AtriCure ACM1 CryoICE Box User Guide

June 11, 2024
AtriCure

AtriCure ACM1 CryoICE Box

AtriCure-ACM1-CryoICE-Box-product

Technical Reference Guide

The AtriCure cryoICE BOX is a medical device used for freezing tissue during surgical procedures. It is designed to be used by healthcare professionals only.

Product Usage Instructions

  1. Set-Up Sequence: Before using the cryoICE BOX, follow the set-up sequence:
  2. Open the tank and verify that the valve assembly is attached to the tank with the orange heater band.
  3. Turn on the machine. Be sure to route the N2O vent tubing to a safe area prior to use.
  4. Verify the tank indicator. If it is flashing, this indicates that the tank is closed. Open the tank valve to resolve.
  5. Operation: To operate the cryoICE BOX:
  6. Connect the probe. Connect the red thermocouple connector to the red connector and the black thermocouple connector to the black connector. Push back the locking collar on the orange exhaust connector prior to connecting, and push the blue inlet connector straight in. Pull back on the orange and blue connectors after connecting to verify a good connection.
  7. Start a freeze cycle by pressing the Start/Stop button. The button will turn blue. Refer to the temperature display for troubleshooting tips.
  8. Wait until the Start/Stop button turns green before beginning a new freeze cycle.
  9. To abort a freeze cycle, press the Start/Stop button again.
  10. Purge and Power Down Sequence: After use, follow the purge and power down sequence:
  11. Disconnect the probe. Disconnect all connections and discard the probe (it is a single-use device). Push the locking collars of the orange and blue connectors towards the cryoICE BOX to disconnect.
  12. Purge gas from the machine.
  13. Turn off the machine.

For more information on using the cryoICE BOX, refer to the AtriCure Cryo Module (ACM) User’s Manual. It is important to follow all safety instructions and instructions for use when operating the cryoICE BOX.

Scan QR code to view AtriCure’s cryoICE BOX Quick Start Guide Video

cryoICE BOX Set-Up Sequence

Open tankAtriCure-ACM1-CryoICE-Box-fig- \(2\)

  • Open valve “all the way open”
  • Active tank should have orange heater band on it. Verify valve assembly is attached to tank with orange heater band.

Turn on machineAtriCure-ACM1-CryoICE-Box-fig- \(3\)

  • Turn on power switch located on back of unit
  • cryoICE BOX should be set up at least 15-minutes prior to the procedure

Note : Be sure to route the N2O vent tubing to a safe area prior to use.

Verify tank indicatorAtriCure-ACM1-CryoICE-Box-fig-
\(4\)

Note : Flashing tank indicates the tank is closed. Open tank valve to resolve
Note : Be sure to route the N2O vent tubing to a safe area prior to use

Connect probe
Connect probes to color matching connectionsAtriCure-ACM1-CryoICE-Box-fig-
\(5\)

  • A. Thermocouple connectors: Red to red and black to black
  • B. Orange exhaust connector: push back locking collar prior to connecting
  • C. Blue inlet connector: pushes straight in
    Note : Pull back on orange and blue connectors after connecting to verify good connection. Refer to ACM Cryo Console ‒ Technical Reference Guide for troubleshooting tips for connecting probe.

Start a freeze cycleAtriCure-ACM1-CryoICE-Box-fig-
\(6\)

  • A. Press the “Start/Stop” button to initiate the freeze cycle (button will turn blue)
  • B. Timer starts when the temperature reaches –40°C
  • C. Temperature display
    Refer to ACM Cryo Console ‒ Technical Reference Guide for troubleshooting tips

Freeze completed, the auto-defrost cycle beginsAtriCure-ACM1-CryoICE-Box-
fig- \(7\)

  • A. When the timer hits “0”, the machine will go into defrost (orange) mode and begin to warm probe to 10°C before returning to green ready status
    Note : Wait until “Start/Stop” button turns green before beginning a new freeze cycle

Aborting a freeze cycleAtriCure-ACM1-CryoICE-Box-fig-
\(8\)

  • A. Freeze can be stopped at any time by pressing the “Start/Stop” button once
    Machine will initiate a defrost cycle (orange) and return to ready status (green)

cryoICE BOX Purge and Power Down Sequence

Note : Leave unit powered ON until purge sequence is completed

  1. Disconnect probe
  2. A. Disconnect all connections and discard probe (single use device)
  3. B. Push locking collars of orange and blue connectors towards cryoICE BOX to disconnectAtriCure-ACM1-CryoICE-Box-fig- \(9\)
  4. Close tank
  5. Close valve completely (turn to right)AtriCure-ACM1-CryoICE-Box-fig- \(10\)
  6. Purge gas from machine
  7. A. Press and hold vent button until gas gauge stops moving
  8. Gas gauge is mounted on cart frameAtriCure-ACM1-CryoICE-Box-fig- \(11\)
  9. Turn off machineAtriCure-ACM1-CryoICE-Box-fig- \(12\)

See AtriCure Cryo Module (ACM) User’s Manual for a complete list of safety related instructions and instructions for use.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

AtriCure User Manuals

Related Manuals