STEVENS HydraProbe SDI-12 Professional FW User Guide

June 9, 2024
STEVENS

SDI-12 Quick Start Guide
Professional FW Version 6.0

HydraProbe SDI-12 Professional FW

SDI-12 communication protocol allows compatible devices to communicate with each other. More information about SDI-12 can be found at http://www.sdi-12.org. 4.1, 4.0, 3.0, 2.8 and 2.7 firmware versions have a different array of commands. Contact Stevens Water for more information.

Model Numbers

Version Part # Suffix

2| Professional, w/25 ft. cable
4| Professional, w/50 ft. cable
6| Professional, w/100 ft. cable

Power

Power Requirements 9 to 16 VDC (12VDC Ideal)
Power Consumption <1 mA Idle, 10 mA for 2s Active

Wiring

Red Wire + Power Input
Black Wire Ground
Blue Wire SDI-12 Data Signal

Addressing

The first character of any command or response on SDI-12 is the sensor address. A lowercase ‘a’ is used to represent the address. Each SDI-12 sensor must have its own unique address. The default address is “0”.
Use SDI-12 “Transparent Mode” to issue commands.

SDI-12 Command Response Description
aAb! b Change Sensor Address

a – Sensor Address
b – New Sensor Address

Identification

A request for identification will return the sensor address, part number, firmware version, sensor version, calibration, and serial number.

SDI-12 Command Response Description
aI! a12STEVENSWnnnnnv.vvvcSNxxxxxxxx Send Identification

a – Sensor address
12 – SDI-12 protocol version
STEVENSW – Manufacturer
nnnnn – Part number
v.vvv – Firmware version
c – Calibration
xxxxxxxxx – Serial number

Measurements

SDI-12 Command Response Description
am! attn. Request Measurement

a – Sensor address
ttt – seconds (000 – 999) until the
measurement is ready
n – number of data fields (1-9) in the
measurement
aD0!| a| Send Measurement Readings
F – Soil Moisture
I – Bulk EC (Temp Corrected)
G – Temperature (C)
aD1!| a| Send Measurement Readings
H – Temperature (F)
J – Bulk EC
L – Real Dielectric Permittivity
aD2!| a| Send Measurement Readings
M – Imaginary Dielectric Permittivity
K – Pore Water EC
O – Dielectric Loss Tangent
aM1!| attn.| Request Measurement
tat – seconds (000 – 999) until the
measurement is ready
n – number of data fields (1-9) in the
measurement
aD0!| a| Send Measurement Readings
L – Real Dielectric Permittivity
M – Imaginary Dielectric Permittivity
N – Imaginary Dielectric Permittivity
(Temperature Corrected)
aD1!| a

| Send Measurement Readings
O – Dielectric Loss Tangent
P – Diode Temperature

The following tables list the values and units:

Selector Order Parameter Unit
F Soil Moisture Water fraction by Volume (wave)
G Soil Temperature Celsius (C)
H Soil Temperature Fahrenheit (F)
I Bulk EC
(Temperature Corrected) Siemens/Meter (S/m)
J Bulk EC Siemens/Meter (S/m)
K Pore Water EC Siemens/Meter (S/m)
L Real Dielectric Permittivity
M Imaginary Dielectric Permittivity
N Imaginary Dielectric Permittivity
(Temperature corrected)
O Dielectric Loss Tangent
P Diode Temperature Celsius (C)

SDI-12 Measurement Sets

Command| P1| P2| P3| P4| P5| P6| P7| P8| P9
aM! and aC!| F| I| G| H| J| L| M| K| O
aM1! and aC1!| L| M| N| O| P| | | |

Pore Water Offset

SDI-12 Command Response Description
aXR_PWOS! a Read Pore Water Offset
aXWPWOS! a Write Pore Water Offset
aXD_PWOS! a+3.4 Reset Pore Water Offset to default 3.4

Calibration

The following extended command will change the coefficients in one of two general formulas that translate the real dielectric permittivity to soil moisture. In many cases, the HydraProbe will not need to be recalibrated. The default General calibration has been heavily reviewed and will provide reasonable accuracy for most applications. If you need to change the calibration or if a custom calibration is required, we recommend referring to the HydraProbe user manual for more information.

SDI-12 Command Response Description
aXR_SOIL! a<G/O/R/C/K> Get Current soil type

G – General
O – Organic
R – Rockwool
C – Custom 1
K – Custom 2
aXRSOIL!| a<G/O/R/C/K>| Write New Soil Type
G – General
O –Organic
R – Rock Wool
C – Custom 1
K – Custom 2
aXR_COEFA!| a| Read coefficient A
aXR_COEFA!| a| Read coefficient B
aXR_COEFA!| a| Read coefficient C
aXR_COEFA!| a| Read coefficient D
aXR_COEFA!| a| Read coefficient E
aXR_COEFF!| a| Read coefficient F
aXR_COEFF!| a
| Read all coefficients
aXWCOEFA
!| a| Write coefficient A
aXWCOEFA!| a| Write coefficient B
aXWCOEFA!| a| Write coefficient C
aXWCOEFA!| a| Write coefficient D
aXWCOEFA!| a| Write coefficient E
aXRCOEFF!| a| Write coefficient F
aXR_COEFF!| a
| Reset all coefficient to default

Accuracy and Ranges

Parameter|
---|---
Soil moisture for inorganic mineral soils| Accuracy*: +/- 0.01 WFV for most soils ( m3,m-3)
+/- <0.03 for fine textured soil (typical)
Range: From Complete Dry to Full Saturation (0% to 100% of saturation)
Bulk EC| Accuracy: +/- 2.0% or 0.02 S/m Whichever is greater
Range: 0 to 1.0 S/m
Temperature| Accuracy:  +/- 0.3 °C
Range: -40 to 75 °C
Inter-Sensor Variability| +/- 0.012 WFV (Typical)
Pore Water EC| Hillforts Equation, depends on soil conditions

*Accuracy of soil moisture depends on the soil and is highly variable.