RCB BP05 Wireless Arm Type Blood Pressure Monitor User Manual

June 10, 2024
RCB

BP05 Wireless Arm Type Blood Pressure Monitor
User ManualRCB BP05 Wireless Arm Type Blood Pressure
Monitor

Remote Care Partners
Wireless Arm Type Blood

Pressure Monitor
Model # BP04
Model # BP05

INSTRUCTION MANUAL

INTRODUCTION

This Blood Pressure Monitor is provided to you by Remote Care Partners (RCP) for use exclusively with the health monitoring program from your provider.
Please read this manual carefully and keep it handy for future reference

NOTES ON SAFETY

Before using the unit, please read this section carefully to familiarize yourself with the features and operations.
Please do not use this monitor on babies, young children, or individuals who cannot express their consent for the readings to be taken.
This device is suitable for use by individuals 12 years old or older.
Individuals with severe arteriosclerosis are not suitable to take blood pressure measurements.
Blood pressure readings may change in the case of pregnancy. Pregnant women can consult their doctor before taking measurements.
Your new digital blood pressure monitor uses the oscilloscope method of blood pressure measurement. The monitor detects your blood’s movement through your brachial  artery and converts the movement into a digital reading. An oscilloscope monitor does not need a stethoscope, so the monitor is simple to use.
Intelligent inflation will reduce the uncomfortable feeling by incorrect inflation, shorten the measurement time, and prolong the cuff’s usage lifetime.
No modification of this equipment is allowed.
Placing the cuff hose around the neck may cause suffocation.
Do not use alcohol or any other flammable cleaning agent on this device. Use only soap and water.
Do not use the device near a cell phone, as it may cause incorrect readings.
Do not use the device in direct sunlight.
Do not use the equipment where flammable gas (such as anesthetic gas, oxygen, or hydrogen)or flammable liquid (such as alcohol)are present.
Do not touch the output of the AC adapter and the patient simultaneously.
Do not touch the live end of the battery and the patient simultaneously when changing the batteries.
Self-measurement is not medical treatment.Normal blood pressure is 120/80 or less
Handling batteries properly
As soon as old batteries run out, replace them with new batteries.

  • Remove dead batteries immediately.
  • Do not use old and new batteries together.
  • Align the polarities of batteries correctly.
  • Remove the batteries when the unit will not be used for more than 3 months. Otherwise, batteries may leak and cause damage to the equipment.

WHEN TAKING A MEASUREMENT

Avoid smoking, eating, or exercising for 30 minutes before taking measurements.
Do not allow other people to use your blood pressure monitor. The blood pressure monitor is specifically assigned to you and you only.
Allowing other people to use your monitor can mislead your provider.
Unless otherwise instructed by your provider, you can take your blood pressure on either arm.
Remove tight-fitting or thick clothing from your upper arm. The cuff should be placed on bare skin.
QUICK START
Avoid smoking, eating, or exercising for 30 minutes before taking measurements.

  1. Sit upright in a chair with both your feet on the floor.
  2. Apply the blood pressure cuff to the bare skin of your upper arm. Do not place the cuff over your clothing.
  3. Pull on the end of the cuff until it wraps securely around your upper arm. Place your arm on a table, so the cuff is at the same level as your heart.
  4. Press the “Stop/Start” button on the device. The cuff will automatically inflate, and the measurement will start. Do not talk or move around while the measurement is taking place.
  5. When the measurement is complete, the cuff will automatically deflate, and your systolic & diastolic pressure values plus the pulse rate will display on the Blood Pressure monitor.
  6. The readings will automatically be transferred to the RCP Smart Hub and sent to your provider.

PROPER POSITIONING FOR TAKING A BLOOD PRESSURE MEASUREMENT

RCB BP05 Wireless Arm Type Blood Pressure Monitor -
fig| RCB BP05 Wireless Arm Type Blood
Pressure Monitor - fig1
---|---

PRODUCT OVERVIEW

Monitor

RCB BP05 Wireless Arm Type Blood Pressure Monitor -
Monitor

a. Display
b. [Forward/Backward] buttons
c. [Connection] button (Date/Time)
d. [Memory] button
e. [START/STOP] button

Display

RCB BP05 Wireless Arm Type Blood Pressure Monitor -
Display

A

| | Memory symbol Appears when viewing readings stored in the memory.
---|---|---

B

| | Average value symbol Flashes when viewing the average of the latest 2 or 3 readings taken within the most recent 10 minute timeframe.

©

| | Sync symbol
Flashes/appears when your data needs to be transferred because the stored memory is either almost, or completely full. Once you pair your monitor with your smart device, transfer your readings immediately before the monitor deletes the oldest reading. Up to 60 readings can be stored in the internal memory of your monitor.

D

| Systolic blood pressure reading

E

| Diastolic blood pressure reading

F

| | Deflation symbol
Appears during cuff deflation.

G

| | Heartbeat symbol
Flashes while taking a measurement.

