Danfoss PFM 100 Measuring Device User Guide
- June 4, 2024
- Danfoss
Table of Contents
Danfoss PFM 100 Measuring Device
Application
The PFM 100 is used for measuring the differential pressure on both sides of a valve in hydronic systems.
The set consists of:
- One PFM 100 Measuring Instrument.
- Two hoses with couplings.
- Two needles incl. ball valves.
Flow and pressure can be displayed in various units, which you can choose from in the menu.
The PFM 100 can communicate with you in one of 10 possible languages.
Please note that the pressure unit must not be exposed to subzero temperatures. The pressure unit must at all times be stored at temperatures above 0° C.
Functions
| ON / OFF
button.
---|---
| Menu button.
Used for navigating through the menu options. Press to enter the menu. Press
and hold to exit the menu.
| Confirmation button.
Press to select the marked item.
Usage
-
Connect the hoses to the valve.
-
Air the hoses.
-
Connect the valve to the PFM 100.
-
Zero adjust.
• From the main menu select the Zeroing option in order to correct the static pressure influence.
• To reset the device to zero follow the instruction on the display. -
Type in the Kv-value.
If the kV value of the measured valve is known, it can be entered in the PFM 100 to display the valve flow.
• From the main menu select the Kv value option.
• Press the Confirmation button to go through the sequence of numbers.
• Press the Menu button to increase / decrease the numbers.
• Press and hold the Confirmation button to save and exit the menu. -
Read off the flow in the display of the PFM 1
Technical Specifications
Type | Specifications |
---|---|
Pressure sensor | Piezo resistive true differential |
Pressure range | 10 bar |
Max. overpressure | Positive side: 15 bar Negative side: 10 bar |
Non-linearity and hysteresis error | 0.15% of pressure range |
Temperature error | 1.5% over ambient and medium temperature |
Medium temperature | -5 to 90°C |
Ambient temperature | -5 to 50°C |
Storage temperature | -5 to 50°C |
Pressure connection | R21 quick couplers |
Batteries | 2 x AA NiMH rechargeable |
Power consumption | Max. 55 mA |
Display | 128 x 64 monochromatic, backlight |
Keyboard | 3 keys |
Pressure units | 11 |
Flow units | 11, incl. US units |
kV-range | 0 to 99999, with 0.1 step |
Dimensions | 94 x 218 x 35 mm |
Weight | 600 g, incl. batteries |
Cover | IP65 |
Calibration validity | 12 months |
Ordering
No. | Type | Set / pcs | Danfoss code no. |
---|---|---|---|
1 | FPM 100 Measuring Instrument | 1 | 003L8260 |
Accessories
No. | Type | Set / pcs | Danfoss code no. |
---|---|---|---|
1 | Measuring hose, 2 x 1.5 m | 1 set | 003L8261 |
2 | Needle | 1 pcs | 003L8262 |
3 | Quick coupling | 1 pcs | 003L8263 |
4 | Filters | 1 set | 003L8264 |
5 | Strap | 1 pcs | 003L8265 |
6 | ASV-I/M measuring KIT for PFM100 | 1 set | 003L8274 |
7 | ∆p tool for pump optimization | 1 pcs | 013G7861 |
DanfossA/S
Climate Solutions• danfoss.com • +45 7488 2222
Any information, including, but not limited to information on selection of
product, its application or use, product design, weight, dimensions, capacity
or any other technical data in product manuals, catalogues descriptions,
advertisements, etc. and whether made available in writing, orally,
electronically, online or via download, shall be considered informative, and
is only binding if and to the extent, explicit reference is made in a
quotation or order confirmation. Danfoss cannot accept any responsibility for
possible errors in catalogues, brochures, videos and other material. Danfoss
reserves the right to alter its products without notice. This also applies to
products ordered but not delivered provided that such alterations can be made
without changes to form, tit or function of the product.
All trademarks in this material are property of Danfoss A/5 or Danfoss group
companies. Danfoss and the Danfoss logo are trademarks of Danfoss A/5. All
rights reserved.
Documents / Resources
|
Danfoss PFM 100 Measuring
Device
[pdf] User Guide
PFM 100, Measuring Device, PFM 100 Measuring Device
---|---
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>