Honeywell VERSAFLOW SONIC 1000 Ultrasonic Clamp On Flowmeter User Guide

June 9, 2024
Honeywell

Honeywell VERSAFLOW SONIC 1000 Ultrasonic Clamp On Flowmeter

Honeywell VERSAFLOW SONIC 1000 Ultrasonic Clamp On
Flowmeter

Symbols Explanation

VERSAFLOW SONIC 1000/TWS 9000 F/…Ex

Ultrasonic clamp-on flowmeter

Installation, assembly, start-up and maintenance may only be performed by appropriately trained personnel. Check the nameplate for correct operating conditions.

For use in hazardous areas, special codes and regulations are applicable. Instruments must not be connected to power supply before reading instructions described in the supplementary manual.

The responsibility as to the suitability, intended use and corrosion resistance of the used materials against the measured fluid of this device rests solely with the operator.

For complete documentation (manuals, supplementary manuals, data sheets and certificates) please refer to:

http://www.honeywellprocess.com/

Special conditions to be observed

  • For ambient and process temperatures, specific product and electrical data, see Ex manual or certificate
  • The painted aluminium enclosures of the flow sensor and the splitter box shall be protected against electrostatic charging
  • For dimensions and details of the flameproof joints, the manufacturer shall be contacted
  • The tensile strength of the special fasteners is at least 700 N/mm² (property class A2-70 / A4 70)
  • The protection class of the signal converter enclosure is according EN 60529: IP65…68
  • All connection cables are fixed and installed correctly so adequate protection against possible damage is guaranteed
  • Only for connection to a separately certified TWS9000 F/…Ex flow converter
  • The instructions provided with the product shall be followed in detail to assure safe operation Ex ► Type Examination Certificate: CSASNe 21 ATEX 2082 X / CSANe 21 ATEX 1081 X

General

Ta = -40…+70°C / -40…+158°F
Tp = -40…+200°C / -40…+392°F

Maximum ambient and process temperatures are depending on version (e.g liner material , DN DN size), temperature and protection class and maximum surface temperature of sensor.

Diameter range and rail versions

Diameter range and rail versions

For initial set up, we strongly recommend the use of the relevant manuals in addition!

General

General

Device nameplate

Device nameplate

Check the device nameplate to ensure that the device is delivered according to your order.

Device nameplate

Electrical connection

All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate! Observe the national regulations for electrical installations!

Observe without fail the local occupational health and safety regulations. Any work done on the  electrical components of the measuring device may only be carried out by properly trained specialists.

Electrical connections signal converter

Electrical connections signal converter

Power supply – grounding

Ⓐ  100…230VAC (-15% / +10%), 22VA
Ⓑ Diameter range and rail versions 24VDC (-55% / +30%), 12W
Ⓒ 24VAC/DC (AC: -15% / +10%; DC: -25% / +30%), 22VA or 12W

The device must be grounded in accordance with regulations in order to protect personnel against electric shocks.

For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
Refer to the manual for connection of Ex (/i ) acc. to NAMUR

① Sensor cable connections
② I/O connections
③ Mains supply connection

Installation

Installation

Change the position of the transducer

Change the position of the transducer

① – ② Choose sensor size
③ – ④ Choose the applicable measuring mode
⑤ Go through menu X1…X5

General installation of the rails

General installation of the rails

Greasing the transducer surfaces

Greasing the transducer surfaces

Check the VERSAFLOW SONIC manual for more details on installation options (e.g. Large version installation or applying solid pads)

General configuration instructions

General configuration instructions

Go through menu X1…X5

① Enter the values for the installation menu, X1…X7.2.8
② Read the advised mounting distance in menu X7.2.3
③ The advised distance (menu X7.2.3) must be > 246 mm / 9.7″ for V-mode
Follow up with menu X7.2.4…X7.2.8 Run the optimization loop
④ Repeat if necessary.
Adjust settings in menu X5 / X7 and finish with X7.2.9…7.2.11

Installation menu X

Installation menu X

Quick Setup

① Status message
② Tag number
③ Measured variable
④ Signal strength
⑤ Operating keys

Quick Setup

Connection diagram

Connection diagram

Contact

Honeywell Process Solutions
2101 City West Blvd
Houston, TX 77042
www.honeywellprocess.com

© Honeywell International Inc.
Subject to change without notice

Honeywell Logo

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>