testoon MCC1010 1000 A AC Current Clamp Owner’s Manual

June 10, 2024
testoon

testoon MCC1010 1000 A AC Current Clamp

Product Information

MCC1010 and MVC1010 1000 A AC Current Clamp
The MCC1010 and MVC1010 1000 A AC Current Clamp is a non-contact current sensing clamp used for measuring AC current up to 1000 A. It has a CAT IV 600 V current measurement rating and a 1 mA/1 A sensitivity with a rated burden of

  1. The clamp has large jaws that can fit cables up to 52 mm and is suitable for leakage currents. The output terminals are protected with a voltage limiting circuit to ensure safe-to-touch voltage under all rated conditions. MCC1010 and MVC1010 are rated to CATIV 600 V, according to the international standard IEC 61010.

Product Usage Instructions

When using the MCC1010 and MVC1010, it is important to note that  the rated terminating resistance (burden) of 1 must be provided by the instrument connecting to the clamp. The clamp is intended to clamp around a single phase of the asset being tested and not phase + neutral, as this could yield incorrect results due to current flowing in opposite directions. Keep the clamps separate wherever possible to improve the quality of multiphase current measurements.

An ammeter measuring up to 1 A can be connected directly to the MCC1010 or MVC1010. When measuring current up to 1000 A, the ammeter reading will be 1 A due to the 1000:1 step-down ratio of the clamp. If using a TRMS multimeter, the voltage reading will be TRMS because of the accuracy and crest factor performance of the clamp.

When completing resistance measurements, both clamps must be used. The MCC1010 induces a current through a conductor while the MVC1010 measures the voltage drop across the conductor. This way, non-contact resistance measurements can be undertaken using the MCC1010 and the MVC1010 clamps.

The MCC1010 and MVC1010 are simple to use and have applications with various instruments including power quality analyzers, oscilloscopes, motor testers, clamp enabled earth testers, attached rod technique (ART), stakeless testing, data loggers, and digital multimeters.

INSTRUCTION

  • CAT IV 600 V current measurement to1000 A AC
  • 1 mA/1 A sensitivity with rated burdenof 1 Ω
  • Large jaws for cables up to 52 mm
  • Overvoltage protection of the outputterminals
  • Suitable for leakage currents

DESCRIPTION
The current probe is designed for measurements of AC currents up to 1000 A with a 1000:1 step-down ratio.
The probe offers excellent accuracy (0.3% error at full range) and small phase shift (0.7° at full range) so it can be used for metering of power, energy and power quality applications. Such high performance is achieved because the magnetic cores are made from high-permeability FeNi alloy.
The rated terminating resistance (burden) of 1 Ω must be provided by the instrument connecting to MCC1010 and MVC1010. The output terminals are protected with a voltage limiting circuit ensuring a safe-to-touch voltage under all rated conditions.
MCC1010 and MVC1010 are rated to CATIV 600 V, according the international standard IEC 61010.

APPLICATIONS
The MCC1010 and MVC1010 are non contact current sensing clamps which is simple to use and has application with the following instruments:

  • Power quality analysers
  • Oscilloscopes
  • Motor testers
  • Clamp enabled earth testers
  • Attached is Rod Technique (ART)
  • Stateless testing
  • Data loggers
  • Digital multimeters

CLAMP USAGE
The MCC1010 and MVC1010 are intended to clamp around a single phase of the asset to be tested and not phase + neutral because this would yield an incorrect result as a consequence of current flowing in opposite directions. Keep the clamps separate wherever possible, to improve quality of multiphase current measurements.

An ammeter measuring up to 1 A can be connected directly to the MCC1010 or MVC1010. A current of 1000 A will result in the ammeter reading 1 A (1000:1 step-down ratio).

The MCC1010 and MVC1010 generate 1 mV output signal across 1 Ω burden resistance for every 1 A of primary current, so if using a multimeter, one with an mV AC range is required. Dedicated earth testers instruments like Megger’s DET range, DET3TC, DET4TC and DET4TCR will need to ensure the input resistance and measurement circuitry are appropriate for the 1 mV – 1 V input range and a primary current limit of 20 A.

If using a TRMS multimeter the voltage reading will be TRMS because of the accuracy and crest factor performance of the MCC1010 and MVC1010 clamps.

Both clamps must be used when completing resistance measurements. The MCC1010 induces a current through a conductor. The MVC1010 measures the voltage drop across the conductor. This way, non-contact resistance measurements can be undertaken using the MCC1010 and the MVC1010 clamps.

ELECTRICAL SPECIFICATIONS

  • Step-down current ratio: 1000:1
  • Sensitivity: 1 mA/A

Amplitude and phase accuracy of the output signal

Primary current * Accuracy of output signal Phase shift of output signal
1 mA – 100 mA

≤1% + 5 μA

| not specified
0.1 A – 1 A|

≤0.7°

1 A – 10 A
10 A – 100 A| ≤0.5%
100 A – 1000 A| ≤0.3%

Unless otherwise specified, reference conditions are: 22 ±3°C, 50% humidity, sinusoidal current at 50/60 Hz, no DC offset, centred conductor, external magnetic field <40 A/m, load impedance (burden) ≤1 Ω.

  • Maximum continuous current: 1000 A at ≤500 Hz 100 A at >500 Hz
  • Frequency bandwidth: 15 Hz – 10 kHz
  • Crest factor: ≥6 for current up to 2000 A peak (300 A rms)
  • Influence of crest factor: ≤1% for CF≤4
  • Nominal load impedance: ≤1 Ω (burden resistance)
  • Maximum output voltage: ≤28 V peak (electronic limiter)
  • Influence of frequency: 30 Hz – 5 kHz ≤0.25%
  • Influence of conductor position in the jaws: ≤0.3% of amplitude
  • Load influence up to 5Ω: amplitude within specification up to 900 A ≤0.25% of amplitude above 900 A ≤0.1° on phase
  • Influence of DC offset: ≤2% up to 20 A DC
  • Working voltage: ≤600 V rms

ENVIRONMENTAL

  • Operating temperature: -20 °C to +50 °C, <85% RH -4 °F to +122 °F, <85% RH (excluding ice or dirt in the jaws)
  • Storage temperature: -40 °C to +70 °C, <85% RH -40 °F to +158 °F, <85% RH
  • Influence of temperature: ≤0.1% on amplitude phase within specification
  • Influence of humidity amplitude: and phase within specification (excludes ice or dirt on the jaws)
  • Max. conductor diameter: 52 mm

GENERAL SPECIFICATIONS

  • Casing protection: IP40 with jaws closed
  • Operating altitude: 2000 m
  • Output terminals: 4 mm shrouded sockets
  • Electrical safety: IEC 61010-1:2010 +IEC 61010-2-030:2010 + IEC61010-2-032:2002
  • EMC: IEC61326-1
  • Safety: CATIV 600 V Pollution degree 2
  • Weight: 700g
  • Dimensions: 45 mm x 110 mm x 218 mm

ORDERING INFORMATION

Item (Qty)| Cat. No.
MCC1010| 1010-516
MVC1010| 1010-518

SALES OFFICE

  • ADDRESS: Megger Limited Archcliffe Road Dover CT17 9EN England
  • T +44 (0) 1304 502101
  • E [email protected]

MCC1010–MVC1010_DS_en_V01

  • www.megger.com
  • ISO 9001
  • The word ‘Megger’ is a registered trademark

L’innovation à sa juste mesure

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

testoon User Manuals

Related Manuals