UNI-T UT801 Bench Type Digital Multimeters User Manual

June 9, 2024
UNI-T

UNI-T UT801 Bench Type Digital Multimeters

Overview

Digital Bench-Type Multimeter Model UT801 is the maximum reading 1999 and 3 1/2 digits and UT802 is the maximum reading 19999 and 4 1/2 digits, both models are in manual range, DC / AC current type digital multimeter, This is also the extra large characters in LCD display backlight with full function, full measurement and full overload protection, as well as a good product design outlook, In addition to all the conventional features, include DC/AC voltage, DC/AC current, resistance, frequency, capacitance, temperature ℃, Transistor hFE、diode and continuity buzzer.

This operating manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly.

Unpacking Inspection
Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part. If you find any missing or damage, please contact your dealer in your country.

  • Operating Manual 1 piece
  • Test Lead 1 pair
  • Alligator Clip 1 pair
  • K Type Temperature Probe 1 piece ( For the temperature under 230℃testing)
  • Multi-Purpose Socket 1 piece
  • Power Cord 1 piece (AC220V 50Hz DC9V/200mA)

Safety Information

This Meter compl ies wi th the standards IEC61010-1 in pollution degree 2, overvoltage category (CAT II 1000V) and double insulation. If you can not follow up this operating instruction to use the meter and it reduces the chance to have an using protection.

  1. Before using the Meter and Test Leads inspect both items. Do not use the Meter and Test Leads if it is damaged or the case (or part of the case) is removed or no reaction on LCD display. Prohibited to use the meter without housing or housing without screw fix up in order to avoid possible electric shock or to avoid possible damage to the meter or to the equipment under test.
  2. If the damage of test leads, use only the same model number or identical electrical specifications replacement parts.
  3. Do not use your finger to touch on any testing cable, connector, unused terminal input or circuit during the testing stage.
  4. When the meter working at an effective voltage over 60V in DC or 30V rms in AC, special care should be taken for there is danger of electric shock.
  5. Selecting the correct terminal input and turn the rotary switch to select the measuring function. In case of no any idea on the value input of the current, just simply test from the high value to low one.
  6. Do not overload voltage or current on EITHER between terminal and terminal OR between terminal and grounding which indicate on meter limitation.
  7. The rotary switch should be placed in the right position and no any changeover of range shall be made during measurement is conducted to prevent damage of the Meter.
  8. Do not use or store the meter in an environment of high temperature, humidi ty, flammable and electromagnetic environment. The performance of the meter may deteriorate after dampened.
  9. The internal circuit of the meter shall not be altered at will to avoid damage of the meter and any accident
  10. Replace the battery as soon as the battery indicator “ ”Appears. With a low battery, the meter might produce false readings that can lead to electric shock and personal injury.
  11. Turn the meter off when it is not is use and take out the battery when not using for a long time.

General Specifications

  1. Maximum Voltage between terminal input and COM: 1000V( except 200mV, 250V)
  2. μA mA terminal input protection: (CE)250mA 265V auto recovery fuse
  3. 10A terminal input protection: (CE)F1 (10A H 1000V) Fast-type melted fuse Φ10.3x38mm
  4. Resistance input protection: PTC/1000V
  5. Capacitance input protection : (CE)F2, F3 (0.5A H 1000V) Fast-type melted fuse Φ6.35×31.8mm
  6. Frequency input protection: PTC/1000V
  7. Temperature input protection: (CE)0.5A 1000V fuse
  8. terminal input protection: PTC/1000V
  9. hFE input protection: (CE)250mA 265V auto recovery fuse, F3 (0.5A H 1000V) Fast type melted fuse Φ6.35×31.8mm
  10. Display: LCD full function signal display, maximum reading is 1999(UT801) , 19999(UT802) Updates 2-3 times/second
  11. Range: Manual
  12. Polarity Display: Auto
  13. Overload indication: 1
  14. Battery Deficiency:
  15. Operating Temperature: 0~40℃(32℉~104℉)
  16. Storing Temperature: -10~50℃(14℉~122℉)
  17. Relative Humidity: 0℃~30℃ below ≤75% 30℃~40℃ ≤50%
  18. Electromagnetic Field: Under 1V/m the influence of radiated radio-frequency electromagnetic field phenomenon, Total accuracy= specific accuracy+ measurement 5%, Over 1V/m radiated radiofrequency electromagnetic which do not have any reference data on this topic.
  19. Power: AC(external power adapter AC220V/ DC9V-200mA) or DC(internal battery type 2 R14/1.5V 6 pieces)
  20. Product size: (300x245x105)mm
  21. Product Net Weight: About1500g(without the accessories)
  22. Safety Compliances: IEC 61010: CATⅡ1000V