H

| Date/time display
| | Bluetooth ON symbol
Appears when your readings are being transferred.

off

| Bluetooth OFF symbol
Appears when Bluetooth is disabled.
J| | High symbol
Appears if your systolic blood pressure is i 130 mmHg or the diastolic blood pressure is 80 mmHg. Refer to sub-section 1.4 for more information.
---|---|---
K|
or
| Irregular heartbeat symbol
Appears when irregular rhythm is detected 2 or more times during a measurement. The symbol with the number appears as below:
Irregular heartbeat is detected 2 times:
 is detected 3 times or more:
An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while your monitor is measuring blood pressure.

Irregular heartbeat symbol may be displayed when the monitor detects arm movements. Refer to section 7 for more information.
If it continues to appear. we recommend you to consult with and follow the directions of your physician.
(L)| | Movement error symbol
Appears with your reading when your body is moving during a measurement. If it appears, remove the arm cuff and wait for 2 to 3 minutes. Then, reapply the cuff: remain still and take another measurement.
M| | OK symbol
Flashes when your monitor is connected to your smart device or readings are transferred successfully.
---|---|---
N| | Low battery symbol
Flashes when batteries are low.
| Depleted battery symbol
Appears when batteries are depleted.
O| Pulse display I Memory number
Pulse rate appears after the measurement.
When pressing the button, the memory number will also appear for approximately 1.5 seconds before the pulse rate shows again on the display.

PREPARATION BEFORE USE

Insert the Batteries
New batteries were installed in your device before shipping. Unless instructed by a support person, skip this section

  1. Push down the hook of the battery cover and pull downward.
  2. Insert 4 “AA” batteries as indicated in the battery compartment.
  3. Close the battery cover. As soon as batteries are correctly installed, the year flashes on the display. Refer to sub-section 2.2 to set the date and time manually, or refer to sub-section 4.1 to pair your monitor to set the date and time automatically.

Note

  • When the symbol flashes on the display, it is recommended you replace batteries.
  • To replace batteries, turn your monitor off and remove all batteries. Then replace with 4 new batteries at the same time.
  • Replacing batteries will not delete previous readings.
  • The supplied batteries may have a shorter life span than new batteries.
  • Disposal of used batteries should be carried out in accordance with local regulations.

Using the AC Adapter
We recommend keeping batteries in your monitor at all times, even when you choose to use the AC adapter. When the monitor is used without batteries installed, you will  lose any history that the monitor is saving if the AC adapter is disconnected.

  1. Insert the AC adapter into the AC adapter jock on the rear of your monitor BEFORE plugging it into an electrical outlet.
  2. Plug the AC adapter into an electrical outletRCB BP05 Wireless Arm Type Blood Pressure Monitor - AC Adapter
  3. To unplug the AC adapter, unplug the AC adapter from the electrical out FIRST, Then Remove the AC adapter plug from the monitor

Setting the Date and Time
It is not necessary to set the Date and Time on the device. The Date and Time of your readings are provided by the Smart Hub.
The Time and Date on the Smart Hub are automatically synchronized with our servers.
Attaching the Arm Cuff
Insert the air plug into the air jack until it clicks.

TAKING A MEASUREMENT

Note:

  • Avoid smoking, eating, or exercising for 30 minutes before taking a measurement.
  • Do not move or talk during measurement.
  • Measurement can be taken on either arm.

Sitting Correctly

  1. Relax
  2. Sit upright in a chair with both your feet on the floor.
  3. Remove tight-fitting or thick clothing from your upper arm. Apply the blood pressure cuff to the bare skin of your upper arm. Do not place the cuff over your clothing.
  4. Place your arm on a table so the cuff will be at the same level as your heart.

Proper Sitting Position For Taking Your Blood Pressure

RCB BP05 Wireless Arm Type Blood Pressure Monitor - Proper
Sitting

Applying the Arm Cuff

  1. Place your elbow on a table, palm facing upward, and wrap the arm cuff around the upper left arm.
  2. Pull on the end of the cuff until it wraps securely around your upper arm. Do not overtighten the cuff. Allow ½ inch between the bottom of the cuff and your elbow joint.
  3. The air tube should run along the inside of your arm.
  4. Make sure the cuff is at the same level as your heart.

Proper Cuff Placement For Taking Your Blood Pressure

RCB BP05 Wireless Arm Type Blood Pressure Monitor - Proper
Cuff

Taking the Measurement

  1. After the cuff is placed correctly on your upper arm, Press the “Start/Stop” Button.
  2. The cuff will automatically inflate, and the measurement will start. The rising pressure in the cuff is shown on the display. Do not talk or move around while the measurement is taking place.RCB BP05 Wireless Arm Type Blood Pressure Monitor - pressure
  3. When the measurement is complete, the cuff will automatically deflate, and your systolic & diastolic pressure values and pulse rate will display on the Blood Pressure monitor.
  4. The readings will automatically be transferred to the RCP Smart Hub and sent to your provider.

You can press the “Start/Stop” button at any time to stop the measurement and deflate the cuff.

Understanding The Results

Normal blood pressure is 120/80 or less..

RCB BP05 Wireless Arm Type Blood Pressure Monitor - Normal
blood

TROUBLESHOOTING

If any of the problems listed below occur during a measurement, make sure that no other electrical device is within 12 inches of the device. If the problem persists, refer to the table below.

Display/Problem Possible Cause Solution

appears or the arm
cuff does not inflate.| Air plug is not completely plugged into the monitor.| Insert the air plug Insert securely. Refer to sub-
The arm cuff is not applied correctly.| Apply the arm cuff correctly, then take another measurement.
Air is leaking from the arm cuff.| Replace the arm cuff with a new one. Refer to
appears or a
measurement
cannot be complete
after the arm cuff
inflates.| You move or talk during a measurement and the arm cuff does not inflate sufficiently.| Remain still and do not talk during a measurement. If “E2” appears repeatedly, inflate the arm cuff manually until the
systolic pressure is 30 to 40 mmHg above your previous readings. Refer t(
The systolic pressure is above 210 mmHg and a measurement cannot be taken.| Inflate the arm cuff manually until the systolic pressure is 30 to 40 mmHg above your previous readings. Refer
appears| The arm cuff is inflated exceeding the maximum allowable pressure.| Do not touch the arm cuff and/or bend the air tube while taking a measurement.
If inflating the arm cuff manually, refer to the end of sub-section 3.1.
appears| You move or talk during a measurement. Vibrations disrupt a measurement.| Remain still and do not talk during a measurement.
Display/Problem| Possible Cause| Solution
---|---|---
appears| The pulse rate is not detected correctly.| Apply the arm cuff correctly, then take another measurement. Refer to sub-section 2.4. Remain still and sit correctly during a measurement.
If the “ ” symbol continues to appear, we recommend you to consult with your physician.
RCB BP05 Wireless Arm Type Blood Pressure Monitor -
icon18
does not flash during
a measurement
| The monitor has malfunctioned.| Press the [START/
STOP] button again. If “Er” still appears, contact Support
855-477-7000
| The monitor cannot connect to a smart device or transmit data correctly.| Contact Support
855-477-7000
| The monitor is waiting for pairing with the smart device.| Contact Support
855-477-7000
| There are more than 40 readings in memory to be transferred.| Contact Support
855-477-7000
The date and time is not set.
| There are 60 readings in memory to be transferred.
flashes| Batteries are low.| Replacing all batteries
with 4 new batteries is

If you cannot solve your issue using the tables above, please contact Support at 855-477-7000. Do not disassemble the unit.

STORAGE AND MAINTENANCE

  • Keep the unit away from direct sunlight, extreme temperature, humidity, or moisture.
  • Use a dry, soft cloth to clean the unit, or if desired, use a cloth lightly dampened with water.
  • Do not use alcohol, benzene, thinner, or other volatile liquids to clean the unit.
  • Do not wash or expose the arm cuff to liquid.
  • Remove the batteries from the unit when it will not be used for more than three months.

NOTE: Please recycle or dispose of batteries per local regulations.
WARNING: Batteries may pose a choking hazard. As with all small items, do not let children handle batteries. If swallowed, seek medical attention immediately.
PRECAUTION:
Do not dispose of batteries in fire. Batteries may explode or leak.

SPECIFICATIONS

Model BP7250 REF HEM-7150T-Z
Display LCD digital display
Cuff pressure range 0 to 299 mmHg
Blood pressure measurement range SYS: 60 to 260 mmHg DIA: 40 to 215 mmHg
Pulse measurement range 40 to 180 beats / min.
Accuracy Pressure: ±3 mmHg

Pulse: ±5% of display reading
Inflation| Automatic by electric pump
Deflation| Automatic pressure release valve
Measurement method| Oscilloscope method
Transmission method| Bluetooth® low energy technology
Wireless communication| Frequency range: 2.4 GHz (2400 – 24815 MHz) Modulation: GFSK
Effective radiated power: < 20 dBm
Operation mode| Continuous operation
IP classification| Monitor: IP 20
Optional AC adapter: IP21
Rating| DC6 V 4.0 W
Power source| 4 “M” batteries 1.5 V or optional AC adapter (INPUT AC100 – 240 V 50 – 60 Hz 0.12 – 0.065 A)
Battery life| Approximately 1,000 measurements (using new alkaline batteries)

CONTACT INFORMATION AND NOTICES
Remote Care Partners
855-477-7000
www.remotecarepartners.com

FCC Compliance Statement

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:

  1. This device may not cause harmful interference, and
  2. This device must accept any interference received, including interference that may cause undesired operation.

FCC WARNING

This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide  reasonable protection against harmful interference in a residential installation.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to  radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception,  which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and the receiver.
  • Connect the equipment into an outlet different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals