careium A150 Eliza S Digital Social Alarm and Smartcare Hub User Guide

June 4, 2024
careium

A150 Eliza S Digital Social Alarm and Smartcare Hub
User Guidecareium A150 Eliza S Digital Social Alarm and Smartcare
Hub

Read first: Safety information

Always read and follow the safety information accompanied by this symbol.

  • Always test the system according to the instructions before use and always test the product after adjustments have been made.
  • The device is not always suitable for all users and should not be seen as a substitute for the caregiver’s monitoring routines.
  • The device is NOT intended for life-sustaining equipment, where any technical failure may result in serious injury or death.
  • Check the unit regularly and replace if necessary.
  • Users should pay particular attention to the potential for interference from other systems operating in the same or adjacent radio frequency bands.
  • The compartment covers may be opened only by authorized persons in accordance with the instructions in the service manual.
  • Battery replacement may be performed only by authorized persons in accordance with the instructions in the Maintenance section. Only use recommended battery type as stated in the Technical Data section.

CAUTION
Only use batteries, power supplies, and accessories that have been approved for use with this particular model. Connecting other accessories may be dangerous and may invalidate the type of approval and guarantee.

  • Only use recommended power supply as stated in the Technical Data section.
  • The power supply can be used as a disconnecting device.
    The wall socket shall be installed near the equipment and shall be easily accessible.

IMPORTANT!
This device uses radio signals, the mobile phone network, and the terrestrial network. This means that connection cannot be guaranteed in all circumstances.

1.1. Medical units
The use of equipment that transmits radio signals as a social alarm, can interfere with insufficiently protected medical equipment. Consult a doctor or the manufacturer of the equipment to determine if it has adequate protection against external radio signals, or if you have any questions.
To avoid potential interference, manufacturers of implanted medical devices recommend a minimum separation of 15 cm between a wireless device and the medical device.
If you have any reason to suspect that interference is taking place, turn off the device immediately. If you have any questions about using your wireless device with an implanted medical device, consult your healthcare provider.

Important information

All systems using radio and network communication are subject to interference beyond the user’s control. Products from Careium are designed to minimize the impact of such interference.
Nevertheless, the user must be aware that the system components can be subjected to interference or other influences that may cause malfunction.
It is therefore important to regularly check that every part of the system works in all areas, especially radio communications. Contact your supplier immediately in case of any suspected malfunction.
Users should pay particular attention to the risk of disruption from products that communicate using the same or adjacent radio frequency bands.
For further information, please contact your supplier or refer to the service manual or the documentation at www.careium.com/support/.
The device has a rechargeable battery that enables the device to be operated even in the event of a power failure.
The device will beep for a short while after power failure.
The alarm receiving center will be notified in the event of a power failure as well as a low charge level. When the battery reaches a low charge level, this is indicated by a blinking status light, see 5.3. Indicator lights on p. 10. The backup battery power supply secures the operation for up to 72 hours.

Package contents

  • Eliza S
  • Power supply
  • Radio trigger with accessories
  • User guide/Quick start guide

Description of functions

Eliza S is a digital social alarm and smart care hub.
Adaptable for future needs and connects to a wide range of accessories and welfare services – enabling the user a tailor-made solution.
Eliza S allows the user to call for assistance if they have an accident or a fall at home. Eliza S can potentially be a lifesaving device if someone has a fall or some other accident when they are alone.
Eliza S has a built-in module for 2G, 3G, and 4G mobile networks, the unit has the possibility to transfer alarms over mobile telephone networks. The unit offers the user the possibility of two-way hands-free voice communication, even if the users are far away from the Eliza S, using the existing standards for VoIP (Voice over Internet Protocol) calling.
The unit uses SIP, a standard protocol to initiate speech and video over digital connections. In addition to voice-connected alarm calls, Eliza S offers a platform for transfer of detailed alarm information, speech messages, online status, etc.
Eliza S is assigned a unique identification number (ID) that is transferred and corresponds to personal information stored in the alarm receiving center (ARC).
The radio trigger is adapted for use with the unit,  see 9. Radio trigger on p 17.

4.1. Activate the alarm
The assistance alarm can be activated:

  • by pressing the button on a radio trigger.
  • by pressing the integral alarm button on the Eliza S.

On activation, Eliza S calls an alarm receiver to summon help.
4.2. Alarm transmitters
50 radio transmitters (triggers/sensors) can be programmed with the Eliza S, e.g. a radio trigger. In the event of an alarm from a radio transmitter, the battery status of the radio transmitter is tested.
4.3. Alarm receivers
Up to 10 alarm receivers can be programmed. The alarm is identified using an alarm code that is programmed into the Eliza S.
The person receiving the alarm can then speak directly with the person raising the alarm and take appropriate action.
4.4. Other functions
Additional functions can be activated or modified – refer to the service manual or contact your supplier for further information.

Overview

5.1. Back view

careium A150 Eliza S Digital Social Alarm and Smartcare Hub -
parts

  1. I/O connector
    3.5 mm audio out or wired input

  2. USB-port (optional)
    For USB expansion

  3. External antenna connector
    For an external antenna that must be adapted for use with Eliza S

  4. Power switch
    Turns Eliza S on or off

  5. Power supply socket 9V, 2A

5.2. Top view

1. Loudspeaker 4. Square button (green)
2. Alarm button (red) 5. Circular button (yellow)
3. Power indicator 6. Status indicator

5.3. Indicator lights

| ALARM
Steady light
Blinking light| RED
Normal operation
Alarm activated
---|---|---
| ON
Steady light
Blinking light| GREEN
Mains power
Mains failure
| SQUARE| GREEN
Press to cancel an
alarm
| CIRCLE| YELLOW
| STATUS
Off
Steady light
Blinking light| YELLOW
Normal operation
Network problem
Low battery

Installation

6.1. Remove the top cover

CAUTION
Only authorized persons may open the cover. Turn off the device and disconnect the external power supply before removing the cover.careium A150 Eliza S
Digital Social Alarm and Smartcare Hub - Figure 1

  • Ensure that Eliza S is switched off.
  • Place your fingers of both hands in the groves on the sides of Eliza S and gently pull outwards and upwards on the top cover.

IMPORTANT
Do not bend or twist the cover excessively as it may get damaged.

6.2. Inside view
The illustration below is shown with the top cover removed.

careium A150 Eliza S Digital Social Alarm and Smartcare Hub - Figure
2

1. Battery 4. Maintenance USB-port
2. Battery connector 5. Service menu button
3. SIM card slot

6.3. Install the SIM card

IMPORTANT
Turn off the device and disconnect the external power supply before removing the cabinet.

  • Carefully move the battery aside to expose the SIM card holder. If the battery already is installed, please note if the battery is connected to not damage the wiring.
  • Install the SIM card according to the picture, and note the markings.
  • Replace or install the battery.

IMPORTANT
Always make sure the battery is connected if the device is new or has been stored.

  • Replace the cover and gently push in into place. Listen for the clicks that tells you that the cover is in place.

CAUTION
1234 or 0000 are not appropriate PIN codes for the SIM card, if any of these PIN codes are used, the code should be changed before installation. Change of PIN code is preferably done using a mobile phone.

Connection
Eliza S is easy to install:

  1. Connect the power supply to the wall socket and the power supply lead to the DC input socket. Keep the device connected to the main power at all times.
  2. Turn the power switch to the I position
  3. Check that the On indicator lamp has a steady red light
  4. The status light shall go off within 90 seconds if the mobile network (GSM/UMTS/LTE) is detected
  5. Carry out a test alarm from the Eliza S

CAUTION
Installation must be performed only by authorized persons in accordance with the instructions in the service manual.

Communication path

Example of a simplified communication path careium A150 Eliza S Digital
Social Alarm and Smartcare Hub - Figure 3

  1. Radio or wired alarm trigger activated.
  2. Connection via the mobile network (GSM/UMTS/LTE) to Voice, IP alarm protocol and additional data
  3. IP connection via the mobile network (GSM/UMTS/LTE)
  4. Voice connection via the mobile network (GSM/UMTS/LTE)
  5. IP connection to a digital Alarm Receiving Centre (ARC)
  6. Optional connection to the i-care® online device management system.

Pairing

Contacts, alarm codes, and settings are programmed using www.icareonline.com or by using a computer via a web browser.
In some cases, the Eliza S can be delivered preprogrammed by the supplier. For help with pairing, please refer to the service manual or contact your supplier.

8.1. Pairing radio triggers
Up to 50 radio triggers/sensors or other radio-alarm transmitters can be paired to Eliza S.

  • Press and hold the Green square button for 4 seconds and release when an audible signal is played and the Power indicator starts to blink.
  • Activate the radio transmitter. When Eliza S receives the radio signal an audible signal is played.
  • Confirm the pairing by pressing and releasing the red integral alarm button once on Eliza S. The successful pairing is confirmed with an audible signal.
  • The radio trigger is now paired and set to a default alarm type. The alarm type can be changed anytime using the optional i-care® online service.
    Note: Repeat the steps above to pair additional radio triggers.

Check the radio coverage
Always check the function after installation and pairing.

  • Press the Green square button for 4 seconds.
  • Press/activate the radio trigger/accessory. When Eliza S receives the radio signal, you will hear a tone. Repeat this in all areas where the radio trigger/accessory shall function.

8.3. Check the mobile network signal strength
When installing the Eliza S, the mobile network signal strength must be verified.
The mobile network signal strength can be checked via www.icareonline.com or by using a local computer via any web browser. For more information, see the service manual.

8.3.1. On-device check
Check the mobile network signal strength on the device.

  1. Remove the compartment top cover of the device to access the Service menu button.
  2. Press the Service menu button, see 6.2. Inside view on p. 12.
  3. A synthetic voice will say, -Service Mode!.
  4. Press the Green square button repeatedly until a synthetic voice will say, -Signal strength!.
  5. Press the Main Integral alarm button once.
  6. The actual signal strength will be presented via a synthetic voice.
  7. Press the Green square button repeatedly until a synthetic voice will say, -Exit!.
  8. Press the Main Integral alarm button once.
  9. Replace the compartment cover.

Note! If an acceptable value can’t be achieved an external antenna must be used.

8.4. Check the functionality
Always check the function after installation and pairing.
Test alarms shall be done regularly and in a way that secures the whole alarm chain from the radio transmitter to the alarm receiver.

8.5. Automatic test
The functionality of the Eliza S and its communication can automatically be checked down to every other minute through the optional i-care® online service. It can also be checked by automatic function monitoring or a periodical test report to the Alarm receiving center (ARC), for example, once a day or once a week.

Radio Trigger

The radio trigger uses two-way radio communication, which means that when Eliza S receives an alarm from a radio trigger the care phone will send a confirmation of receipt to the radio trigger. This is confirmed on the radio trigger by the indicator light change from red to green. If the radio trigger doesn’t get any confirmation, the alarm is repeated.
The radio trigger can be worn using a wristband or neckband.
The radio trigger is waterproof according to IP67 by an integrated gasket and can be carried in the shower and in wet environments.

General information

  • Verification of mobile network coverage should always be done as a part of the installation.
  • In order for an alarm to be sent, at least one alarm receiver/response center needs to be programmed.
  • The Eliza S must not come into contact with water or any other liquid.
  • Used batteries shall be disposed of in an environmentally-responsible manner.
  • Eliza S needs to be equipped with a mini-SIM (2FF) and have an appropriate mobile subscription including voice and data.

Maintenance

11.1. Cleaning
The Eliza S and the radio trigger should be cleaned using a slightly damp cloth.
Do not use other solvents when cleaning.
Note! By ordering a complete housing set you can upgrade your used radio trigger to look as brand new.
11.2. Reinstatement
If the Eliza S is reinstated at a new user, it shall be reset to default settings and re-programmed according to the data and requirements of the new user.
11.3. Replace the battery
Battery should be replaced within three years, or a.s.a.p after battery alarm that is not related to a power cut. It’s normal that a long-term power cut generates a battery alarm.
11.4. Replace the radio trigger battery
To replace the radio trigger’s battery, please refer to the service manual or contact your supplier.

CAUTION
Battery replacement may be performed only by authorized persons in accordance with the instructions in the service manual.