LCD Display

Functions

Functional Buttons

Operational Measurement Guide

  1. Selecting the correct terminal input and turns the rotary switch to &elect the measuring function. If fail& to do that, the buzzer beep and the warning signal nash on.
    Range| Alarm Alert On False Terminal Input
    ---|---
    V Hz 0·c| 10A mAµA
    mAµA     hFE F| 10A
    10A| mAµA
  2. DC or AC Voltage Measurement
    • To avoid harms to you or damages to the Meter from electric shock, please do not attempt to measure voltagas higher than 1000 V although reading may be obtained.
    • The Meter has an in put impedance of 10MO(exoeptUT802/ACV input impedance of 2MO) This loading effect can cause measurament errors in high impedance circuits and so you need to 1ake a note on it.
  3. DC or AC Current Measurement
    • Before connecting the Meter in serial with the tasted return circuit, closed the return circuit current to avoid the danger of sparking.
    • Do not use over >10A current measurement. Although the mater can work on below 20A current testing but for Iha avoiding any harms to you or damages to the Meter.
  4. Measuring Resistance, Diodes, Continuity or capacitance
    • To maintain the measuremen1 accuracy, diS00unt circuit power and discharge all Iha high voltage capacftors during Iha measuring resistance.
    • When measuring high resistance on 1MO or above, it is normal to take several seconds to obtain a stable reading. In order lo obtain stable reading, choose shorter test lead to carrying out measurement.
    • The test leads and the Meter inside wire will bring around 0.10~0.20 of error 1o resistance measurement when measuring low resistance. To obtain accurate readings in low-resistance, short circuit the test leads beforehand and record the reading obtained, call this reading as X. Then use the equation: measured resistance value (Y} – {X) = accurate readings of resistance.
    • During maasuramant, Diodes is in a good silicon junction drops between 500mV~800mV &Iii the normal value. The continuity measuremen1, the poles between resistance is >1000. it is a short circutt, but on the poles between resistance is s100, it is a good connection, buzzer is con1inually beep on, and the reading value is nearly to the circuit resistance value, Untt is C.

Accuracy Specifications

  • Accuracy: ±(% reading + digits), guarantee for 1 year
  • Operating temperature: 18’C~28’C
  • Environmental humidity: Less than 75%RH
  1. DC Voltage
    • Input Impedance: is average on 1 OMO Maximum
    • Voltage Input: 1000V (Except 200mV, 250V)
  2. AC Voltage
    • Input Impedance: UT801 is average on 10MO UT802 is about 2MO.
    • Maximum Voltage Input: 1000\frm&
    • Fraquency: 45Hz~400l-lz
    • Display: True RMS
  3. DC Current****
    Range| Resolution| Accuracy Tolerance:±(% Reading+Oigits)
    ---|---|---
     | UT801| UT802| UT801| UT802
    200µA| 0.1µA| 0.01µA| ±(0.8%+2)| ±(0.5%+20)
    2mA| 1µA| 0.1µA
    20mA| 10µA| 1µA
    200mA| 0.1mA| 0.01mA
    10A| 10mA| 1mA| ±(2.0%+4)| ±(1.5%+40)
    • When l!5A, Continuou& mea&Uremen1 lass than 10 159COnds at an interval more than 15 minutes.
  4. AC Current****
    Range| Resolution| Accuracy Tolerance:±(% Reading+Digits)
    ---|---|---
     | UT801| UT802| UT801| UT802
    2mA| 1µA| 0.1µA| ±(1.0%+3)| ±(0.8%+40)
    20mA| 10µA| 1µA
    200mA| 0.1mA| 0.01mA
    10A| 10mA| 1mA| ±(2.5%+5)| ±(2.0%+40)
    • Frequency: 45Hz~400l-lz
    • When 5A, Continuous measuramen1 less than 10 seconds at an interval more than 15 minutes.
  5. Resistance****
    Range| Resolution| Accuracy Tolerance:±(% Reading+Digits)
    ---|---|---
     | UT801| UT802| UT801| UT802
    2000| 0.10| 0.010| ±(0.8%+3)| ±(0.5%+10)
    2k0| 10| 0.10
    20k0| 100| 10
    200k0| 1000| 100
    2MO| 1k0| 1000
    20MO| 10k0|  | ±(1.5%+5)|
    200MO|  | 10k0|  | ±(5%+40)
    • When >100MC resistance measuremen1 as reference purpuse.
  6. Capacitance****
    Range| Resolution| Accuracy Tolerance:±(% Reading+Digits)
    ---|---|---

20nF

| UT801| UT802| UT801| UT802
10pF| 1pF| ±(4%+3)| ±(4%+10)
2µF| 1nF| 100pF
200µF’| 100nF| 10nF| ±(5%+5)| ±5%+10)

  • “>40μF capacitance measurement as reference purpuse.
  1. Frequency****
    Range| Resolution| Accuracy Tolerance:±(% Reading+Digits)
    ---|---|---
     | UT801IUT802| UT801| UT802
    2kHz| 1Hz  0.1Hz| ±(1.5%+5)| ±(1.2%+10)
    200kHz| 100Hz 10Hz
    • Input Amplitude a:
    • (2kl-lz range) 50mV: sas30Vrms
    • (200kHz range)150mV: sa:530Vrms
  2. Temperature
    Range| Resolution| Accuracy Tolerance:±(% Reading+Digits)
    ---|---|---
     | UT801| UT802| UT801| UT802
    -40–20°C| 1·c| o.1·c| -(8%+5)| -(8%+40)

    -2o·co·c| ±(1.2%+4)| ±(1.2%+30)
    0-10o·c| ±(1.2%+3)| ±(1.2%+25)
    100~1000″C| ±(2.5%+2)| ±(2.5%+20)

    • Thermocouple: It is suitable to use K type of thermocouple. This include point contact K-type thermocouple can only be used on less than 230-C temperature measurements.
  3. Diode Test
  4. Continuity Test
    • When circuit is disconnected with resistance value >1000, the buzzer does not beep. When circuit is in good connection with resistance value s100, buzzer beeps continuously.
  5. Transistor hFE****
    Range| Resolution| Remarks
    ---|---|---
    hFE| UT801 IUT802| lbO is about 10µA, Vee is about 2.5V
    w ow

Replacing the battery(see Diagram 4)
Specifications and other information shown on this instruction manual are subject to change without notice:

UNI-TREND TECHNOLOGY (CHINA) GO., LTD.
No6, Gong Ye Bei 1st Road,
Songshan Lake National High-Tech Industrial
Development Zone, Dongguan City,
Guangdong Province, China
Tel: (86-769) 8572 3888
http://www.uni-trend.com

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals