ENERGY SUPPORT TOGA GC 2 Dissolved Gas Analysis with Automated Full Vacuum Degassing Instruction Manual

June 10, 2024
ENERGY SUPPORT

ENERGY Logo the ECH advantage in-lab | mobile | on-line | process
Gas-in-oil-analysis TOGA 2 made by GC
Dissolved gas analysis with automated full vacuum degassingENERGY SUPPORT
TOGA GC 2 Dissolved Gas Analysis with Automated Full Vacuum
Degassing

TOGA GC 2 Dissolved Gas Analysis with Automated Full Vacuum Degassing

Gas-in-oil-analysis
Dissolved gas analysis with automated full vacuum degassing
Product description
The TOGA GC 2 has been designed for a fast and reliable measurement routine. The analysis – done in 2 steps – follows the ASTM and IEC standard guidelines. For many
decades Dissolved Gas Analysis has proven to be a favourable method for insulation diagnostics. The TOGA GC 2 has been developed to be user-friendly and delivers reliable and accurate results. The analysis system allows a calibration of the complete system following the measurement routine.

ENERGY SUPPORT TOGA GC 2 Dissolved Gas Analysis with Automated Full Vacuum
Degassing - Total Gas Content

Total Gas Content
Equally important for insulation diagnostics is the Total Gas Content (TGC).
The TOGA GC 2 is currently the only system which measures the TGC during the degassing process accurately. The increase in the TGC is a clear indication of existing faults in the transformer, which can be thermal, electrical or mechanical in nature.
The analysis is done in 2 steps with high selectivity and accuracy, fully compliable with laboratory procedure. Degassing of the oil sample is done through multi-periodic vacuum extraction (according to ASTM D 3612 and IEC 60567). The gas sample is automatically transferred to the GC. Gas chromatographic analysis is done with a Thermal Conductivity Detector (TCD), Flame Ionisation Detector (FID) and with a Methanizer. All results are displayed in ppm, sensitivity is as low as 0.1 ppm for hydrocarbons. Therefore, new transformer oils and transformers following their periodic maintenance can also be analysed. Besides the results, all other operating conditions are registered and stored in a file together with all remaining parameters. The graphics of the results is displayed in real time on the monitor, in the table or/and as a test report in a print file. The integrated Expert System will evaluate the results according to international standards like „Duval“ or „IEC“.
Applications

  • Insulating oil
  • Gear oil
  • Mineral oil
  • Hydraulic fluids
  • Lubrication fluids
  • Biodiesel
  • Engine oil

Advantages

  • One-touch operation
  • Accurate measurement of total gas content
  • Most sensitive detection for Acceptance-, Service- and Routine tests (0.1 ppm)
  • Adaptable for Factory Transformer Heat-Run tests
  • Ideal solution for laboratories and workshops

ENERGY SUPPORT TOGA GC 2 Dissolved Gas Analysis with Automated Full Vacuum
Degassing - TOGA GC 2

Features and Results

  • Fully automatic procedure
  • Multi-periodic vacuum degassing
  • Full Dissolved Gas Analyses (11 gases) and Total Gas Content
  • Automatic transfer of gas sample
  • Easy calibration of the whole system with traceable calibration gas
  • Auto sampler available for 16 or 32 samples for unattended operation
  • Expert System for quick and easy result evaluation
  • Evaluation according to: IEC 60599, Roger ratios, Duval etc.

ENERGY SUPPORT TOGA GC 2 Dissolved Gas Analysis with Automated Full Vacuum
Degassing - T result

Automated version
The auto sampler technology has been designed to increase the efficiency of the TOGA system to make it more user-friendly and cost-efficient. The auto sampler is available for syringes with 30 or 50 mL and with 16 or 32 positions. It provides optimum value, time-efficiencies and excellent performance. There is no longer a need to change sample containers once the oil sample is taken from the transformer. Simply plug your syringe into the sample tray – absolutely contamination free.

ENERGY SUPPORT TOGA GC 2 Dissolved Gas Analysis with Automated Full Vacuum
Degassing - Auto sampler

Technical specifications

Vacuum degassing: Oil- and mercury-free
Vacuum sensor: 0.1 – 200 mbar
Sample volume: 30 – 50 mL
Gas chromatograph: Thermal Conductivity Detector (TCD), Flame Ionisation

Detector (FID), Methanizer, 2 micro packed
columns; furnace cassette with column and temperature programmed column heating system, separate
heating and cooling of each column
Autosampler:| Sample tray interchangeable for 30 mL or 50 mL glass syringes; available for 16 or 32 samples;
for all types of transformer oils (new and used); procedure fully-automated, contamination-free injection
of oil samples, integrated rinsing steps to avoid carry-overs
Calibration:| With standard gas for the complete measurement procedure
Power Supply:| 110 V / 60 Hz or 230 V / 50 Hz
Gas requirements:| 99.999 % argon, 99.999 % hydrogen, synthetic air (optional)
Dimensions:| 530 x 355 x 370 mm (W x D x H)
Weight:| 28 kg
Device control:| PC software (PC not included in the scope of delivery)

Distribution by

ENERGY LogoENERGY SUPPORT GmbH
Sperberweg 47
D-41468 Neuss Germany
T +49 (0) 2131 4039607
F +49 (0) 2131 4039608
[email protected]
www.energy-support.d

Documents / Resources

| ENERGY SUPPORT TOGA GC 2 Dissolved Gas Analysis with Automated Full Vacuum Degassing [pdf] Instruction Manual
TOGA GC 2 Dissolved Gas Analysis with Automated Full Vacuum Degassing, TOGA GC 2, Dissolved Gas Analysis with Automated Full Vacuum Degassing, Automated Full Vacuum Degassing, Vacuum Degassing
---|---