Technical data

12.1. Carephone Eliza S

Model no: A150
Dimensions: 170 x 138 x 43 mm (L x W x D)
Weight incl. power supply and cables: 0.6 kg
Power supply: 9 VDC power supply
Power consumption: 2W (Nominal)
Backup battery: 3.7 V, 2000 mAh, Li-Ion
Backup time: Up to 72 hours (new and fully charged battery)
Inputs: 1 Normally Open (NO)
Output: Audio
Communication: GSM 2G, UMTS/HSDPA 3G, LTE 4G, Wi-Fi (optional with USB-

version)
Radio frequency band:| 869,2-869,25 MHz
A number of radio transmitters:| Up to 50
Equipment class:| Class 1 radio equipment

12.2. Mobile networks
(MHz) [maximum radio-frequency power/dBm]

Network bands:
2G GSM| (900) [33], (1800) [30]
---|---
3G UMTS| 1 (2100) [24], 8 (900) [24]
4G LTE FDD| 1 (2100) [24], 3 (1800) [24],
7 (2600) [24], 8 (900) [24],
20 (800) [24] , 28 (700) [24]
SIM card interface:| 1,8V and 3V mini-SIM (2FF)

12.3. Other radio
(MHz) [maximum radio-frequency power/dBm]

SRD: (869.200-869.250) [10]
Bluetooth, version: 5, BLE, long-range (2402 – 2480) [10]
ZigBee: IEEE 802.15.4  (868.0 – 868.6) [20],  (2400 – 2483.5) [20]

Wi-Fi (optional with USB version and Wi-Fi
adapter):| IEEE 802.11 b/g/n, (2412 – 2472) [19] IEEE 802.11 a/n, (4915 – 5835) [18]

12.4. Portable radio trigger, Enzo

Dimensions: 39 x 32 x 11 mm (L x W x D)
Weight 12 g (excluding wristband or necklace)
Battery: 3V lithium battery CR 2032
Battery life: Up to five years
Water resistance: Complies with IP67
Radio frequency band: 869,2-869,25 MHz
Radiofrequency power: max. 10 mW (10 dBm)
Radio range: indoors ≥ 40 meters (normal environment)
outdoors ≥ 300 meters (typical, line of sight)

12.5. Environment
This product is intended for indoor use in a normal residential environment.

Operating temperature: 5° C (41° F) – 35° C (95° F)
Humidity: 0% to 75% relative humidity (non-condensed)
Environmental class: 1 (EG-I)

Others

13.1. EU Declaration of conformity
Hereby, Careium declares that the radio equipment type A150 and 9150 is in compliance with Directives: 2014/53/EU and 2011/65/EU.
The full text of the EU declaration of conformity is available at the following internet address: www.careium.com/dofc.

13.2. Ecodesign declaration, energy efficiency for external power supply
Hereby, Careium declares the external power supply for this device is in compliance with Commission Regulation (EU) 2019/1782 regarding ecodesign requirements for external power supplies pursuant to Directive 2009/125/EC.
The full information regarding the ecodesign requirements is available at the following internet address: www.careium.com/ecodesign

13.3. Environmental information
This product complies with the requirements of the EU directives, 2012/19/EU (WEEE) and 2006/66/EC (batteries). These directives regulate the product liability for electrical and battery recycling with the purpose of increasing recycling and minimizing waste. The unit is marked with the crossed-out wheeled bin logo, which indicates that it shall be handed in for recycling.
The product can be returned free of charge to a recycling station that is connected, directly or via a recycling system, to Careium or to your distributor. For detailed instructions, please check with your distributor or visit our website, www.careium.com.

Note! The WEEE information and recycling instructions apply to European Union member states only. For other countries please check local legislation or contact your distributor. Manufactured in accordance with the EU directive, 2011/65/EU (RoHS2). The materials used in the radio trigger wristband and neckband meet the textile safety requirements of Oeko-Tex standard 100.

13.4. Legal notices
The information contained here is subject to change without notice. The only warranties for Careium products and services are set forth in the express warranty statements accompanying such products and services.
Nothing herein should be construed as constituting an additional warranty.
Cardium shall not be liable for technical or editorial errors or omissions contained herein.
13.5. UK Declaration of Conformity
Hereby, Careium declares that the radio equipment type A150 and 9150 is in compliance with the relevant UK legislation.
The full text of the UK declaration of conformity is available at the following internet address: www.careium.com/dofc

Manual_Careium_Eliza_S_A150_100x148_en_v10
©2021 Careium AB. All rights reserved.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals