Bittium Respiro HSAT Recording with 2 RIP Belts User Guide

June 10, 2024
Bittium

Bittium Respiro HSAT Recording with 2 RIP Belts

Bittium HSAT™ Recording with 2 RIP belts

Equipment

Bittium-Respiro-HSAT-Recording-01

Equipment needed in recording:

  1. Bittium Respiro™

  2. Pulse oximeter

  3. Quick Guide/ Notes and Warnings

  4.  Nasal cannula

  5.  Two RIP belts

  6.  Medical tape

  7. Respiro™ Patch for 2 RIP Belts and a spare patch
    Equipment needed for charging Respiro (if required):

  8.  Charger

  9. Charger dock

Do not use a broken device or cannula whose packing is opened.

Starting a recording

Attach the RIP belt to the patch as shown.

Bittium-Respiro-HSAT-Recording-02

  • Adjust the belts if they are too loose.
  • Check that the belts are not crossed. One belt for thorax and one for abdomen.

Bittium-Respiro-HSAT-Recording-03

Attach the cannula to Respiro and rotate the connector slightly clockwise until you feel it connect.
Bittium-Respiro-HSAT-Recording-04

  • Move the slider up the tubes to adjust the fit.

  • Choking hazard! Position the nasal cannula carefully as shown in the figures.

  • In case a cannula with a mouthpiece is used, the oral part should be pointed towards your mouth.
    Bittium-Respiro-HSAT-Recording-05

  • Choking hazard! Use medical tape for securing the cannula.
    Bittium-Respiro-HSAT-Recording-06

  • Press the power button for about 3 seconds until the LED indicators are lit in blue.

  • Continue to the next step. Bittium-Respiro-HSAT-Recording-07

  • Press the power button again for about 8 seconds until the device vibrates and the LEDs start blinking blue.

  • The LEDs are lit in green to indicate a successful recording start.

  • The LEDs will turn off automatically after 30 seconds if everything is OK and you are lying down.
    Bittium-Respiro-HSAT-Recording-09

  • If any LED indicator blinks red or yellow, check that the item corresponding to the indicator is connected properly. Explanations for the indicators can be found on the next page.Bittium-Respiro-HSAT-Recording-10

During a recording

You can shortly press the power button to mark times when you are not sleeping, such as going to bed, going to the bathroom, waking up… (Optional)

Stopping a recording

  • Press the power button for about 3 seconds until the device vibrates 3 times and the LED indicators are lit in blue one by one from left to right.
  • Pulse oximeter turns off automatically when you remove the sensor from your finger.
  • The LED indicators are lit in blue for 5 minutes but you can move to the next step.

Bittium-Respiro-HSAT-Recording-12

  • Detach the equipment from yourself. Place all the items back into the carry case and return it to the clinic
    Bittium-Respiro-HSAT-Recording-13

LED indicators

If any LED indicator blinks red or yellow, check that the item corresponding to the indicator is connected properly. Once connections are OK, the LEDs will be lit in green. Contact your healthcare professional if LEDs are lit solid red or there are no lights at all.

Bittium-Respiro-HSAT-Recording-14

Connect Respiro to an outlet using the charger dock and the charger if you want to see the battery charge level. The LED indicators represent the device battery charge level. If at least two green LED indicators are blinking or all the LEDs are continuously lit, the battery charge is sufficient.

Bittium Biosignals 5800670A01, 27.6.2022