95% RH
-20 60°C, 10 95% RH
Operating Environment| 5 40°C, 30 85% RH
Weight| 30 g (without batteries)
Dimensions| 93 x 38 x 11 mm
Air Pressure| 700-1060hPa
EMC INFORMATION
Guidance and manufacturer’s declaration – electromagnetic emissions
The Dr Trust Portable ECG Smart Connect- 1201 is intended for use in the
electromagnetic environment specified below. The customer or the user of the
Dr Trust Portable ECG Smart Connect-1201 should assure that it is used in such
an environment.
Emissions| Compliance| Electromagnetic environment– guidance
RF emissions CISPR 11| Group 1| The Dr.Trust ECG Pen uses RF energy only for
its internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The Dr.Trust ECG Pen is suitable for use in
all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations/ flicker emissions IEC 61000-3-3| Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Dr Trust Portable ECG Smart Connect- 1201 is intended for use in the
electromagnetic environment specified below. The customer or the user of the
Dr Trust Portable ECGSmartConnect-1201should assure that it is used in such an
environment.
Immunity test| EN 60601 test level| Compliance level| Electromagnetic
environment –guidance
Electrostatic discharge (ESD) IEC 61000-4-2| ±6 kV contact ±8 kV air| ±6 kV
contact ±8 kV air| Floors should be wood, concrete or ceramic tile. If floors
are
covered with synthetic material, the relative humidity should be at least 30
%.
Electrical fast transient/burst IEC 61000-4-4| ±2 kV for power supply lines|
±2 kV for power supply lines| Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5| ±1 kV line(s) to line(s)
±2 kV line(s) to earth| ±1 kV line(s) to line(s)
±2 kV line(s) to earth| Mains power quality should be that of a typical
commercial or hospital environment.
Voltage caps. | <5 % Ur | <5 % Ur | Mains power quality should be |
---|---|---|---|
short | (>95 % dip in UT) | (>95 % dip in Ur) | that of a typical commercial or |
interruptions and | for 0,5 cycle | for 0,5 cycle | hospital environment. If the |
user
voltage variations| 40 % UT| 40 % UT| of the Dr.Trust ECG Pen requires
on power supply| (60 % dip in UT)| (60 % dip in UT)| continued operation
during
input lines| for 5 cycles| for 5 cycles| power mains interruptions, it is
IEC 61000-4-11| 70 % UT| 70 % UT| recommended that the
| (30 % dip in UT) for 25 cycles| (30 % dip in UT) for 25 cycles| 870E be
powered from
<5 % UT| <5 % UT| an uninterruptitie power supply or
(>95 % dip in UT)| (>95 % dip in UT)|
for 5s| for 5s| a battery.
Power frequency| 3 A/m| Not applicable| Not applicable
(50/60 Hz) magnetic field| | |
NOTE Ur is the a.c. mains voltage prior to application of the test level
Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3| 3 Vrrns 150 kHz to 80
MHz
3 V/m 80 MHz to 2.5 GHz
3 Vrms
3 V/m| Portable and mobile RF communications equipment should be used no
closer to any part of the 870E , including cables, than the recommended
separation distance calculated
from the equation applicable to the frequency of the transmitter. Recommended
separation distance
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
Distance in metres (m).
---|---|---
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Dr trust Portable ECG
Smart Connect – 1201 is used exceeds the applicable RF compliance level above,
the Dr trust Portable ECG Smart Connect- 1201 should be observed to verify
normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the Dr trust Portable ECG
Smart Connect-
1201. Recommended separation distances between portable and mobile RF
communications equipment and the Dr Trust Portable ECG Smart Connect- 1201.
Dr trust Portable ECG Smart Connect- 1201 is intended for use in an
electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the Dr trust Portable ECG Smart Connect- 1201 can
help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Dr trust Portable ECG Smart
Connect- 1201 as recommended below, according to the maximum output power of
the communications equipment.
For transmitters rated at a maximum output power not listed above,
the recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 : At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 : These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
SYMBOL DEFINITIONS
| Type CF applied part| | The product is with Bluetooth function
---|---|---|---
| ATTENTION, should read the Instruction for Use| | Protection against solid
foreign objects of 125 mm 0 and greater and against vertically falling water
drops when enclosure tilted up to 15°
| CE Mark ( made in compliance with Directive 93/42/EEC on Class Ila medical
devices)
| Consult Instruction for Use| | This way up
| Manufacturer| | Fragile
| Date of manufacture| | Keep dry
Reference number| | Keep away from sunlight
| Serial number| | Permissible operating temperature and humidity
| Batch number| | Permissible storage/transport temperature and humidity
CONTACT US
CONTACT ADDRESS
USA
Nureca INC.USA
276 5th Avenue, Suite 704-397,
New York (NY) – 10001, USA
INDIA
Corporate Office (Mumbai)
Nureca Limited
128 Gala Number Udyog Bhavan,
1st Floor Sonawala Lane, Goregaon East
Mumbai City Maharashtra 400063
Contact us
India: +91-7527013265 /+91-9356658436
Website: www.drtrust.in
Corp Website:
www.nureca.com
Customer Care Executive Email:
customercare@nureca.com
Connect with us on social networks
Facebook: @drtrust
Instagram: @drtrustisin
Youtube: NurecaUsa
COPYRIGHT © 2021 NURECA LTD ALL RIGHTS RESERVED
Scan to View
Product Demo Video
www.drtrustusa.com/1201
References
- Dr Trust - India's Most Trusted Healthcare & Wellness brand | Dr Trust
- Nureca Limited - Specialists in Health and wellness products
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>